K Number
K232814

Validate with FDA (Live)

Date Cleared
2023-12-05

(83 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Electronic Blood Pressure Monitor intended for use in measuring blood pressure and pulse rate in adult patient population. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the arm. It can be used by medical professionals or at home. The cuff circumference is limited to 22-42 cm.

Device Description

Electronic Blood Pressure Monitor mainly consist of the main body (include screen display, air tube connector, memory button and start/stop button), cuff, USB cable, air tube, and AA batteries. Electronic Blood Pressure Monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The radial artery in the arm changes from blocked to open as the pressure in the cuff tied around the arm changes from high to low, causing the pressure in the cuff to be superimposed on a series of small pressure pulses. The sphygmomanometer senses these signals and, after certain calculations, finds the systolic and diastolic pressures of the radial artery in the body. Electronic Blood Pressure Monitor can be divided into three models (YJ320, YJ321E, YJ326E) according to their appearance and functions.

AI/ML Overview

The provided text is for an FDA 510(k) clearance for an Electronic Blood Pressure Monitor. It details non-clinical and clinical testing performed to demonstrate substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document provides a comprehensive table of non-clinical performance and safety acceptance criteria, along with the reported "Conclusion" (Pass/Fail). For the clinical validation, it states the acceptance criteria are met based on deviations from a reference standard.

Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):

Inspecting ItemAcceptance Standard (Simplified)Reported Device Performance (Conclusion)
1. Identification requirementMeets IEC 60601-1:2005 in 7.2.Pass
2. Life SpanAfter at least 10,000 full scale cycles, the sphygmomanometer should still meet safety and performance requirements. A full scale cycle: pressure rises from <= 2.67kPa (20mmHg) to max, then falls to <= 2.67kPa (20mmHg).Pass
3. Safety Requirements3.1 Maximum cuff pressure: Overpressure protection function: vent valve opens when display exceeds 39.33kPa (295mmHg), pressure reduces to < 2kPa (15mmHg) within 10s. Cuff pressure above 2kPa (15mmHg) for no more than 3 minutes.Pass
3.2 Venting: Simple/clearly marked measure to deflate cuff. Pressure drops from 34.67kPa (260mmHg) to 2kPa (15mmHg) in <= 10s when valve fully open.Pass
4. Performance Index4.1 Range: At least 0kPa (0mmHg) to 39.3kPa (295mmHg).Pass
4.2 Resolution ratio: Display resolution 0.1kPa/1mmHg.Pass
4.3 Repeatability: Max difference between repeated readings at each point <= 0.533kPa (4mmHg) for static continuous low pressure. All readings comply with 4.4.Pass
4.4 Pressure sensor accuracy: Max error of cuff pressure measurement <= +/- 0.4kPa (+/- 3mmHg) at any point, increasing or decreasing pressure.Pass
4.5 Pulse: (a) Pulse measurement range (40 ~ 199) times/min, resolution 1 time/min. (b) Pulse accuracy: +/- 5%.Pass
5. Requirements for charging sources and pressure control valves5.1 Aeration source: Enough air within 10s to reach 40kPa (300mmHg) in a 200 (12 cubic inch) container.Pass
5.2 Pressure controlled air valve: 5.2.1 Air leakage: Max pressure drop <= 0.133kPa (1mmHg) for 10s from 33.33kPa (250mmHg), 20kPa (150mmHg), and 6.67kPa (50mmHg) in a container <= 80 volume. 5.2.2 Valve/cuff bleed rate: Pressure drop rate from 33.33kPa (250mmHg) to 6.67kPa (50mmHg) >= 0.267kPa/s (2mmHg/s). 5.2.3 Venting: Rapid venting from 34.67kPa (260mmHg) to 2kPa (15mmHg) in <= 10s.Pass
6. Cuff with air bag6.1 Size: Air bag length ~0.8 * limb circumference, width ~0.5 * length.Pass
6.2 Compression resistance: Cuff, air bag, and pipeline withstand max expected internal pressure.Pass
6.3 Cuff interface, structure: After 1,000 opening/closing & 10,000 40kPa (300mmHg) pressure cycles, closure/sealing remain intact to meet other standard requirements.Pass
7. System air leakageRate of pressure drop from air leakage <= 0.133kPa/s (1mmHg/s).Pass
8. Function8.1 Display function: LCD displays systolic, diastolic BP, pulse rate, with "kPa" or "mmHg" units.Pass
8.2 Automatic zeroing function: Automatically returns to zero after turn on, automatically opens for BP test.Pass
8.3 Error prompt function: Displays incorrect indication if fails to measure BP/pulse rate correctly.Pass
8.4 Low voltage prompt function: Battery symbol "☐" flashes when battery voltage <= 4.2V +/- 0.2V.Pass
8.5 Automatic shutdown function: Auto-shutdown if no operation within 2 minutes after measurement.Pass
8.6 Memory function: Stores 2*99 groups of data for users A and B. Clears data via [Start/Stop] + [memory/read] for 3s (displays CLR).Pass
8.7 Clock setting function: Displays/sets time (year, month, day, hour, minute).Pass
8.8 Arrhythmia prompt function: Indicates irregular heartbeat with "♥" symbol.Pass
8.9 Voice broadcast prompt function: Voice broadcast in "mmHg" unit display mode.Pass
8.10 Cuff detection prompt function: Displays " "[check] symbol when cuff connected correctly, " "[error] symbol if exception.Pass
9. Appearance and structure9.1: Shape correct, surface bright, clean, color uniform.Pass
9.2: Characters and symbols clear, accurate, firm.Pass
9.3: Function keys flexible, reliable, fasteners not loose.Pass
9.4: Cuff not damaged.Pass
9.5: No missing strokes in LCD display.Pass
10. Electrical safety requirementsMeets IEC 60601-1.Pass
11. Electromagnetic compatibilityMeets IEC60601-1-2.Pass
12. Environmental requirementsMeets IEC 60601-1-11.Pass

