(177 days)
The Augment Off-Axis Instrument System consists of specialized instruments intended to prepare anatomy prior to implantation of Comprehensive® Reverse Shoulder System components in reverse total shoulder arthroplasty procedures or Alliance® Augmented Glenoid System components in anatomic total shoulder arthroplasty procedures in accordance with their respective cleared indications for use and contraindications.
The instruments for this system are specifically indicated for use with the Comprehensive ® Reverse Shoulder System (K193373) or the Alliance® Augmented Glenoid System (K193180).
The Augment Off-Axis Instrument System consists of specialized instruments that are designed for preparation of anatomy prior to implantation of Comprehensive® Reverse Shoulder System or Alliance® Augmented Glenoid System components. The instruments in this system are intended to provide an alternate approach in ensuring proper preparation of the glenoid bone and tissue for receipt of a glenoid implant augment in reverse and anatomic total shoulder arthroplasty procedures.
A drop-in instrument tray contains the Augment Off-Axis reusable surgical instruments for secure and ergonomic storage, sterilization, and transportation between uses.
The provided FDA 510(k) document for the "Augment Off-Axis Instrument System" (K233148) does not describe an AI/ML device or software that requires acceptance criteria based on performance metrics such as sensitivity, specificity, or accuracy.
Instead, this submission is for a system of specialized surgical instruments intended to prepare anatomy for shoulder implant procedures. The document focuses on demonstrating substantial equivalence to existing predicate devices (K193373 and K193180) based on non-clinical performance data.
Therefore, the requested details regarding acceptance criteria, study design, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set information are not applicable to this specific device submission.
The document explicitly states:
- "Clinical testing was not required as a basis for substantial equivalence."
- "Substantial equivalence is supported by the results of verification testing to confirm features, geometry, and performance of the Augment Off-Axis instruments, which support that the subject device performs as well as the predicate system instruments."
The "acceptance criteria" here are implicitly tied to the verification testing and the demonstration that the new instruments perform as intended and are safe and effective, similar to the predicate devices. These criteria would relate to engineering specifications, material properties, sterility, and functional performance of the surgical tools themselves, not statistical performance metrics for an AI algorithm.
{0}------------------------------------------------
March 22, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Fix Surgical % Jennifer Palinchik President JALEX Medical 27865 Clemens Road, Suite #3 Westlake, Ohio 44145
Re: K233148
Trade/Device Name: Augment Off-Axis Instrument System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, KWS, KWT Dated: February 16, 2024 Received: February 20, 2024
Dear Jennifer Palinchik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely, Joseph P. Russell Digitally signed by Joseph P. Russell -S -ਟ Date: 2024.03.22 13:19:54 -04'00' for: Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K233148
Device Name Augment Off-Axis Instrument System
Indications for Use (Describe)
The Augment Off-Axis Instrument System consists of specialized instruments intended to prepare anatomy prior to implantation of Comprehensive® Reverse Shoulder System components in reverse total shoulder arthroplasty procedures or Alliance® Augmented Glenoid System components in anatomic total shoulder arthroplasty procedures in accordance with their respective cleared indications for use and contraindications.
The instruments for this system are specifically indicated for use with the Comprehensive ® Reverse Shoulder System (K193373) or the Alliance® Augmented Glenoid System (K193180).
| Type of Use (Select one or both, as applicable) | |
|---|---|
| 区 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Fix Surgical. The logo consists of a circular icon to the left and the text "FIX SURGICAL" to the right. The icon features a stylized letter "S" with curved lines inside a circle. The text "FIX SURGICAL" is in a sans-serif font and is colored in the same blue as the icon.
510(k) Submission -K233148 Augment Off-Axis Instrument System
510(k) Summary
| Applicant: | Fix Surgical425 Fayette Street, #617Conshohocken, PA 19428 |
|---|---|
| Date: | 03/21/2024 |
| Applicant Contact: | Brian Karpinski, Presidentbrian@fixsurgical.com |
| Contact Person:Contact Email:Contact Telephone:Contact Fax: | Jennifer Palinchik, Presidentjpalinchik@jalexmedical.com(440) 935-3282(440) 933-7839 |
| Regulation Number: | 21 CFR §888.3660 |
| Classification Product Code: | PHX, KWS, KWT |
| Device Trade Name:Device Classification Name:Device Common Name:Device Class:Reviewing Panel: | Augment Off-Axis Instrument SystemShoulder joint metal/polymer semi-constrained cemented prosthesisShoulder ProsthesisIIOrthopedic |
| Primary Predicate Device:Additional Predicates: | K193373 – Comprehensive® Reverse Shoulder SystemK080642 – Comprehensive® Reverse Shoulder SystemK193180 – Alliance® Augmented Glenoid System |
Device Description:
The Augment Off-Axis Instrument System consists of specialized instruments that are designed for preparation of anatomy prior to implantation of Comprehensive® Reverse Shoulder System or Alliance® Augmented Glenoid System components. The instruments in this system are intended to provide an alternate approach in ensuring proper preparation of the glenoid bone and tissue for receipt of a glenoid implant augment in reverse and anatomic total shoulder arthroplasty procedures.
A drop-in instrument tray contains the Augment Off-Axis reusable surgical instruments for secure and ergonomic storage, sterilization, and transportation between uses.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for FIX SURGICAL. The logo consists of a circular icon on the left and the text "FIX SURGICAL" on the right. The icon contains a stylized "S" shape formed by two curved lines within a circle. The text "FIX SURGICAL" is in a sans-serif font, with "FIX" being slightly larger than "SURGICAL". The entire logo is in a light blue color.
510(k) Submission -K233148 Augment Off-Axis Instrument System
Indications for Use:
The Augment Off-Axis Instrument System consists of specialized instruments intended to prepare anatomy prior to implantation of Comprehensive® Reverse Shoulder System components in reverse total shoulder arthroplasty procedures or Alliance® Augmented Glenoid System components in anatomic total shoulder arthroplasty procedures in accordance with their respective cleared indications for use and contraindications.
The instruments for this system are specifically indicated for use with the Comprehensive ® Reverse Shoulder System (K193373) and Alliance® Augmented Glenoid System (K193180).
Summary of Technological Characteristics:
The potential impact on substantial equivalence for each of the technological differences of the Augment Off-Axis Instrument System and the predicate Comprehensive® Reverse Shoulder System (K193373) and the Alliance® Augmented Glenoid System (K193180) instrument systems was addressed through verification and validation processes.
Collectively, the instruments of the subject Augment Off-Axis Instrument System have the same intended use and similar indications for use as predicate system instruments. Technological characteristics of the subject system are similar to those of the information provided herein demonstrates that any differences do not impact safety or effectiveness.
These aspects of the subject device were determined to be substantially equivalent to the predicates as the systems compare similarly in:
- Regulatory Characteristics and Intended Use/Indications for Use ●
- Device Function/Performance
- Materials and Manufacturing Processes
- Design Features/Dimensions
- Post-Processing Procedures, including Sterility and Shelf-Life Characteristics
Performance Data - Nonclinical:
Substantial equivalence is supported by the results of verification testing to confirm features, geometry, and performance of the Augment Off-Axis instruments, which support that the subject device performs as well as the predicate system instruments.
Clinical Testing:
Clinical testing was not required as a basis for substantial equivalence.
Conclusion:
Based upon a comparison of indications for use, designs, materials and sterilization method, performance characteristics, and operational principles, the components of the Augment Off-Axis System are substantially equivalent to those of the predicate devices identified in this premarket notification.
N/A