K Number
K233263

Validate with FDA (Live)

Date Cleared
2024-04-09

(193 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope.

Device Description

The Disposable Trocars includes Normal trocars (model A), Balloon trocars (model B), Thread trocars (model C) and Single-port trocars (model E). They are generally composed of cannula and obturator, with or without filter/cleaning ring/protector. The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope. If needed, the CO2 can be infused through the valve. The product is sterilized by Ethylene oxide and is a single-use product.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Disposable Trocars) seeking FDA clearance. It outlines the substantial equivalence to predicate devices based on various characteristics and non-clinical testing. However, it does not contain information about a study proving the device meets acceptance criteria related to AI/algorithm performance, human reader improvement with AI assistance, or the establishment of ground truth by expert consensus.

The document primarily focuses on bench testing and material biocompatibility for traditional medical device components, not software or AI performance. The details requested in the prompt, such as sample sizes for test sets, data provenance for AI studies, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, and training set details, are not present in this type of FDA submission for a device like a trocar.

Therefore, I cannot provide the detailed information requested in the prompt based on the given text. The text describes the testing of the physical device's performance properties (e.g., obturator compatibility, air leakage, balloon rigidity) and biocompatibility, not the performance of an AI or algorithmic component.

If the request refers to the performance criteria for the physical device itself (the trocars), then I can extract some relevant information, though it won't align perfectly with the AI/algorithm-focused questions in the prompt.


Based on the provided text, and assuming "device performance" refers to the physical attributes and functionalities of the Disposable Trocars, here's what can be extracted, acknowledging that many of your specific questions are not addressed by this type of document:

The Disposable Trocars are physical medical instruments used in endoscopic procedures. The "acceptance criteria" and "device performance" discussed in this document relate to their physical and biological properties, not AI or algorithmic performance.

1. A table of acceptance criteria and the reported device performance:

The document states that "All the test results were 'PASS'." This implies that the device met pre-defined acceptance criteria for each test. However, the specific quantitative acceptance criteria values are not explicitly listed, only the type of tests performed.

Acceptance Criterion (Test Type)Reported Device Performance
AppearancePASS
Obturator CompatibilityPASS
Insertion & Cannula StabilityPASS
Air LeakagePASS
Balloon Rigidity (for relevant models)PASS
Fixation Device Retention (for relevant models)PASS
Cytotoxicity (ISO 10993-5:2009)Passed (Biocompatible)
Sensitization (ISO 10993-10:2021)Passed (Biocompatible)
Irritation (ISO 10993-23:2021)Passed (Biocompatible)
Acute Systemic toxicity (ISO 10993-11:2017)Passed (Biocompatible)
Pyrogenicity (USP<151>)Passed (Biocompatible)
EO sterilization validation (EN ISO 11135-1, 11737-1,11737-2)Validated
Transportation test (ASTM D4169)Validated
Packaging validation (ISO 11607-1/-2)Validated
Shelf life (3 years determined by stability study/ageing test)Validated
Bacterial Endotoxin Testing (USP-NF:2023 <85>)Passed

2. Sample sized used for the test set and the data provenance:

  • Sample Size: Not specified in the provided text. The document mentions "performance studies and bench testing" but does not give the number of units tested.
  • Data Provenance: Not applicable in the context of clinical data for AI. The tests are bench tests performed by the manufacturer, Hangzhou Kangji Medical Instrument Co., Ltd., which is based in China. The data would be prospective in the sense that the company specifically conducted these tests for the submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This document describes testing of a physical medical device. Ground truth as understood in AI/image analysis (e.g., expert radiologist consensus) is not relevant here. The ground truth for these tests would be established by validated test methods and measurement standards for physical properties. The tests were evaluated by their "Design Engineer."

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. This is for AI/image analysis. The "adjudication" for these physical tests would be the measurement against a pre-defined specification or standard.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is a physical medical device, not an AI or software device that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. This is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For biocompatibility: ISO and USP standards (e.g., cell cultures for cytotoxicity, animal models for systemic toxicity).
  • For performance: Engineering specifications and test methods for physical properties (e.g., force measurements for insertion, pressure measurements for air leakage).

8. The sample size for the training set:

  • Not applicable. This is a physical medical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 9, 2024

Hangzhou Kangji Medical Instrument Co., Ltd. % Esther Zhang Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, Shanghai 201102 China

Re: K233263

Trade/Device Name: Disposable Trocars Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 28, 2023 Received: September 29, 2023

Dear Esther Zhang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Mark Trumbore -S Date: 2024.04.09 12:02:16 -04'00 Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K233263

Device Name

Disposable Trocars

Indications for Use (Describe)

The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) summary K233263

| Submitter

Device submitter: Hangzhou Kanqji Medical Instrument Co., Ltd.

