K Number
K233321

Validate with FDA (Live)

Date Cleared
2024-06-13

(258 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Double Balloon Endoscope Model EN-840T: This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine and rectum. Never use this product for any other purposes.

Over-Tube TS-1214C: This product is intended to be used as an accessory with the FUJIFILM Double Balloon Endoscope cleared for use with an Over-tube is used to assist with the movement of the scope inside the upper or lower digestive tract. This product is not intended for use for any neonates, infants or children.

Device Description

a. Double Balloon Endoscope Model EN-840T: The insertion portion of the device has a mechanism (hereinafter "the bends the tip from right to left and up and down, and a flexible tube (hereinafter") consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the tip is arranged inside the insertion portion for inserting the surgical instrument. The insertion portion of the endoscopes comes into contact with the mucosal membrane. The tip of the insertion portion is called the "Distal end" which contains the Imaging section, Balloon air feed outlet , Distal cap, Objective lens, Air/water nozzle, Water jet nozzle, Instrument channel outlet, Objective lens, and Light guide. The bending portion is controlled by knobs on the control portion section to angulate the distal end to certain angles. The Flexible portion refers to the long insertion area between the Control portion (a part of Non-insertion portion). This portion contains light guides), air/water channels, a forceps/suction channel, a CMOS image sensor, and cabling. The class fiber bundles alow light to travel through the body cavity, thereby providing enough light to the CMOS sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction. The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connects the endoscopes to the light source.

b. Over-tube TS-1214C: The over-tube TS-1214C is introduced in the patient's anatomy with the pairing endoscope. TS-1214C is assembled over the outer diameter of endoscope. The endoscope and the over-tube are not advanced both at the same time, but alternatingly and successively. TS-1214C is provided sterile and single-patient use only. This accessory is a Class 2 device.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device (Double Balloon Endoscope EN-840T and Over-tube TS-1214C). It describes the device, its intended use, and a comparison to predicate devices, and lists various non-clinical tests performed to demonstrate substantial equivalence. However, it does not include information about specific acceptance criteria, reported device performance in a table, sample sizes, ground truth establishment, expert qualifications, or any human reader studies (MRMC or standalone AI performance).

Therefore, based on the provided text, I cannot answer most of your detailed questions about acceptance criteria and the study proving the device meets those criteria. The document focuses on regulatory compliance through substantial equivalence, primarily established through non-clinical testing against recognized standards.

Here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Cannot be extracted. The document mentions various tests were conducted (EMC, Electrical safety, Biocompatibility, Endoscope specific testing, Sterility, Field of view, Bending capability, Rate of suction, Working length, Diameter of forceps channel, Viewing direction, Resolution, LG output) and states "The subject device met performance specifications" or "Bench testing data demonstrated that the subject devices are substantially equivalent in performance." However, it does not provide specific numerical acceptance criteria or reported performance values in a table format.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Cannot be extracted. The document mentions "Bench testing data" and "Endoscope specific testing" but does not specify the number of samples or units tested for any of these non-clinical tests, nor the data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This device is an endoscope and over-tube, which are hardware devices for visualization and intervention, not an AI system that requires expert-established ground truth for diagnostic accuracy. The testing described is non-clinical (electrical, mechanical, biocompatibility, sterility).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. As above, this is for non-clinical hardware testing, not diagnostic AI performance assessment requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. The document describes a traditional medical device (endoscope and over-tube) and its non-clinical testing for safety and effectiveness, based on substantial equivalence to predicate devices. There is no mention of AI or assistant features, nor any MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This device is not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Cannot be extracted and not applicable in the usual sense. For non-clinical tests, "ground truth" would correspond to the established measurement standards and specifications (e.g., a caliper for diameter, a force gauge for bending, a flow meter for suction rate, etc.). The document indicates these tests were performed according to recognized consensus standards (e.g., ISO, AAMI).

8. The sample size for the training set

  • Not applicable. This is not an AI/ML device requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable. This is not an AI/ML device requiring a training set.

