K Number
K233233

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-10-11

(13 days)

Product Code
Regulation Number
888.3353
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EMPHASYS Femoral Stems are intended for use in total and partial hip arthroplasty. The stems are intended only for uncemented use.

Total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Partial hip arthroplasty (hip hemiarthroplasty) is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of a satisfactory natural acetabulum and sufficient femoral bone to seat and support the femoral stem.

EMPHASYS Femoral Stems are intended for single use only.

Total hip replacement is indicated in the following conditions:

  1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
  2. Avascular necrosis of the femoral head.
  3. Acute traumatic fracture of the femoral head or neck.
  4. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
  5. Certain cases of ankylosis.

Partial hip replacement is indicated in the following conditions:

  1. Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated with internal fixation.
  2. Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.
  3. Avascular necrosis of the femoral head.
  4. Non-union of femoral neck fractures.
  5. Certain high subcapital and femoral neck fractures in the elderly.
  6. Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.
  7. Pathology involving only the femoral head/neck and/or proximal femur that can be adequately treated by hip hemiarthroplasty.

The EMPHASYS Femoral Stems are indicated for cementless use only.

Device Description

The EMPHASYS Femoral Stems include HA-coated femoral stems in standard and high offsets and in collared and collarless configurations.

The stems are manufactured from titanium alloy forgings per ASTM F620-20 and coated with a layer of plasma-sprayed HA per ASTM F1185-14 and ISO 13779-6: 2015.

AI/ML Overview

The provided FDA 510(k) summary (K233233) describes an application for an extension of the shelf life of the EMPHASYS Femoral Stems from 5 years to 10 years, not a new device or an AI/software-based device.

Therefore, the requested information regarding acceptance criteria for device performance, sample size for test sets, expert ground truth establishment, MRMC studies, standalone algorithm performance, and training set details are not applicable to this submission. This document focuses on the equivalence of properties related to an extended shelf life.

However, I can extract and present the acceptance criteria and reported device performance related to the shelf-life extension from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Related to Shelf-Life Extension)Reported Device Performance (for 10 years shelf life)
Maintenance of HA coating properties (Crystallinity, Foreign Phases, CaP ratio) as per ISO-13779-3 for aged products.HA coating properties on aged products were tested and confirmed to maintain specifications.
Maintenance of HA coating thickness as per ASTM F1854 for aged products.HA coating thickness on aged products was tested and confirmed to maintain specifications.
Maintenance of HA coating porosity as per ASTM E2109 for aged products.HA coating porosity on aged products was tested and confirmed to maintain specifications.
Packaging integrity and sterility validation for 10 years.Packaging shelf-life validation has been carried out to confirm a shelf-life of 10 years.

The submission explicitly states:

  • No clinical tests were conducted to demonstrate substantial equivalence for this shelf-life extension.
  • The shelf life extension will not impact the fit, form, or function of these components as they are manufactured to the same specifications and utilize the same materials as the predicate devices.
  • The nonclinical tests reviewed as part of K211657 and K203167 (previous clearances for the device and a predicate) remain current.

Given that this is a 510(k) for a shelf-life extension of an existing medical device, rather than the initial clearance of a device that relies on complex performance metrics or AI, the detailed performance study requirements (like those for AI/software devices) are not relevant here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

October 11, 2023

DePuy Ireland UC Sarah Matamisa Regulatory Affairs Specialist Loughbeg, Ringaskiddy Co. Cork, Ireland

Re: K233233

Trade/Device Name: EMPHASYS Femoral Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, KWL, KWL, KWY, MEH Dated: September 27, 2023 Received: September 28, 2023

Dear Sarah Matamisa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K233233

Device Name

EMPHASYS Femoral Stems

Indications for Use (Describe)

EMPHASYS Femoral Stems are intended for use in total and partial hip arthroplasty. The stems are intended only for uncemented use.

Total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Partial hip arthroplasty (hip hemiarthroplasty) is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of a satisfactory natural acetabulum and sufficient femoral bone to seat and support the femoral stem.

EMPHASYS Femoral Stems are intended for single use only.

INDICATIONS

Total hip replacement is indicated in the following conditions:

  1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
    1. Avascular necrosis of the femoral head.
    1. Acute traumatic fracture of the femoral head or neck.
    1. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis,
  • hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
    1. Certain cases of ankylosis.

Partial hip replacement is indicated in the following conditions:

  1. Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated with internal fixation.

  2. Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.

    1. Avascular necrosis of the femoral head.
    1. Non-union of femoral neck fractures.
    1. Certain high subcapital and femoral neck fractures in the elderly.
  1. Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.

  2. Pathology involving only the femoral head/neck and/or proximal femur that can be adequately treated by hip hemiarthroplasty.

