K Number
K233155

Validate with FDA (Live)

Date Cleared
2024-01-12

(107 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClearPointer is intended to be used in conjunction with the SmartFrame OR System and a compatible stereotactic optical navigation system for patient registration and navigation.

Device Description

The ClearPointer Optical Navigation Wand is an optical tracker intended to be used with the SmartFrame OR Stereotactic System and ClearPoint Bone Screw Fiducials, both of which are subject of pending 510(k) submission. The ClearPointer is used in conjunction with a Medtronic StealthStation Navigation System to align the SmartFrame OR tower to the desired target. It is composed of 4 optical spheres and utilizes a sphere positional geometry that is recognizable by the Medtronic StealthStation S8 Optical navigation System. It is equipped with a Pointer attachment to allow for image registration and entry point location. The Pointer Attachment can also be removed from the Optical Tracker component, so that the Tracker can be mounted to a SmartFrame OR. The SmartFrame OR Hardware Kit is provided with the following components:

  • 1x ClearPointer™ ●
  • 1x Pointer Attachment ●
  • 1x Reference Array Bracket Arm ●
  • 1x Torx Screwdriver ●
  • 10x Reflective Spheres
AI/ML Overview

Here's a summary of the acceptance criteria and study information for the ClearPointer Optical Navigation Wand, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Instrument Error0.11 – 0.30 mm
Registration Error0.2 – 0.6 mm

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample size used for the accuracy testing (benchtop testing). It only states that the testing was performed, but not how many measurements were taken or how many units of the device were tested.

The data provenance is retrospective bench testing, as it describes the testing performed after the device was developed to verify design specifications. The country of origin for the data is not explicitly stated, but given that ClearPoint Neuro, Inc. is based in the US and is filing with the FDA, it can be inferred that the testing aligns with US regulatory standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the accuracy testing. The accuracy measurements (instrument and registration error) were likely determined through objective, quantifiable methods using precise measurement tools in a controlled benchtop environment, rather than expert interpretation of medical images or patient outcomes.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

The document does not specify any adjudication method. Given that the listed tests are objective accuracy measurements (instrument error, registration error) conducted during bench testing, adjudication by multiple experts is not typically applicable in this context.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or reported in this document. The ClearPointer Optical Navigation Wand is a physical navigation instrument, not an AI-assisted diagnostic or interpretative tool for medical images, so an MRMC study involving human readers with and without AI assistance is not relevant to its stated purpose.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance evaluation was conducted in the form of "benchtop accuracy testing." This testing measured the inherent accuracy of the device (instrument error and registration error) independent of human variability in a clinical setting, simulating a workflow with Stealthstation S8 software.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The ground truth used for the benchtop accuracy testing was based on objective, verifiable measurements of instrument and registration error against known or precisely controlled spatial references. It is not an expert consensus, pathology, or outcomes data, as these are typically associated with diagnostic or prognostic medical devices.

8. The Sample Size for the Training Set:

The document does not mention a training set because this device is a physical medical instrument, not a machine learning or AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

This question is not applicable as there is no training set for this device.

{0}------------------------------------------------

January 12, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

ClearPoint Neuro, Inc. Brennan Sullivan Regulatory Affairs Manager 120 S. Sierra Avenue, Suite 100 Solana Beach, California 92075

Re: K233155

Trade/Device Name: ClearPointer Optical Navigation Wand Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: December 8, 2023 Received: December 11, 2023

Dear Brennan Sullivan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Image /page/1/Picture/6 description: The image shows a digital signature. The name "Adam D. Pierce -S" is written on the left side of the image and also appears as the signer on the right side. The date of the signature is 2024.01.12, and the time is 08:21:19 -05'00'.

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233155

Device Name ClearPointer Optical Navigation Wand

Indications for Use (Describe)

The ClearPointer is intended to be used in conjunction with the SmartFrame OR System and a compatible stereotactic optical navigation system for patient registration and navigation.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

A. Device Information

CategoryComments
Sponsor:ClearPoint Neuro, Inc.6349 Paseo Del LagoCarlsbad, CA 92011
Correspondent ContactInformation:Brennan Sullivan120 S. Sierra Ave.Solana Beach, CA 92075617-678-1028bsullivan@clearpointneuro.com
Device Common Name:Neurological Stereotactic Instrument
Device Regulation & Name:Neurological Stereotaxic Instrument, 21CFR882.4560
Classification & Product Code:Class IIHAW
510(k) Number:K233155
Device Proprietary Name:ClearPointer Optical Navigation Wand

Primary Predicate Device Information:

Predicate Device:NexProbe Image-Guided Probe(Navigus Passive Probe, Model NP-1000)
Predicate Device Manufacturer:Medtronic, Inc.
Predicate Device Common Name:Neurological Stereotactic Instrument
Predicate Device Premarket Notification #K040334
Predicate Device Classification & NameNeurological Stereotaxic Instrument,21CFR 882.4560
Predicate Device Classification &Product Code:Class IIHAW

