K Number
K233008

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-06-05

(257 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The aap Cannulated Headless Bone Screws are indicated for use in non-spinal fragment fixation, osteotomy and arthrodesis of human bones appropriate for the size of the screw. The devices are intended for single use only. The aap Cannulated Headless Bone Screws are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.

K-Wires are indicated for use as guide wire for osteosynthesis implants and for application as implant according to the AO/ASIF principles of fracture management.

The K-Wires are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.

Device Description

The subject device Cannulated Headless Bone Screws are cannulated at both the leading end (tip) and at the trailing end. The leading/trailing diameters are: 2.5/3.4 mm. 3.5/4.5 mm. 4.0/4.8 mm. 4.5/5.3 mm, 5.5/6.2 mm, 6.5/7.2 mm, and 7.5/7.9 mm., The leading thread have different pitches which facilitates closure of any fracture gap and generates compression across the bone fragments. The subject screws are provided in overall lengths ranging from 12 mm, and with threaded lengths (at the leading end or tip) ranging from 4 mm to 40 mm. The subject screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and ISO 5832-3.

The subject K-Wire is provided in one size with a diameter of 2.4 mm and a length of 250 mm. The subject K-Wire has a trocar point, is threaded, and is manufactured from stainless steel conforming to ASTM F138 and ISO 5832-1.

AI/ML Overview

This document describes the aap Cannulated Headless Bone Screws, a medical device for bone fixation. Since this is a 510(k) summary, the core of the submission is to demonstrate substantial equivalence to existing legally marketed devices, rather than establishing brand-new performance criteria for a novel device. Therefore, the "acceptance criteria" here are largely defined by demonstrating that the new device performs at least as well as and shares similar characteristics with the predicate devices.

Here's an analysis of the provided information based on your questions:

1. Table of Acceptance Criteria and Reported Device Performance

Instead of explicit acceptance criteria with numerical targets as one might see for a novel device, the "acceptance criteria" here are met by demonstrating similarity and comparable performance to the predicate devices.

Acceptance Criteria (Implied by Substantial Equivalence to Predicates)Reported Device Performance (as demonstrated in the submission)
Intended Use: The device must have the same or similar intended use as the predicate devices.The aap Cannulated Headless Bone Screws have the same intended use as the primary predicate K111316 and reference devices K130590 and K131459, and similar technological characteristics for bone fracture fixation (bone screws and K-Wire).
Technological Characteristics: The device must have similar technological characteristics (e.g., design, materials, operating principles) as the predicate devices.The subject device bone screws have similar ranges of thread diameters, threaded lengths, and screw designs to the primary predicate K111316 and reference device K130590. The K-Wire dimensions are within the range of K-Wires cleared in K131459.
Material Composition: The device materials must be identical or equivalent to those of the predicate devices.The screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and ISO 5832-3, using materials and processes identical to predicate devices. The K-Wire is manufactured from stainless steel conforming to ASTM F138 and ISO 5832-1, also using processes identical to predicate devices.
Biocompatibility: The device must be biocompatible, demonstrated by using identical/equivalent materials and manufacturing processes as predicate devices.Biocompatibility for the subject device materials was referenced from K111316, K130590, and K131459, indicating identical materials and processes as previously cleared devices.
Sterilization: The device must be able to be sterilized to a sterility assurance level (SAL).The devices are provided non-sterile and are to be sterilized to a SAL of 10-6 by the end user using the same sterilization method (moist heat) and parameters as previously cleared devices. Analysis showed the subject devices do not create a new worst case for moist heat sterilization.
Mechanical Performance: The device must meet recognized industry standards for mechanical performance.Mechanical testing of the subject device screws was performed according to ASTM F543. While specific quantitative results aren't provided in this summary, the implication is that they meet the standard to support substantial equivalence.
Packaging and Shelf Life: The device packaging and shelf-life characteristics must be consistent with predicate devices.Packaging and shelf life information for the subject devices was referenced from K111316, K130590, and K131459.
Manufacturing Process: The device must be manufactured using identical or similar processes as predicate devices.All subject device final finished components are manufactured in the same facilities using identical manufacturing processes as used for previously cleared aap device components.
Pediatric Population Use Support: The device's use in pediatric populations must be supported by comparison to predicate devices.Reference device K161616 is used to support the use of the subject screws in the pediatric population.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a number of devices or clinical cases for this submission's testing. The non-clinical performance data mentions "mechanical testing of the subject device screws according to ASTM F543." ASTM F543 outlines test methods, but the specific number of screws tested is not provided in this summary.
  • Data Provenance: The data is primarily non-clinical performance data which includes mechanical testing, sterilization validation, and references to previous submissions (K111316, K130590, K131459, K161616) for biocompatibility, packaging, and shelf-life. There is no clinical data included in this submission. The origin of the reference data for predicate devices is not specified, but previous 510(k) submissions would have their own established provenance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This submission relies on non-clinical performance testing and substantial equivalence to predicate devices, not on establishing ground truth via expert review of clinical cases.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

