K Number
K233013

Validate with FDA (Live)

Date Cleared
2024-01-23

(123 days)

Product Code
Regulation Number
876.5130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 2-Way 100% Silicone ClearTract Catheter is intended for drainage of the urinary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder, but also suprapubic placement or by nephrostomy. Intended population is adults and pediatrics.

Device Description

The 2-Way 100% Silicone ClearTract Catheter is a standard single-use 2-way Foley catheter that is constructed of medical grade silicone with a surface modification. It incorporates two (2) lumens, one for inflation/deflation of the balloon and the other for drainage of the urinary tract. The drainage inlet is located distal to the catheter's balloon. The connection between the catheter and the urinary bag is accomplished by using a standard non- interconnector. Catheterization is accomplished by inserting the catheter through the urethra. Drainage can also be accomplished by suprapubic placement or nephrostomy tract placement of the catheter.

AI/ML Overview

The provided text is an FDA 510(k) clearance letter for a medical device: the "2-Way 100% Silicone Cleartract Catheter." This type of document primarily focuses on establishing substantial equivalence to a legally marketed predicate device, rather than providing detailed acceptance criteria and study results for software-based medical devices (like AI algorithms).

Therefore, the document does not contain the information requested regarding:

  • A table of acceptance criteria and reported device performance for a software component.
  • Sample sizes used for a test set or training set, or their provenance.
  • Number of experts and their qualifications for ground truth establishment.
  • Adjudication methods.
  • MRMC comparative effectiveness study results or effect sizes.
  • Standalone algorithm performance.
  • Type of ground truth used (e.g., pathology, outcomes data).
  • How ground truth for the training set was established.

The document states that "performance testing was conducted on the subject device," and "All pre-determined acceptance criteria were met" for various bench performance tests related to the physical catheter (flow rate, balloon integrity, etc.). These are physical and mechanical tests for the catheter itself, not a study evaluating software or AI performance.

In summary, this document describes the clearance of a physical medical device (a catheter) and not a software/AI-driven device, thus the specific information requested about acceptance criteria and study details for an AI/software component is not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 23, 2024

Silg Technologies Corporation % Aaron Rogers Director of Regulatory and Quality Pathway, LLC. 8779 Cottonwood Avenue, Suite 105 Santee, CA 92071

Re: K233013

Trade/Device Name: 2-Way 100% Silicone Cleartract Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: December 18, 2023 Received: December 19, 2023

Dear Aaron Rogers:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen -S

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233013

Device Name

2-Way 100% Silicone Cleartract Catheter

Indications for Use (Describe)

The 2-Way 100% Silicone ClearTract Catheter is intended for drainage of the urinary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder, but also suprapubic placement or by nephrostomy. Intended population is adults and pediatrics.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

TRADITIONAL 510(K) SUMMARY

1. SUBMITTER INFORMATION

ApplicantSILQ Technologies,Corporation 323 Sunny Isles Blvd., 7thFloor Sunny Isles Beach,FL 33160
Official CorrespondentAaron RogersPathway, LLC.8779 Cottonwood Ave., Suite 105Santee, CA 92071619-415-0103 x704arogers@pathwaynpi.com
Date PreparedJanuary 16, 2024
2. DEVICE NAME
Trade Name of the Device2-Way 100% Silicone Cleartract Catheter
Common Name:Foley Catheter
Classification Name:Urological Catheter and Accessories
Classification Regulation:21 CFR 876.5130
Device Class:II
Product Code:EZL
Panel:Gastroenterology/Urology
3. PREDICATE AND REFERENCEDEVICE IDENTIFICATIONPredicate: 2-Way 100% Silicone Cleartract Catheter (K222118)Reference: Silicone Foley Catheter for single use (K172807)

4. DEVICE DESCRIPTION:

The 2-Way 100% Silicone ClearTract Catheter is a standard single-use 2-way Foley catheter that is constructed of medical grade silicone with a surface modification. It incorporates two (2) lumens, one for inflation/deflation of the balloon and the other for drainage of the urinary tract. The drainage inlet is located distal to the catheter's balloon. The connection between the catheter and the urinary bag is accomplished by using a standard non- interconnector. Catheterization is accomplished by inserting the catheter through the urethra. Drainage can also be accomplished by suprapubic placement or nephrostomy tract placement of the catheter.

