(146 days)
Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning.
Leksell GammaPlan® is a powerful, computer-based treatment planning system specifically designed for the simulation and planning of stereotactic Leksell Gamma Knife® radiosurgery based on tomographic and projectional images.
The basis of treatment planning is the acquisition and processing of digital images by a computer workstation running the treatment planning application software. The program is capable of handling a range of different imaging modalities. Images from tomographic sources such as Computer Tomography (CT), Magnetic Resonance (MR) and Positron Emission Tomography (PET) scanners can be used as well as projectional images from angiograms (AI). This allows the direct comparison between vascular structures in projectional images and tissue structures in CT and MR.
Digital images can be imported into the system via the computer network.
The treatment planning application has the ability to plan a patient's treatment protocol based on a single target or multiple targets.
The basic elements of treatment planning are:
- defining the cranial target or targets .
- . devising the configuration of the collimators to be used during treatment
- determining the parameters of the radiation shots to be delivered by Leksell • Gamma Knife®.
The provided FDA 510(k) summary for the Leksell GammaPlan® (LGP) (K232854) does not contain specific details regarding acceptance criteria and the comprehensive study results in the way requested for a typical AI/ML device.
This submission is for an updated version of a treatment planning system, which is generally software for medical device operation, rather than an AI/ML algorithm that performs autonomous diagnostic or prognostic tasks. As such, the performance testing focuses on software verification and validation against requirements, rather than clinical performance metrics in terms of sensitivity, specificity, etc., with a ground truth established by experts.
However, I can extract the information that is present and indicate where the requested information is not available in the document.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| For software functionality and performance: | "Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new and existing features against requirement specifications." |
| "Formal design and usability validation has been performed on a clinically equivalent device by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs." | "Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification and user needs have been met and that the system is confident and stable." |
| "The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety for the new functionality was effective." | "The results of verification and validation as well as conformance to relevant safety standards demonstrate that LGP v11.4 meets the established safety and performance criteria..." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified in the document. The testing described is software verification and validation, not a clinical trial with a "test set" in the sense of patient data for AI model evaluation.
- Data Provenance: Not specified, as it's not a clinical data-driven study. The document mentions "clinically equivalent device" for validation, implying internal testing and possibly simulated or anonymized data, but details are absent.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of Experts: Not specified.
- Qualifications of Experts: The document mentions "competent and professionally qualified personnel" performed design and usability validation, but no specific professional qualifications (e.g., radiologist, medical physicist) or experience levels are provided. Ground truth in the context of a treatment planning system primarily relates to the accuracy of computational models and adherence to clinical guidelines, rather than expert interpretation of images for diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not specified. It's unlikely this type of adjudication was performed, as the evaluation focused on software functionality and adherence to specifications, not on resolving disagreements in expert clinical assessment.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC study was not performed. This device is a treatment planning system, not an AI diagnostic or assistance tool intended to improve human reader performance in interpreting medical images. The document states, "No animal or clinical tests were performed to establish substantial equivalence with the predicate device."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: The device itself is software for treatment planning. Its "standalone" performance would be its ability to correctly calculate and display treatment plans based on inputs, which is covered by the mentioned verification and validation testing. However, this is not "standalone AI algorithm performance" in the typical sense of a diagnostic AI. The system is inherently "human-in-the-loop" as it requires a clinician to define targets and parameters.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: For the type of software described (treatment planning), the "ground truth" for verification and validation typically involves:
- Specification Compliance: Verification that the software performs according to its written requirements and design specifications.
- Industry Standards: Conformance to relevant medical device software (e.g., IEC 62304) and medical physics standards.
- Phantom/Physical Measurement Comparisons: For dose calculations, comparison with physical measurements in phantoms or known analytical solutions (though not explicitly detailed for this specific 510(k)).
- Clinical Equivalence/Usability: Validation that the software supports intended clinical use, as performed by "competent and professionally qualified personnel."
The document does not specify "expert consensus," "pathology," or "outcomes data" as ground truth for this submission, as these are more relevant for diagnostic or prognostic AI.
8. The sample size for the training set
- Sample Size for Training Set: Not applicable and not mentioned. This device is a treatment planning system, not a machine learning model that requires a training set in the conventional sense.
9. How the ground truth for the training set was established
- Ground Truth for Training Set Establishment: Not applicable and not mentioned, as there is no machine learning "training set" for this device.
{0}------------------------------------------------
February 8, 2024
Image /page/0/Picture/1 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
Elekta Solutions AB % Simon Sjohage Regulatory Affairs Engineer Kungstensgatan 18 Box 7593 Stockholm, Stockholm SE-10393 SWEDEN
Re: K232854
Trade/Device Name: Leksell GammaPlan® (LGP) Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system. Regulatory Class: Class II Product Code: IWB, MUJ Dated: September 13, 2023 Received: September 15, 2023
Dear Simon Sjohage:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lora Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K232854
Device Name
Leksell GammaPlan® (LGP)
Indications for Use (Describe)
Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning.