No. 1668 Chuniiang East Road. Economic Development Zone. Tongly. Hangzhou, 311501, China.

Contact person: Martin Sun

Manager of Regulatory Affairs Phone: +86-0571-69901712 Fax: +86-0571-69901712 Email: martin.sun@kanqji.com

Date: April 9, 2024

II Device

Trade Name of Device: Disposable Trocars Common Name: Trocars Regulation Number: 21 CFR 876.1500 Regulation Name: Laparoscope, General & Plastic Surgery Requlatory Class: II Product code: GCJ Review Panel: General and Plastic Surgery

III Correspondent

Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District, Shanghai, P.R.China Contact: Esther ZHANG Email: Esther.zhang@llins-tech.com

IV Device description

The Disposable Trocars includes Normal trocars (model A), Balloon trocars (model B), Thread trocars (model C) and Single-port trocars (model E). They are generally composed of cannula and obturator, with or without filter/cleaning ring/protector.

The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope. If needed, the CO2 can be infused through the valve. The product is sterilized by Ethylene oxide and is a single-use product.

Normal trocars have 3 different cannulas (normal cannula, normal cannula with filter, normal cannula with cleaning ring) and 5 different obturators (Ingot-type, Optical obturator A. Optical obturator B. Auto-shielded obturator, Blunt obturator). Normal cannula consists of Cannula Body, Cannula Tube Seat, Choke Valve, Sheath Cap and Stopcock. It may be equipped with a filter or cleaning ring as an accessory. Cleaning ring is equipped inside the cannula and used for removing blood and foreign matter from the laparoscope. Filter is detachable, installed in stopcock, and used for filtering out any

{4}------------------------------------------------

foreign matter that spills out of the body during surgery. Stopcock can be used for insufflation if needed. Ingot-type obturator is easy to insertion which consists of Needle Tip, Needle Tube and Needle Seat. Optical obturator A has a transparent needle tip while Optical obturator B has a transparent needle tip and needle tube. Needle tube of these two optical obturators is hollow which allows laparoscope go inside to see if it has passed the abdominal wall. Auto-shielded obturator has a hided blade which only showing up in puncturing process. Blunt obturator has a blunt needle tip which can protect internal organ during the procedure. Normal trocars are available from 3-15mm of diameter and 65-160mm of length.

Balloon Trocars are equipped an inflatable balloon with a fixed (Balloon cannula A) or movable holder (Balloon cannula B) to provide stabilization during a laparoscopic procedure. The Cannula consists of Balloon, Cannula body, Holder A, One-way vale, Cannula Tube Seat, Stopcock, Sheath Cap and Choke Valve. Stopcock can be used for insufflation if needed. Movable holder can be slid down to adjust its height at a desired position for stabilization. There are five different obturators as same as normal trocars: lngot-type, Optical obturator A, Optical obturator B, Auto-shielded obturator, Blunt obturator. Balloon Trocars are available in from 3-15mm of diameter and 55-140mm of length.

Thread trocars consists of a thread obturator and a Thread cannula A (thoracic trocars) or B (Orthopedics trocars). Threaded cannula enhances stability when exchanging instruments. Thread cannula B consists of Cannula body, Cannula Tube Seat, Stopcock, Sheath Cap and Choke Valve while Thread cannula A does not have a stopcock. Stopcock can be used for insufflation if needed. Thoracic trocars are available in from 5-20mm of diameter and 65-85mm of length. Orthopedics trocars are designed for Orthopedics laparoscopic surgery and available in from 5-20mm of diameter and 65-120mm of length.

Single-port Trocars are designed to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive endoscopic surgery. It consists of cannula, universal seal system, a variable height protector or fixed height protector, with or without obturator. Single port cannula A and B use same materials and only different in shapes. Universal seal system A supports the free combination and arrangement of trocar with different numbers and sizes to meet different surgical needs. Obturator is used with Single port cannula A. Universal seal system B is fully transparent and visible for surgical operation. Besides, soft material makes it good for handling the endoscopic instruments and reducing surgical compression congestion to the patient skin. Universal seal system C has a larger handling space than system B. Protector, which has fixed height or variable height, is used for protect surgical incision.

Due to the different structure and function, diameter, length of the Cannula and Obturator, and package amount, there are many combinations for final product.

V Indications for use

The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope.