The EMPHASYS Femoral Stems are indicated for cementless use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) SUMMARY

(As required by 21 CFR 807.92 and 21 CFR 807.93)

Submitter Information
NameDePuy Ireland UC
AddressLoughbeg, RingaskiddyCo. Cork, IRELAND
Establishment RegistrationNumber3015516266
Name of contact personSarah Matamisa
e-mail addresssmatamis@its.jnj.com and DePuySynthesJointsRegulatoryAffairs@its.jnj.com
Alternative contact personElaine Pears
e-mail addressepears@its.jnj.com
Work mobile+44 7876 217532
Date prepared11 October 2023
Name of device
Trade or proprietary nameEMPHASYS Femoral Stem
Common or usual nameTotal or Hemi-Hip Arthroplasty Prosthesis
Classification nameProsthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, UncementedProsthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium PhosphateProsthesis, Hip, Hemi-, Femoral, MetalProsthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
ClassII
Classification panel87 Orthopedics
Regulation21 CFR 888.335321 CFR 888.336021 CFR 888.3390
Product Code(s)LZO, MEH, KWL, KWY
Legally marketed device(s) towhich equivalence is claimedPrimary Predicate:- EMPHASYS Femoral Stems (K211657), cleared August 5, 2021Secondary Predicate:- DePuy CORAIL AMT Hip Prosthesis (K203167), cleared November 19, 2020
Reason for 510(k) submissionThe purpose of this submission is to extend the currently approved shelf life of EMPHASYS Femoral Stems (K211657) from 5 years to 10 years.
Device descriptionThe EMPHASYS Femoral Stems include HA-coated femoral stems in standard and high offsets and in collared and collarless configurations.
The stems are manufactured from titanium alloy forgings per ASTM F620-20 andcoated with a layer of plasma-sprayed HA per ASTM F1185-14 and ISO13779-6: 2015.
Intended use of the deviceTotal and Partial Hip Arthroplasty
Indications for useEMPHASYS Femoral Stems are intended for use in total and partial hip arthroplasty.The stems are intended only for uncemented use.Total hip arthroplasty is intended to provide increased patient mobility and reduce painby replacing the damaged hip joint articulation in patients where there is evidence ofsufficient sound bone to seat and support the components.Partial hip arthroplasty (hip hemiarthroplasty) is intended to provide increased patientmobility and reduce pain by replacing the damaged hip joint articulation in patientswhere there is evidence of a satisfactory natural acetabulum and sufficient femoralbone to seat and support the femoral stem.EMPHASYS Femoral Stems are intended for single use only.Total hip replacement is indicated in the following conditions:1. A severely painful and/or disabled joint from osteoarthritis, traumaticarthritis, rheumatoid arthritis, or congenital hip dysplasia.2. Avascular necrosis of the femoral head.3. Acute traumatic fracture of the femoral head or neck.4. Failed previous hip surgery including joint reconstruction, internal fixation,arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hipreplacement.5. Certain cases of ankylosis.Partial hip replacement is indicated in the following conditions:1. Acute fracture of the femoral head or neck that cannot be appropriatelyreduced and treated with internal fixation.2. Fracture dislocation of the hip that cannot be appropriately reduced andtreated with internal fixation.3. Avascular necrosis of the femoral head.4. Non-union of femoral neck fractures.5. Certain high subcapital and femoral neck fractures in the elderly.6. Degenerative arthritis involving only the femoral head in which theacetabulum does not require replacement.7. Pathology involving only the femoral head/neck and/or proximal femur thatcan be adequately treated by hip hemiarthroplasty.
The EMPHASYS Femoral Stems are indicated for cementless use only.

{6}------------------------------------------------

K233233 Page 2 of 4

{7}------------------------------------------------

SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THEPREDICATE DEVICE
CharacteristicsSubject Device:EMPHASYS Femoral StemsPredicate Device:EMPHASYS Femoral Stems(K211657)Predicate Device:DePuy CORAIL AMT Prosthesis(K203167)
Intended UseTotal and Hemi-Hip ArthroplastyTotal and Hemi-Hip ArthroplastyTotal and Hemi-Hip Arthroplasty
MaterialTitanium Alloy (Ti6Al4V) withHA coatingTitanium Alloy (Ti6Al4V) withHA coatingTitanium Alloy (Ti6Al4V) HAcoating
FixationUncementedUncementedUncemented
Sterile MethodGammaGammaGamma
PackagingNylon inner and outer pouchNylon inner and outer pouchNylon Inner Pouch andouter PETG blisterwith Tyvek peel lid
Shelf life10 Years5 Years10 Years

{8}------------------------------------------------

PERFORMANCE DATA

SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The subject HA coated EMPHASYS Femoral Stems are manufactured to the same production specifications as the predicate devices (K211657 and K203167). As the subject devices have been validated to be manufactured to the same specifications and utilize the same materials, the shelf life extension will not impact the fit, form or function of these components. Therefore, the nonclinical tests reviewed as part of K211657 and K203167 remain current.

Packaging shelf-life validation has been carried out to confirm a shelf-life of 10 years.

HA coating properties on aged products were tested as per ISO-13779-3 (Crystallinity, Foreign Phases and CaP ratio), ASTM F1854 (Thickness) and ASTM E2109 (Porosity) to confirm a shelf-life of 10 years.

SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION

No clinical tests were conducted to demonstrate substantial equivalence.

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

The subject device EMPHASYS Femoral Stems is equivalent to the predicates EMPHASYS Femoral Stems (K211657) and DePuy CORAIL AMT Prosthesis (K203167). There have been no modifications made to the intended use, design, biocompatibility, materials, sterilization or packaging of the existing EMPHASYS Femoral Stems.

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.