K Number
K232670

Validate with FDA (Live)

Device Name
HiCardi+ H100
Manufacturer
Date Cleared
2024-09-20

(385 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HiCardi+ H100 is intended to record symptomatic cardiac events and continuous electrocardiogram (ECG) for 72 hours from ambulatory patients by attaching HiCardi+ H100 SmartPatch to the skin surface. ECG records are stored in SmartPatch for review by the clinician after the recording period (up to 72 hours) is completed. It is indicated for use on 22 years or older who may be symptomatic and/or asymptomatic or suffer from transient symptoms as palpitations, shortness, of breath, dizziness, light-headedness, fatique, or anxiety. The reported ECG metrics include preliminary indications of the irreqular beat and/or rhythm and heart rate measurement based on a single-lead ECG basis. The arrhythmia information addressed in the report does not contain diagnostic interpretation; the reported indication is provided for review by the intended user to make a diagnosis based on clinical judgment and experience.

Device Description

The HiCardi+ H100 consists of two components:

  • . Patient-worn ECG patch (Sensor, SmartPatch);
  • . Review & Report Generation Software (LiveStudio).

The SmartPatch is a wearable and reusable patch device intended to measure ECG signals from a patient for 72 hours. The SmartPatch is intended to record ECG signals and automatically indicate the irregular beat and/or rhythm of the recorded ECG data to aid the clinician in finalizing the ECG interpretation. The SmartPatch has an Event Button the patient can press to record symptoms, such as heart palpitations, dizziness, or chest pain, etc.

The LiveStudio is a review software operating on a personal computer (PC). The software visualizes the ECG data and preliminary indications from the SmartPatch for review, editing, and interpretation by the clinician. The LiveStudio provides tools to assist the clinician with their review. The final report includes the summary results of the review and can be documented by PDF formatted documents and printed by a printer connected to the PC.

AI/ML Overview

The provided document details the FDA 510(k) clearance for the HiCardi+ H100 device. However, it does not contain specific acceptance criteria or an analytical study that proves the device meets those criteria in the format requested.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (ZIO® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service) through various performance data and compliance with standards. It lists various bench tests and clinical data to support substantial equivalence but does not provide details on specific acceptance criteria for a performance study of the arrhythmia detection algorithm or the results of such a study.

Here's a breakdown of what can be extracted and what is missing from your request based on the provided text:

Missing Information:

  • A table of acceptance criteria and the reported device performance: This detailed table is not present. While standards are listed (e.g., IEC 60601-2-47, EC57 for validation of heart rate detection and arrhythmia detection algorithm), the specific performance metrics (e.g., sensitivity, specificity, accuracy for arrhythmia detection) and their associated acceptance thresholds are not provided, nor are the actual numerical results for these metrics.
  • Sample size used for the test set and the data provenance: Not specified for any performance evaluation directly comparing the device's algorithmic performance to a ground truth.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified.
  • Adjudication method: Not specified.
  • Multi-reader multi-case (MRMC) comparative effectiveness study: Not mentioned or detailed.
  • Standalone (algorithm only without human-in-the-loop performance) study: While "Validation of heart rate detection and development of arrhythmia detection algorithm" is mentioned as a bench test, the specifics of this "validation" are not provided, including the study design, acceptance criteria, methods, or results of a standalone performance evaluation.
  • Type of ground truth used: Not explicitly stated for performance validation studies.
  • Sample size for the training set: Not specified.
  • How the ground truth for the training set was established: Not specified.

What is mentioned that is relevant to performance data for substantial equivalence, but doesn't fully answer your questions:

  • Performance Data: The document states that "Verification and validation activities established the safety and performance characteristics of the subject device with respect to the predicate."
  • Bench Testing: Several bench tests were conducted, including:
    • "Validation of heart rate detection and development of arrhythmia detection algorithm (IEC 60601-2-47, EC57)"
    • "Disposable ECG electrodes Performance (ANSI/AAMI EC12)"
    • "Software Verification and Validation (IEC 62304)"
  • Compliance to Standards: The device complies with various standards, including IEC 60601-2-47 (EC57), which is relevant to ambulatory ECG performance and includes requirements for heart rate detection and arrhythmia detection. However, the specific performance requirements within these standards and how the device met them numerically are not detailed here.

