K Number
K232756

Validate with FDA (Live)

Device Name
Colibrí
Manufacturer
Date Cleared
2023-12-27

(110 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.

The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/lonization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.

The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

Bacterial suspensions for AST and purity plates are identified by barcode label.

The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.

The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria.

Device Description

The Colibrí is an instrument which automates the picking of selected colonies from plated media and prepares MALDI target slides for the bioMérieux VITEK MS systems or the Bruker MALDI Biotyper CA systems that are used in clinical laboratories for identification and differentiation of organisms grown on plated media by Matrix-Assisted Laser Desorption/Jonization Time-of Flight Mass Spectrometry (MALDI-TOF MS). The Colibri automates the preparation of microbial suspensions at known concentration for bioMérieux VITEK 2 systems and Beckman Coulter MicroScan WalkAway systems that are used in clinical laboratories for AST analyses. Moreover, the Colibrí is used for Purity Plates preparation for purity assessments.

The Colibrí includes the following components:

  • Colibrí instrument and software with on-board pipetting system and nephelometer .
  • Colibrí Primary Tubes
  • Colibrí Spreader
  • Colibrí Daily Verification kit.

Colibri is designed to be used in conjunction with the WASPLab device for culture plate incubation and image analysis. After appropriate plate incubation, the operator selects the colonies from a digital image of culture media plate streaked with microbiological human specimen, available through WebApp software, the WASPLab User Interface.

The operator assigns the automatic ID or AST tasks to the isolated colonies to be processed. Then, the operator loads the plates in the Collbri where colonies are automatically picked, spotted on the target slide and overlayed with the matrix or suspended into the dedicated solution for the preparation of the microbial suspension for AST purposes (Secondary Tube).

When used in conjunction with the bioMérieux VITEK MS systems, the Colibri can prepare the 48-spot target slides by performing the direct spotting of colonies. The calibrator used for quality control is manually applied by the operator at the end of the automated colony spotting. When used in conjunction with the Bruker MALDI Biotyper CA systems, the Colibri can prepare either reusable 48-spot or disposable 96-spot targets by performing the Direct Transfer Sample Procedure. The BTS used for quality control is manually applied by the operator at the end of the automated colony spotting.

When used in conjunction with the bioMérieux VITEK 2 systems or the Beckman Coulter MicroScan WalkAway systems, the Colibri can prepare the microbial suspension at the proper concentration by direct colony suspension method. The onboard nephelometer allows the preparation of Secondary Tubes (AST suspensions) at the correct concentration and the Colibri Spreader is used for Purity Plates preparation.

The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

The traceability of prepared Secondary Tube and Purity Plates is maintained by dedicated labels applications.

Colibrí requires four different calibrations, one on the nephelometer, three on the cameras. None of these calibration activities require user intervention if not in terms of periodical cleaning of the mechanical component as described in the dedicated section of the User Manual. The Set-up calibration of nephelometer and camera units are performed during the device initial setup. Auto-calibration is performed at the end of the initial set-up and periodically during the preventive maintenance to check that all the mechanical references can be found inside the positioning tolerances, that the I/Os are responsive. Runtime calibration is performed during the normal usage to automatically check the proper functioning of the Colibrí.

Colibrí requires a daily nephelometer verification to check the proper reading of suspensions at different turbidity values.

AI/ML Overview

The Colibrí device is an automated in vitro diagnostic specimen preparation system. The provided text describes the acceptance criteria and the study that proves the device meets these criteria for preparing microbial suspensions for Antimicrobial Susceptibility Testing (AST) using Beckman Coulter MicroScan WalkAway systems.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance:

The acceptance criteria are implicitly derived from the successful outcomes of the analytical studies. The performance is reported as the percentage of successful outcomes for each metric.

MetricAcceptance Criteria (Implied)Reported Device Performance
Preparation of Microbial Suspensions for ASTMicrobial concentration within acceptable limits: - E. coli ATCC 25922: 3-7 x 105 CFU/mL - Other bacteria: 2-8 x 105 CFU/mL98.5% of prepared suspensions had microbial concentration within acceptable limits. - E. coli: 100% (36/36) - Pseudomonas aeruginosa: 96.7% (29/30) - Staphylococcus aureus: 97.6% (41/42) - Enterococcus faecalis: 100% (30/30)
AST Challenge Test (Agreement with Manual Preparation)Essential Agreement (EA) of MICs: High agreement Category Agreement (CA): High agreement Discrepancies (vmj, maj): Low/noneOverall EA: 100% (1232/1232 evaluable MIC results within 1 two-fold dilution) Overall CA: 98.4% (4187/4254 SIR categorizations in agreement) Very Major discrepancy (vmj): 0 Major discrepancy (maj): 0
Reproducibility StudyBest-case reproducibility: ≥95% (implied) Worst-case reproducibility: ≥89% (implied)Best-case reproducibility: ≥99.8% (all panels combined) Worst-case reproducibility: ≥94.3% (all panels combined)
Sample preparation for Quality Control100% of MIC values within CLSI/panel IFU QC range100% (all tested organisms and antimicrobial agents)
Purity Plates Evaluation (Cross-contamination)Absence of cross-contamination (100% monomicrobial growth)100% (453/453 Purity Plates showed monomicrobial growth)