Clinical Validation Acceptance Criteria (Simplified, based on ISO 81060-2:2018):

For the clinical study, the acceptance criteria are implicit in the statement: "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018."

  • Reported Device Performance: The device (YJ320) met these limits, indicating it performed within the required accuracy for blood pressure measurement against a reference standard.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size:
    • Clinical Data: 86 adult subjects (49 females, 37 males).
    • Non-Clinical Data: Not explicitly stated as a "test set" in terms of subject count, but performance was evaluated based on the device itself and its components against established standards.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study participants or if it was retrospective or prospective. Given it's a 510(k) submission from a Chinese company, common practice would be for the clinical study to be prospective and likely conducted in China or a region where the standard is recognized.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Clinical Ground Truth: The clinical validation was conducted according to ISO 81060-2:2018, which is "Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type." This standard outlines specific methods for clinical validation, typically involving simultaneous measurements by trained observers (auscultation) against the automated device being tested. The number of experts (observers) involved in establishing the reference measurement (ground truth) is usually prescribed by this standard, but not explicitly stated in this summary. The standard generally requires at least two trained observers. Their qualifications would typically involve being trained and validated to perform accurate auscultatory blood pressure measurements.
  • Non-Clinical Ground Truth: For the non-clinical tests, the "ground truth" is defined by the technical specifications and performance limits set by the referenced international standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 10993, IEC 80601-2-30). Test equipment and calibrated reference standards are used by technical personnel to verify compliance.

4. Adjudication method for the test set

  • Clinical Data: The "Same Arm Sequential Method" was used as per ISO 81060-2:2018. This method involves taking sequential measurements from the same patient's arm using both the test device and a reference method (typically auscultation by trained observers). The standard itself dictates how discrepancies are handled and how the mean error and standard deviation of differences are calculated, which serves as the "adjudication" against the standard's limits. No explicit "expert adjudication" process like 2+1 or 3+1 is mentioned, as the standard relies on the statistical agreement of measurements.
  • Non-Clinical Data: Not applicable in the context of expert adjudication for defining a "ground truth" in the same way as clinical or image-based studies. Tests are performed against objective criteria and measured by calibrated equipment.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or assistive devices where human interpretation is involved. This device is an automated blood pressure monitor, not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, the clinical validation study (ISO 81060-2:2018) is essentially a standalone performance assessment of the automated blood pressure monitor against a reference standard. The device's algorithm performs the measurement autonomously, and its output (systolic and diastolic pressure) is compared directly to the ground truth.

7. The type of ground truth used

  • Clinical Ground Truth: The ground truth for blood pressure measurements in the clinical study was established by comparison with a reference method as defined by ISO 81060-2:2018. This standard typically refers to invasive measurements or, more commonly for non-invasive validation, simultaneous auscultatory measurements performed by trained observers. The document also states, "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018," which confirms adherence to this specific standard for ground truth establishment.
  • Non-Clinical Ground Truth: The "ground truth" for non-clinical tests (e.g., accuracy, range, safety features) is based on the objective technical specifications and performance requirements outlined in referenced international standards, tested using appropriate calibrated measuring equipment.

