K Number
K233240

Validate with FDA (Live)

Device Name
GERDX-System
Manufacturer
Date Cleared
2024-06-21

(267 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GERDX®-System is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease (GERD) in patients who require and respond to pharmacological therapy.

Device Description

The GERDX-System is a sterile application and suture system indicated for the treatment of symptomatic chronic gastroesophageal reflux disease (GERD) in patients who require and respond to pharmacological therapy.

The GERDX-System consist of following procedural components: GERDX-Applicator, GERDX-Tissue Retractor, GERDX-Suture System and the GERDX-Guide Wire.

The Suture System and the Tissue Retractor are loaded onto the Applicator. The Guide Wire is placed within the antrum of the stomach by standard procedure. By use of the Guide Wire, the Applicator is introduced orally and advanced through the esophagus into the stomach. After removing the Guide Wire a thin profile gastroscope is inserted into a channel within the Applicator and the Applicator is positioned under direct endoscopic visualization at the target location within the gastrointestinal (GI) tract. The Tissue Retractor is advanced out of the distal tip of the Applicator. Gastric tissue is engaged by the Tissue Retractor and withdrawn into the Applicator, thereby pulling the tissue into the open arms of the Applicator. The arms of the Applicator are then closed to grasp and approximate the target tissue. The Suture System is then deployed to transmurally affix the GI tissue. The procedure is repeated to deploy additional Suture Systems if necessary. The number of Suture Systems that are placed will vary based on patient anatomy and physician's judgement. Parts of the Suture System remain within the patient.

AI/ML Overview

The provided text is a 510(k) summary for the GERDX-System, an endoscopic suture/plication system for treating GERD. Medical devices in this category (Class II) are reviewed for substantial equivalence to existing predicate devices, rather than requiring extensive clinical trials for approval. Thus, the information available is focused on demonstrating equivalence rather than a detailed efficacy study with defined acceptance criteria for device performance in the same way a novel drug or high-risk medical device might require.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative "acceptance criteria" and "reported device performance" in the typical sense of a clinical trial for efficacy (e.g., target reduction in symtom score or a specific diagnostic sensitivity/specificity). Instead, the studies aim to demonstrate equivalence in performance and safety to a predicate device.

Acceptance Criteria (Implied by Equivalence)Reported Device Performance (Summary of Findings)
Bench Tests: Equivalent technological characteristics, designs, and operational properties to the predicate device (Plicator EPS) for full-thickness plication forces.Side-by-side bench tests comparing the GERDX-System and Plicator EPS confirmed equivalent technological characteristics, designs, and operational properties, supporting performance and safety.
Biocompatibility: Meets acceptance criteria according to ISO 10993-1:2018.Biocompatibility of GERDX-System met the acceptance criteria according to ISO 10993-1:2018.
Clinical Efficacy/Safety: Demonstrated effectiveness as a minimally invasive endoscopic intervention to control GERD, comparable to predicate.Two peer-reviewed articles published where GERDX-System was proven as an effective minimally invasive endoscopic intervention to control GERD.
Animal Testing (Safety/Feasibility): Demonstrated safety and feasibility in a GLP-conform laboratory setting.Animal studies on a porcine model in a GLP-conform laboratory demonstrated the safety and feasibility of GERDX-System.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Bench Tests: The specific sample size for the side-by-side bench tests is not mentioned. Data provenance is not specified but would typically be from the manufacturer's internal testing.
  • Clinical Testing: The document refers to "Two peer-reviewed articles have been published." The details of these studies, including sample size, country of origin, and whether they were retrospective or prospective, are not available in this summary.
  • Animal Testing: The sample size for the porcine model studies is not mentioned. Data provenance would be from a GLP-conform laboratory.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not explicitly provided in the 510(k) summary. For a device like this, ground truth would likely refer to clinical outcomes (symptom reduction, physiological measures of reflux) as assessed by clinicians or researchers, rather than an "expert read" of images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided. It's improbable that an adjudication method like 2+1 or 3+1 (typically used for image interpretation in AI studies) would be relevant for the type of clinical or bench tests described for this device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC study was not done. This device is not an AI-powered diagnostic or assistive tool, but rather a surgical device (endoscopic suture/plication system). Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Bench Tests: The ground truth for bench tests would be defined by engineering standards and measurements of physical forces and properties.
  • Biocompatibility: Ground truth is established by adherence to ISO 10993-1:2018 standards and associated laboratory testing.
  • Clinical Testing: The "effectiveness" mentioned in the peer-reviewed articles would likely be based on patient-reported outcomes (e.g., symptom scores, quality of life), physiological measurements (e.g., pH monitoring), or endoscopic findings, representing "outcomes data."
  • Animal Testing: Ground truth for safety and feasibility in animal studies would be based on direct observation, histological analysis, and other relevant physiological measurements in the porcine model.

8. The sample size for the training set

This device does not involve a "training set" in the context of machine learning or AI models. It is a physical medical device.