2. Sample sizes used for the test set and the data provenance:

  • Preparation of Microbial Suspensions for AST:

    • Test Set Size: 132 microbial suspensions (36 E. coli, 30 Pseudomonas aeruginosa, 42 Staphylococcus aureus, 30 Enterococcus faecalis).
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study involved three Colibrí instruments, suggesting internal validation. Retrospective or prospective is not specified, but the nature of the validation suggests prospective testing.
  • AST Challenge Test:

    • Test Set Size: Different species: Enterobacterales (n=50 isolates), Staphylococcus (n=20 isolates), Streptococcus (n=12 isolates), Enterococcus (n=18 isolates), non-fermenters (n=10 isolates). Each processed by three Colibrí instruments, yielding varying numbers of MIC results and SIR categorizations across different panels (e.g., 2454 total MIC results for Enterobacterales on NM-NF50 panel).
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study design implies prospective testing within the manufacturer's validation process.
  • Reproducibility Study:

    • Test Set Size: 9 gram-positive and 9 gram-negative strains, processed on 3 Colibrí instruments over 3 days, with each condition tested in triplicate (total of 27 replicates for each strain-antimicrobial agent combination).
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). Implied prospective.
  • Sample preparation for Quality Control:

    • Test Set Size: CLSI-recommended reference strains (Escherichia coli ATCC 25922, Pseudomonas aeruginosa ATCC 27853, Staphylococcus aureus ATCC 29213, Enterococcus faecalis ATCC 29212). The "No. MIC within QC range" indicates a total of 30 or 36 or 42 tests for each drug-organism combination, per three instruments.
    • Data Provenance: Not explicitly stated. Implied prospective.
  • Purity Plates Evaluation:

    • Test Set Size: 453 purity plates (150 from AST Challenge, 162 from AST Reproducibility, 141 from Quality Control studies).
    • Data Provenance: Not explicitly stated. Implied prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The ground truth for AST results (MICs and SIR categories) generally refers to the results obtained from a reference method. In this case, "Manual suspension was used as comparative method" for the AST Challenge Test. This implies that manually prepared suspensions, processed by the MicroScan WalkAway, served as the reference standard.
  • The text does not specify the number of experts or their qualifications for establishing this manual ground truth. It mentions that three different technicians operated the Colibrí machines, but it doesn't detail the personnel for the manual comparative method or for interpreting the results as ground truth beyond the "FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria" and CLSI guidelines.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • The document does not describe any expert adjudication process for the test set results. The comparison is made against a "manual result" (ground truth). The discrepancies (vmj, maj, min) are simply categorized and reported, implying a direct comparison without further expert review for resolving initial disagreements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC study was not conducted. This device (Colibrí) is an automated system for sample preparation and does not involve human "readers" or "AI assistance" in the typical sense of image analysis for diagnosis. Its role is to automate a laboratory process, and the performance is measured against reference methods, not human interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The studies presented are primarily standalone (algorithm only without human-in-the-loop performance) in terms of the Colibrí device's automated functions. The device picks colonies, prepares suspensions, and records data automatically. The performance metrics (inoculum density, MIC accuracy, reproducibility, purity) assess the device's output against established standards and manual methods.
  • While an operator designates colonies for picking, the act of preparation itself is automated and evaluated for its accuracy. The "manual suspension" used for comparison acts as the reference for the "algorithm only" performance of the Colibrí in producing the suspension.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • The ground truth for the analytical studies combines reference methods/standards and established guidelines:
    • For microbial suspension concentration: Viable cell count (CFU/mL) against CLSI and FDA guidelines.
    • For AST results: Manual suspension preparation as the comparative method, and comparison of MICs and SIR categories against FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria and CLSI guideline M07.
    • For reproducibility: Comparison to the "mode result" (most frequent MIC value) and established reproducibility criteria (e.g., within one doubling dilution).
    • For Quality Control: CLSI-recommended QC ranges and MicroScan panel IFU values.
    • For Purity Plates: Visual assessment (implied) to confirm monomicrobial growth.

8. The sample size for the training set:

  • This document describes performance validation studies for a medical device (Colibrí), not a machine learning model. Therefore, there is no "training set" in the context of data used to train an AI algorithm. The Colibrí is an automated instrument with pre-programmed functions, not a learning algorithm that requires a training dataset.

