K Number
K232742

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-11-03

(57 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

· Shoulder: Rotator cuff repair, Bankart repair, biceps tenodesis, acromio-clavicular separation repair, deltoid repair, capsular shift or capsulolabral reconstruction

· Foot/Ankle: Lateral stabilization, medial stabilization, achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, and digital tendon transfers

· Knee: Medial collateral ligament repair, lateral ligament repair, patellar tendon, posterior oblique ligament repair, iliotibial band tenodesis, and joint capsule closure, medial patellofemoral ligament repair/reconstruction, and anterolateral ligament repair/reconstruction

• Hand/Wrist: Scapholunate ligament reconstructions, repair/reconstructions, repair/reconstruction of collateral ligaments digital tendon transfers, and carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)

· Elbow: Biceps tendon reattachment, ulnar/radial collateral ligament reconstruction, and lateral epicondy.itis repair

· Hip: Acetabular labral repair and gluteal tendon repair

Device Description

The Arthrex FiberTak suture anchor is an 'all-suture' soft-tissue device intended to be used for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip. The anchor is constructed from a hollow braid of polyester with a single loaded suture component composed of UHWMPE or a polyblend of UHMWPE and polyester. The anchor is pre-loaded on a disposable inserter and will be sold sterile for single-use.

AI/ML Overview

The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone. The submission is to expand indications of the 2.6 mm FiberTak Suture Anchor to include Anterolateral Ligament Repair/Reconstruction.

Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Pull-out (tensile) strength of the proposed Arthrex FiberTak Suture Anchor should meet criteria established by published literature for anterolateral ligament repair/reconstruction.Mechanical testing demonstrated that the pull-out (tensile) strength of the proposed Arthrex FiberTak Suture Anchor met the criteria established by published literature for anterolateral ligament repair/reconstruction.

2. Sample Size for the Test Set and Data Provenance:

The document does not explicitly state the sample size (number of anchors tested) for the mechanical testing. It also doesn't specify the country of origin of the data or whether the study was retrospective or prospective. It only mentions "mechanical testing."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This is not applicable as the study involved mechanical testing against established criteria from published literature, not expert human assessment to establish a "ground truth" for diagnostic or prognostic purposes.

4. Adjudication Method for the Test Set:

Not applicable, as this was mechanical testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. This type of study is not relevant for a mechanical device like a suture anchor. The assessment focuses on the physical properties of the device, not its performance in interpretation by human readers.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

Not applicable. This device is a physical medical implant, not an AI algorithm. Its performance is inherent in its mechanical properties.

7. Type of Ground Truth Used:

The ground truth for the acceptance criteria was "criteria established by published literature for anterolateral ligament repair/reconstruction." This implies a consensus or widely accepted standard derived from previous research on the biomechanical requirements for such repairs.

8. Sample Size for the Training Set:

Not applicable. As this is a mechanical medical device, there is no "training set" in the context of machine learning or AI models.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. There is no training set for this type of medical device.

{0}------------------------------------------------

November 3, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Arthrex Inc. Kristi Frisch Regulatory Affairs Specialist, Principal 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K232742

Trade/Device Name: Arthrex FiberTak Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: September 6, 2023 Received: September 7, 2023

Dear Kristi Frisch:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Jesse Muir Digitally signed by
Jesse Muir Jesse Muir -S -5 Date: 2023.11.03 11:13:44 -04'00' Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232742

Device Name Arthrex FiberTak Suture Anchor

Indications for Use (Describe)

The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

· Shoulder: Rotator cuff repair, Bankart repair, biceps tenodesis, acromio-clavicular separation repair, deltoid repair, capsular shift or capsulolabral reconstruction

· Foot/Ankle: Lateral stabilization, medial stabilization, achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, and digital tendon transfers

· Knee: Medial collateral ligament repair, lateral ligament repair, patellar tendon, posterior oblique ligament repair, iliotibial band tenodesis, and joint capsule closure, medial patellofemoral ligament repair/reconstruction, and anterolateral ligament repair/reconstruction

• Hand/Wrist: Scapholunate ligament reconstructions, repair/reconstructions, repair/reconstruction of collateral ligaments digital tendon transfers, and carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)

