K Number
K232903

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-10-17

(29 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Legend X Platform and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (Applicator VO).

It is also intended for use in dermatologic and general surgical procedures for the non-invasive treatment of mild to moderate facial wrinkles and rhytides when using TriPollar RF Energy and for muscle conditioning to stimulate healthy muscles (Applicators 1-3).

Legend X Platform is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. Legend X Platform is intended to be operated by a trained professional who is present to monitor treatment.

Device Description

The Legend X Platform ("Proposed Device") is a software-controlled capital equipment platform that enables application of radiofrequency onto the skin for ablation, resurfacing, or noninvasive treatment of wrinkles and rhytides or muscle activation. The application of radiofrequency or muscle activation only occurs under the continuous and direct control of a user with direct visualization, via the user inputs into the Console, Applicator indicators and subsequent operation of the Foot Switch and Applicators.

The proposed device consists of a Console with connected Foot Switch. Patient Controlled Manual Switch, Applicators, and Applicator accessories (disposable tips) needed to perform some of Legend X dermatological and general surgical procedures.

AI/ML Overview

The provided text is a 510(k) summary for the Legend X Platform, a medical device. It does not contain information about acceptance criteria or a study that proves the device meets those criteria in the context of diagnostic or screening performance.

Instead, the document details various performance data related to the device's safety and effectiveness, focusing on engineering, manufacturing, and software validation. This is typical for a 510(k) submission, which aims to demonstrate substantial equivalence to a predicate device rather than conducting a de novo clinical efficacy study.

Here's a breakdown of the performance data provided, addressing what can be inferred from your request, and indicating what is not available in this document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't provide a table of acceptance criteria with specific numerical targets and corresponding performance results for diagnostic accuracy, sensitivity, specificity, etc. This is because the device (Legend X Platform) is an electrosurgical cutting and coagulation device, and a muscle stimulator, not a diagnostic or screening tool with such performance metrics.

However, it outlines the areas of performance testing and implies that the device met the requirements of the standards and internal specifications used.

CategoryAcceptance Criteria (Implied)Reported Device Performance
ReprocessingAdherence to AAMI TIR-30:2011 for cleaning and disinfection of non-disposable components.Cleansing and disinfection instructions validated against AAMI TIR-30:2011.
SterilityCompliance with ISO 11135:2014 for EO sterilization and ISO 10993-7:2008/AMD-1:2019 for EO residuals.Single-use disposable sterile (EO) devices and EO residuals validated per specified ISO standards.
Shelf Life & Sterile Barrier PackagingCompliance with ASTM F1980-16, ASTM F1929-15, ASTM F88/F88M-15, and ISO 11607.Evaluated per specified ASTM and ISO standards.
BiocompatibilityCompliance with ISO 10993-1, 10993-5, 10993-10, 10993-11, and 10993-23, and FDA guidance.Patient-contacting components evaluated and validated according to specified FDA guidance and ISO standards.
Electrical Safety, Performance & EMCCompliance with IEC 60601-1, 60601-1-2, 60601-2-10, and 60601-2-2.Device fully evaluated for electrical safety and EMC compliance to specified IEC standards.
SoftwareCompliance with FDA guidance for software in medical devices; conformance to design specifications and user needs.Software developed, tested, and verified per FDA guidance document; V&V testing confirms conformance to design specifications and user needs.
V&V Testing (Device Functionality)Device operates as specified by design input requirements, including applicator, switch operations, safety features, and general functionality; adherence to regulatory standards.Performance testing executed to verify overall functionality, controls, safety features, and adherence to regulatory standards. Results confirm conformance to design specifications and needs.
Animal TestingCompliance with "General Considerations for Animal Studies Intended to Evaluate Medical Devices" FDA guidance.Animal performance testing executed according to "General Considerations for Animal Studies Intended to Evaluate Medical Devices" guidance.

2. Sample size used for the test set and the data provenance:

  • Not Applicable in the traditional sense for diagnostic/screening performance. The device is not a diagnostic tool that produces "test sets" of patient data for performance evaluation like an AI algorithm for image analysis would.
  • The validation activities mentioned (reprocessing, sterility, biocompatibility, electrical safety, software V&V, animal testing) would involve specific sample sizes or populations relevant to those engineering and biological tests (e.g., number of sterilization cycles, quantity of materials for biocompatibility, number of software test cases, number of animals in the animal study). However, these details are not provided in this summary.
  • Data provenance is not specified for these tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This device review does not involve establishing ground truth from expert interpretations, as it's not an AI/ML-driven diagnostic device. The "ground truth" for the engineering performance tests would be defined by the technical specifications, standards, and established scientific methods (e.g., chemical analysis for residuals, physical testing for seal strength, etc.).