9. How the ground truth for the training set was established:

  • As explained above, there is no training set for this device in the context of AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 27, 2023

Copan WASP S.r.l. Chiara Congiu Regulatory Affairs Manager Via A. Grandi, 32 Brescia, Brescia 25125 Italy

Re: K232756

Trade/Device Name: Colibrí Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON, QQV, QBN Dated: September 8, 2023 Received: September 8, 2023

Dear Chiara Congiu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM) Branch Chief, General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232756

Device Name Colibrí

Indications for Use (Describe)

The Colibri is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of gramnegative and gram-positive bacterial species grown on solid culture media.

The Colibri is an automated processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/lonization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimiorobial Susceptibility Testing and purity assessment.

The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

Bacterial suspensions for AST and purity plates are identified by barcode label.

The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.

The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submitter I.

Applicant Name:Copan WASP Srl
Via A. Grandi 32
25125 Brescia, Italy
+39 030 2687211
copan.regulatory@copangroup.com
Contact PersonChiara Congiu
Copan WASP Srl
Via A. Grandi 32
25125 Brescia, Italy
+39 338 6904942
copan.regulatory@copangroup.com
Establishment Registration Number:3009288740
Date Prepared:September 08, 2023

II. Device Name

Proprietary NameColibrí
Common/Usual NameColibrí
Classification NameClinical mass spectrometry microorganism identifica-tion and differentiation system (21 CFR 866.3378)Fully automated short-term incubation cycle antimicro-bial susceptibility system (21 CFR 866.1645)
Device ClassII
Product CodeQQV, QBN, LON
PanelMicrobiology

{4}------------------------------------------------

III. Legally Marketed Predicate Device

Device NameColibrí
510(K) NumberK223245

No reference Devices were used in this submission.

Device Description IV.

The Colibrí is an instrument which automates the picking of selected colonies from plated media and prepares MALDI target slides for the bioMérieux VITEK MS systems or the Bruker MALDI Biotyper CA systems that are used in clinical laboratories for identification and differentiation of organisms grown on plated media by Matrix-Assisted Laser Desorption/Jonization Time-of Flight Mass Spectrometry (MALDI-TOF MS). The Colibri automates the preparation of microbial suspensions at known concentration for bioMérieux VITEK 2 systems and Beckman Coulter MicroScan WalkAway systems that are used in clinical laboratories for AST analyses. Moreover, the Colibrí is used for Purity Plates preparation for purity assessments.

The Colibrí includes the following components:

  • Colibrí instrument and software with on-board pipetting system and nephelometer .
  • Colibrí Primary Tubes
  • Colibrí Spreader
  • Colibrí Daily Verification kit.

Colibri is designed to be used in conjunction with the WASPLab device for culture plate incubation and image analysis. After appropriate plate incubation, the operator selects the colonies from a digital image of culture media plate streaked with microbiological human specimen, available through WebApp software, the WASPLab User Interface.

The operator assigns the automatic ID or AST tasks to the isolated colonies to be processed. Then, the operator loads the plates in the Collbri where colonies are automatically picked, spotted on the target slide and overlayed with the matrix or suspended into the dedicated solution for the preparation of the microbial suspension for AST purposes (Secondary Tube).

When used in conjunction with the bioMérieux VITEK MS systems, the Colibri can prepare the 48-spot target slides by performing the direct spotting of colonies. The calibrator used for quality control is manually applied by the operator at the end of the automated colony spotting. When used in conjunction with

{5}------------------------------------------------

the Bruker MALDI Biotyper CA systems, the Colibri can prepare either reusable 48-spot or disposable 96-spot targets by performing the Direct Transfer Sample Procedure. The BTS used for quality control is manually applied by the operator at the end of the automated colony spotting.

When used in conjunction with the bioMérieux VITEK 2 systems or the Beckman Coulter MicroScan WalkAway systems, the Colibri can prepare the microbial suspension at the proper concentration by direct colony suspension method. The onboard nephelometer allows the preparation of Secondary Tubes (AST suspensions) at the correct concentration and the Colibri Spreader is used for Purity Plates preparation.

The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

The traceability of prepared Secondary Tube and Purity Plates is maintained by dedicated labels applications.

Colibrí requires four different calibrations, one on the nephelometer, three on the cameras. None of these calibration activities require user intervention if not in terms of periodical cleaning of the mechanical component as described in the dedicated section of the User Manual. The Set-up calibration of nephelometer and camera units are performed during the device initial setup. Auto-calibration is performed at the end of the initial set-up and periodically during the preventive maintenance to check that all the mechanical references can be found inside the positioning tolerances, that the I/Os are responsive. Runtime calibration is performed during the normal usage to automatically check the proper functioning of the Colibrí.

