K Number
K232969

Validate with FDA (Live)

Date Cleared
2024-06-07

(260 days)

Product Code
Regulation Number
876.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1) Endoscopic marking; 2) Hemostasis for: A) Mucosal/sub-mucosal defects < 3cm. B) Bleeding ulcers. C) Polyps < 1.5cm in diameter. D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel

Device Description

The proposed device is consisted of a clip section. The delivery section. The delivery section include: slide bar, slip ring, protective sleeve, rotation head, coated spring tube, slip ring holder, traction assembly, connecting tube, decoupled piece, decoupled shrapnel, rotation sleeve. And the clip section include: clip, sliding block, shaft and rocker arm. The clip is made of stainless steel. The clip can be opened and closed more than five times prior to deployment, aiding in repositioning of the lesion site. The proposed device is available in single ring and three ring models, and each case is divided into rotatable and non-rotatable clip. The effective length of the proposed device includes 1250mm, 1650mm, 2350mm and the opening width is 9mm, 11mm. 13mm and 16mm. The proposed devices are provided in sterile and single use.

AI/ML Overview

The provided text describes the 510(k) submission for a medical device, the "Disposable Hemostatic Clips." This submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. Crucially, the document explicitly states: "No clinical study is included in this submission."

Therefore, I cannot provide information regarding:

  • Acceptance Criteria and Reported Device Performance (Table): There are no clinical performance metrics of the device against "acceptance criteria" presented from a clinical study. The document details non-clinical tests and their pass/fail status, indicating they met product requirements, but not against specific acceptance criteria for clinical performance.
  • Sample size and data provenance (test set): No clinical test set.
  • Number of experts and qualifications (ground truth for test set): Not applicable for a clinical test set.
  • Adjudication method (test set): Not applicable.
  • MRMC comparative effectiveness study: Not conducted.
  • Standalone (algorithm only) performance: Not applicable as this is a mechanical device, not an AI algorithm.
  • Type of ground truth used (clinical): No clinical data to establish ground truth.
  • Sample size for training set: No clinical training set.
  • How ground truth for training set was established: No clinical training set.

However, I can provide information based on the non-clinical testing summarized in the document, framed as how the device "meets the acceptance criteria" for functional and safety performance, as demonstrated by bench testing and biocompatibility assessments.

Non-Clinical Acceptance Criteria and Reported Device Performance

The device's performance was evaluated through a series of non-clinical tests to verify it met design specifications and standards, demonstrating substantial equivalence to the predicate device. The "acceptance criteria" for these tests are implicitly that the device passes the specified tests and complies with the relevant ISO and ASTM standards.

Here's a summary of the non-clinical tests and their outcomes:

Acceptance Criteria (Implicit: Pass Test & Comply with Standards)Reported Device Performance (Summary)
Material Biocompatibility (ISO 10993 series)Clip Component:
CytotoxicityNo cytotoxicity reported.
IrritationNo irritation reported.
Skin SensitizationNo skin sensitization reported.
Acute Systemic ToxicityNo acute toxicity reported.
Sub-acute Systemic ToxicityNo sub-acute toxicity reported.
GenotoxicityNo genotoxicity reported.
Pyrogenicity (Bacterial Endotoxins)Levels below 2.15 EU/device (in accordance with USP <85>).
Implantation TestNo abnormal reaction reported.
Material Biocompatibility (ISO 10993 series)Delivery Component:
CytotoxicityNo cytotoxicity reported.
IrritationNo irritation reported.
Skin SensitizationNo skin sensitization reported.
Acute Systemic ToxicityNo acute toxicity reported.
PyrogenicityLevels below 2.15 EU/device.
Sterilization (ISO 11135:2014 & ISO 10993-7:2008)Achieved Sterility Assurance Level (SAL) of 10⁻⁶. EO and Ethylene Chlorohydrin (ECH) residuals were below specified limits.
Packaging Integrity (ASTM F1929-15, F88/F88M-15, F1886/F1886M-16)Passed tests after environmental conditioning and simulated transportation (ASTM D4169-22). Included seal strength, dye penetration, and visual inspection.
Shelf-life (ASTM F1980 -16)Validated for a 3-year shelf-life through baseline and accelerated testing.
Performance Bench TestsAll tests passed and "demonstrated the result can meet the product requirements." These tests include: - Appearance - Rotation performance - Release force - Clamping strength - Relocation - Tensile strength - Dimension - Mechanical integrity of clip assembly - Scope compatibility/usability - Endoscope damage - Torque - Clip approach - Three ring handle strength Specifically noted that the three-ring model's tension tests showed results similar to the predicate device, able to withstand clinical use forces.

