K Number
K232845

Validate with FDA (Live)

Date Cleared
2023-11-14

(61 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IPL Cooling Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.

Device Description

The subject device IPL Cooling Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The IPL emission activation is by finger switch. The device contains a Xenon lamp and its built-in skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the treated skin, the device will not emit the light pulse. Also, the subject device is equipped with cooling function to cool down the skin temperature and give a user a better experience, but this function has no other therapeutic use.

AI/ML Overview

The provided text is a 510(k) Pre-market Notification for an IPL Cooling Hair Removal Device. It details the device's characteristics and compares it to a predicate device to establish substantial equivalence. However, it does not contain the specific information required to complete the table and answer all the questions about acceptance criteria and the study proving the device meets those criteria.

The document primarily focuses on:

  • Device Description: What the device is and how it works (IPL technology for hair reduction).
  • Comparison to Predicate Device: Technological characteristics like wavelength range, energy density, spot size, pulse duration, and indications for use.
  • Performance Data: This section outlines types of testing performed (Biocompatibility Safety, Electrical Safety and EMC Safety, Eye Safety, Software Verification and Validation) and states that the tests passed.

Missing Information:

The document explicitly states that "Performance data supports that the device is safe and as effective as the predicate device for its intended use." However, it does not provide quantitative acceptance criteria or the specific numerical results of those tests (reported device performance) for comparison. It only lists the types of tests performed and the standards they adhered to.

Specifically, the document does not contain any information regarding clinical performance studies, human reader studies, or the establishment of ground truth for such studies. The performance data provided are related to electrical safety, biocompatibility, eye safety, and software validation, which are typically assessed against published standards rather than through clinical effectiveness studies with explicit acceptance criteria for diagnostic or treatment accuracy.

Therefore, I cannot fully complete the table or answer questions 2-9 based on the provided text.

Here's how I can address what is available and highlight the missing parts:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Biocompatibility: Conformance to ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Skin Sensitization)Passed: Tests performed and passed in accordance with ISO 10993-5 and ISO 10993-10 for patient-directly contacting components.
Electrical Safety & EMC Safety: Conformance to IEC 60601-1-2, IEC 60601-1, IEC 60601-1-11, IEC 60601-2-83Passed: Tests performed and passed against specified IEC standards.
Eye Safety: Conformance to IEC 62471Passed: Test performed and passed against IEC 62471.
Software Verification and Validation: Software documentation consistent with moderate level of concern; all software requirement specifications met; all software hazards mitigated to acceptable risk levels.Passed: System validation testing demonstrated all software requirement specifications met and hazards mitigated.
Clinical Effectiveness for Hair Removal: (This information is missing from the document for the subject device. The document only states it's "as effective as the predicate device" without providing specific acceptance criteria or results from a clinical study for the subject device.)(Specific performance data for hair removal effectiveness is not provided in this document.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document describes engineering and biocompatibility tests, not clinical studies with patient test sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided as no clinical effectiveness study with human assessment of ground truth is detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided as no clinical effectiveness study requiring expert adjudication is detailed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not provided. The device is an IPL hair removal device, not an AI-assisted diagnostic tool, so an MRMC study is not applicable in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not provided. This is not an AI algorithm, but a physical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the safety and performance tests mentioned (biocompatibility, electrical safety, eye safety, software), the "ground truth" is defined by the technical standards (e.g., ISO, IEC). For clinical effectiveness (hair removal), no specific ground truth methodology is described for the subject device, although it is asserted to be "as effective as" the predicate.

8. The sample size for the training set
This information is not provided as it's not an AI/machine learning device requiring a training set.

9. How the ground truth for the training set was established
This information is not provided as it's not an AI/machine learning device requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

November 14, 2023

Shenzhen Siken 3D Technology Development Co., Ltd. % Rain Yip Registration Engineer Feiying Drug and Medical Consulting Technical Service Group Rm 2401 ZhenYe International Business Center, No. 3101-90, Qianhai Road, Nanshan District Shenzhen, Guangdong 518000 China

Re: K232845

Trade/Device Name: IPL Cooling Hair Removal Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: September 6, 2023 Received: September 14, 2023

Dear Rain Yip:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to. 21 CFR 820.30. Design controls: 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha L. Hithe Tanisha L. s Hithe -S 2023.11.14 08:21:23 -05'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232845

Device Name

IPL Cooling Hair Removal Device, model: SKB-1808, SKB-2308, SKB-2128N, SKB-2188N

Indications for Use (Describe)

IPL Cooling Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

"510(k) Summary" as required by 21 CFR Part 807.92.