Colibrí requires a daily nephelometer verification to check the proper reading of suspensions at different turbidity values.

V. Intended Use / Indications For Use

The Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.

The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/lonization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.

The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

Bacterial suspensions for AST and purity plates are identified by barcode label.

{6}------------------------------------------------

The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.

The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria.

VI. Comparison to Predicate

According to its intended use, Colibrí supports the performance of the connected IVD Analyzers by facilitating sample preparation for Gram-Negative and Gram-Positive bacteria ID or AST according to their intended use.

This submission aims to demonstrate the suitability of AST microbial suspensions prepared by Colibrí for processing with MicroScan WalkAway analyzers.

Similarities
ItemNew DevicePredicate Device
Device Name(K number)ColibríColibrí (K223245)
Device ClassificationClass II (special controls)Class II (special controls)
Regulation Number21 CFR 866.3378 and 21 CFR 866.164521 CFR 866.3378 and 21 CFR 866.1645
Product CodeQQV, Automated System For Sample Preparation And Identification Of MicroorganismsFrom Cultured Isolates By Mass SpectrometryQBN, Mass Spectrometry, Maldi Tof, Microorganism Identification, Cultured IsolatesLON, System, Test, Automated, Antimicrobial Susceptibility, Short IncubationQQV, Automated System For Sample Preparation And Identification Of MicroorganismsFrom Cultured Isolates By Mass SpectrometryQBN, Mass Spectrometry, Maldi Tof, Microorganism Identification, Cultured IsolatesLON, System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Indications for UseThe Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.Bacterial suspensions for AST and purity plates are identified by barcode label.The Colibrí is intended for use by trained healthcare professionals in clinical laboratoriesThe Colibrí™ is an automated in vitro diagnostic specimen preparation system for use with WASPLab® to prepare MALDI-TOF targets for the bioMérieux VITEK® MS or Bruker MALDI Biotyper® CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK® 2 Antimicrobial Susceptibility Testing (AST) system for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.The Colibrí™ is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.The Colibrí™ software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.Bacterial suspensions for AST and purity plates are identified by barcode label.The Colibrí™ is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory
Similarities
ItemNew DeviceColibríPredicate DeviceColibrí (K223245)
Device Name(K number)in conjunction with other clinical and labora-tory findings, including Gram staining, to aid inthe diagnosis of bacterial infections.The Colibrí has not been validated for use in theidentification or processing of yeast species,molds, Nocardia, or mycobacteria.findings, including Gram staining, to aid in thediagnosis of bacterial infections.The Colibrí™ has not been validated for use inthe identification or processing of yeast species,molds, Nocardia, or mycobacteria.
Method of testingDirect testing from isolated colonies for IDpurposes in conjunction with bioMérieuxVITEK MS systems or Bruker MALDI Bio-typer CA systems.Direct testing from isolated colonies for ASTpurposes in conjunction with bioMérieuxVITEK 2 systems or Beckman Coulter Mi-croScan WalkAway systems.Direct testing from isolated colonies for IDpurposes in conjunction with bioMérieuxVITEK MS systems or Bruker MALDI Bio-typer CA systems.Direct testing from isolated colonies for ASTpurposes in conjunction with bioMérieuxVITEK 2 systems.
Sample/Media TypeIsolated bacterial colonies from any patientsource grown on plates included in K193138,K220546 and K223245.Same
Plate managementAutomatic image capturing management andmanual loading into instrument for picking ac-tivities.Same
Colonies selectionThe colony to be picked is selected by an oper-ator on a digital plate using a Graphical UserInterface on a dedicated workstation.Same
Method of Colony PickingAutomatic picking of colonies using pipettetips.Same
ID Target preparationWhen connected with VITEK MS systems, aportion of microbial colony from an agar plateis automatically spotted on a VITEK MS-DStarget slide (VITEK MS DS Target Slides, 48positions disposable plastic targets) by usingthe pipetting system. 1µL of matrix is automat-ically applied to the spot using the pipettingsystem. The dried target slide is then manuallyloaded into the VITEK MS system.When connected with MALDI Biotyper CAsystems, a portion of microbial colony from anagar plate is automatically spotted on a BrukerTarget Plate (IVD 48 Spot Target plate orMBT Biotarget 96 US IVD) by using the pi-petting system. 1µL of matrix is automaticallyapplied to the spot using the pipetting system.The dried target slide is then manually loadedinto the MALDI Biotyper CA systems.Same
AST Suspension Preparation for VITEK 2Using a pipetting system, a predefined numberof morphologically similar colonies are trans-ferred into Primary Tube containing saline so-lution (0.45% NaCl Saline Solution pH 4.5 to7.0). A homogenous heavy suspension of or-ganisms is prepared and checked by using on-board Colibrí nephelometer. In the SecondaryTube containing 3.0mL of the same saline solu-tion, a variable aliquot of the heavy suspensionis automatically transferred to obtain the finalmicrobial concentration according to IVDpackage insert indications. The suspensionsprepared by Colibrí must be tested inMANUAL MODE on the VITEK 2 systems.Same
CalibrationColibrí requires four different calibrations, oneon the nephelometer, three on the cameras.Same
Similarities
ItemNew DevicePredicate Device
Device Name(K number)ColibríColibrí (K223245)
Preparatory activitiesNone of these calibration activities requireuser intervention.Nephelometer verification by check usingDaily verification Kit.Same
Quality controlCompletely dependent on next-step IVD analyzers.Same