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not Applicable. Adjudication methods like 2+1 or 3+1 refer to how discrepancies in expert readings are resolved to establish a ground truth for a diagnostic task. This is not relevant to the performance data presented for the Legend X Platform.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No such study was done or reported. This device is an electrosurgical and muscle stimulation platform, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study with AI assistance is outside the scope of this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable in the context of diagnostic/screening performance. While the device does have "Legend X Software" which was developed, tested, and verified, this software primarily controls the device's operation (energy delivery, user interface, safety features). There is no indication that this software performs a standalone diagnostic or interpretative function that would be assessed in terms of "algorithm only" performance for identifying medical conditions. The device explicitly states it "is intended to be operated by a trained professional who is present to monitor treatment."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For the technical and biological performance tests, the "ground truth" is based on:
    • Validated Standards: AAMI TIR-30, ISO 11135, ISO 10993 series, ASTM F series, ISO 11607, IEC 60601 series.
    • Internal Design Specifications: As mentioned in the software V&V section, "Results of verification and validation testing confirm that the proposed device conforms to design specifications and meets the needs of the intended users."
    • FDA Guidance: "Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "General Considerations for Animal Studies Intended to Evaluate Medical Devices."
  • There's no mention of ground truth established via expert consensus, pathology, or outcomes data for the device's operational performance, as it's not a diagnostic device.

8. The sample size for the training set:

  • Not Applicable. The document describes a medical device, not an AI/ML algorithm that requires a "training set" of data for learning. The software mentioned is likely traditionally programmed and tested rather than machine learning based.

9. How the ground truth for the training set was established:

  • Not Applicable, as there is no mention of a training set or an AI/ML algorithm that learns from data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.

October 17, 2023

Pollogen Ltd % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K232903

Trade/Device Name: Legend X Platform Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, NGX Dated: September 18, 2023 Received: September 18, 2023

Dear Prithul Bom:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.10.17
15:50:52 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023

See PRA Statement below.

Indications for Use

510(k) Number (if known) K232903

Device Name Legend X Platform

Indications for Use (Describe)

The Legend X Platform and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (Applicator VO).

It is also intended for use in dermatologic and general surgical procedures for the non-invasive treatment of mild to moderate facial wrinkles and rhytides when using TriPollar RF Energy and for muscle conditioning to stimulate healthy muscles (Applicators 1-3).

Legend X Platform is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. Legend X Platform is intended to be operated by a trained professional who is present to monitor treatment.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K232903

5.1. General information

510(k) SubmitterPollogen Ltd
6 Kaufman Street
Tel Aviv, IL 6801298
FDA Registration Number3008753275
Primary CorrespondentKaren Smith
Vice President, Regulatory & Quality
Lumenis Be, Inc., a Pollogen Ltd Family Company
Contact InformationEmail: karen.smith@lumenis.com
Secondary CorrespondentLola Fajinmi
Regulatory Affairs Consultant
MedTeCheck
Contact InformationEmail: lola@MedTeCheck.com
Date PreparedSeptember 17, 2023

5.2. Device Identification

The Legend X Platform consists of a Console with connected components including: a Foot Switch, Patient Controlled Manual Switch, Applicators, and Applicator accessories (disposable tips) needed to perform some of Legend X dermatologic procedures. The Legend X Platform can be defined as follows:

Proprietary NameLegend X Platform
Device Classification Name:Electrosurgical, Cutting & Coagulation & Accessories
Regulation Description:Electrosurgical Cutting and Coagulation Device and Accessories
Regulation Number:878.4400
Device Class:Class II
Product Code:GEI, NGX

5.3. Predicate Devices

Primary Predicate:

Proprietary NamePollogen Legend+™ System
Device Classification Name:Electrosurgical, Cutting & Coagulation & Accessories
Regulation Description:Electrosurgical Cutting and Coagulation Device andAccessories
510(k) Number:

{5}------------------------------------------------

Pollogen Ltd

Regulation Number:878.4400
Device Class:Class II
Product Code:GEI

Secondary Predicate:

Proprietary NameLegend Pro DMA
Device Classification Name:Stimulator, Muscle, Powered for Muscle Conditioning
Regulation Description:Powered Muscle Stimulator
510(k) Number:K200545
Regulation Number:890.5850
Device Class:Class II
Product Code:NGX

5.4. Device Description

The Legend X Platform ("Proposed Device") is a software-controlled capital equipment platform that enables application of radiofrequency onto the skin for ablation, resurfacing, or noninvasive treatment of wrinkles and rhytides or muscle activation. The application of radiofrequency or muscle activation only occurs under the continuous and direct control of a user with direct visualization, via the user inputs into the Console, Applicator indicators and subsequent operation of the Foot Switch and Applicators.