B. Date Summary Prepared

December 8, 2023

C. Description of Device

The ClearPointer Optical Navigation Wand is an optical tracker intended to be used with the SmartFrame OR Stereotactic System and ClearPoint Bone Screw Fiducials, both of which are subject of pending 510(k) submission. The ClearPointer is used in conjunction with a Medtronic StealthStation Navigation System to align the SmartFrame OR tower to the desired

{5}------------------------------------------------

ClearPoint Neuro, Inc. ClearPointer

target. It is composed of 4 optical spheres and utilizes a sphere positional geometry that is recognizable by the Medtronic StealthStation S8 Optical navigation System. It is equipped with a Pointer attachment to allow for image registration and entry point location. The Pointer Attachment can also be removed from the Optical Tracker component, so that the Tracker can be mounted to a SmartFrame OR. The SmartFrame OR Hardware Kit is provided with the following components:

  • 1x ClearPointer™ ●
  • 1x Pointer Attachment ●
  • 1x Reference Array Bracket Arm ●
  • 1x Torx Screwdriver ●
  • 10x Reflective Spheres

D. Indications for Use

The ClearPointer is intended to be used in conjunction with the SmartFrame OR System and a compatible stereotactic optical navigation system for patient registration and navigation.

E. Comparison of Technological Characteristics

The ClearPointer is substantially equivalent to the predicate device NexProbe, subject of K040334. The devices are similar in design in that they consist of a consist of a reference frame in a similar geometry and are assembled with 4 optical tracking spheres. They can both be attached to a stereotactic frame for registration and alignment using the Medtronic StealthStation Software and they utilize similar workflows to achieve alignment.

The differences between the devices primarily relate to the difference in stereotactic frame with which they are compatible. The ClearPointer is intended to be used with the ClearPoint SmartFrame OR System and the NexProbe is intended to be used with NexFrame. Any differences between the ClearPointer and NexProbe do not present any additional risks of safety or efficacy.

ClearPointerNexProbe® Image-GuidedProbe(Navigus Passive Probe,Model NP-1000)Comparisonon
Classification21CFR 882.456021CFR 882.4560Identical
Product CodeHAWHAWIdentical
NexProbe® Image-Guided
ProbeComparis
ClearPointer(Navigus Passive Probe,Model NP-1000)on
Indications forUseThe ClearPointer is intended to beused in conjunction with theSmartFrame OR System and acompatible stereotactic opticalnavigation system for patientregistration and navigation.The Navigus Passive Probe isintended to be used inconjunction with the Navigusfamily of trajectory guides andthe Medtronic StealthStation®Treatment Guidance System forpatient registration andnavigation. The device isprovided sterile and for singleuseIdentical
Intended UseThe ClearPointer is intended foruse by a Neurosurgeon in astandard operating roomenvironment to place and orient theSmartFrame OR Tower towards atarget in the brain along a plannedtrajectory during stereotacticfunctional neurosurgicalprocedures.The NexProbe Image-guidedProbe is intended for use by aNeurosurgeon in a standardoperating room environment toplace and orient the NexframeTower towards a target in thebrain along a planned trajectoryduring stereotactic functionalneurosurgical procedures.Identical
MaterialsTitanium Ti 6Al-4VSiliconePEEKNylon370HR composite20% Glass-filled polycarbonatePolymer materialStainless SteelSimilar
ComponentsClearPointer Optical TrackerPointer AttachmentOptical Tracker Reflective SpheresReference Array Bracket ArmT8 Torx ScrewdriverNexProbe subassembly11.5mm passive spheresSimilar
CompatibilityMedtronic StealthStation S8O-Arm® RegistrationPointMerge RegistrationMedtronic StealthStation S8O-Arm® RegistrationPointMerge RegistrationIdentical
Single Use orReusableSingle-UseSingle-useIdentical
Sterile?EtO SterilizedSterileIdentical

Table 1: Side-by-side comparison of ClearPointer with Predicate Device

{6}------------------------------------------------

ClearPoint Neuro, Inc. ClearPointer

F. Bench Testing

ClearPoint Neuro has performed extensive testing to demonstrate that the ClearPointer is safe and effective for its intended use. The development of the ClearPointer was conducted in conformance with the company's design control procedures. Design inputs provided the requirements for the respective product specifications. Design Verification testing was performed to confirm that the ClearPointer meets all design specifications. These tests included

{7}------------------------------------------------

ClearPoint Neuro, Inc. ClearPointer

verification of physical, performance, and safety requirements, as well as benchtop accuracy testing. Accuracy testing was performed by measuring instrument error and registration error of a simulated workflow with Stealthstation S8 software. The results are summarized in Table 2 below:

TestError
Instrument Error0.11 – 0.30 mm
Registration Error0.2 – 0.6 mm

Table 2: ClearPointer Accuracy Results

Risk analysis was also performed with mitigation of all identified risks to acceptable levels.

G. Conclusion

The subject ClearPointer and the predicate Medtronic NexProbe have identical indications for use and similar technological characteristics and principles of operation. ClearPoint Neuro has performed extensive bench testing to demonstrate that the differences between ClearPointer and the predicate device do not raise any risks of safety or efficacy. ClearPointer has been demonstrated to meet all test specifications and has been shown to be as safe, as effective, and to perform as well as the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).