{7}------------------------------------------------

{8}------------------------------------------------

Differences
ItemNew DevicePredicate Device
Device Name(K number)ColibríColibrí (K223245)
AST Suspension Preparation for Beckman CoulterMicroScan WalkAwayUsing a pipetting system, a predefined number of morphologically similar colonies are transferred into Primary Tube containing water or saline solution (0.9% NaCl). A homogenous heavy suspension of organisms is prepared and checked by using an on-board Colibrí nephelometer. In the Secondary Tube containing 25 ml of MicroScan Inoculum Water with PLURONIC or 25 ml of Mueller-Hinton Broth with 3% lysed horse blood, a variable aliquot of the heavy suspension is automatically transferred to obtain the final microbial concentration according to IVD package insert indications. The suspension prepared by Colibrí must be used to rehydrate MicroScan AST panels.Not present

These differences do not affect substantial equivalence of Colibrí and the Predicate Device.

VII. Performance Data

The performance studies which demonstrate the ability of the predicate device Colibrí to prepare MALDI-TOF targets and microbial suspensions for VITEK 2 have been conducted under K193138, K220546 and K223245.

Please, refer to the Decision Summary for K193138 for the following performance studies on the MALDI-TOF targets preparation:

  • Precision/Reproducibility -
  • -Colony Picking Accuracy
  • Accuracy of Bacterial Identification -
  • -Positional Accuracy
  • Carry-Over -
  • Colony Stability -
  • -Spot Stability

Please, refer to the Decision Summary for K220546 for the following performance studies on the microbial suspension preparation for AST:

  • -Precision/Reproducibility of results for VITEK 2
  • Colony Picking Accuracy -

Copan WASP S.r.l., Colibrí - 510(k) Summary

{9}------------------------------------------------

  • Microbial Suspension Accuracy for VITEK 2 -
  • Accuracy of AST Results for VITEK 2 -
  • -Carry-Over
  • -Pipettor Trueness and Precision
  • Accuracy of the Onboard Nephelometer -
  • E. coli Suspensions Preparation Verification Study -
  • Purity Plate Growth -
  • -Reagent Hold Time

Please, refer to the Decision Summary for K223245 for the validation of the full workflow of Colibri in conjunction with the WASPLab.

Analytical Studies

The Analytical studies performed with the Colibrí support its use for the preparation of microbial suspension used in conjunction with the MicroScan WalkAway AST analyzer. The Analytical studies demonstrated that microbial suspensions prepared from isolated colonies by the Colibri can be used to hydrate MicroScan panels for the determination of susceptibility of organisms to certain drugs included on the panels. The used methodology (direct colony suspension) and claimed prerequisites for sample preparation are in line with the IVD analyzer manufacturer IFU.

Preparation of Microbial Suspensions for AST

The accuracy of the inoculum density in the Secondary Tubes was determined by viable cell count of the microbial suspensions prepared by three Colibrí instruments and compared to the acceptance range 3-7 x 105 CFU/mL for E. coli ATCC 25922- and 2-8 x 105 CFU/mL2 for other bacteria.

The percentage of prepared suspension with microbial concentration within the acceptable limits was 98.5% and has been compared between instruments with x-test, resulting in no statistically significant difference among the instruments (p-value = 0.34). The results provided evidence that the microbial concentration of AST suspensions prepared by Colibrí is accurate.