K232845

Date: 2023-10-12

I. Submitter

Shenzhen Siken 3D Technology Development Co., Ltd. 4th Building,No.33 Left East Avenue,Songgang, Bao'an District, Shenzhen, Guangdong, China Post code: 518105 Tel .: +86 755 2769 7523

Jingquan Liu R&D director Tel .: +86 150 1255 5680 Email: liudirector@siken3d.com

II. Device

Name of Device: IPL Cooling Hair Removal Device Model(s): SKB-1808, SKB-2308, SKB-2128N, SKB-2188N Common or Usual Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

III. Predicate Device

Predicate device: 510(k) number: K230097 Manufacturer: Shenzhen BSX Technology Electronics Co., Ltd. Trade name: IPL Hair Removal Device, BSXT101 Product code: OHT Approval date: April 6, 2023

IV. Device Description

The subject device IPL Cooling Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.

{5}------------------------------------------------

The IPL Cooling Hair Removal Device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and its built-in skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the treated skin, the device will not emit the light pulse. Also, the subject device is equipped with cooling function to cool down the skin temperature and give a user a better experience, but this function has no other therapeutic use.

There are SKB-1808, SKB-2128N, and SKB-2188N four models in this application. Their work principle, function, intended use, structure, and composition are the same, with differences being product appearance, and wavelength range, also their energy outputs are slight differences, but these parameters are within the predicate device.

V. Indications for Use

IPL Cooling Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.

VI. Comparison of Technological Characteristics With the Predicate Device

The subject device IPL Cooling Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the subject device may be found substantially equivalent to its predicate device.

ComparisonElementsSubject DevicePredicate device K230097
K NumberPendingK230097
Trade nameIPL Cooling Hair RemovalDevice, SKB-1808, SKB-2308,SKB-2128N, SKB-2188NIPL Hair Removal Device,BSXT101
Wavelength rangeSKB-1808: 5901200nmSKB-2308: 5301200nmSKB-2128N: 5501200nmSKB-2188N: 5501200nm470-1200nm
Energy mediumXenon Arc FlashlampXenon Arc Flashlamp
Energy densitySKB-1808: Max. 4.8J/cm² (rangein 2.14.8J/cm²)SKB-2308: Max. 5.2J/cm² (rangein 1.35.2J/cm²)SKB-2128N: Max. 4.4J/cm² (rangein 1.64.4J/cm²)SKB-2188N: Max. 4.4J/cm² (rangein 1.64.4J/cm²)Max.5.0J/cm²

The subject device is compared with the following Predicate Device in terms of intended use, design, specifications, and performance:

{6}------------------------------------------------

ComparisonElementsSubject DevicePredicate device K230097
Spot size$3\text{cm}^2$$(3\pm0.5)\text{cm}^2$
Pulse duration$10\pm5\text{ms}$4~10ms
Pulsing controlFinger switchFinger switch
Delivery deviceDirect illumination to tissueDirect illumination to tissue
Indication for use/Intended useIPL Cooling Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair.
Location for useOTCOTC

VII.Performance Data

The following performance data were provided in support of the substantial equivalence determination.

1) Biocompatibility Safety

The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the "Use of International Standard ISO 10993-1. 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document issued on Sep. 4, 2020", as recognized by FDA. The battery of testing was performed to, and passed, including:

  • ISO 10993-5 Biological Evaluation of Medical Devices -Part 5: Tests for In Vitro Cytotoxicity

  • ISO 10993-10 Biological Evaluation of Medical Devices –Part 10: Tests for Irritation and Skin Sensitization

2) Electrical Safety and EMC Safety

Electrical safety and Eye safety testing was performed to, and passed, the following standards:

  • IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility

  • IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment

  • IEC 60601-2-83 Medical Electrical Equipment Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment

3) Eye Safety

  • IEC 62471 Photobiological safety of lamps and lamp systems

4) Software Verification and Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Summary

{7}------------------------------------------------

Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device.

VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL Cooling Hair Removal Device is to be concluded substantial equivalent to its predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.