The proposed device consists of a Console with connected Foot Switch. Patient Controlled Manual Switch, Applicators, and Applicator accessories (disposable tips) needed to perform some of Legend X dermatological and general surgical procedures. The summary descriptions of each component are outlined below.

Legend X Console: The Legend X Console (also known as the Console or Main Control Unit) contains the graphic display interface for the user that is provided by a touchscreen monitor for viewing and a computer running the Legend X software. The monitor allows for user input during initial setup and throughout the session. The console also provides power and connectivity for the applicators, foot switch, and patient manual-controlled switch.

Legend X Foot Switch: The Legend X Foot Switch (Foot Switch) is connected to the Legend X Console via a connector cord. The Foot Switch enables or disables the energy delivery through the applicators to treatment area.

Legend X Applicators: The Legend X Applicators (1, 2, 3 and VO) are handheld handpieces which deliver energy to the treatments area. Depending on the applicator and selected user inputs, the user can utilize the applicators to deliver either muscle energy onto the muscles of the face and body or RF energy for ablation and resurfacing or noninvasive treatment of wrinkles and rhytides onto the skin. Each applicator is connected to the Console via an Applicator Connector cord.

{6}------------------------------------------------

Legend X Patient Controlled Manual Switch: The Legend X Patient Controlled Manual Switch is also a handheld device that serves as the patient interface that allows the patient to stop the operation of the chosen treatment mode upon the press of a button. In this situation, the system does not allow operation to continue. A Patient Controlled Manual Switch cord connects the Patient Controlled Manual Switch to the Console.

Legend X Software: The Legend X Software provides the user and patient with the ability to safely commence, drive, and stop the operation of the Applicators on the skin or muscle area of interest throughout the face and body. It receives user input from the Legend X Console, Foot Switch, and Patient Manual Controlled Switch, and computes the appropriate output to the chosen applicator connected to the Console. It provides a graphical user interface where the treatment timeline as well as status of operation is shown in real time and displays important system status information.

Legend X Disposable Tips: The Legend X Disposable Tips set is part of the Legend X Applicator VO kit. The Disposable Tips Set includes:

gen12 disposable tip gen 36 disposable tip gen 36L disposable tip gen 100 disposable tip H7X7 disposable tip

The patient contacting portions of the disposable tips are constructed of stainless steel and gold plating. The disposable tips are connected to handheld Applicator VO for delivery of radiofrequency electrical current via an array of multi-electrode pins onto the skin surface.

5.5. Intended Use/Indications for Use

The Legend X Platform and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (Applicator VO).

It is also intended for use in dermatologic and general surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides when using TriPollar RF Enerqy and for muscle conditioning to stimulate healthy muscles (Applicators 1-3).

Legend X Platform is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. Legend X Platform is intended to be operated by a trained professional who is present to monitor treatment.

5.6. Comparison of Technological Characteristics with the Predicate Device

Overall, the proposed and predicate devices are based on the following similar basic technological elements:

{7}------------------------------------------------

Device contains a central console that connects each of the main components to facilitate dermatologic and general surgical procedures and facilitates the selection of either RF, muscle stimulation, or fractional RF treatment modes via the graphic user interface

Device contains applicators to allow energy delivery and application of the selected mode on to the area of interest on the skin or muscles

Device contains similar components between the subject and predicate devices (i.e., console, foot switch, applicators, patient manual-controlled switch, and disposable tips)

Device requires continuous direct control by the user to operate the Applicators Device mode operation is only at the command of the user

Device mode operation may be halted at the command of either the user or the patient

The main difference between the proposed device and the predicates, Pollogen Legend+ TM System (K173503) and Legend Pro DMA (K200545), is that the Legend X Platform includes two of the five disposable tips (gen100 and H7x7) as an additional option. The Legend X Platform is an upgraded version of the predicate devices. For this reason, additional testing for the disposable tips was performed and is captured as part of this submission.