Not applicable. There's no test set requiring human adjudication for diagnostic or interpretive tasks. The evaluation is based on engineering specifications, material standards, and mechanical test results.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This is a medical device (bone screws and K-wires), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC study or AI assistance evaluation was performed.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a physical medical device, not an algorithm or software.

7. The Type of Ground Truth Used

The "ground truth" for this submission is established through:

  • Engineering Specifications and Material Standards: Compliance with recognized standards like ASTM F136, ISO 5832-3, ASTM F138, ISO 5832-1, ASTM F543, ISO 17665-1, ISO 10993-1. These standards define the acceptable properties and performance for such devices.
  • Bench Testing: Mechanical testing data according to ASTM F543.
  • Referencing Predicate Devices: The established regulatory clearances and performance data of the cited predicate devices (K111316, K130590, K131459, K161616) serve as a de-facto "ground truth" for comparison under the substantial equivalence pathway.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not a machine learning model, so there is no training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 5, 2024

aap Implantate AG % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K233008

Trade/Device Name: Cannulated Headless Bone Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: May 14, 2024 Received: May 14, 2024

Dear Kevin Thomas:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices

{2}------------------------------------------------

OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233008

Device Name

Cannulated Headless Bone Screws

Indications for Use (Describe)

Cannulated Headless Bone Screws

The aap Cannulated Headless Bone Screws are indicated for use in non-spinal fragment fixation, osteotomy and arthrodesis of human bones appropriate for the size of the screw. The devices are intended for single use only. The aap Cannulated Headless Bone Screws are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.

K-Wire

K-Wires are indicated for use as guide wire for osteosynthesis implants and for application as implant according to the AO/ASIF principles of fracture management.

The K-Wires are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K233008 Cannulated Headless Bone Screws aap Implantate AG

May 30, 2024

ADMINISTRATIVE INFORMATION

Manufacturer Nameaap Implantate AG
Lorenzweg 5
12099 Berlin Germany
Telephone +49 30 750 19 129
Fax +49 30 750 19 111
Official ContactAgnieszka Mierzejewska, Chief Operating Officer
Representative/ConsultantKevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
12264 El Camino Real, Suite 400
San Diego, CA 92130
Telephone +1 858-792-1235
Fax +1 858-792-1236
Email kthomas@paxmed.com
flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Device NameCommon NameCannulated Headless Bone ScrewsScrew, fixation, bone
Regulation Number21 CFR 888.3040
Regulation NameSmooth or threaded metallic bone fixation fastener
Regulatory ClassClass II
Primary Product CodeHWC
Additional Product CodesHTY
Classification PanelOrthopedic
Reviewing OfficeOffice of Health Technology 6 (OHT 6: Orthopedic Devices)
Reviewing DivisionDivision of Health Technology 6 C (Restorative, Repair and Trauma Devices)

PREDICATE DEVICE INFORMATION

Primary Predicate Device K111316, aap Cannulated Screws, aap Implantate AG

Reference Devices K130590, aap Cannulated Screw 2.0, aap Implantate AG K131459, aap Wire Bone, aap Implantate AG K 161616, DePuy Synthes 4.0 mm and 4.5 mm Cortex Screws, DePuy Synthes 2.4 mm Cannulated Screws, DePuy Synthes 3.5 mm and 4.0 mm Cannulated Screws, DePuy Synthes 4.5 mm Cannulated Screws, DePuy Synthes 6.5 mm Cannulated Screws, DePuy Synthes 7.0 mm Cannulated Screws, DePuy Synthes 1.5 mm Headless, Synthes USA Products, LLC

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Cannulated Headless Bone Screws

The aap cannulated headless bone screws are indicated for use in non-spinal fracture and fragment fixation, osteotomy and arthrodesis of human bones appropriate for the screw. The devices are intended for single use only. The aap Cannulated Headless Bone Screws are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.