{5}------------------------------------------------

VI Substantial Equivalence

Model ofproposeddevicePredicate deviceManufacturer510(k) No.
K112358Unimax Trocar SystemUnimax MedicalSystems Inc.GCJ
K211577SoftFix™ Balloon TrocarUnimax MedicalSystems Inc.GCJ
K141715Glove PortNELISOTJ
K093372SILS™ PortCOVIDIEN LPGCJ
K073719ASC TriPort LaparoscopicAccess DeviceAdvanced SurgicalConcepts.GCJ

able Trocars are substantially equivalent to the predicate devices: osed Dien

Comparison of technological characteristics with the predicate devices

{6}------------------------------------------------

DevicefeatureSubject Device(Model A/B/C)Predicate Device K112358(Unimax Trocar System)Predicate DeviceK211577( SoftFixTMBalloon Trocar)Comment
ClassIIIIIIEquivalent
Product codeGCJGCJGCJEquivalent
Regulationnumber21 CFR 876.150021 CFR 876.150021 CFR 876.1500Equivalent
Intended useThe Disposable Trocarshave application in a varietyof endoscopic procedures toprovide a port of entry forendoscopic instrumentsand/or endoscope.The Unimax Trocar System, Model:Auto-Locking Trocar, BladelessTrocar, Visible Trocar, HassonTrocar, Dilating Trocar SecondaryTrocar, and Thoracic Trocar hasapplication in a variety of endoscopicprocedures to provide a port ofentry for endoscopic instruments.The SoftFixTM BalloonTrocar have application in avariety of endoscopicprocedures to provide a portof entry for endoscopicinstruments into the bodycavity of patients.Equivalent.
ModelsNormal trocars (Model A)Bladeless, Visible and auto-locking,Secondary trocars/DifferentComment
Balloon trocars (Model B)Hasson trocarsBalloon trocar (bladedobturator with safety lock forthe shield, bladelessobturator, optical obturator,hasson obturator, cannulasleeve with a stopcock forinsufflation)1
Thread trocars (Model C)Thoracic trocars/
Dilating trocars
Device structureCannula with or withoutcleaning ring/filter, ballooncannula, cannula withstopcock for insufflation,obturatorCannula with stopcock for insufflation,obturatorCannula with a stopcock forinsufflation, obturator
SpecificationDiameter: 3-15mmLength: 60-160mmDiameter: 3-15mmLength: 65-150mmDiameter: 5-12mmLength: 70-100mmSimilarComment 2
Reuse durabilitySingle useSingle useSingle useEquivalent
SterilizationEOEOEOEquivalent
Single useYesYesYesEquivalent
PerformanceObturator Compatibility;Insertion & Cannula Stability;Air Leakage; balloon rigidity;fixation device retentionObturator Compatibility;Insertion&Cannula Stability;Air LeakageObturator Compatibility;Insertion & Cannula Stability;AirLeakage; balloon rigidity;fixation device retentionEquivalent
BiocompatibilityISO 10993-1, ISO 10993-5,ISO 10993-10, ISO 10993-7, ISO 10993-12ISO 10993-1, ISO 10993-5, ISO10993-10, ISO 10993-7, ISO 10993-12ISO 10993-1, ISO 10993-5,ISO 10993-10, ISO 10993-7, ISO 10993-12Equivalent

Table 1 Substantial equivalence discussion- Model A/B/C

{7}------------------------------------------------

{8}------------------------------------------------

Discussion:

Comment 1

The subject device and the predicate device have intended use, to applicate in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. The obturator and cannula have miner difference on needle tip and cannula accessories (e.g. cleaning, filter), while these differences have been validated by performance test and do not influence use. This difference does not affect the clinical safety of the subject device.

Comment 2

The Disposable Trocars are available in a range of Length is only to allow entry of other endoscopic instruments and does not have any effect on device performance. Additionally, length information is printed on the Obturator. Choice of Trocar specification depends on user's preference and clinical need. The differences on length do not raise new and effectiveness.

Conclusion:

The Disposable Trocars (Model A/B/C) have the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the subject device and predicate devices do not alter suitability of the proposed device for its intended use.

