K Number
K233300

Validate with FDA (Live)

Date Cleared
2023-12-26

(88 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use is for drilling, reaming, and decorticating of bone and other bone-related tissue during general surgical orthopedic procedures in a variety of surgical procedures. It is also usable in the placement of screws, wires, pins, and other fixation devices.

This attachment is designed for general surgical use where hard tissue and/or bone must be cut, reamed, drilled, and/or fixated with screws, including but not limited to, the hand, wrist, elbow, sternum, shoulder, foot, ankle, knee, and hip.

The intended use is also to facilitate the placement of the pedicle screws using the power technique (cordless). When the attachment is attached, the Stryker Power Tools provide power to rotate the screwdrivers for the insertion of pedicle screws, in non-cervical spine surgical procedures.

Device Description

The Stryker POWEReam ¼" Drive is an attachment used with a Stryker battery powered handpiece that drives accessories to achieve their intended function. The design reduces the speed and increases the torque output from the handpiece by a ratio of 5:1. The accessories are intended for drilling, reaming, and decorticating of bone and other bonerelated tissue during general orthopedic procedures in a variety of surgical procedures.

The purpose of this premarket notification is to seek clearance for a new indication for the Stryker POWEReam 1/4" Drive (part number 4405-250-000). This indication is for facilitating the placement of pedicle screws in non-cervical spine surgical procedures.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called "POWEReam 1/4" Drive" manufactured by Stryker Instruments. It is a traditional 510(k) submission, meaning it aims to demonstrate substantial equivalence to a legally marketed predicate device.

The crucial point for your request is that this device is a mechanical surgical instrument, not an AI/software-based medical device. Therefore, the concepts of acceptance criteria for AI performance, training sets, test sets, expert ground truth, MRMC studies, or standalone algorithm performance are not applicable to this submission.

The document explicitly states:

  • "Summary of Non-Clinical Testing: A suite of non-clinical testing was executed to demonstrate substantial equivalence to the predicate device. Testing included: Performance Design Verification Testing, Human Factors and Usability Engineering, Simulated Use Design Validation Testing. All pre-defined acceptance criteria for the above tests have been met." (Section 7.7)
  • "Summary of Clinical Testing: Clinical testing was not required for this traditional 510(k)." (Section 7.8)

This means the acceptance criteria and performance data are related to the mechanical and functional aspects of the device (e.g., torque, speed reduction, human factors of use), not diagnostic accuracy or AI performance on medical images.

Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria and study details for an AI-based device, as this document describes a traditional mechanical surgical instrument submission.

{0}------------------------------------------------

December 26, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size.

Stryker Instruments Marie Gildea Senior Staff Regulatory Affairs Specialist 1941 Stryker Way Portage, Michigan 49002

Re: K233300

Trade/Device Name: POWEReam 1/4" Drive Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, HWE Dated: December 7, 2023 Received: December 7, 2023

Dear Marie Gildea:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2023.12.26
09:00:55 -05'00'
for

Colin O'Neill, MBE Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K233300

Device Name

POWEReam 1/4" Drive

Indications for Use (Describe)

The intended use is for drilling, reaming, and decorticating of bone and other bone-related tissue during general surgical orthopedic procedures in a variety of surgical procedures. It is also usable in the placement of screws, wires, pins, and other fixation devices.

This attachment is designed for general surgical use where hard tissue and/or bone must be cut, reamed, drilled, and/or fixated with screws, including but not limited to, the hand, wrist, elbow, sternum, shoulder, foot, ankle, knee, and hip.

The intended use is also to facilitate the placement of the pedicle screws using the power technique (cordless). When the attachment is attached, the Stryker Power Tools provide power to rotate the screwdrivers for the insertion of pedicle screws, in non-cervical spine surgical procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

K233300 Page 1 of 1

{3}------------------------------------------------

7.1 Submitter Information

This Premarket Notification is submitted by:

Stryker Instruments 1941 Stryker Way Portage MI 49002 USA

Contact Information

Marie Gildea Senior Staff Regulatory Affairs Representative Ph: 011-353-87-9907249 Email: marie.gildea@stryker.com Date Prepared: December 2023

7.2 Device Name

Table 7.1: Subject Device Information
Subject Device Information
Trade/ Proprietary NamePOWEReam 1/4" Drive
Regulation NameThoracolumbosacral pedicle screw system
Review PanelOrthopedic
Product CodeNKB
Regulatory ClassClass II

7.3 Predicate Device

The legally marketed predicate for the subject device is detailed in Table 7.2.