· Elbow: Biceps tendon reattachment, ulnar/radial collateral ligament reconstruction, and lateral epicondy.itis repair

· Hip: Acetabular labral repair and gluteal tendon repair

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Date PreparedSeptember 6, 2023
SubmitterArthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945
Contact PersonKristi Frisch
Regulatory Affairs Specialist, Principal
Tel 239-598-4302 x73849
Kristi.Frisch@Arthrex.com
Trade NameArthrex FiberTak Suture Anchor
Common NameSuture Anchor
Product CodeMBI
Classification Name21 CFR 888.3040: Smooth or threaded metallic bone fastener
Regulatory ClassII
Primary Predicate DeviceK231113: Arthrex FiberTak Suture Anchor
Reference Device(s)K221364 MITEK TIGHT-N™ Tendon Docking Anchor
Purpose of SubmissionThis Special 510(k) premarket notification is submitted to expand
indications of the 2.6 mm FiberTak Suture Anchor to include
Anterolateral Ligament Repair/Reconstruction.
Device DescriptionThe Arthrex FiberTak suture anchor is an 'all-suture' soft-tissue device
intended to be used for fixation of suture (soft tissue) to bone in the
shoulder, foot/ankle, knee, hand/wrist, elbow, and hip.
The anchor is constructed from a hollow braid of polyester with a single
loaded suture component composed of UHWMPE or a polyblend of
UHMWPE and polyester.
The anchor is pre-loaded on a disposable inserter and will be sold sterile
for single-use.
Indications for UseThe Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
Shoulder: Rotator cuff repair, Bankart repair, SLAP lesion repair, biceps tenodesis, acromio-clavicular separation repair, deltoid repair, capsular shift or capsulolabral reconstruction Foot/Ankle: Lateral stabilization, medial stabilization, achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, and digital tendon transfers Knee: Medial collateral ligament repair, lateral collateral ligament repair, patellar tendon, posterior oblique ligament repair, iliotibial band tenodesis, joint capsule closure, medial patellofemoral ligament repair/reconstruction, and anterolateral ligament repair/reconstruction Hand/Wrist: Scapholunate ligament reconstruction, carpal ligament reconstructions, repair/reconstruction of collateral ligaments digital tendon transfers, and carpometacarpal joint arthroplasty (basal thumb joint arthroplasty Elbow: Biceps tendon reattachment, ulnar/radial collateral ligament reconstruction, and lateral epicondylitis repair Hip: Acetabular Labral Repair and gluteal tendon repair

{5}------------------------------------------------

Performance DataMechanical testing demonstrated that the pull-out (tensile) strength ofthe proposed Arthrex FiberTak Suture Anchor met the criteriaestablished by published literature for anterolateral ligamentrepair/reconstruction.
Technological ComparisonThe proposed Arthrex Suture Anchor device and predicate device(K231113) have the same scientific technology, packaging, sterility,shelf-life, and MRI safety labeling. The proposed device modificationconsists of the addition of the anterolateral ligamentrepair/reconstruction indication. The reference device DePuy SynthesMitek Sports Medicine MITEK TIGHT-N Tendon Docking Anchor iscomparable in design, intended use and inclusion of the anterolateralligament repair/reconstruction indication.The Arthrex FiberTak Suture Anchor is substantially equivalent to thepredicate device in which the design features and intended uses are thesame.Any differences between the proposed device and the predicate deviceare considered minor and do not raise questions concerning safety oreffectiveness.
ConclusionThe Arthrex FiberTak Suture Anchor is substantially equivalent to thepredicate devices cleared under K231113 in which the basic designfeatures and intended use are the same. Any differences between theArthrex FiberTak Suture Anchor and the predicate devices areconsidered minor and do not raise different questions of safety oreffectiveness.The submitted mechanical testing data demonstrates that the ArthrexFiberTak Suture Anchor is substantially equivalent to that of thepredicate devices for the desired indication.Based on the indications for use, technological characteristics, and thesummary of data submitted, Arthrex Inc. has determined that theproposed device is substantially equivalent to the currently marketed

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.