Study Proving the Device Meets Acceptance Criteria (Non-Clinical)

The "study" proving the device meets its non-clinical acceptance criteria consists of the comprehensive set of Non-Clinical Tests and Biocompatibility Testing.

  1. Sample sizes used for the test set and the data provenance:

    • The document does not specify the exact sample sizes for each non-clinical test (e.g., number of units tested for tensile strength or rotation performance). It states "all tests were passed."
    • Data Provenance: The tests were performed by Ningbo Xinwell Medical Technology Co., Ltd. (China) or their designated testing facilities, as per the submission. This is internal testing data for the 510(k) submission.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of non-clinical, bench testing. The "ground truth" for these tests is defined by the established industry standards (ISO, ASTM, USP) and the device's design specifications. Conformance to these standards and specifications is evaluated through laboratory testing.
  3. Adjudication method for the test set:

    • Not applicable as this refers to human review of clinical data. Non-clinical tests have pass/fail criteria based on objective measurements against defined standards.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No, a MRMC comparative effectiveness study was not done. The submission states: "No clinical study is included in this submission." This implies no clinical performance data involving human readers or patients, only bench and biocompatibility testing. The comparison to the predicate device is based on functional equivalence and meeting, or not raising new questions about, safety and effectiveness through non-clinical data.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a mechanical hemostatic clip and delivery system, not an algorithm or AI-powered device.
  6. The type of ground truth used:

    • For non-clinical performance and safety: The ground truth is defined by established international and national standards (ISO 10993, ISO 11135, ASTM F1929, ASTM F88/F88M, ASTM F1886/F1886M, USP <85>, ASTM D4169, ASTM F1980) and the device's internal design specifications and requirements. The "truth" is whether the device's physical properties and performance characteristics meet these objective, measurable standards.
  7. The sample size for the training set:

    • Not applicable as there is no clinical study data, training set, or machine learning component cited.
  8. How the ground truth for the training set was established:

    • Not applicable as there is no clinical study data, training set, or machine learning component cited.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 7, 2024

Ningbo Xinwell Medical Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co, Ltd. P.O. Box 120-119 Shanghai. 200120 CHINA

Re: K232969

Trade/Device Name: Disposable Hemostatic Clips Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: September 21, 2023 Received: May 9, 2024

Dear Diana Hong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sivakami Venkatachalam -S

for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K232969

Device Name

Disposable Hemostatic Clips

Indications for Use (Describe)

The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1) Endoscopic marking; 2) Hemostasis for: A) Mucosal/sub-mucosal defects < 3cm. B) Bleeding ulcers. C) Polyps < 1.5cm in diameter. D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K232969

    1. Date of Preparation: 06/03/2024
    1. Sponsor Identification

Ningbo Xinwell Medical Technology Co., Ltd. No.188 Binjiang Road, Cixi High-tech Industrial Development Zone, Cixi City, Zhejiang,P.R. China

Establishment Registration Number: 3013526170

Contact Person: Lei Xi Position: RA manager Tel: +86-021-34693382 Email: xilei@xinhaigroup.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

    1. Identification of Proposed Device
      Trade Name: Disposable Hemostatic Clips Common Name: Hemostasis Clipping Device

Regulatory Information Classification Name: Hemorrhoidal Ligator;

{5}------------------------------------------------

Classification: II, Product Code: PKL, Regulation Number: 21CFR 876.4400 Review Panel: Gastroenterology/Urology,

Indications for Use:

The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of:

    1. Endoscopic marking;
    1. Hemostasis for:
  • A) Mucosal/sub-mucosal defects < 3cm.
  • B) Bleeding ulcers.
  • C) Polyps < 1.5cm in diameter.
  • D) Diverticula in the colon.

E) Arteries < 2 mm.

F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection.