9. How the ground truth for the training set was established

As there is no "training set" for this device, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 21, 2024

G-Surg GmbH % Anna Pocsai Consultant novineon CRO GmbH Friedrich-Miescher-StraBe 9 Tuebingen, Baden-Wuerttemberg 72076 Germany

Re: K233240

Trade/Device Name: GERDX-System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: ODE Dated: May 24, 2024 Received: May 24, 2024

Dear Anna Pocsai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K233240

Device Name

GERDX-System

Indications for Use (Describe)

Indications for Use (Describe) The GERDX®-System is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease (GERD) in patients who require and respond to pharmacological therapy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Table of Content of Volume 5: Premarket Notification 510(k) Summary

TABLE OF CONTENT OF VOLUME 5: PREMARKET NOTIFICATION 510(K) SUMMARY5-1
5. PREMARKET NOTIFICATION 510(K) SUMMARY5-1
1. CONTACT DETAILS5-1
2. DEVICE NAME5-1
3. LEGALLY MARKETED PREDICATE DEVICES5-1
4. DEVICE DESCRIPTION SUMMARY5-1
5. INTENDED USE / INDICATIONS FOR USE:5-2
6. COMPARISON OF CHARACTERISTICS5-2
6.1 INDICATIONS FOR USE COMPARISON5-2
6.2 TECHNOLOGICAL COMPARISON5-2
7. NON-CLINICAL AND CLINICAL TESTS SUMMARY & CONCLUSION5-3

5. Premarket Notification 510(k) Summary

1. CONTACT DETAILS

G-SURG GmbH Weinbergstraße 28, 83370 Seeon-Seebruck Telephone: +49-8624 409160 E-Mail: j.mair@g-surg.com Contact: Julian Mair Date prepared: June 17, 2024

2. DEVICE NAME

Trade Name: GERDX-System Common Name: Endoscope accessory Classification Name: Endoscopic Suture/Plication System, Gastroesophageal Reflux Disease (Gerd), Class: II Regulation Number: 876.1500 Product Code: ODE

3. LEGALLY MARKETED PREDICATE DEVICES

Endoscopic Plication System; Plicator™ (K071553); Product Code: ODE

4. DEVICE DESCRIPTION SUMMARY

{5}------------------------------------------------

The GERDX-System is a sterile application and suture system indicated for the treatment of symptomatic chronic gastroesophageal reflux disease (GERD) in patients who require and respond to pharmacological therapy.

The GERDX-System consist of following procedural components: GERDX-Applicator, GERDX-Tissue Retractor, GERDX-Suture System and the GERDX-Guide Wire.

The Suture System and the Tissue Retractor are loaded onto the Applicator. The Guide Wire is placed within the antrum of the stomach by standard procedure. By use of the Guide Wire, the Applicator is introduced orally and advanced through the esophagus into the stomach. After removing the Guide Wire a thin profile gastroscope is inserted into a channel within the Applicator and the Applicator is positioned under direct endoscopic visualization at the target location within the gastrointestinal (GI) tract. The Tissue Retractor is advanced out of the distal tip of the Applicator. Gastric tissue is engaged by the Tissue Retractor and withdrawn into the Applicator, thereby pulling the tissue into the open arms of the Applicator. The arms of the Applicator are then closed to grasp and approximate the target tissue. The Suture System is then deployed to transmurally affix the GI tissue. The procedure is repeated to deploy additional Suture Systems if necessary. The number of Suture Systems that are placed will vary based on patient anatomy and physician's judgement. Parts of the Suture System remain within the patient.

5. INTENDED USE / INDICATIONS FOR USE:

The GERDX®-System is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease (GERD) in patients who require and respond to pharmacological therapy.

6. COMPARISON OF CHARACTERISTICS

6.1 INDICATIONS FOR USE COMPARISON

With the exception of the device name, the indications for use statements are identical.

6.2 TECHNOLOGICAL COMPARISON

The proposed GERDX-System is clinically, biologically, and technologically identical of the proposed predicate device (K071553). The devices differ only in some technological details. Summary of differences:

  • GERDX-Applicator: By the predicate device force is transmitted manually via cables and . pulls, whereas for the GERDX-System force is transmitted via a manually operated hydraulic system. However, this modification shows several improvements in safety-relevant aspects.
  • . GERDX-System: All components of GERDX-System are sterile, while some components of predicate device are non-sterile and reusable. There are no labelling regarding of cleaning and disinfection is present on the label of the GERDX-System
  • . GERDX-System contains an additional component, a Guide Wire, which had to be purchased separately by the predicate device. The component Guide Wire is identical for the two devices.
  • . GERDX-Suture: same design, except of the material of the surgical suture used. However, the suture for the GERDX-System is cleared by FDA and its biocompatibility has been confirmed.

{6}------------------------------------------------

7. NON-CLINICAL AND CLINICAL TESTS SUMMARY & CONCLUSION

Bench tests:

Procedure-related bench tests were provided to assess the forces associated with a full thickness plication. Side by side tests were performed with GERDX-System and with the proposed predicate. The results of these tests confirm that the GERDX-System and Plicator EPS have equivalent technological characteristics, designs, and operational properties and support the performance and safety of the GERDX-System.

Biocompatibility of GERDX-System met the acceptance criteria according to ISO 10993-1:2018.

Clinical testing

Two peer-reviewed articles have been published whereas GERDX-System was proved as an effective minimally invasive endoscopic intervention to control GERD.

Animal testing

Animal studies performed on a porcine model in a GLP-conform laboratory demonstrated the safety and feasibility of GERDX-System.

Based on indications for use, clinical evaluation, performance- and biocompatibility tests presented in this premarket notification, the GERDX-System is substantially equivalent to the Plicator and do not raise new issues of safety and effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.