8. The sample size for the training set

  • This device is an Electronic Blood Pressure Monitor, which likely uses traditional signal processing and algorithms (oscillometric principle) rather than deep learning or AI that requires a "training set" in the modern machine learning sense. Therefore, a distinct "training set" of data for algorithm development, as seen in AI/ML medical devices, is not referenced or applicable here. The algorithms are based on established physiological principles and engineering.

9. How the ground truth for the training set was established

  • As explained in point 8, the concept of a "training set" as it applies to establishing ground truth for machine learning models is not relevant to this traditional medical device. The underlying algorithms are based on established physical and biological principles, and their accuracy is demonstrated through the clinical and non-clinical validation studies described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 5, 2023

Zhuhai Yueja Medical Device Technology Co.,Ltd. Feng Yan Department Manager Room 201, Building 2, No.2 Liushi Road, Tangjiawan Town, High-tech Zone Zhuhai, Guangdong 519000 China

Re: K232814

Trade/Device Name: Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 30, 2023 Received: September 13, 2023

Dear Feng Yan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232814

Device Name Electronic Blood Pressure Monitor

Indications for Use (Describe)

Electronic Blood Pressure Monitor intended for use in measuring blood pressure and pulse rate in adult patient population. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the arm. It can be used by medical professionals or at home. The cuff circumference is limited to 22-42 cm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Section 3 510(k) Summary

l 510(k) Submitter

Device Submitter: Zhuhai Yueja Medical Device Technology Co.,Ltd. Room 201, Building 2, No.2 Liushi Road, Tangjiawan Town, High-tech Zone, Zhuhai, 519000, China

Contact Person: Feng yan Department Manager Phone: +86-18826906029 E-mail: fengy@myzr.com.cn

II Device

Trade Name of Device:Electronic Blood Pressure Monitor
Regulation Number:21 CFR 870.1130
Classification Name:Electronic Blood Pressure Monitor
Product Code:DXN
Regulation Number:Electronic Blood Pressure Monitor
Regulatory ClassII
Review PanelCardiovascular

III Predicate Devices

510k NumberK222994
Trade Name of Device:Electronic Blood Pressure Monitor
Regulation Number:21 CFR 870.1130
Regulation Name:Electronic Blood Pressure Monitor
Regulatory ClassII
Product Code:DXN

IV Device Description

Electronic Blood Pressure Monitor mainly consist of the main body (include screen display, air tube connector, memory button and start/stop button), cuff, USB cable, air tube, and AA batteries.

Electronic Blood Pressure Monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The radial artery in the arm changes from blocked to open as the pressure in the cuff tied around the arm changes from high to low, causing the pressure in the cuff to be superimposed on a series of small pressure pulses. The sphygmomanometer senses these signals and, after certain calculations, finds the systolic and diastolic pressures of the radial artery in the body.

Electronic Blood Pressure Monitor can be divided into three models (YJ320、YJ321E、

{5}------------------------------------------------

YJ326E) according to their appearance and functions. The differences among these models are shown as below:

ModelAir pumps and airvalvesCuffsSoftware FunctionHardware
YJ320SameSameSameBlack and whitebacklit screen
YJ321ESameSameRemoved cuff detectionand misoperation promptfunctionsBlack and whitebacklit screen
YJ326EDifferentDifferentSameColor backlitscreen

V Indications for use

Electronic Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the arm. lt can be used by medical professionals or at home. The cuff circumference is limited to 22-42 cm.