Conclusion based on provided text:

The provided document, an FDA 510(k) clearance letter and summary, confirms that the HiCardi+ H100 device received clearance based on substantial equivalence to a predicate device. It indicates that performance testing, including "Validation of heart rate detection and development of arrhythmia detection algorithm," was conducted in accordance with relevant standards (IEC 60601-2-47, EC57). However, the document does not provide the detailed acceptance criteria or the specific results of such a performance study (e.g., sensitivity, specificity, accuracy values) which would prove the device quantitatively meets those criteria. It notes that "Clinical data was reviewed to demonstrate the substantial equivalence of the subject device," but does not elaborate on the nature or results of this clinical review in terms of a performance study against specific acceptance criteria for its arrhythmia detection capability.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

09/20/2024

Mezoo Co., Ltd. Jessica Kim Regulatory Affairs Manager Room 808, 200 Gieopdosi-ro, Jijeong-myeon Gangwon-do, Seoul 26354 Korea, South

Re: K232670

Trade/Device Name: HiCardi+ H100 Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MLO, DQK Dated: August 23, 2024 Received: August 23, 2024

Dear Jessica Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Kimberly N. Crowley -S

For: Jennifer Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232670

Device Name HiCardi+ H100

Indications for Use (Describe)

The HiCardi+ H100 is intended to record symptomatic cardiac events and continuous electrocardiogram (ECG) for 72 hours from ambulatory patients by attaching HiCardi+ H100 SmartPatch to the skin surface. ECG records are stored in SmartPatch for review by the clinician after the recording period (up to 72 hours) is completed. It is indicated for use on 22 years or older who may be symptomatic or suffer from transient symptoms as palpitations, shortness, light-headedness, fatigue, or anxiety. The reported ECG metrics include preliminary indications of the irregular beat and/or rhythm and heart rate measurement based on a singlelead ECG basis. The arthythmia information addressed in the report does not contain diagnostic interpretation; the reported indication is provided for review by the intended user to make a diagnosis based on clinical judgment and experience.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

SUBMITTER

Applicant:

MEZOO CO., LTD. #808, Medical Device Complex Center, 200 Gieopdosi-ro, Jijeong-myeon, Wonju-si, Gangwon-do, 26354 Republic of Korea

Phone number: +82-33-761-2006 Fax number: +82-33-761-2036

Official Correspondent:

Boram Kim RA manager of MEZOO brkim@me-zoo.com #808, Medical Device Complex Center, 200 Gieopdosi-ro, Jijeong-myeon, Wonju-si, Gangwon-do, 26354 Republic of Korea

Date Prepared: August 30, 2023

DEVICE

Trade Name (Proprietary Name)HiCardi+ H100
Common NameWearable ECG ambulatory recorder
Classification Regulation870.2800: Recorder, Magnetic Tape, Medical870.1425: Computer, Diagnostic, Programmable
Device ClassClass II
Product CodeMLO, DQK

PREDICATE DEVICE (Legally Marketed Device)

Trade NameZIO® SkyRunner (SR) Electrocardiogram (ECG) MonitoringService
510(k) Submitter/HolderiRhythm Technologies, Inc.
510(k) NumberK143513

{5}------------------------------------------------

DEVICE DESCRIPTION

The HiCardi+ H100 consists of two components:

  • . Patient-worn ECG patch (Sensor, SmartPatch);
  • . Review & Report Generation Software (LiveStudio).

The SmartPatch is a wearable and reusable patch device intended to measure ECG signals from a patient for 72 hours. The SmartPatch is intended to record ECG signals and automatically indicate the irregular beat and/or rhythm of the recorded ECG data to aid the clinician in finalizing the ECG interpretation. The SmartPatch has an Event Button the patient can press to record symptoms, such as heart palpitations, dizziness, or chest pain, etc.