Table 1: Percentage suspensions with acceptable microbial concentration, stratified per microor-
ganism and instrument
MicroorganismAvg. Calcu-lated concen-tration(CFU/mL)Expectedconcentra-tion(CFU/mL)Colibrí#1Colibrí#2Colibrí#3Overall
Escherichia coli4.8 x 1053-7 x 10512/1212/1212/1236/36(100%)

1 Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems

2 CLSI guideline - M07 Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically, 11th Edition

{10}------------------------------------------------

MicroorganismAvg. Calculated concentration (CFU/mL)Expected concentration (CFU/mL)Colibrí #1Colibrí #2Colibrí #3Overall
Pseudomonas aeruginosa$2.9 x 10^5$$2-8 x 10^5$10/109/1010/1029/30 (96.7%)
Staphylococcus aureus$5.7 x 10^5$$2-8 x 10^5$13/1414/1414/1441/42 (97.6%)
Enterococcus faecalis$4.2 x 10^5$$2-8 x 10^5$10/1010/1010/1030/30 (100%)
Total$4.5 x 10^5$$2-8 x 10^5$43/44 (97.7%)43/44 (97.7%)44/44 (100%)130/132 (98.5%)

AST Challenge Test

The accuracy of MICs obtained by MicroScan WalkAway with microbial suspensions prepared by Colibrí from isolated colonies was evaluated using representative isolates of different species of Enterobacterales (n=50), Staphylococcus (n=20), Streptococcus (n=12), Enterococcus (n=18) and non-fermenters (n=10). The strains included in this study were both susceptible and resistant strains, exhibiting a range of on-scale MIC values toward at least 4 antibiotics representative for the major classes of drugs. Each strain was grown on Trypticase Soy Agar + 5% Sheep Blood or MacConkey Agar (only Enterobacterales and non-fermenters) at specific incubation times and then processed by three Colibrí operated by three different technicians. Manual suspension was used as comparative method.

Microbial suspensions were processed by the MicroScan WalkAway using the appropriate antibiotic panel in Table 2.

MicroScan PanelIndicated Organisms GroupsAntimicrobial Agents
MSP6Streptococcus spp.Ampicillin
Cefepime
Cefotaxime
Ceftriaxone
Clarithromycin
Clindamycin
Daptomycin
Erythromycin
Levofloxacin
Linezolid
Meropenem
Penicillin G
Tetracycline
Vancomycin
Table 2: MicroScan Panels Tested in AST Challenge Study
0

{11}------------------------------------------------

MicroScan PanelIndicated Organisms GroupsAntimicrobial Agents
NM-NF50Enterobacterales,Non-fermentersAmikacinGentamicinTobramycinCefepimeCeftazidimeCefotaximeMeropenemImipenemAmpicillin/SulbactamPiperacillin/TazobactamTicarcillin/K ClavulanateAztreonamLevofloxacinCiprofloxacinMinocyclineTetracyclineTrimethoprim/Sulfamethoxazole
PM-E37Enterococcus spp.,Staphylococcus spp.AmpicillinCiprofloxacinClindamycinDaptomycinErythromycinLevofloxacinLinezolidMinocyclineMoxifloxacinNitrofurantoinPenicillin GTetracyclineTigecyclineVancomycin

The SIR category was reported according to the FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria. Essential Agreement (EA) of the MICs and Category Agreement (CA) were calculated in comparison with the manual result. The discrepant SIR results were categorized as Very Major discrepancy (vmj), Major discrepancy (maj) and Minor discrepancy (min).

MICs obtained using Colibrí for microbial suspensions preparation showed very high agreement with the manual preparation for all the microorganisms group; overall, 1232/1232 evaluable MIC results were within 1 two-fold dilution of the reference result and 4187/4254 SIR categorizations were in agreement. The overall Essential Agreement of the evaluable MIC results was 100% and the Category Agreement was 98.4%.

Table 3: AST Challenge Study summary of results, stratified by MicroScan panel

{12}------------------------------------------------

MicroScanpanelOrganismGroupTotaltested# EA% EATotalEvalua-ble# EA ofEvalua-ble% EAof Eval-uableTotalcat.# CA% CA# S# R#vmj#maj#min
NM-NF50Enterobacterales24542454100%577577100%2352230498.0%11979960048
NM-NF50Non-fermenters294294100%165165100%29128296.9%126120009
PM-E37Staphylococcus690690100%157157100%66365698.9%465159007
PM-E37Enterococcus525525100%258258100%51951699.4%354102003
MSP6Streptococcus429429100%7575100%429429100%38139000

Reproducibility Study

The Reproducibility Study was performed to demonstrate consistency of AST results given by Beckman Coulter MicroScan WalkAway system on microbial suspensions prepared by different Colibrí instruments in different test days.

Culture media showing isolated colonies of 9 gram-positive and 9 gram-negative strains were processed on 3 Colibrí over 3 days to prepare microbial suspensions that were tested for purity and loaded in the appropriate antibiotic panel following the MicroScan instructions for use. Each condition was tested in triplicate for a total number of 27 replicates for each combination strain-antimicrobial agent.