VI Technological Characteristics Comparison

VI-1: Comparison of Electronic Blood Pressure Monitor

DeviceCharacteristicSubject DevicePredicate Device(K222994)Discussionn
Type of BloodPressureMonitorElectronic Blood PressureMonitorYJ320、YJ321E、YJ326EElectronic Blood PressureMonitorAOJ-33 series-
Product CodeDXNDXNSame
Regulation No.21 CFR 870.113021 CFR 870.1130Same
ClassIIIISame
Indications forUseElectronic Blood PressureMonitor is a digital monitorintended for use inmeasuring blood pressureand pulse rate in adultThe Arm Blood PressureMonitor is intended to measurethe systolic and diastolic bloodpressure as well as the pulserate of adult person viaSame
patient population. It is non-invasive oscillometric intended to measure the technique in which an inflatable diastolic, systolic blood CUFF is wrapped around the pressures and pulse rate upper arm at medical facilities or through an inflatable cuff at home. wrapped around the arm. It can be used by medical professionals or at home. The cuff circumference is limited to 22-42 cm.
Measurement siteUpper armUpper armSame
Patient PopulationAdultAdultSame
Measurement ItemSYS, DIA, Pulse RateSYS, DIA, Pulse RateSame
PrincipleOscillometricOscillometricSame
Blood pressure measurement rangepressure:0 to 295mmHgSYS: 55 to 255 mmHgDIA: 25 to 200 mmHg30 ~ 255 mmHgDifferent 1
Accuracy$\pm$ 3 mmHg$\pm$ 3 mmHgSame
Pulse Rate Range40~199beat/min40-199 bpmSame
Accuracy$\pm$ 5% of reading$\pm$ 5% of readingSame
Cuff size22~42cm22~42cmSame
Power supply4 "AA" batteries(d.c.6v) orUSB-C INPUT 5V d.c~6V d.c/600mALithium-ion battery, D.C. 3.7VDifferent 2
Operation condition+5°C~+40°C,15%RH85%RH80kPa105kPa+5°C~+40°C15%RH-90%RH70 kPa -106 kPaDifferent 3
Storage condition-20°C~+55°C,15%RH85%RH,80kPa105kPa-20°C~+55°C10%RH-93%RH70 kPa -106 kPaDifferent 4
Patient ContactingSurface-contacting,Less than 24 hSurface-contacting,Less than 24 hSame
Biocompatibility evaluationCytotoxicity, skin sensitization and irritationCytotoxicity, skin sensitization and irritationSame
ElectricalSafetyIEC 60601-1IEC 60601-1-11ISO 80601-2-30IEC 60601-1IEC 60601-1-11ISO 80601-2-30Same
EMCIEC 60601-1-2IEC 60601-1-2Same
BiocompatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-23ISO 10993-1ISO 10993-5ISO 10993-10Different 5
Clinical dataISO 81060-2: 2018ISO 81060-2: 2018Same

{6}------------------------------------------------

{7}------------------------------------------------

Different 1- Blood pressure measurement range

The Blood Pressure Range of the subject device includes the blood pressure range of predicate devices and meets the requirements of IEC 80601-2-30. Therefore, the difference will not affect the safety and effectiveness of the subject device.

Different 2- Power supply

The difference between subject device and predicate device (K222994) is the batteries, the predicate device use Lithium-ion battery, D.C. 3.7V while the subject device use 4 "AA" batteries (d.c.6v) or USB-C INPUT 5V d.c-6V d.c/600mA, but the power supply safety of subject device is justified by the IEC60601-1 electricity test reports. Thus this difference will not affect the safety and effectiveness, therefore this item is considered as substantial equivalence.

Different 3- Operation condition

The difference between subject device and predicate device (K222994) is the humidity and Atmospheric pressure, the predicate device is 15%RH-90%RH, 70 kPa -106 kPa, while the subject device is 15%RH-85%RH, 80kPa~105kPa, it is included in predicate device. Therefore, this item is considered as substantial equivalence.

Different 4- Storage condition

The difference between subject device and predicate device (K222994) is the humidity

Section 3-4

{8}------------------------------------------------

K232814

and Atmospheric pressure, the predicate device is 15%RH-90%RH, 70 kPa -106 kPa, while the subject device is 15%RH-85%RH, 80kPa~105kPa, it is included in predicate device. Therefore, this item is considered as substantial equivalence.

Different 5- Biocompatibility

The predicate device (K222994) was tested by ISO 10993-5, ISO 10993-10, and the testing part included the battery.

While the main part of subject device that comes into contact with the patient is the cuff, the part that comes into contact with the operator is mainly the cover and cuff. The material of cover is ABS, which is a commonly used material in the market. It has Limited contact with intact skin. Per FDA's 2020 guidance document, Use of International Standard ISO 10993-1, ABS materials are exempt from biological testing. In addition, the cuff has tested by ISO 10993-5 and ISO 10993-10.

Based on the above description, these differences do not affect the achievement of the product's intended use, nor do they compromise the safety of patients and operators.

{9}------------------------------------------------

VII Summary of Non-Clinical Testing and Clinical Testing

Non-Clinical Data:

Non-clinical testing for Electronic Blood Pressure Monitor was performed to demonstrate verification testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below.

