K Number
K232749

Validate with FDA (Live)

Device Name
DOOPANG
Manufacturer
Date Cleared
2024-06-11

(277 days)

Product Code
Regulation Number
882.5891
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DOOPANG(YPS-301BD) is indicated for:

  • The acute treatment of migraine with or without aura in patients 18 years of age or older.
  • The prophylactic treatment of episodic migraine in patients 18 years of age or older.
Device Description

This product is a small, portable product that uses the principle of transcutaneous nerve stimulation. By attaching an electrode to the forehead, it transmits electrical stimulation to the upper branch of the trigeminal nerve transdermally. The electrode surface transmits electric current to the skin to generate action potentials on the nerves and adjusts the output mode (Program 1 and Program 2) to preventative treatment of episodic migraine(60Hz) and acute treatment of migraine(100Hz). Specially designed electrode of this product allows contact with the trigeminal nerve endings associated with migraine headaches. The electrodes give nerve stimulation to the orbital upper branch (V1) inside the forehead skin and transmit it to the trigeminal nervous system. This electrical stimulation is transmitted percutaneously through a bipolar adhesive electrode located on the forehead to preventative treatment of episodic migraine(60Hz) or acute treatment of migraine(100Hz).

AI/ML Overview

This FDA 510(k) summary (K232749) describes the "DOOPANG (YPS-301BD)" device, a transcutaneous electrical nerve stimulator for headache treatment. The summary demonstrates substantial equivalence to a predicate device, "Cefaly Dual" (K201895).

However, the provided text does not contain information about acceptance criteria for performance metrics, nor does it detail a study that proves the device meets such criteria in terms of clinical effectiveness. This type of device, a transcutaneous electrical nerve stimulator, typically relies on established efficacy of its mechanism and comparative safety and technical specifications rather than a novel clinical performance study with defined acceptance criteria for diagnostic output.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and reported device performance.
  2. Sample size used for the test set and data provenance.
  3. Number of experts used to establish ground truth and their qualifications.
  4. Adjudication method.
  5. MRMC comparative effectiveness study results.
  6. Standalone performance.
  7. Type of ground truth used.
  8. Sample size for the training set.
  9. How ground truth for the training set was established.

The document focuses on demonstrating substantial equivalence in technological characteristics, indications for use, and safety, as opposed to proving a specific clinical performance metric with a dedicated pivotal study.

Here's what the document does provide, structured according to your request where applicable:

1. A table of acceptance criteria and the reported device performance

  • Not explicitly provided in terms of clinical acceptance criteria like sensitivity, specificity, accuracy, or similar outcome measures for a diagnostic device. The comparison focuses on technical specifications to demonstrate equivalence.
FeatureProposed Device (DOOPANG)Predicate Device (Cefaly Dual)Note
Indications for Use- Acute treatment of migraine with or without aura in patients 18 years or older.- Acute treatment of migraine with or without aura in patients 18 years or older.Similar
- Prophylactic treatment of episodic migraine in patients 18 years or older.- Prophylactic treatment of episodic migraine in patients 18 years or older.
Intended PopulationPatients 18 years of age or olderPatients 18 years of age or olderSame
Intended Use EnvironmentHome & Healthcare settingsHome & Healthcare settingsSame
Power SourceRechargeable 3.7V Lithium-Polymer Battery1 rechargeable LiPo 3.7 V batterySame
Channels11Same
Software Programs2 fixed programs: Program 1 (acute migraine, 100Hz), Program 2 (prophylactic migraine, 60Hz)2 fixed programs: Program 1 (acute migraine, 100Hz), Program 2 (prophylactic migraine, 60Hz)Same
Program 1: Max. output current16 mA16 mASame
Program 1: Pulse width250 μs, fixed 100 Hz250 μs, fixed 100 HzSimilar
Program 1: Session durationfixed 61 minutesfixed 60 minutesSlight difference (1 minute), deemed not significant.
Program 2: Max. output current16 mA16 mASame
Program 2: Pulse width250 μs, fixed 60 Hz250 μs, fixed 60 HzSimilar
Program 2: Session durationfixed 21 minutesfixed 20 minutesSlight difference (1 minute), deemed not significant.
WaveformBiphasic, rectangular, symmetricalBiphasic, rectangular, symmetricalSame
Net charge (µC) per pulse00Same
Maximum output current (mA)16 mA16 mASame
Maximum current density (mA/cm² @ 500 ohms)2.322.37Similar, proposed device is slightly lower and considered less risky as it falls within the predicate's performance range.
Maximum average power density (W/cm² @ 500 ohms)0.0000470.000047Same