Catheter sizes 14Fr., 16Fr., and 18Fr. were cleared under the predicate, K222118. Subject catheter of size 24 Fr. is new and being requested for clearance as part of this submission.

{5}------------------------------------------------

5. INDICATIONS FOR USE:

The 2-Way 100% Silicone ClearTract Catheter is intended for drainage of the urinary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder, but also suprapubic placement or by nephrostomy. Intended population is adults and pediatrics.

K233013

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES

Subject Device:Comparison Element 2-Way 100% Silicone ClearTractCatheterPredicate Device:2-Way 100% SiliconeClearTract CatheterReference Device:Silicone Foley Catheter for singleuse
510(k) NumberK233013K222118K172807
Indications forUseThe 2-Way 100% SiliconeClearTract Catheterisintended for drainage ofthe urinary tract.Catheterizationisaccomplished by insertingthe catheter throughthe urethra and intothe bladder, but alsosuprapubic placement orby nephrostomy. Intendedpopulation is adults andpediatrics.The 2-Way 100%Silicone ClearTractCatheter is intended fordrainage of the urinarytract. Catheterization isaccomplished byinserting the catheterthrough the urethra andinto the bladder, but alsosuprapubic placement orby nephrostomy.Intended population isadults and pediatrics.Two-way DisposableSilicone Foley Catheter:Urethral catheterization forbladder drainage forurological use only; theindwelling time is no morethan 30 days. Three-wayDisposable Silicone FoleyCatheter:Urethralcatheterization for bladderdrainage and bladderirrigation for urological useonly; the indwelling time isno more than 30 days.
Device Type2-Way Foley Catheter withinflation and drainagelumens2-Way Foley Catheterwith inflation anddrainage lumens2-Way with inflation anddrainage lumens3-Way with inflation,drainage and irrigationlumens
Size24Fr14Fr, 16Fr, 18Fr6Fr, 8Fr, 10Fr, 12Fr, 14Fr,16Fr, 18Fr, 20Fr, 22Fr, 24Fr
Balloon Volume30ml5ml, 10ml3ml, 5ml, 10ml, 15ml, 30ml
Length405 mm405 mm400 mm
Materials ofConstructionSilicone, Medical GradeSilicone, Medical GradeSilicone, Medical Grade
Polymer SurfaceModificationYesYesNo
Indwelling TimeUp to 30 daysUp to 30 daysUp to 30 days
Single Use?YesYesYes
Prescription Use?YesYesYes
Sterile?YesYesYes

As evidenced by the above table, both the subject and the predicate devices have same intended use, but the subject and predicate devices have different technological characteristics. However, performance testing was conducted on the subject device, and it was established that the differences in technological characteristics between the subject and the predicate does not raise different questions of safety or effectiveness.

{6}------------------------------------------------

7. SUMMARY OF NONCLINICAL TESTING:

Below is a list of the tests that were performed and successfully completed for the subject device per the specified guidance and standards:

K233013

  • Biocompatibility testing according to ISO 10993-1:2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Guidance "Use of International Standard ISO 10993-1" (2016).
  • The EO sterilization process was validated in accordance with ANSI/AAMI/ISO 11135:2014. ●
  • . The following bench performance testing were conducted per ASTM F623-19 and the FDA guidance document, 'Conventional Foley Catheters - Performance Criteria for Safety and Performance Based Pathway'-
    • Flow Rate through Drainage Lumen O
    • Balloon Integrity (Resistance to Rupture) O
    • O Inflated Balloon Response to Traction
    • Balloon Volume Maintenance O
    • Balloon Size and Shaft Size O
    • Deflation Reliability (Failure to Deflate) o

All pre-determined acceptance criteria were met.

8. CONCLUSIONS

Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.

§ 876.5130 Urological catheter and accessories.

(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.