Type of Use (Select one or both, as applicable) | Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Elekta, a company that specializes in medical technology. The logo consists of a circular graphic with three dots of varying sizes inside, followed by the word "Elekta" in a sans-serif font. The graphic and the text are both in a teal color.
510(K) SUMMARY (21 CFR § 807.92)
| I. | SUBMITTER | Elekta Solutions ABKungstensgatan 18 Box 7593Stockholm, Stockholms län [SE-01] SE SE1039 |
|---|---|---|
| Contact: | Simon SjöhageRegulatory Affairs EngineerPhone: +46 (8) 58725637 | |
| Establishment Registration #: | 3015232217 | |
| 510(k) Number: | K232854 | |
| Date Prepared: | 13 September 2023 | |
| II. | DEVICE | |
| Trade Name: | Leksell GammaPlan® (LGP) | |
| Release Version: | Release 11.4 | |
| Product Classification: | Class II | |
| Common Name: | Radionuclide RT Treatment planning system | |
| Regulation Number: | 21 CFR § 892.5750 | |
| Regulation Description: | Radionuclide radiation therapy system | |
| Product Code: | IWB, MUJ | |
| III. | PREDICATE DEVICE | Leksell GammaPlan® (K173791) |
IV
IV. INTENDED USE / INDICATIONS FOR USE
Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning.
V. DEVICE DESCRIPTION
Leksell GammaPlan® is a powerful, computer-based treatment planning system specifically designed for the simulation and planning of stereotactic Leksell Gamma Knife® radiosurgery based on tomographic and projectional images.
The basis of treatment planning is the acquisition and processing of digital images by a computer workstation running the treatment planning application software. The program is capable of handling a range of different imaging modalities. Images from tomographic sources such as Computer Tomography (CT), Magnetic Resonance (MR) and Positron Emission Tomography (PET) scanners can be used as well as projectional images from angiograms (AI). This allows the direct comparison between vascular structures in projectional images and tissue structures in CT and MR.
Digital images can be imported into the system via the computer network.
The treatment planning application has the ability to plan a patient's treatment protocol based on a single target or multiple targets.
The basic elements of treatment planning are:
- defining the cranial target or targets .
- . devising the configuration of the collimators to be used during treatment
- determining the parameters of the radiation shots to be delivered by Leksell • Gamma Knife®.
{4}------------------------------------------------
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE
The major changes in this release, as compared to the previously cleared version, are primarily the ability to compare alternative plans simultaneously and the possibility to import DICOM RT Doses from external DICOM compliant systems. The fundamental technical characteristics and the principles of operation of the subject device Leksell GammaPlan version 11.4 have not changed and are substantially equivalent to its predicate device Leksell GammaPlan K173791.
VII. SUMMARY OF PERFORMACE TESTING (NON-CLINICAL)
Development, verification, and validation activities for the modified SaMD were carried out in accordance with design controls as required by FDA's Quality System Requlation (21 CFR §820.30), applicable ISO 13485 Quality Management System requirements, ISO 14971 Risk Management requirements, and IEC 62304 requirements for software life-cycle processes. Non-clinical testing was performed to evaluate device performance and functionality in accordance with design and risk management requirements at subsystem, integration and system levels including interoperability.
Documentation of software development and verification testing activities for LGP is maintained in accordance with FDA's "Guidance for the Content of Premarket Submissions for Device Software Functions," June 2023, for an Enhanced Documentation Device (Class C per IEC 62304).
Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new and existing features against requirement specifications.
Formal design and usability validation has been performed on a clinically equivalent device by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs. The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety for the new functionality was effective. Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification and user needs have been met and that the system is confident and stable.
VIII. SUMMARY OF PERFORMACE TESTING (CLINICAL)
No animal or clinical tests were performed to establish substantial equivalence with the predicate device. The performance data demonstrate that the Leksell GammaPlan is as safe and effective and performs as well as the predicate device.
IX. SUBSTANTIAL EQUIVALENCE CONCLUSION
Leksell GammaPlan v11.4 is substantially equivalent (SE) to the predicate device. Leksell GammaPlan (K173791). The intended use and indications for use are identical to the predicate device and the principles of operation remain unchanged.
The technological characteristics are substantially equivalent to the predicate device; enhancements to existing functionality do not affect the fundamental scientific technology or raise different questions of safety or effectiveness of the device. The device safety and performance have been addressed by non-clinical testing in conformance with predetermined performance criteria, FDA quidance, and recognized consensus standards.
The results of verification and validation as well as conformance to relevant safety standards demonstrate that LGP v11.4 meets the established safety and performance criteria and is substantially equivalent to the predicate device.
N/A