In summary, the provided FDA 510(k) summary focuses on demonstrating the substantial equivalence of the new endoscope and over-tube through engineering and biocompatibility testing against established standards and predicate devices. It does not contain the kind of performance data (e.g., sensitivity, specificity, AUC, or reader performance) typically associated with AI/ML-driven diagnostic devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 13, 2024

FUJIFILM Healthcare Americas Corporation Chaitrali Kulkarni Sr. Regulatory Affairs Specialist 81 Hartwell Avenue Suite 300 Lexington, Massachusetts 02421

Re: K233321

Trade/Device Name: Double Balloon Endoscope EN-840T, Over-tube TS-1214C Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDA Dated: May 14, 2024 Received: May 14, 2024

Dear Chaitrali Kulkarni:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K233321

Device Name

Double Balloon Endoscope EN-840T; Over-tube TS-1214C

Indications for Use (Describe)

Double Balloon Endoscope Model EN-840T

This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine and rectum.

Never use this product for any other purposes.

Over-Tube TS-1214C

This product is intended to be used as an accessory with the FUJIFILM Double Balloon Endoscope cleared for use with an Over-tube is used to assist with the movement of the scope inside the upper or lower digestive tract. This product is not intended for use for any neonates, infants or children.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K233321Page 1 of 3
510(k) #:K233321
510(k) Summary
Prepared on: 2024-06-12
Contact Details21 CFR 807.92(a)(1)
Applicant NameFujifilm Healthcare Americas Corporation
Applicant Address81 Hartwell ave Suite 300 Lexington MA 02421 United States
Applicant Contact Telephone704-517-4886
Applicant ContactMs. Chaitrali Kulkarni
Applicant Contact Emailchaitrali.kulkarni@fujifilm.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameDouble Balloon Endoscope EN-840T;Over-tube TS-1214C
Common NameEndoscope and accessories
Classification NameEnteroscope And Accessories
Regulation Number876.1500
Product CodeFDA
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K143556Fujifilm Double Balloon Endoscopes Models EN-530T and EN-58FDA
K221551FUJIFILM Endoscope Model El-740D/SFDS
K183032FUJIFILM Double Balloon Endoscope El-580BTFDA
K223295Endoscope Model EN-580T, Over-tube TS-1314B, Balloon BS-4FDA
Device Description Summary21 CFR 807.92(a)(4)
Technological characteristics

a. Double Balloon Endoscope Model EN-840T

Technological characteristics

The insertion portion of the device has a mechanism (hereinafter "the bends the tip from right to left and up and down, and a flexible tube (hereinafter") consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the tip is arranged inside the insertion portion for inserting the surgical instrument.

The insertion portion of the endoscopes comes into contact with the mucosal membrane.

The tip of the insertion portion is called the "Distal end" which contains the Imaging section, Balloon air feed outlet , Distal cap, Objective lens, Air/water nozzle, Water jet nozzle, Instrument channel outlet, Objective lens, and Light guide.

The bending portion is controlled by knobs on the control portion section to angulate the distal end to certain angles.

{5}------------------------------------------------

The Flexible portion refers to the long insertion area between the Control portion (a part of Non-insertion portion). This portion contains light guides), air/water channels, a forceps/suction channel, a CMOS image sensor, and cabling. The class fiber bundles alow light to travel through the body cavity, thereby providing enough light to the CMOS sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction.

The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connects the endoscopes to the light source.

b. Over-tube TS-1214C

The over-tube TS-1214C is introduced in the patient's anatomy with the pairing endoscope. TS-1214C is assembled over the outer diameter of endoscope. The endoscope and the over-tube are not advanced both at the same time, but alternatingly and successively. TS-1214C is provided sterile and single-patient use only. This accessory is a Class 2 device.

