K Number
K232762

Validate with FDA (Live)

Date Cleared
2023-10-03

(25 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The dS Breast Coil 7ch 1.5T is intended to be used in conjunction with a Philips Prodiva 1.5T CX and a MR 5300 1.5T Magnetic Resonance Scanner to produce diagnostic images of the breast, chest wall, and axillary tissues that can be interpreted by a trained physician.

Device Description

The dS Breast Coil 7ch 1.5T is a 7-element phased array receive only coil to be used on a 70cm bore Philips 1.5T MR Systems. The coil is designed to work in unison with the Body Coil of the MRI system, which will transmit the radio frequency (RF) signals, so that the coil may receive the resultant RF signal from the excited nuclei. The subject coil is designed to be used in conjunction with Philips Prodiva 1.5T CS and 1.5T CX MR System (K173507) and with Philips MR 5300 System (K212673) with dStream interfaces. The subject coil is designed for optimum coverage and high-resolution visualization of detailed cartilage structures of the breast anatomy. The coil is used independently and cannot be combined with any other coils. The coil is only available in 1.5T version. The open coil housing design of the dS Breast Coil 7ch 1.5T allows for lateral-medial and cranial-caudal access to the breast for both diagnostic breast imaging and interventional exams.

AI/ML Overview

The device in question is the dS Breast Coil 7ch 1.5T.

Based on the provided text, the acceptance criteria and the study that proves the device meets them are outlined as follows:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Non-Clinical Performance Testing (aligned with FDA-recognized consensus standards and FDA guidance documents):The device underwent and met standards for:
- Electrical/Mechanical Safety (IEC 60601-1)Testing conducted.
- EMC Immunity, Electrostatic Discharge Testing (IEC 60601-1-2)Testing conducted.
- Basic Safety and Essential Performance of Magnetic Resonance Equipment (IEC 60601-2-33)Testing conducted.
- Image Uniformity and Signal-to-Noise Ratio (NEMA-MS-1.3.9.14)Testing conducted. Acknowledged as sufficient quality for diagnostic use by clinical experts (see below).
- Determination of Essential Image Quality Parameters (IEC62464-1)Testing conducted. Acknowledged as sufficient quality for diagnostic use by clinical experts (see below).
- Biological Safety Evaluation (ISO 10993-1)Testing conducted.
- Cleaning and Disinfection Validations (ISO 17664)Testing conducted to support reprocessing instructions.
Clinical Performance Testing (Acquired Image Quality):Images produced on the subject coil were confirmed to be of sufficient quality for diagnostic use.
Substantial Equivalence (Safety and Performance Based Pathway) as outlined in the guidance "Magnetic Resonance (MR) Receive-only Coil - Performance Criteria for Safety and Performance Based Pathway" (December 11, 2020).All criteria in the guidance were met, demonstrating that the device is as safe, as effective, and performs as well as or better than the predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not explicitly stated in terms of the number of patient images or specific cases. The clinical performance testing mentions "images produced on the subject coil."
  • Data Provenance: Not explicitly stated as retrospective or prospective, nor the country of origin. However, the clinicians involved were "U.S. Board Certified radiologists."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: Two.
  • Qualifications of Experts: U.S. Board Certified radiologists. No specific years of experience are mentioned.

4. Adjudication Method for the Test Set:

  • The text states, "Acquired Image quality was assessed by two U.S. Board Certified radiologists to confirm images produced on the subject coil are sufficient quality for diagnostic use." This implies a consensus or agreement was reached between the two readers, but the explicit adjudication method (e.g., 2+1, 3+1, majority vote, etc.) is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a MRMC comparative effectiveness study was not done. The study focuses on confirming the diagnostic quality of images produced by the new coil, rather than comparing human reader performance with and without AI assistance, or comparing human readers with another device. This is a breast coil, not an AI diagnostic software.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Not applicable. The dS Breast Coil 7ch 1.5T is a hardware device for image acquisition, not an algorithm or AI software for interpretation. Its performance is evaluated on the quality of the images it produces for human interpretation.

7. Type of Ground Truth Used:

  • For the clinical performance assessment, the "ground truth" was established by expert consensus (assessment by two U.S. Board Certified radiologists) on the "sufficient quality for diagnostic use" of the images produced by the coil. This is effectively expert opinion regarding image quality for diagnostic purposes.

8. Sample Size for the Training Set:

  • Not applicable. The dS Breast Coil 7ch 1.5T is a passive hardware device (receive-only coil) and does not involve AI or machine learning algorithms that require a "training set" in the conventional sense. The "training" for the device would be its engineering design and manufacturing process.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

October 3, 2023

Invivo Corporation (Business Trade Name: Philips) % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K232762

Trade/Device Name: dS Breast Coil 7ch 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: September 8, 2023 Received: September 8, 2023

Dear Prithul Bom:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

DlyK

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232762

Device Name dS Breast Coil 7ch 1.5T

Indications for Use (Describe)

The dS Breast Coil 7ch 1.5T is intended to be used in conjunction with a Philips Prodiva 1.5T CX and a MR 5300 1.5T Magnetic Resonance Scanner to produce diagnostic images of the breast, chest wall, and axillary tissues that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K232762)