2. Sample size used for the test set and the data provenance

  • Not applicable / Not provided. The summary focuses on non-clinical testing (electrical safety, EMC, software validation, biocompatibility, cybersecurity, performance, usability) to demonstrate equivalence based on technical specifications and regulatory standards. There is no mention of a clinical test set with patient data for performance evaluation in the context of diagnostic accuracy or therapeutic outcome metrics.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable / Not provided. No such clinical performance study is described.

4. Adjudication method for the test set

  • Not applicable / Not provided. No such clinical performance study is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a direct-to-patient neuromodulation device, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a direct-to-patient neuromodulation device. The software validation mentioned refers to the proper functioning of the embedded software control, not an AI algorithm for clinical interpretation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable / Not provided for clinical performance. For the non-clinical tests, "ground truth" would be defined by the specifications of the standards (e.g., IEC 60601-1 for electrical safety).

8. The sample size for the training set

  • Not applicable / Not provided. No machine learning model or training set is mentioned for clinical performance.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. No machine learning model or training set is mentioned for clinical performance.

Summary of Device Rationale from the Document:

The DOOPANG (YPS-301BD) device demonstrates substantial equivalence to the Cefaly Dual (K201895) predicate device primarily through:

  • Identical Indications for Use: Both devices are indicated for acute treatment and prophylactic treatment of episodic migraine with or without aura in patients 18 years or older.
  • Similar Technological Characteristics: The devices share comparable design, intended population, use environment, power source, channels, software programs (two fixed programs for acute and prophylactic treatment with similar frequencies and pulse widths), waveform, and output parameters.
  • Minor Differences Not Adversely Affecting Safety or Effectiveness: Slight differences in session duration (1 minute) and maximum current density are considered insignificant. The slightly lower current density of the proposed device is even considered less risky.
  • Compliance with Regulatory Standards: The device underwent non-clinical testing for electrical safety (AAMI/IEC 60601-1, 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6, 62366-1), home healthcare environment compatibility (IEC 60601-1-11), software validation, biocompatibility (ISO 10993-1), and cybersecurity. These tests confirm the device meets established safety and performance benchmarks for its classification.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

June 11, 2024

Ybrain, Inc. % Jonghyun Kim CEO, GMS Consulting #612, DE Riverwork Building B, 66, Cheongcho-ro, Deogyang-gu Goyang-si, Gyeonggi-do 10543 Korea. South

Re: K232749

Trade/Device Name: Doopang Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous electrical nerve stimulator to treat headache Regulatory Class: Class II Product Code: PCC Dated: May 24, 2024 Received: May 24, 2024

Dear Jonghyun Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jitendra V. Virani -S

CDR Jitendra Virani Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K232749

Device Name

DOOPANG (YPS-301BD)

Indications for Use (Describe)

The DOOPANG(YPS-301BD) is indicated for:

  • The acute treatment of migraine with or without aura in patients 18 years of age or older.
  • The prophylactic treatment of episodic migraine in patients 18 years of age or older.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

  1. 510(K) Summary

{4}------------------------------------------------

510(k) Summary

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

May 23, 2024

2. Submitter's Information [21 CFR 807.92(a)(1)]

  • Name of Manufacturer: Ybrain, Inc. ●
  • Address: 228, 54, Changeop-ro, Sujeong-gu, Seongnam-si, Gyeonggi-do, Republic of Korea (zip code 13449) ● Kiwon, Lee/President Contact Name:
  • Telephone No.: +82-70-4190-1899
  • Email Address: ra@ybrain.com
  • PCC Registration No.:

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

510(k) NumberK232749
Trade/Device/Model NameDOOPANG/DOOPANG/YPS-301BD
Product NameTranscutaneous Electrical Nerve Stimulator to TreatHeadache
Device Classification NameStimulator, Nerve, Electrical, Transcutaneous, For Migraine
Regulation Number21 CFR 882.5891
Classification Product CodePCC
Device ClassClass II
510(k) Review PanelNeurology

{5}------------------------------------------------

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate device within this submission is shown as follow;

Predicate Device

510(k) NumberK201895
Trade/Device/Model NameCefaly Dual
Product NameTranscutaneous Electrical Nerve Stimulator to TreatHeadache
Device Classification NameStimulator, Nerve, Electrical, Transcutaneous, For Migraine
Requlation Number21 CFR 882.5891
Classification Product CodePCC
Device ClassClass II
510(k) Review PanelNeurology

These predicate devices have not been subject to a design-related recall

{6}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a)(4)]

This product is a small, portable product that uses the principle of transcutaneous nerve stimulation. By attaching an electrode to the forehead, it transmits electrical stimulation to the upper branch of the trigeminal nerve transdermally.

The electrode surface transmits electric current to the skin to generate action potentials on the nerves and adjusts the output mode (Program 1 and Program 2) to preventative treatment of episodic migraine(60Hz) and acute treatment of migraine(100Hz). Specially designed electrode of this product allows contact with the trigeminal nerve endings associated with migraine headaches. The electrodes give nerve stimulation to the orbital upper branch (V1) inside the forehead skin and transmit it to the trigeminal nervous system. This electrical stimulation is transmitted percutaneously through a bipolar adhesive electrode located on the forehead to preventative treatment of episodic migraine(60Hz) or acute treatment of migraine(100Hz).

6. Indications for use [21 CFR 807.92(a)(5)]

The DOOPANG(YPS-301BD) is indicated for:

  • The acute treatment of migraine with or without aura in patients 18 years of age or older.
  • The prophylactic treatment of episodic migraine in patients 18 years of age or older.

7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

There are no significant differences in the technological characteristics of these devices compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the DOOPANG(YPS-301BD) and the predicate device:

Proposed DevicePredicate DeviceNote
K NumberK232749K201895-
ManufacturerYbrain, Inc.CEFALY Technology-
Trade Name(Model Name)DOOPANG(YPS-301BD)Cefaly Dual-
Product NameTranscutaneous ElectricalNerve Stimulator to TreatHeadacheTranscutaneous ElectricalNerve Stimulator to TreatHeadache-
Product CodePCCPCCSame
Regulation Number21 CFR 882.589121 CFR 882.5891Same
Regulatory ClassIIIISame
Proposed DevicePredicate DeviceNote
510(k) Review PanelNeurologyNeurologySame

[Table 1. Comparison of Proposed Device to Predicate Devices]

{7}------------------------------------------------

DOOPANG(YPS-301BD) isThe Cefaly® Dual is
Indications for Useindicated for:- The acute treatment ofmigraine with or withoutaura in patients 18 yearsof age or older.- The prophylactictreatment of episodicmigraine in patients 18years of age or older.indicated for:- The acute treatment ofmigraine with or withoutaura in patients 18 yearsof age or older.- The prophylactictreatment of episodicmigraine in patients 18years of age or older.Similar
Intended PopulationPatients 18 years of ageor olderPatients 18 years of ageor olderSame
Intended Use EnvironmentHome & HealthcaresettingsHome & HealthcaresettingsSame
Power SourceA rechargeable 3.7VLithium-Polymer Battery1 rechargeable LiPo3.7 V batterySame
Channels11Same
Software provided2 fixed programs:- 1 fixed program for theacute treatment ofmigraine attacks(Program1)- 1 fixed program forprophylactic treatment ofmigraine(Program 2)2 fixed programs:- 1 fixed program for theacute treatment ofmigraine attacks(Program1)- 1 fixed program forprophylactic treatment ofmigraine(Program 2)Same
Program 1:
Max. output current16 mA16 mA
Pulse width250 μs,fixed 100 Hz,250 μs,fixed 100 Hz,Similar
Pulse frequencyfixed 61 minutesfixed 60 minutes
Session duration
Program 2:
Max. output current16 mA16 mA
Pulse width250 μs,fixed 60 Hz,250 μs,fixed 60 Hz,Similar
Pulse frequencyfixed 21 minutesfixed 20 minutes
Session durationBiphasic, rectangular,symmetricalBiphasic, rectangular,symmetricalSame
Waveform
Net charge (µC) per pulse00Same
Maximum output current (mA):
16 mA16 mASame
Maximum current density (mA/cm2,r.m.s) at 500 ohms2.322.37Similar
Maximum average power density(W/cm2, r.m.s) at 500 ohms0.0000470.000047Same

