K Number
K232988

Validate with FDA (Live)

Date Cleared
2023-11-21

(60 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diagnostic Ultrasound System Aplio flex Model CUS-AFL00 and Aplio go Model CUS-AGG00 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, pediatric, small organ (thyroid, breast and testicle), neonatal cephalic, trans-rectal, transvaginal, musculoskeletal (both conventional and superficial), cardiac, peripheral vascular, and thoracic/pleural.

This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler , Power Doppler as well as Tissue Harmonic Imaging, Combined Modes and Acoustic attenuation mapping.

This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.

Device Description

The Aplio flex, Model CUS-AFL00 and Aplio go, Model CUS-AGG00, V2.0 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ an array of probes including flat linear array, convex, and sector array with frequency ranges between approximately 2.5MHz to 12MHz.

AI/ML Overview

The provided text is a 510(k) summary for the Aplio flex and Aplio go Software V2.0 Diagnostic Ultrasound System. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed performance study with explicit acceptance criteria and corresponding results for specific device performance metrics.

Therefore, the document does not contain the detailed information required to fully answer your request, especially regarding:

  • A table of acceptance criteria and reported device performance.
  • Sample sizes, data provenance, number of experts, adjudication methods for a test set.
  • MRMC comparative effectiveness study results (effect size).
  • Details on standalone performance.
  • Training set details and ground truth establishment for the training set.

The document states generally that "Risk Analysis and verification and validation activities demonstrate that the established specifications for these devices have been met" and "Additional performance testing included in the submission was conducted in order to demonstrate that the requirements for the new features were met." However, it does not elaborate on what those specific specifications or requirements were, nor does it quantify the performance against them.

The information that can be extracted or inferred is as follows:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with specific, quantifiable acceptance criteria and corresponding performance metrics for the Aplio flex and Aplio go Software V2.0. The overall "acceptance criterion" implied throughout the document is demonstrating substantial equivalence to the predicate device (Xario 200G/100G Diagnostic Ultrasound System, Software Version 1.1, K182596) in terms of safety and effectiveness, and meeting established specifications in verification and validation activities.

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of clinical performance evaluation (e.g., patient cases, images). The testing mentioned is broadly "Risk Analysis and verification and validation activities" and "Additional performance testing for new features." This suggests engineering and system-level testing rather than a clinical performance study with a distinct test set. Therefore, sample size and data provenance are not reported.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided, as the document does not describe a clinical study with a ground truth established by experts.

4. Adjudication method for the test set

This information is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention an MRMC comparative effectiveness study or any AI components that would assist human readers in a way that would require an "AI vs without AI assistance" comparison. The device is a diagnostic ultrasound system, not an AI-powered diagnostic aide in the sense of an algorithm interpreting images for a human.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided, and based on the device description as a diagnostic ultrasound system, a standalone "algorithm only" performance would not be the primary focus for its type of application.

7. The type of ground truth used

This is not explicitly stated in relation to a clinical performance study. For the verification and validation, the "ground truth" would generally be established by engineering specifications, validated test methods, and potentially physical phantoms or established reference measurements for image quality, functional performance, and safety parameters.

8. The sample size for the training set

The document does not mention a "training set," which would typically be relevant for machine learning or AI-driven devices. This device is presented as an ultrasound system, not an AI model that requires a training set in the typical sense.

9. How the ground truth for the training set was established

As no training set is mentioned, information on how its ground truth was established is not provided.

{0}------------------------------------------------

November 21, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.

Canon Medical Systems Corporation % Yoshiaki Cook Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K232988

Trade/Device Name: Aplio flex and Aplio go Software V2.0 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: September 21, 2023 Received: September 22, 2023

Dear Yoshiaki Cook:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K232988

Device Name

Aplio flex and Aplio go Software V2.0 Diagnostic Ultrasound System

Indications for Use (Describe)

The Diagnostic Ultrasound System Aplio flex Model CUS-AFL00 and Aplio go Model CUS-AGG00 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications:

fetal, abdominal, pediatric, small organ (thyroid, breast and testicle), neonatal cephalic, trans-rectal, transvaginal, musculoskeletal (both conventional and superficial), cardiac, peripheral vascular, and thoracic/pleural.

This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler , Power Doppler as well as Tissue Harmonic Imaging, Combined Modes and Acoustic attenuation mapping.

This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

AL SYSTEMS USA, INC.

Made For life

510(k) SUMMARY

K232988

SUBMITTER'S NAME 1. Fumiaki Teshima Sr. Manager, Quality Assurance Dept. Quality, Safety and Regulation Center Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi-ken, Japan 324-8550

  • ESTABLISHMENT REGISTRATION 2. 9614698

OFFICIAL CORRESPONDENT/CONTACT PERSON 3.

Yoshiaki Cook Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive Tustin, CA 92780 ycook@us.medical.canon +1 (657) 270-5595

4. DATE PREPARED

September 21, 2023

5. DEVICE NAME/TRADE NAME

Aplio flex and Aplio go Software V2.0 Diagnostic Ultrasound System

COMMON NAME 6.

System, Diagnostic Ultrasound

7. DEVICE CLASSIFICATION

Class II

Ultrasonic Pulsed Doppler Imaging System – Product Code: 90-IYN [per 21 CFR 892.1550] Ultrasonic Pulsed Echo Imaging System – Product Code: 90-IYO [per 21 CFR 892.1560] Diagnostic Ultrasonic Transducer – Product Code: 90-ITX [per 21 CFR 892.1570]

{4}------------------------------------------------

8. PREDICATE DEVICE

ProductMarketed by510(k) NumberClearance Date
Xario 200G/100G DiagnosticUltrasound System, SoftwareVersion 1.1 (Primary predicate)Canon MedicalSystems USA, Inc.K182596November 02, 2018
Aplio i900/i800/i700 DiagnosticUltrasound System, SoftwareV7.0 (Reference device)Canon MedicalSystems USA, Inc.K223017March 31, 2023

9. REASON FOR SUBMISSION

New device.