K-Wire

K-Wires are indicated for use as guide wire for osteosynthesis implants and for application as implant according to the AO/ASIF principles of fracture management.

The K-Wires are indicated for adults, and children (2 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.

SUBJECT DEVICE DESCRIPTION

The subject device Cannulated Headless Bone Screws are cannulated at both the leading end (tip) and at the trailing end. The leading/trailing diameters are: 2.5/3.4 mm. 3.5/4.5 mm. 4.0/4.8 mm. 4.5/5.3 mm, 5.5/6.2 mm, 6.5/7.2 mm, and 7.5/7.9 mm., The leading thread have different pitches which facilitates closure of any fracture gap and generates compression across the bone fragments. The subject screws are provided in overall lengths ranging from 12 mm, and with threaded lengths (at the leading end or tip) ranging from 4 mm to 40 mm. The subject screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and ISO 5832-3.

The subject K-Wire is provided in one size with a diameter of 2.4 mm and a length of 250 mm. The subject K-Wire has a trocar point, is threaded, and is manufactured from stainless steel conforming to ASTM F138 and ISO 5832-1.

PERFORMANCE DATA

Recognized standards used in the non-clinical performance testing included:

ASTM F136 (2013) Standard Specification for Wrought Titanium-4 Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications (UNS R56401);

ISO 5832-3 (2021) Implants for surgery - Metallic materials - Part 3: Wrought titanium 6-aluminium 4vanadium alloy;

ASTM F138 (2019) Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673);

ISO 5832-1 (2016) Implants for surgery - Metallic materials - Part 1: Wrought stainless steel;

ISO 17665-1 (2006) Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices:

ISO 17665-2 (2009) Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1;

ISO 10993-1 (2018) Biological evaluation of medical devices – Part I : Evaluation and testing within a risk management process; and

ASTM F543 (2017) Standard Specification and Test Methods for Metallic Medical Bone Screws.

Non-clinical data submitted or referenced to demonstrate substantial equivalence included:

  • referenced from for the primary predicate K111316 and the reference devices K130590 and K131459 . was moist heat sterilization validation to a sterility assurance level of 10° by the overkill method according to

{6}------------------------------------------------

ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO TIR 17665-2, and ANSI/AAMI/ISO 14937; analysis showed that the subject devices do not create a new worst case for moist heat sterilization;

  • referenced from K111316, K130590, and K131459 was packaging and shelf life information for the subject . devices:
  • referenced from K111316, K130590, K131459 was biocompatibility for the subject device materials; ●
  • provided in this submission was mechanical testing of the subject device screws according to ASTM F543. ●

No clinical data were included in this submission.

EQUIVALENCE TO MARKETED DEVICES

The subject device, the primary predicate device K111316, and the reference devices K130590 and K131459 have the same technological characteristics and use the same operating principles for bone fracture fixation (bone screws and K-Wire) or for use as an instrument (K-Wire to facilitate placement of cannulated screws). The reference device K161616 is to support the use of the subject screws in the pediatric population.

The subject device bone screws are provided in ranges of thread diameters, threaded lengths that are similar to the primary predicate K111316 and the reference device K130590. The subject screw designs also are similar to the designs of screws cleared in K111316 and K130590.

The subject device K-Wire has a diameter of 2.4 mm, an overall length of 250 mm, and a trocar point. These dimensions are within the range of K-Wires (with trocar point) cleared in K131459.

The subject device screws are manufactured from titanium alloy conforming to ASTM F136 and ISO 5832-3 using materials and processes identical to that used for screws cleared in the primary predicate device K111316 and the reference device K130590.

The subject device K-Wire is manufactured from stainless steel confirming to ASTM F138 or ISO 5832-1 using processes identical to that used for K-Wires cleared in the reference device submission K131459.

All subject device final finished components are manufactured in the same facilities using identical and identical manufacturing processes as used for the previously cleared aap device components. Therefore, the subject devices are substantially equivalent to predicate devices regarding biocompatibility.

The subject device components are provided non-sterile in the same packaging as in the primary predicate K111316 and the reference devices K130590 and K131459, and are to be sterilized to a sterility assurance level (SAL) of 10 the end user using the same sterilization method (moist heat) and parameters as the previously cleared devices. The subject devices do not represent a new worst-case for the sterilization.

In support of substantial equivalence of the subject device screws mechanical testing was performed according to ASTM F543.

CONCLUSION

The subject devices, the primary predicate device, and the reference devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.