The reported MIC values were used to calculate within-instrument and between-instruments reproducibility. Reproducibility was calculated as the total number of results that were within one doubling dilution of the mode result divided by total number of results that were one doubling dilution from the mode were considered reproducible for best-case reproducibility calculations but considered not reproducible for worst-case calculations.

Best-case reproducibility was ≥95% and worst-case reproducibility was ≥89% for all antimicrobial agents tested across all panels with microbial suspensions prepared from isolated colonies by each instrument. The reproducibility results are acceptable.

MicroScan PanelColibríBest Casea (%)Worst Caseb (%)
MSP6Instrument 1117/117 (100%)113/117 (96.6%)
Instrument 2117/117 (100%)110/117 (94.0%)
Instrument 3117/117 (100%)108/117 (92.3%)
Combined351/351 (100%)331/351 (94.3%)
NM-NF50Instrument 1459/459 (100%)453/459 (98.7%)
Instrument 2458/459 (99.8%)453/459 (99.7%)
Instrument 3459/459 (100%)452/459 (98.5%)
Combined1376/1377 (99.9%)1358/1377 (98.6%)
PM-E37Instrument 1297/297 (100%)297/297 (100%)
Instrument 2297/297 (100%)295/297 (99.3%)
Instrument 3297/297 (100%)297/297 (100%)

Table 4: Summary of reproducibility results -stratified by MicroScan panel

Copan WASP S.r.l., Colibrí - 510(k) Summary

{13}------------------------------------------------

MicroScan PanelColibríBest Casea (%)Worst Caseb (%)
Combined891/891 (100%)889/891 (99.8%)

ªCalculated assuming the off-scale results are within one well from the mode. bCalculated assuming the off-scale results are greater than one well from the mode.

Sample preparation for Quality Control

The study was performed to demonstrate the reproducibility of MIC results for AST Quality Control organisms using the Colibrí as suspension preparator. CLSI-recommended reference strains Escherichia coli ATCC 25922, Pseudomonas aeruginosa ATCC 27853, Staphylococcus aureus ATCC 29213 and Enterococcus faecalis ATCC 29212 were grown on trypticase soy agar + 5% sheep blood.

The sample preparation for Quality Control was conducted daily at the beginning of the working session on each instrument involved in the Analytical Studies. All instruments were used to prepare bacterial suspensions then tested using the appropriate AST panel. MIC values for each drug/organism combination were compared to the established ranges reported in the MicroScan panels labeling in 100% acceptable MIC values. Purity of all the suspensions was confirmed by Purity Plates prepared by Colibrí. The summary of results is in the table below.

QC organismAntimicrobial agentQC range(CLSIM1006)On panel val-ues (panelIFU)No. MIC withinQC range
Escherichia coliATCC 25922Amikacin0.5 - 4≤4 - 836/36 (100%)
Ampicillin/Sulbactam2/1 - 8/4≤4/236/36 (100%)
Aztreonam0.06 - 0.25≤0.536/36 (100%)
Ceftazidime0.06 - 0.5≤136/36 (100%)
Cefotaxime0.03 - 0.12≤436/36 (100%)
Ciprofloxacin0.004 - 0.016≤0.1236/36 (100%)
Cefepime0.016 - 0.12≤136/36 (100%)
Gentamicin0.25 - 1≤1 - 236/36 (100%)
Imipenem0.06 - 0.5≤0.536/36 (100%)
QC organismAntimicrobial agentQC range(CLSIM1006)On panel val-ues (panelIFU)No. MIC withinQC range
Levofloxacin0.008 - 0.06≤0.536/36 (100%)
Meropenem0.008 - 0.06≤0.1236/36 (100%)
Minocycline0.25 - 1≤236/36 (100%)
Piperacillin/Tazobactam1/4 - 8/4≤8/436/36 (100%)
Tetracycline0.5 - 2≤236/36 (100%)
Ticarcillin/K Clavulanate4/2 - 16/2≤16/236/36 (100%)
Tobramycin0.25 - 1≤236/36 (100%)
Trimethoprim/Sulfame-thoxazole≤0.5/9.5≤1/1936/36 (100%)
Ciprofloxacin0.12 - 0.5≤0.542/42 (100%)
Clindamycin0.06 - 0.25≤0.2542/42 (100%)
Daptomycin0.12 - 1≤0.25 - 142/42 (100%)
Erythromycin0.25 - 1≤0.25 - 142/42 (100%)
Levofloxacin0.06 - 0.5≤142/42 (100%)
StaphylococcusaureusATCC 29213Linezolid1 - 4≤1 - 442/42 (100%)
Minocycline0.06 - 0.5≤142/42 (100%)
Moxifloxacin0.016 - 0.12≤0.2542/42 (100%)
Nitrofurantoin8 - 32≤3242/42 (100%)
Penicillin G0.25 - 20.25 - >842/42 (100%)
Tetracycline0.12 - 1≤142/42 (100%)
Tigecycline0.03 - 0.25≤0.2542/42 (100%)
Vancomycin0.5 - 20.5 - 242/42 (100%)
Amikacin1 - 4≤4 - 830/30 (100%)
Aztreonam2 - 82 - 830/30 (100%)
Cefepime0.5 - 4≤1 - 430/30 (100%)
Pseudomonas ae-ruginosaATCC 27853Cefotaxime8 - 328 - 3230/30 (100%)
Ceftazidime1 - 4≤1 - 430/30 (100%)
Ciprofloxacin0.12 - 1≤0.12 - 130/30 (100%)
Gentamicin0.5 - 2≤1 - 430/30 (100%)
Imipenem1 - 41 - 430/30 (100%)
QC organismAntimicrobial agentQC range(CLSIM1006)On panel val-ues (panelIFU)No. MIC withinQC range
Levofloxacin0.5 - 4≤0.5 - 430/30 (100%)
Meropenem0.12 - 1≤0.12 - 130/30 (100%)
Piperacillin/Tazobactam1/4 - 8/4≤8/430/30 (100%)
Tetracycline8 - 328 ->830/30 (100%)
Ticarcillin/K Clavulanate8/2 - 32/2≤16/2 - 32/230/30 (100%)
Tobramycin0.25 - 1≤230/30 (100%)
EnterococcusfaecalisATCC 29212Ampicillin0.5 - 2≤130/30 (100%)
Ciprofloxacin0.25 - 2≤0.5 - 130/30 (100%)
Daptomycin1 - 41 - 430/30 (100%)
Erythromycin1 - 41 - 430/30 (100%)
Levofloxacin0.25 - 2≤1 - 230/30 (100%)
Linezolid1 - 4≤1 - 430/30 (100%)
Moxifloxacin0.06 - 0.5≤0.25 - 0.530/30 (100%)
Nitrofurantoin4 - 16≤3230/30 (100%)
Penicillin G1 - 40.5 - 230/30 (100%)
Tetracycline8 - 324 ->830/30 (100%)
Tigecycline0.03 - 0.12≤0.2530/30 (100%)
Vancomycin1 - 41 - 430/30 (100%)