Inspecting itemInspection content and acceptance standardConclusion
1. Identification requirementMeets the requirements of IEC 60601-1:2005 in 7.2Pass
2. life spanAfter at least 10,000 full scale cycles, the sphygmomanometer should still meet the safety requirements and performance requirements in the standard.Pass
A full scale cycle is when the pressure rises from 2.67kPa(20mmHg) or less to a maximum pressure value and then falls to 2.67kPa(20mmHg) or less.
3. Safety requirementsThe sphygmomanometer should have overpressure protection function. When the pressure display exceeds 39.33kPa(295mmHg), the sphygmomanometer's vent valve should be opened, and the gas path pressure should be reduced to less than 2kPa(15mmHg) within 10s. In addition, the equipment shouldPass
3.1 Maximum cuff pressure
4.Performanceindex
ensure that the cuff pressure isabove 2kPa(15mmHg) for no morethan 3 minutes.
3.2 ventingThe sphygmomanometer shouldprovide a simple and clearlymarked measure allowing the userto deflate the cuff. When the valveof the charging system is fully openand quickly deflated, the time forthe pressure to drop from34.67kPa(260mmHg) to2kPa(15mmHg) should not exceed10s.Pass
4.1 RangeA blood pressure monitor has arange of at least 0kPa(0mmHg) to39.3kPa(295mmHg).Pass
4.2 ResolutionratioThe display resolution of thesphygmomanometer should be0.1kPa/1mmHg.Pass
4.3RepeatabilityFor a sphygmomanometer,measured at static continuous lowpressure, the difference betweenrepeated readings measured ateach point within the scale rangeshould be no greater than 0.533kPa(4mmHg). All readings shallcomply with the requirements in 4.4Pass
4.4 PressuresensoraccuracyThe maximum error of the pressuremeasurement in the cuff should be$\pm$ 0.4kPa( $\pm$ 3mmHg) at anymeasuring point in the range,regardless of whether the pressureis increased or decreased.Pass
(a) The pulse measurement rangePass
4.5 Pulseof the sphygmomanometer should
be (40 ~ 199) times/min, and the
resolution should be 1 time/min.
(b) Pulse accuracy: ±5%.
5.1 AerationsourceThe sphygmomanometer inflationPass
source should provide enough air
within 10 seconds to reach
40kPa(300mmHg) in a 200(12
cubic inch) container.
5Requirementsfor chargingsources andpressurecontrolvalves.5.2 Pressurecontrolled airvalve5.2.1 air leakagePass
The sphygmomanometer valve is
closed, and the maximum pressure
drop in a container with a volume
not exceeding 80 at initial
pressures of 33.33kPa(250mmHg),
20kPa (150 MMHG) and
6.67kPa(50mmHg) should not
exceed 0.133kPa(1mmHg) for 10s.
5.2.2 Valve/cuff bleed rate
When the valve is in the pressurecontrol position, the pressure droprate from 33.33kPa(250mmHg) to6.67kPa(50mmHg) should not beless than 0.267kPa/s (2mmHg/s);5.2.3 VentingThe rapid venting of a gas-filled
system when the valve is fully openshould not take longer than 10s forthe pressure to drop from34.67kPa(260 mmHg) to2kPa(15mmHg).
6 Cuff with airbag6.1 SizeThe length of the cuff air bag isabout 0.8 times the circumferenceof the limb measured at the medianline of the cuff's intended use area,and the width of the cuff air bag isabout half the length;Pass
6.2compressionresistanceThe cuff, the integrated air bag andthe entire pipeline should be able towithstand the internal pressureequal to the maximum pressureexpected to be used by the cuff;Pass
6.3 Cuffinterface,structureAfter 1,000 opening and closingcycles and 10,00040kPa(300mmHg) pressure cycles,the closure and sealing off the cuffand the integrated airbag shallremain intact enough to meet theother requirements of this standard.Pass
7 System air leakageThe rate of pressure drop causedby air leakage of the entire systemof the sphygmomanometer shouldnot be greater than0.133kPa/s(1mmHg/s).
8 function8.1 displayfunctionThe liquid crystal display of thesphygmomanometer should beable to display systolic blood
pressure, diastolic blood pressureand pulse rate, and there should be"kPa" or "mmHg" blood pressuredisplay form.
8.2 AutomaticzeroingfunctionThe sphygmomanometer should beable to automatically return to zeroafter each turn on, andautomatically open to test bloodpressure.Pass
8.3 ErrorpromptfunctionThe sphygmomanometer shoulddisplay an incorrect indication if itfails to measure blood pressurecorrectly or fails to measure pulserate correctly.Pass
8.4 LowvoltagepromptfunctionWhen the sphygmomanometer isturned on, when the battery voltageis low to 4.2V±0.2V, the batterysymbol "☐" should flash on thedisplay screen.Pass
8.5 AutomaticshutdownfunctionThe sphygmomanometer willautomatically shut down if nooperation is performed within 2minutes after the measurement iscompleted.Pass
8.6 memoryfunctionThe sphygmomanometer canmemory store 2*99 groups ofmeasurement data, 99 groups ofusers A and B, in the memory state,press the [Start/Stop] key and the[memory/read] key at the sametime, 3 seconds later the screen willdisplay CLR, release the doublekey can clear all the memory data.Pass
8.7 ClocksettingfunctionThe sphygmomanometer should beable to display time and set time,including year, month, day, hourand minute.Pass
8.8The sphygmomanometer should bePass
Arrhythmiapromptfunction8.9Voicebroadcastpromptfunctionable to indicate the irregularheartbeat of the human body andcan display the "♥" symbol.In the "mmHg" unit display mode,sphygmomanometer shouldthehave a voice broadcast promptfunction.Pass
Cuff8.10detectionpromptfunctionThe sphygmomanometer has cuffdetection function, and the display" "screen displays "symbolcuff is connectedthewhencorrectly, If an exception occurs,4 " symbol is displayedthe "Pass
9.1 Sphygmomanometer shape should be correct,the surface should be bright, clean, color should beuniform.9.2 The characters and symbols of the bloodPassPass
9 Appearanceand structurepressure monitor should be clear, accurate and firm.9.3 The function keys of the sphygmomanometershould be flexible and reliable, and the fastenersshould not be loose.Pass
9.4 The cuff should not be damaged.Pass
9.5 There should be no missing strokes in the LCDdisplay process.Pass
Electrical10safetvrequirementsThesphygmomanometer shallmeet the requirements specified inIEC 60601-1.Pass
11compatibilityelectromagneticsphygmomanometer shouldThemeet the requirements specified inIEC60601-1-2 Medical electrical1-2:Generalequipment-Part - Partrequirements for basic safety andperformance-CollateralessentialPass
standard: Electromagneticcompatibility-Requirements andtests.
12 EnvironmentalrequirementsThe environmental test of thesphygmomanometer shall complywith IEC 60601-1-11 Part1-11:General requirements forbasic safety and essentialperformance-Collateral Standard:Requirements for medical electricalequipment and medical electricalsystems used in the homehealthcare environment.Pass