{9}------------------------------------------------

Device featureSubject Device(Model E Single-portTrocars)PredicateDeviceK141715(Glove Port)PredicateDeviceK093372(SILS™Port)PredicateDeviceK073719(ASC TriPortLaparoscopicAccessDevice)Comment
ClassIIIIIIIIEquivalent
Regulation number21 CFR 876.150021 CFR 876.150021 CFR 876.150021 CFR 876.1500Equivalent
Intended useThe Disposable Trocars haveapplication in a variety ofendoscopic procedures toprovide a port of entry forendoscopic instrumentsand/orendoscope.Single-port Trocars aredesigned to provide access formultiple instruments and/orendoscope to the abdominalcavity through a single incisionduring minimally invasiveendoscopic surgery.The Glove Port isintended to provideaccess for multipleinstruments and/orendoscope to theabdominal cavitythrough a singleincision duringminimally invasivelaparoscopicsurgery.The SILS™ Port isindicated formultiple instrument orcamera access to theabdominal cavitythrough a singleincisionfor performingminimallyinvasive laparoscopicprocedures.The ASC TriPortLaparoscopic AccessDevice is intendedforuse as a multipleinstrument and/orcamera port duringminimally invasiveabdominallaparoscopic surgery.Equivalent

Table 2 Substantial equivalence discussion-Single-port trocars (Model E)

{10}------------------------------------------------

ModelE01, E02, E03, E04, E05, E06, E07, E08Glove Port Original; Glove Port H; Glove Port A; Glove Port ATSILS™ PortASCLaparoscopic AccessDeviceTriPort Similar Comment 1
Port number3~43~43~43
Absolute size10~2510~25mm10~25mm25mm
MaterialPolycarbonate, siliconePolyurethaneSiliconeSiliconeDifferentComment2
Installation andabdominal wallinner fixationInner ring of the protectorbeing fixed inside theabdominal wall inside.Wound retractorring being fixedinside theabdominal wallinside.Distal ring being fixed inthe abdominal wallinside.Non-fixedEquivalent
SterilizationEO gas sterilizationEO gas sterilizationGamma sterilizationGamma sterilizationDifferentComment3

{11}------------------------------------------------

Discussion:

Comment 1

The subject device and the predicate devices are endoscopic instrument access ports that are used to perform the same function as a trocar. They are all retracts a small abdominal incision to allow endoscopic instruments to pass through to abdomen, and maintains pneumoperitoneum in the abdomen during the surgical procedure, whether or not endoscopic instruments are passing through the port. The difference in models, port number and size will not influence their intended use. Besides, surgeons will select port number and size based on actual needs during procedure. Furthermore, these differences have been addressed by performance test and do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

Comment 2

The materials of the subject device and the predicate devices are partially different. However, the materials have been tested according to ISO 10993 series standards and performance test. The results showed they're biocompatible with human body and no impact of air leakage, connection firmness, Cannula stabilization and other functions. This difference does not affect the effectiveness and safety.

Comment 3

The sterilization method of the subject device and the predicate devices are partially different. However, the subject device was ensured sterility by sterilization validation. Therefore, the differences on sterilization do not raise new questions about safety and effectiveness.

Conclusion:

The Disposable Trocars (Model E) have the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the subject device and predicate devices do not alter suitability of the proposed device for its intended use.

Comparison Summary:

The subject and predicate devices have the same intended use and application. The subject and predicate device designs are similar. They are single-use devices. The differences in technological characteristics between the subject and predicate devices (i.e., needle tip, different size, port number) do not raise different questions of safety and effectiveness.

VII Summary of non-clinical testing Performance testing

{12}------------------------------------------------

The Disposable Trocars has been evaluated by our Design Engineer and through performance studies and bench testing, as attached in Appendix C. Testing encompassed appearance, Obturator Compatibility, Insertion & Cannula Stability, Air Leakage, balloon rigidity, fixation device retention, etc. All the test results were "PASS". The performance of Disposable Trocars meets the technical standard requirements of Hangzhou Kangji Medical Instrument Co., Ltd. as compared to the predicate.

Biocompatibility testing

Biocompatibility of the Disposable Veress needles was evaluated in accordance with ISO 10993-1:2018 for the body contact category. The following tests were performed, as recommended:

CytotoxicityISO 10993-5:2009
SensitizationISO 10993-10:2021
IrritationISO 10993-23:2021
Acute Systemic toxicityISO 10993-11:2017
PyrogenicityUSP<151>

Sterilization and shelf life testing

  • EO sterilization validation per EN ISO 11135-1, 11737-1,11737-2 A
  • A Transportation test per ASTM D4169
  • A Packaging validation per ISO 11607-1/-2
  • The 3 years shelf life of the device is determined based on stability study which A includes ageing test.
  • Bacterial Endotoxin Testing per USP-NF:2023 <85>

VIII Conclusion

The Disposable Trocars is substantially equivalent to its predicate devices as listed in VI Substantial Equivalence. The differences between the predicate and subject device do not raise any new or different questions of safety or effectiveness. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.