10. DEVICE DESCRIPTION

The Aplio flex, Model CUS-AFL00 and Aplio go, Model CUS-AGG00, V2.0 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ an array of probes including flat linear array, convex, and sector array with frequency ranges between approximately 2.5MHz to 12MHz.

11. INDICATIONS FOR USE

The Diagnostic Ultrasound System Aplio flex Model CUS-AFL00 and Aplio go Model CUS-AGG00 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, pediatric, small organ (thyroid, breast and testicle), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculoskeletal (both conventional and superficial), cardiac, peripheral vascular, and thoracic/pleural.

This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler as well as Tissue Harmonic Imaging, Combined Modes and Acoustic attenuation mapping.

This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.

12. SUBSTANTIAL EQUIVALENCE

The subject devices, Aplio flex and Aplio go Software V2.0 Diagnostic Ultrasound System, are substantially equivalent to the predicate devices, Xario 200G/100G Diagnostic Ultrasound System, Software Version 1.1 (K182596).

The subject devices employ the same fundamental scientific technology as the predicate devices and function in a manner similar to, and are intended for the same use as the predicate devices. Differences between the subject devices and the cleared predicate devices do not raise any new questions about the safety and effectiveness of the subject devices. This submission includes evidence to demonstrate the substantial equivalence of the subject devices to the predicate devices.

{5}------------------------------------------------

  • The subject and predicate devices have the same clinical intended use and imaging modes, which differ only by the availability of Pencil Scan and CW Pencil Mode with the predicate devices
    • The indicated uses available with the subject devices are pre-existing with the O predicate devices, with the exception of the additional availability of thoracic/pleural imaging with the subject devices
  • The transducers newly developed for use with the subject devices employ the same ● fundamental scientific technology and are substantially equivalent to those existing with the predicate devices, while the remaining transducers supported by the subject devices have been previously cleared with the predicate devices
  • . The software features supported in the subject and predicate devices are largely identical.
    • o The following features, previously cleared with reference devices, Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V7.0 (K223017), and which are implemented into the subject devices, are not available with the predicate devices:
      • Attenuation Imaging, which enables the calculation and visualization of the ultrasound frequency dependent attenuation coefficient
      • . Vascularity Index, which provides measurements based upon the calculation of pixels within a Power image
      • . Lateral Gain, which enables the ability to adjust gain control from image edge to edge
      • l ApliGate, which enables sharing of de-identified images from Aplio flex and Aplio go
      • Network Storage, which enables raw data to be stored to a Network Attached Server (NAS)
      • . Tricefy Access, which enables DICOM images to be sent from Aplio flex and Aplio go through a firewall to the Tricefy cloud server, a Class 1 medical device
    • The following features and/or functionality, previously cleared with the predicate O devices, are improved in the subject devices as follows:
      • I Trapezoid scan is now enabled for B mode and tilted scan for CDI, PDI, and ADF
      • . QuickScan now enables the setting of PWD gate within the ROI, as well as Doppler angle correction marker and angle oblique scan, in relation to flow location and direction
      • l Mecha4D now enables OmniView image, a display of the transverse plane, along the cutline set on an MPR image
      • l Luminance now supports transducer model PVU-674MVS
      • " Protocol Assistant now enables any cine/still image to be displayed as a guide, previously acquired examination images to be displayed as reference, and any user input text to be displayed
      • Capability to hold an additional transducer, now up to a total of four
    • The following features and/or functionality, which are available with the predicate O devices, are not implemented into the Aplio flex and Aplio go
      • . Superb Micro vascular Imaging (SMI); other Color Doppler image processing features are equivalent between the subject and predicate devices
      • . Auto IMT, a vascular measurement feature; other application measurements are identical between the subject and predicate devices

{6}------------------------------------------------

  • ' 2D Wall Motion Tracking, which enables myocardial wall motion analysis by regional pattern matching in 2D cardiac images
  • . Contrast Imaging, Stress Echo, Shear Wave, and Elastography

14. SAFETY

The subject devices are designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. These devices are in conformance with the applicable parts of the AAMI/ANSI ES60601-1: 2005/(R)2012 & A1:2012, C1:2009/(R)2012& A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021], IEC 60601-1-2:2020, IEC 60601-2-37:2015, IEC 62304:2015, IEC 62359:2017, and ISO 10993-1:2018 standards.

15. TESTING

Risk Analysis and verification and validation activities demonstrate that the established specifications for these devices have been met. Additional performance testing included in the submission was conducted in order to demonstrate that the requirements for the new features were met. The results of all of these studies demonstrate that the subject devices meet established specifications and perform as intended and in accordance with labeling.

FDA guidance document "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers", issued February 21, 2023 was referenced for this submission, along with "Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023.

Additionally, cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 02, 2014, is included in this submission.

Testing of these devices was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and UL systems.

16. CONCLUSION

The Aplio flex, Model CUS-AFL00 and Aplio go, Model CUS-AGG00, V2.0 are substantially equivalent to the Xario 200G/100G Diagnostic Ultrasound System, Software Version 1.1, K182596. The subject devices function in a manner similar to and are intended for the same use as the predicate devices, as described in labeling. The evidence provided in this submission demonstrate that Aplio flex, Model CUS-AFL00 and Aplio go, Model CUS-AGG00, V2.0 are safe and effective for their intended use and perform with substantial equivalence to the predicate devices.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.