Proposed DeviceLegend X PlatformPrimary PredicatePollogen Legend+TM(K173503)SecondaryPredicateLegend Pro DMA(K200545)TechnologicalCharacteristicsComparison
System Console/Main Control UnitSystem Console/Main Control UnitSystem Console/Main Control UnitIdentical Characteristics:Functionality: To providecentral connection to alldevice components; aswell as a user interfacefor user inputs duringsetupSame specifications andmaterialsDifferencesMinor updates toconsole, GUI
Foot SwitchFoot SwitchFoot SwitchSame

Comparison of the technological characteristics with the Predicate Device for each of the Legend X Platform components are described below:

{8}------------------------------------------------

Pollogen Ltd

Traditional 510(k) Notification

Patient-controlledmanual switchPatient-controlledmanual switchPatient-controlledmanual switchSame
Applicators (1, 2, 3, and VO)Applicators (1, 2, 3, and VO)Applicators (1, 2, and 3)Same
Disposable Tips:Gen12Gen36Gen36LGen100H7x7Disposable Tips:Gen12Gen36Gen36LN/AIdentical Characteristics:Functionality: To providedelivery of fractional RFenergySimilar specificationsand intendedfunctionality.DifferencesTwo additional optionaldisposable tips (samematerials with additionalpin configurations forgen100 and H7x7) forapplicator VOMinor external colordifferences

5.7. Performance Data

The Legend X Platform was assessed for performance in accordance with internal design specification with the applicable performance standards to demonstrate safety and effectiveness. The testing identified no new issues of safety or effectiveness. The testing performed are summarized below:

Reprocessing

Cleaning: The cleaning and disinfection instructions provided in labeling for nondisposable components were validated against the following standard: AAMI TIR-30:2011 – A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices

Sterility: Single- use disposable sterile (EO) devices and EO residual were validated in accordance with the following standards:

ISO 11135:2014 - Sterilization of healthcare products - Ethylene Oxide -Requirements for development validation and routine control of a sterilization process for medical devices

ISO10993-7:2008/AMD-1:2019 – Biological evaluation of medical devices – Ethylene Oxide sterilization residuals

Shelf Life and Sterile Barrier Packaging: The shelf life and sterile barrier packaging of

{9}------------------------------------------------

K232903

the single-use disposable devices were evaluated per the following standards:

ASTM F1980-16 – Standard quide for accelerated aging of sterile barrier systems for medical devices

ASTM F 1929-15 – Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration

ASTM F88/F88M-15: Standard Test Method for Seal Strength of Flexible Barrier Materials

ISO 11607 - Packaging for terminally sterilized medical devices

Biocompatibility

The final finished form of the subject device has been used for the biocompatibility evaluation. Biocompatibility for patient contacting components has been evaluated and validated in accordance with the provision of the following FDA Guidance document(s) and standards:

Guidance for Industry and Food and Drug Administration Staff – Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process"

ISO 10993-1:2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

ISO 10993-5:2009 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 - Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

ISO 10993-23:2021 - Biological evaluation of medical devices - Part 23: Tests for irritation

Electrical Safety, Performance and EMC

The proposed device has been fully evaluated for electrical safety and EMC compliance to the following standards:

IEC 60601-1:2005 (Third edition) + CORR. 1: 2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint) – Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) IEC 60601-1-2:2014 – Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 60601-2-10:2016 – Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators

{10}------------------------------------------------

IEC 60601-2-2:2017 - Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

Software

Software was developed, tested, and verified per the FDA quidance document: Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices

Results of verification and validation testing confirm that the proposed device conforms to design specifications and meets the needs of the intended users.

V&V Testing

Performance testing was executed to verify the overall functionality of the proposed device to operate as specified by the design input requirements including applicator, patient-manual controlled-switch, and foot switch operation and controls, various functional safety features, and other general functionality. Requirements for safety and efficacy of the system. including but not limited to the adherence to requlatory standards were verified. Results of verification testing confirm that the proposed device conforms to design specifications and requirements and meets the needs of the intended user.

Animal Testing

Performance testing - animal was executed according to General Considerations for Animal Studies Intended to Evaluate Medical Devices issued on March 28, 2023.

5.8. Conclusion

Based on the indication for use, technological characteristic, performance testing and comparison to the predicate devices, the Legend X Platform raises no new questions of safety and effectiveness as compared to the predicate devices and is substantially equivalent to the predicate devices in terms of safety, efficacy, and performance.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.