Table 5: List of QC organisms and antimicrobial agent tested

3 C29871-AD MicroScan Gram Positive Procedural Manual, Multi-regional PU-06 panels

4 C29884-AB MicroScan MICroSTREP plus 6 Procedural Manual, Multi-regional

5 C80530-AB MicroScan Gram Negative Procedural Manual, Multi-regional LabPro ≥V5.00 (11-digit biotype)

6 CLSI guideline M100. 33rd ed Performance Standard for Antimicrobial Susceptibility Testing. Wayne, PA: Clinical and Laboratory Standards Institute, 2023.

{14}------------------------------------------------

Copan WASP Colibrí™ - 510(k) Summary

Copan WASP S.r.l., Colibrí – 510(k) Summary

{15}------------------------------------------------

Copan WASP Colibrí™ - 510(k) Summary

  • Due to MicroScan panel design, the full QC range was not available for all organism/drug combinations. For the drugs where the concentrations on the MicroScan panel do not cover the full CLSI/FDA-recommended QC range, the reported value was taken as an indication of acceptable QC.

Purity Plates Evaluation

All the microbial suspensions prepared during the analytical studies were tested for purity by Purity Plate preparation using Colibri. The purity assessment was performed to demonstrate the absence of cross contamination during the preparation of Secondary Tubes and the efficiency of the on-board incinerator in sterilizing the spreader. A total of 453 Purity Plates were prepared resulting in 100% acceptable plates without evidence of growth of other organisms than the target one.

{16}------------------------------------------------

StudyNumber of PurityPlates showing mo-nomicrobial growthNumber of PurityPlates testedPercentage of PurityPlates showing mo-nomicrobial Growth
AST Challenge150150100%
AST Reproducibility162162100%
Quality Control141141100%
Total453453100%

VIII. Non-Clinical and/or Clinical Tests Summary & Conclusions

Conclusions:

The predicate device (existing device) and the new device both utilize the same technology for the preparation of the MALDI-TOF targets and the AST microbial suspensions.

The Analytical Studies results demonstrated that the Colibrí when used in conjunction with MicroScan WalkAway device is as safe, as effective, and performs equivalently to the predicate device. The minor differences between the devices do not adversely affectiveness. The used methodology and claimed prerequisites for sample preparation are in line with the IVD analyzer manufacturer IFU and with the relevant CLSI guidance. Analytical study result summaries have been updated in the package insert to harmonize data presentation between the MicroScan WalkAway and bioMérieux VITEK 2 AST systems.

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”