Table VII-1: Performance testing was conducted on the subject device

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

Biocompatibility Testing

The main part of subject device that comes into contact with the patient is the cuff, the part that comes into contact with the operator is mainly the cover and cuff. The material of cover is ABS, which is a commonly used material in the market. It has Limited contact with intact skin. Per FDA's 2020 guidance document, Use of International Standard ISO 10993-1, ABS materials are exempt from biological testing.

In addition, the material of cuff is Nylon, it has Limited contact with intact skin and has tested by ISO 10993-5 and ISO 10993-10.

  • Cytotoxicity
  • Skin Sensitization
  • Skin Irritation

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Electronic Blood Pressure Monitor.

{16}------------------------------------------------

The device complies with the IEC60601-1. IEC60601-1-2. IEC 60601-1-8 and IEC60601-1-11.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.

Performance testing

The Electronic Blood Pressure Monitor were test according to the standard which is IEC 80601-2-30, and the cuff to monitor connector was tested based on IEC 80369-5:2016 + C1:2017, the test result meets the requirements.

Usability

Usability was conducted on the Electronic Blood Pressure Monitor. The device complies with the IEC60601-1-6.

Clinical data:

YJ320 was tested to ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. the clinical validation data on YJ320 can cover YJ321E、YJ326E. The Same Arm Sequential Method was chosen and performed on YJ320. This study included 86 adult subjects (49 females, 37 male) with an age range of 17 to 76 years. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018. No adverse effect and/or complication is found in this study.

VIII Conclusion

The conclusions drawn from the non-clinical tests and clinical test demonstrate that the Electronic Blood Pressure Monitor is as safe as effective, and performs as well as or better than the legally marketed device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).