K Number
K232966

Validate with FDA (Live)

Device Name
FISIOWARM 7.0
Manufacturer
Date Cleared
2024-12-26

(462 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The massage devices are intended to provide a temporary reduction in the appearance of cellulite.

Device Description

FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are active and therapeutic medical devices that perform their function by releasing energy to the human body.

AI/ML Overview

The provided text is a 510(k) summary for the medical device Fisiowarm 7.0. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML device performance study (e.g., diagnostic accuracy, reader performance).

The performance testing mentioned in the document is for electrical safety, electromagnetic compatibility, and a comparative thermal test on human skin to verify energy transfer, not related to an AI/ML device's diagnostic or analytical performance.

Therefore, I cannot provide the requested information from the given text as it does not pertain to an AI/ML-driven medical device and its associated performance study.

The document covers:

  • Device Name: Fisiowarm 7.0
  • Regulation Number: 21 CFR 878.4400 (Electrosurgical Cutting And Coagulation Device And Accessories)
  • Product Code: PBX
  • Intended Use: Topical heating to elevate tissue temperature for the treatment of pain, muscle spasms, increased local circulation, and temporary reduction in cellulite appearance.
  • Predicate Device: Winback Back 3SE (K162828)
  • Performance Testing: Electrical compliance (EN 60601-1, EN 60601-1-2) and tissue testing to compare thermal spread. This is essentially showing the device performs its physical function (heating) comparably to the predicate.
  • No mention of AI/ML components or studies measuring diagnostic/analytical performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 26, 2024

Golden Star Srl Tiziana Borrelli Product Manager Via Enrico Pallini 9 Roma, 00149 Italy

Re: K232966

Trade/Device Name: Fisiowarm 7.0 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX Dated: September 18, 2024 Received: November 27, 2024

Dear Tiziana Borrelli:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Francisco Delgado -S 12:01:43 2024.12.26

Image /page/2/Picture/4 description: The image contains some text on the left side, including '00'05- and 2024.12.26. The text 'elgado -S' is also visible. On the right side, there is a faded watermark of the letters 'FDA'.

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K232966

Device Name

FISIOWARM 7.0 (FISIOWARM 7.0)

Indications for Use (Describe)

FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The massage devices are intended to provide a temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

SECTION 5 510 (k) SUMMARY

{5}------------------------------------------------

510(k) Notification

510 (k) SUMMARY

SUBMITTER INFORMATION

A. Company Name:
----------------------
  • B. Company Address:
  • C. Company Phone: Company e-mail:

D. Contact person:

GOLDEN STAR S.r.l.

Via Enrico Pallini 9 00149 Roma (RM) - Italy

+39 06 5897627 amministrazione@goldenstar.it

Stefano Giordano CEO Golden Star S.r.l.

E. Date Summary Prepared: 18th September 2023

DEVICE IDENTIFICATION

A. Device Name:FISIOWARM 7.0 300 & FISIOWARM 7.0400
B. Common Name:Electrosurgical, Cutting & Coagulation &Accessories
C. Classification:Class II
D. Product Code:PBX
E. Regulation Number:878.4400
E. Submission Type:510k (Original Submission)

LEGALLY MARKETED PREDICATE DEVICE

Primary predicate device510 (k) Holder510 (k) No.Date cleared
Winback Back 3SEWINBACK USA CorpK162828September 18th, 2017

{6}------------------------------------------------

DEVICE DESCRIPTION

FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are active and therapeutic medical devices that perform their function by releasing energy to the human body.

Performance Testing

The following performance testing was conducted to prove compliance with performance requirements and support substantial equivalence:

TestObjectiveResult
ElectricalCompliance with EN 60601-1Pass
Compliance with EN 60601-1-2Pass
Tissue TestingCompare thermal spread ofdevicesEquivalent

INDICATIONS FOR USE STATEMENT

FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The massage devices are intended to provide a temporary reduction in the appearance of cellulite.

SUBSTANTIAL EQUIVALENCE

All information provided with the present submission supports the substantial equivalence for FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 with the predicate device and its accessories that have identical characteristics and intended use and similar indication statement. In addition all clinical data and all performance tests that have been performed in accordance with the Standards for the Software Evaluation and for the Electrical and Electromagnetic Safety test, demonstrate FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 safety and effectiveness for its intended use.