Table 7.2: Predicate Device Information

Predicate Device Name510(k)Product CodeManufacturer
Stryker Xia 3 Power AdaptorK200666NKBStryker Spine

{4}------------------------------------------------

7.4 Indications for Use

The intended use is for drilling, reaming, and decorticating of bone and other bone-related tissue during general surgical orthopedic procedures in a variety of surgical procedures. It is also usable in the placement of screws, wires, pins, and other fixation devices.

This attachment is designed for general surgical use where hard tissue and/or bone must be cut. reamed, drilled, and/or fixated with screws, including but not limited to, the hand, wrist, elbow, sternum, shoulder, foot, ankle, knee, and hip.

The intended use is also to facilitate the placement of the pedicle screws using the power technique (cordless). When the attachment is attached, the Stryker Power Tools power to rotate the screwdrivers for the insertion of pedicle screws, in the non-cervical spine using powered instrumentation.

7.5 Device Description

The Stryker POWEReam ¼" Drive is an attachment used with a Stryker battery powered handpiece that drives accessories to achieve their intended function. The design reduces the speed and increases the torque output from the handpiece by a ratio of 5:1. The accessories are intended for drilling, reaming, and decorticating of bone and other bonerelated tissue during general orthopedic procedures in a variety of surgical procedures.

The purpose of this premarket notification is to seek clearance for a new indication for the Stryker POWEReam 1/4" Drive (part number 4405-250-000). This indication is for facilitating the placement of pedicle screws in non-cervical spine surgical procedures.

7.6 Comparison of Technological Characteristics

The subject device has the same intended use as the predicate device. Both are attachments used with a Stryker battery-powered handpiece that drives accessories to achieve their intended function using the power technique. Specifically, both are used to facilitate the placement of pedicle screws in non-cervical spine procedures.

The design of both the subject and predicate device are also similar. Both transmit torque and speed provided from the handpiece to associated accessories and both connect to handpieces and provide the torque to accessories under power. A difference in the design between the subject and predicate is that the predicate device attaches to a Hudson Modified Trinkle attachment that reduces the speed and increases the torque output from the handpiece at either a 3:1 or 5:1 gear ratio reduction. The subject device reduces the speed itself (from the Powered handpiece) and increases the torque output from the handpiece at a 5:1 gear ratio reduction. There are also differences in dimensions (Diameter, Speed and Mass) between the two devices.

Risk evaluation and performance testing have shown that the differences between the predicate and subject device do not introduce new issues of safety and effectiveness. The evaluation and testing have demonstrated that the subject device is substantially equivalent to the predicate device.

7.7 Summary of Non-Clinical Testing

Traditional 510(k) POWEReam 1/4" Drive

{5}------------------------------------------------

A suite of non-clinical testing was executed to demonstrate substantial equivalence to the predicate device. Testing included:

  • Performance Design Verification Testing
  • . Human Factors and Usability Engineering
  • Simulated Use Design Validation Testing o

All pre-defined acceptance criteria for the above tests have been met. Results from this testing confirm that the subject device performs as intended and supports a determination of substantial equivalence to the predicate device.

7.8 Summary of Clinical Testing

Clinical testing was not required for this traditional 510(k).

7.9 Conclusion

The Stryker POWEReam ¼" Drive is substantially equivalent to the predicate device currently cleared under K200666 as it has the same intended use, similar design, principle of operation and technological characteristics. The addition of the indication for the facilitating of the placement of pedicle screws in the non-cervical does not raise new questions of safety or effectiveness. Performance testing has demonstrated that the Stryker POWEReam ¼" Drive is substantially equivalent to the predicate device.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.