Principles of Operation

a. Double Balloon Endoscope Model EN-840T

The Double Balloon Endoscope Model EN-840T function on the same principles of operation as the predicate device. EN-840T is a gastrointestinal endoscope for observation, diagnosis, and treatment of the upper and lower digestive tract. After connecting to the light source, the devices guide the light through the fiber bundles situated inside the insertion of the device. The light that emits from the distal end of the insertion portion reflects from the target region and forms an image on through a group of object lenses placed at the distal end of the devices. An electric signal from the image sensor is transmitted to the video processor connected to the device. The video processor converts the electric signal into a video signal and displays an image on a monitor.

b. Over-tube TS-1214C

An over-tube is attached to the insertion of the endoscope, and then the scope balloon is mounted to the distal end of the endoscope using the specified setting tool.

By inflating the balloon at distal end of the balloon at the distal end of the endoscope, the distal end of the over-tube and the distal end of the endoscope are maintained in the body cavity.

By inserting an over-tube after the endoscope, it is possible to correct the hollow viscera. And then by inflating a balloon attached to the distal end of the over-tube, the distal end consequently stronger straightening of the hollow viscera can be obtained.

Intended Use/Indications for Use

Double Balloon Endoscope Model EN-840T

This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine and rectum. Never use this product for any other purposes.

Over-Tube TS-1214C

This product is intended to be used as an accessory with the FUJIFLM Double Balloon Endoscope cleared for use with an Over-tube. The Over-tube is used to assist with the movement of the upper or lower digestive tract. This product is not intended for use for any neonates, infants or children.

Indications for Use Comparison

The proposed overtube and double balloon have the same intended use as the predicate devices.

Technological Comparison

The proposed Over-tube has two side holes on the insertion portion. The holes were added to be able to discharge air trapped between

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

{6}------------------------------------------------

K233321

the Over-tube and the body cavity from the endoscope insertion inlet to the body. The addition of the two side holes does not affect the safety or efficacy of the device.

The proposed Over-tube has an additional compatible endoscope to use in combination with the over-tube. The endoscope has been tested to be compatible with the over-tube and there are no new concerns for safety or efficacy.

The balloon material on the proposed Over-tube is different than the predicate devices. Biocompatibility testing has been conducted, there are no new concerns for safety or efficacy.

The proposed endoscope has a CMOS image sensor, the predicate devices have CCD. The difference in the image sensor does not affect the safety or efficacy of the proposed endoscope.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

EMC and Electrical safety of the subject device was evaluated using following standards: ANS/AAM ES ES60601-1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 and A2:2010(R)2012 (Consolidated Text), IEC 60601-1-6:2013, and IEC 60601-1-6:2013, and IEC 60601-2-18:2009.

Biocompatibility of the subject device was evaluated using the following consensus standards: ISO 10993-52009, and ISO 10993-10:2010. Biocompatibility testing was performed in accordance with FDA's quidance, Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', issued September 4, 2020.

Endoscope specific testing was conducted using the following consensus standards: ISO 8600-3:1997, and ISO 8600-4:2014.

The proposed EN-840T is provided non-sterile and must be reprocessed prior to each use. Sterility of the other new accessories TS-13148 and BS-4 were evaluated using the following consensus standards: ISO 11135:2014, ISO 11607-1:2019, ISO 11607-2:2019, ISO 11737-1:2018, and ISO 1:2017. Sterilization was performed in accordance with FDA's guidance, Submission and Review of Sterility Information in Premarket Notification for Devices Labeled as Sterile, issued January 21, 2016.

The subject device met performance specifications in the following additional testing:

• Field of view• Bending capability• Rate of suction• Working length
• Diameter of forceps channel• Viewing direction• Resolution• LG output

The subject devices Double Balloon Endoscope EN-840T and Over-tube TS-1214C are substantially equivalent to the predicate devices based on the same intended use, indiar technological characteristics and materials. The differences in technological characteristics and materials between the subject and predicate devices rafety or effectiveness. Bench testing data demonstrated that the subject devices are substantially equivalent in performance to the difference in materials between subject and predicate devices has been validated through biocompatibility testing. Thus, Double Balloon Endoscope EN-840T and Over-tube TS-1214C the subject devices are substantially equivalent to the predicate devices, Fujifilm Double Balloon Endoscope Model EN-580T and Over-tube TS1314B (K223295).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.