The following matrix illustrates the equivalencies of FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400, as well as the substantial equivalent predicate device.

PREDICATE DEVICES COMPARISON CHART Table 1

{7}------------------------------------------------

GOLDENSTAR S.r.l.FISIOWARM 7.0 300FISIOWARM 7.0 400New DeviceWINBACK USA Corp WinbackBack 3SEPredicate device
"K" NUMBERSK232966K162828
Proprietary nameFISIOWARM 7.0 300FISIOWARM 7.0 400Winback Back 3SE
CFR Section878.4400878.4400
Pro-codePBXPBX
Intended /Indications foruseFISIOWARM® 7.0 300 andFISIOWARM® 7.0 400 areintended to provide topicalheating for the purpose ofelevating tissue temperature forthe treatment of selected medicalconditions such as relief of pain,muscle spasms, and increase inlocal circulation. The massagedevices are intended to provide atemporary reduction in theappearance of cellulite.The Winback Back 3SE is intended toprovide topical heating for the purposeof elevating tissue temperature for thetreatment of selected medicalconditions such as relief of pain,muscle spasms, and increase in localcirculation. The Winback Back 3SEmassage devices are intended toprovide a temporary reduction inthe appearance of cellulite.
Mode ofoperationRadiofrequency energyRadiofrequency energy
OutputUnipolar/Bipolar/Multifunction*Unipolar/Multipolar
Power Supply100-240VAC50/60Hz110-250 VAC50/60 Hz
Frequency300kHz – 1 MHz300kHz – 1 MHz
Max RF Power300W - 400W**300W
ConfigurationConsole with accessoriesCart mounted console withaccessories
Electrical SafetyStandardsComplies with IEC60601-1,IEC60601-1-2, IEC60601-2-2Complies with IEC60601-1,IEC60601-1-2, IEC60601-2-2
HandpiecesResistive, capacitiveResistive, capacitive
ElectrodesResistive, capacitive, neutral,dermato-functional, staticautomatic, neutral dynamicResistive, capacitive, neutral,dermato-functional, staticautomatic, neutral dynamic
  • Bipolar and Multifunction Outputs are the Multipolar Output splitted in two separated connectors.

** The Fisiowarm 400 can handle 30% more power for very demanding and intensive usage.

{8}------------------------------------------------

TECHNICAL CHARACTERISTICS

A comparison of the technological characteristics of FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 and the predicate device has been performed. The results of this comparison demonstrate that the technologic characteristics and the operating principle of FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are the same or very similar to those of the claimed predicate device. Where any differences arise from the analysis of the predicate device characteristics and those of the device subject of this submission, the clinical data evaluated from scientific publications give scientific evidence of the safety and the effectiveness of FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400. The system was evaluated and found compliant with IEC 60601-1 for electrical safety, IEC 60601-1-2 for electromagnetic compatibility, and ISO 10993-1 for biocompatibility of the handpieces and electrodes. Verification and validation data show that the device meets all product specifications.

PERFORMANCE DATA

Laboratory and performance tests were executed to ensure that the device functioned as intended and met design specifications. A comparative thermal test was performed on human skin in order to experimentally verify the transfer of energy to the biological tissue of both devices. This verification was carried out by measuring the trend of the temperature increase in the biological tissue over time, at different energy levels.

Sufficient data were obtained to show that the device is substantially equivalent to the predicate device and meets safety and effectiveness criteria.

CLINICAL DATA

The devices have been designed and validated in such a way that, when used under the conditions and for the purposes intended, they will not compromise the clinical condition or the safety of patients, or the safety and health of users or other people, provided that any risk which may be associated with its use constitute acceptable risks when weighed against the benefits to the patient and is compatible with a high level of protection of health and safety.

CONCLUSION

Based on the foregoing, FISIOWARM® 7.0 300 and FISIOWARM® 7.0 400 are substantially equivalent to the legally marketed, claimed predicate devices for the purposes of this 510(k) submission. Safety and effectiveness were reasonably assured, justifying 510 (k) clearance.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.