The 510(k) Summary was prepared in accordance with 21 CFR

§807.92(c).
Preparation date:August 17, 2023
510(k) Owner:Philips Medical Systems Nederland B.V.Veenpluis 6
5684 PC Best
The Netherlands
Establishment Registration #3003768277
Contact person:Liselotte Kornmann (primary)
Associate Director Regulatory Affairs
Philips Medical Systems Nederland B.V.
Phone: +31 611621238
Trade Name:dS Breast Coil 7ch 1.5T
Classification Name:Coil, Magnetic Resonance, Specialty
Regulation Number:21 CFR 892.1000
Review Panel:Radiology
Device Class:Class II
Product Code:MOS
Predicate Device Trade Name:Biopsy Breast Coil BBC (K032576 - October 23, 2003)
Predicate Classification Name:Coil, Magnetic Resonance
Predicate Regulation Number:21 CFR 892.1000
Predicate Device Class:Class II
Predicate Product Code:MOS

{4}------------------------------------------------

Device description

The dS Breast Coil 7ch 1.5T is a 7-element phased array receive only coil to be used on a 70cm bore Philips 1.5T MR Systems. The coil is designed to work in unison with the Body Coil of the MRI system, which will transmit the radio frequency (RF) signals, so that the coil may receive the resultant RF signal from the excited nuclei. The subject coil is designed to be used in conjunction with Philips Prodiva 1.5T CS and 1.5T CX MR System (K173507) and with Philips MR 5300 System (K212673) with dStream interfaces. The subject coil is designed for optimum coverage and high-resolution visualization of detailed cartilage structures of the breast anatomy. The coil is used independently and cannot be combined with any other coils. The coil is only available in 1.5T version. The open coil housing design of the dS Breast Coil 7ch 1.5T allows for lateral-medial and cranial-caudal access to the breast for both diagnostic breast imaging and interventional exams.

Indications for use

The dS Breast Coil 7ch 1.5T is intended to be used in conjunction with a Philips Prodiva 1.5T CS, Prodiva 1.5T CX, and a MR 5300 1.5T Magnetic Resonance Scanner to produce diagnostic images of the breast, chest wall, and axillary tissues that can be interpreted by a trained physician.

Fundamental Scientific Technology:

Based on the information provided above, the subject dS Breast Coil 7ch 1.5T is considered substantially equivalent to the primary predicate device Biopsy Breast Coil BBC (K032576 - October 23, 2003) in terms of fundamental design, material and scientific technology. At a high level the dS Breast Coil 7ch 1.5T and the predicate coil are based on the following equivalent elements:

  • Same Indications for use .
  • Prescription Use Only
  • Anatomy of interest is the breast .
  • Same magnetic field strength (1.5T) .
  • Rigid housing design that allows each imaging element to be independently . positioned and configured for each patient
  • . Energy source from the MR scanner
  • Designed to support access to the breast anatomy for both diagnostic breast . imaging and interventional exams

The following technological differences exist between the subject and predicate device:

  • The design of the housing is more simplified than the marketed primary predicate device
  • The subject device is compatible with Philips Prodiva and MR 5300 MR systems, . whereas the predicate is compatible with the Ingenia family of MR systems.
  • . There is no 3.0T version of the subject breast coil.

Clinical and non-clinical testing demonstrates that the safety and effectiveness requirements as outlined in FDA guidance Magnetic Resonance (MR) Receive-only Coil - Performance Criteria for Safety and Performance Based Pathway, issued December 11, 2020 were met. No new safety or efficacy concerns are raised as a result of these differences.

{5}------------------------------------------------

Summary of Non- Clinical and Clinical Performance Data:

The subject dS Breast Coil 7ch 1.5T has undergone the following testing in accordance with FDA-recognized consensus standards and as recommended in FDA guidance documents Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016 and Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway, issued December 11, 2020:

Performance Testing - Non-Clinical:

  • IEC 60601-1 General electrical/mechanical safety -
  • IEC 60601-1-2 EMC Immunity, electrostatic discharge testing -
  • IEC 60601-2-33 Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis
  • -NEMA-MS-1. 3. 9. 14 Image uniformity and signal-to-noise ratio testing
  • IEC62464-1 International Standard: Magnetic resonance equipment for medical imaging -Part 1: Determination of essential image quality parameters
  • -ISO 10993-1 Biological safety evaluation
  • ISO 17664 Cleaning and disinfection validations to support reprocessing instructions -

Performance Testing - Clinical:

Acquired Image quality was assessed by two U.S. Board Certified radiologists to confirm images produced on the subject coil are sufficient quality for diagnostic use.

Substantial Equivalence Conclusion:

Substantial equivalence of the dS Breast Coil 7ch 1.5T is demonstrated through the Safety and Performance Based Pathway for magnetic resonance (MR) receive-only coils.

The subject device has the same indications for use and technological characteristics as the predicate device. Substantially equivalent performance is demonstrated by meeting all criterion in the quidance "Magnetic Resonance (MR) Receive-only Coil -Performance Criteria for Safety and Performance Based Pathway" issued on December 11, 2020.

The summary includes the conclusions drawn from the nonclinical and clinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device according to 807.92(b)(3).

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.