  1. As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively

  2. Anchoring to affix jejunal feeding tubes to the wall of the small bowel.

    1. Device Description
      The proposed device is consisted of a clip section. The delivery section. The delivery section include: slide bar, slip ring, protective sleeve, rotation head, coated spring tube, slip ring holder, traction assembly, connecting tube, decoupled piece, decoupled shrapnel, rotation sleeve. And the clip section include: clip, sliding block, shaft and rocker arm. The clip is made of stainless steel. The clip can be opened and closed more than five times prior to deployment, aiding in repositioning of the lesion site. The proposed device is available in single ring and three ring models, and each case is divided into rotatable and non-rotatable clip. The effective length of the proposed device includes 1250mm, 1650mm, 2350mm and the opening width is 9mm, 11mm. 13mm and 16mm. The proposed devices are provided in sterile and single use.
  • Identification of Predicate Device 6.
    Predicate Device 510(k) Number: K202333 Product Name: Lockado™ Repositionable Hemostasis Clip

{6}------------------------------------------------

Non-Clinical Test Conclusion 7.

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ♪ ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
  • ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity

  • ISO 10993-6:2016 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation

  • ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals

  • ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization

  • ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

  • ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation

  • ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.

  • ASTM F88/F88M-15 standard method for seal strength of flexible barrier materials

  • ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection

  • USP <85> Bacterial Endotoxins Test

  • A ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems

Performance testing

The following bench tests were performed on the Disposable Hemostatic Clips:

  • . Appearance
  • Rotation performance •
  • Release force
  • . Clamping strength
  • . Relocation
  • . Tensile strength
  • . Dimension
  • . Mechanical integrity of clip assembly
  • Scope compatibility/usability
  • . Endoscope damage
  • . Torque
  • Clip approach .
  • Three ring handle strength

Above all tests were passed and demonstrated the result can meet the product requirements.

{7}------------------------------------------------

Biocompatibility testing

The biocompatibility evaluation for the proposed device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on September 4, 2020. The following tests were conducted for the clip component of Disposable Hemostatic Clips:

  • . Cytotoxicity
  • Irritation .
  • Skin Sensitization .
  • Acute Systemic Toxicity
  • . Sub-acute Systemic Toxicity
  • . Genotoxicity
  • . Pyrogenicity
  • Implantation Test

The following tests were conducted for the delivery component of Disposable Hemostatic Clips:

  • . Cytotoxicity
  • Irritation ●
  • . Skin Sensitization
  • . Acute Systemic Toxicity
  • . Pyrogenicity

Sterility, Shipping, and Shelf -life

The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 106. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 2.15 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 3 year shelf-life.

  • . Package integrity test - after environmental conditioning, simulated transportation testing in accordance to ASTM D4169-22 on final, packaged, and sterile device.
  • Sterile Barrier Packaging performed on the proposed device: .
    • Seal Strength ASTM F88/F88-15 o
    • Dye penetration ASTM F1929-15 O
    • Visual Inspection ASTM F1886/F1886M-16 O
  • . Shelf-life of 3-years is validated using FDA recognized standard ASTM F1980 -16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

{8}------------------------------------------------

Clinical Test Conclusion 8.

No clinical study is included in this submission.