It is substantially equivalent to these devices in design, function, and intended use. The proposed device, DOOPANG(YPS-301BD) has been tested about electrical safety, EMC, and performance, and the software has been validated. Differences and Risks associated with that:

  • There appears to be a slight difference in session duration between Program 1 and Program . 2 for both the predicate device and the proposed device. However, this discrepancy is not

{8}------------------------------------------------

expected to pose any significant additional risk, as the variation is within a minute or less.

  • The Maximum current density (mA/cm2, r.m.s) at 500 ohms of the proposed device is 2.32, . while the predicate device has a value of 2.37. Although there is a slight difference, it can be arqued that the performance of the proposed device is less risky because it falls within the performance range of the predicate device.
    There are no significant differences between the DOOPANG(YPS-301BD) and the predicate device that would adversely affect the product's use according to the test results. Therefore, these differences do not raise different questions of safety and effectiveness than the predicate.

8. Non-Clinical Test summary

The DOOPANG(YPS-301BD) complies with voluntary standards for electrical safety,

electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:

  1. Electrical Safety, Electromagnetic Compatibility and Performance:

The DOOPANG(YPS-301BD) complies with the electrical safety and electromagnetic compatibility requirements established by the standards.

StandardsNo.StandardsOrganizationStandard TitleVersionPublicationYear
60601-1AAMI/IECMedical Electrical Equipment - Part 1: GeneralRequirements for Basic Safety and EssentialPerformance (IEC 60601-1:2005, MOD)ES60601-1:2005(R)2012 andA1:20122014
60601-1-2IECMedical Electrical Equipment - Part 1-2: GeneralRequirements for Safety – Collateral Standard:Electromagnetic Compatibility - Requirementsand Tests60601-1-2Edition 4.02014-022016
60601-1-6IECMedical electrical equipment - Part 1-6: Generalrequirements for basic safety and essentialperformance - Collateral standard: Usability60601-1-6Edition 3.02012-072012
60601-1-11IECMedical electrical equipment - Part 1-11: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Requirementsfor medical electrical equipment and medicalelectrical systems used in the home healthcareenvironment60601-1-11Edition 2.02015-012016
60601-2-10IECMedical electrical equipment - Part 2-10:Particular requirements for the basic safety andessential performance of nerve and musclestimulators60601-2-10Edition 2.12016-042018
62366-1IECMedical devices - Part 1: Application of usabilityengineering to medical devices62366-1Edition 1.12020-06CONSOLIDATEDVERSION2020

{9}------------------------------------------------

2) Software Validation

The DOOPANG(YPS-301BD) contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance:

  • The content of premarket submissions for software contained in medical devices, on May 11, ● 2005
    1. Biocompatibility
  • ISO 10993-1 and series, Biological evaluation of medical devices
    1. Cybersecurity
  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, on October2, 2014
    1. Label
  • CFR Part 801
    1. Performance
  • Performance Evaluation Test
  • Real Time Stability Test
    1. Usability
  • . Usability Test (IEC 62366-1:2015+AMD1:2020 CSV)

9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

There are no significant differences between the proposed device and the predicate device, K201895 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.

10. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the DOOPANG(YPS-301BD) is substantially equivalent in safety and effectiveness to the predicate device as described herein.

§ 882.5891 Transcutaneous electrical nerve stimulator to treat headache.

(a)
Identification. A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.