    1. Substantially Equivalent (SE) Comparison
Table 1. General Comparison
ItemProposed DevicePredicate DeviceRemark
Product CodePKLPKLSame
Regulation Number21CFR 876.440021CFR 876.4400Same
Indications for UseThe DisposableHemostatic Clips is indicatedfor Endoscopic clip placementwithin the Gastrointestinal tractin adult populations only via astraight or side viewing flexibleendoscope for the purpose of :1) Endoscopic marking;2) Hemostasis for:A) Mucosal/sub-mucosaldefects < 3cm.B) Bleeding ulcers.C) Polyps < 1.5cm indiameter.D) Diverticula in thecolon.E) Arteries < 2 mm.F) Prophylactic clipping toreduce the risk of delayedbleeding post lesion resection.3) As a supplementarymethod, closure of GI tractluminal perforations <20mmthat can be treatedconservatively4) Anchoring to affix jejunalfeeding tubes to the wall of thesmall bowel;The Lockado™ RepositionableHemostasis Clip is indicated forEndoscopic clip placementwithin the Gastrointestinal tractin adult patients only via astraight or side viewing flexibleendoscope for the purpose of:(1) Endoscopic marking,(2) Hemostasis for(a)Mucosal / sub-mucosaldefects < 3cm,(b) Bleeding ulcers,(c) Polyps < 1.5cm in diameter,(d) Diverticula in the colon,(e) Arteries < 2 mm(f) Prophylactic clipping toreduce the risk of delayedbleeding post lesion resection;(3) As a supplementary method,closure of GI tract luminalperforations <20mm that can betreated conservatively(4) Anchoring to affix jejunalfeeding tubes to the wall of thesmall bowel;Same
ConfigurationDelivery system and clipDelivery system and clipSame
assemblyassembly
Open width9mm, 11mm, 13mm and 16mm8mm, 11mm and 16mm, 22mmDifferent
Minimal working channel2.8mm2.8mmSame
Working length1250mm, 1650mm, 2350mm and 2700mm1650mm 1950mm, 2350mm and 2700mmDifferent
Ring numberSingle ring and three ringsSingle ringDifferent
Single UseSingle UseSingle UseSame
LabelingComply with 21 CFR Part 801Comply with 21 CFR Part 801Same
Material
Clip assemblyStainless Steel (SUS316L)Stainless Steel (SUS631)Stainless Steel (SUS316F)
Traction assemblyStainless Steel (SUS304)
Outer tube assemblyStainless Steel (SUS304)Stainless Steel (SUS316F)Polyethylene (PE)
Slide barAcrylonitrile Butadiene Styrene (ABS)UnknownDifferent
Slip ringThermoplastic Elastomer (TPE)
Rotation headAcrylonitrile Butadiene Styrene (ABS)
Thermoplastic Elastomer (TPE)
Protective sleeveHigh pressure-low density polyethylene (LDPE)
Biocompatibility for Delivery system
CytotoxicityNo cytotoxicity
Skin SensitizationNo skin sensitizationSame
IrritationNo irritationComply with ISO 10993 standards
Acute Systemic ToxicityNo acute toxicity
Sub-acute Systemic ToxicityNo sub-acute toxicity
Biocompatibility for clip assembly
CytotoxicityNo cytotoxicity
Skin SensitizationNo skin sensitizationSame
IrritationNo irritationComply with ISO 10993 standards
Acute Systemic ToxicityNo acute toxicity
Sub-acute Systemic ToxicityNo sub-acute toxicity
ImplantationNo abnormal reaction
GenotoxicityNo genotoxicity
Sterilization
MethodEthylene oxideEthylene oxideSame
SAL$10^{-6}$$10^{-6}$Same
Endotoxin Limit20EU20EUSame

Table 1. General Comparison

{9}------------------------------------------------

{10}------------------------------------------------

Different - Open width

The open width for proposed device is available in 9mm, 11mm, 13mm and 16mm four different specifications. The predicate device is available in 8mm, 11mm 16mm and 22mm four different specifications. The 9mm and 13mm specification is different from predicate device. However, these specifications can be covered in the open width range of predicate device. Different working length devices will be selected by physician per patient's condition. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.

Different - Working length

The working length for proposed device is available in 1250mm, 2350mm and 2700mm. The working length of predicate device is available in 1650mm 1950mm and 2700mm. Different working length devices will be selected by physician per patient's condition. The length for proposed device of 1650mm, 2350mm and 2700mm can be covered by predicate device length range. Although 1250mm effective length device are not included in the effective length range of the predicate device, it is very close to the minimum length of predicate device. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.

Different- Ring number

The proposed device has two models of single ring and three rings, while the predicate device only has single ring model. Single ring models can be covered by predicate device. The sponsor also has performed ring tension tests on the three-ring model of the proposed device, and the test results show that three-ring model is similar as predicate device, and it can withstand the forces generated during clinical use. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.

Different-Material

The material for the predicate device is unknown. However, biocompatibility test has been conducted on the proposed device for contact materials and the test result can meet the requirements of ISO 10993 series standard. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.

10. Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device in 510(k) submission, Disposable Hemostatic Clips, is as safe, as effective, and performs as well as or better than the legally marketed

{11}------------------------------------------------

predicate device cleared under K202333.

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.