The LiveStudio is a review software operating on a personal computer (PC). The software visualizes the ECG data and preliminary indications from the SmartPatch for review, editing, and interpretation by the clinician. The LiveStudio provides tools to assist the clinician with their review. The final report includes the summary results of the review and can be documented by PDF formatted documents and printed by a printer connected to the PC.

INDICATIONS FOR USE

The HiCardi+ H100 is intended to record symptomatic cardiac events and continuous electrocardiogram (ECG) for 72 hours from ambulatory patients by attaching HiCardi+ H100 SmartPatch to the skin surface. ECG records are stored in SmartPatch for review by the clinician after the recording period (up to 72 hours) is completed. It is indicated for use on 22 years or older who may be symptomatic and/or asymptomatic or suffer from transient symptoms as palpitations, shortness, of breath, dizziness, light-headedness, fatique, or anxiety. The reported ECG metrics include preliminary indications of the irreqular beat and/or rhythm and heart rate measurement based on a single-lead ECG basis. The arrhythmia information addressed in the report does not contain diagnostic interpretation; the reported indication is provided for review by the intended user to make a diagnosis based on clinical judgment and experience.

Contraindications

  • The HiCardi+ H100 is contraindicated for patients with known allergies or hypersensitivity to ● medical-grade adhesives gel or hydrogel.

  • . The HiCardi+ H100 is contraindicated for patients with potentially life-threatening arrhythmias, or who require inpatient monitoring or immediate analysis of their ECG.

  • . The HiCardi+ H100 is contraindicated for a handicapped person who cannot carry a patch-type electrocardiograph.

  • The HiCardi+ H100 is contraindicated for a person who shows a skin disease or functional disorder when the patch-type electrocardiograph comes into contact with it.

  • . The HiCardi+ H100 is contraindicated for subjects who cannot read the consent form (illiterate, foreigners, etc.).

{6}------------------------------------------------

HiCardi+ H100 510(k) Submission

Substantial Equivalence Discussion

The Device Comparison Table in Section 11. Executive Summary/Predicate Comparison and the discussion of similarities and differences in Section 12. Substantial Equivalence Discussion show that the indications for use, technology, and performance of (MEZOO CO., LTD., HiCardi+ H100) are substantially equivalent with the predicate devices (iRhythm Technologies, Inc., ZIO® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service).

CharacteristicsSubject DevicePredicate Device
ECG Recording Characteristics
ECG Channel1 channel1 channel
ECG Lead VectorModified Lead II recorder andModified Lead II analysisModified Lead II recorder andModified Lead II analysis
Recorder Power SupplyRechargeable batteryNon-rechargeable battery
Recording DurationUp to 3 days (The internal batteryof the SmartPatch is continuouslyrecharged when it is in thecharger.)Up to 14 days
Electrode DetachableDetachable, Non-integrated intothe deviceNon-detachable, integrated into thedevice
Electrode MaterialsSilver-Silver Chloride conventionalelectrodes/hydrogel, and skinadhesiveSilver-Silver Chloride conventionalelectrodes/hydrogel, and skinadhesive
Recorder User InterfaceTwo Buttons- Power On/Off Button for on/offcontrol- Event Button for patient eventmarkingSingle button for on/off control/Patient event marking
Wearbody-wornbody-worn
SterileNoNo
Recording TypeContinuousContinuous
Data Transmission Characteristics
Data TelemetryNoYes
Data Transfer / AnalysisSoftware InterfaceWireless technology for datatransfer is not applied.Software interface suitable for thehost computer environment.Bluetooth technology and cellularphone technology & softwareinterface suitable for the hostcomputer environment.
Performance Characteristics
Performance SafetyStandards ComplianceIEC 60601-1IEC 60601-1-2IEC 60601-2-47ISO 10993-1ANSI/AAMI EC12ANSI/AAMI EC57IEC 60601-1IEC 60601-1-2IEC 60601-2-47ISO 10993-1ANSI/AAMI EC12ANSI/AAMI EC57
Documentation Level(or Software Level ofConcern before 2023)Enhanced DocumentationModerate
Physician Report OutputYes, printable, with analysismetrics for recorded ECG signalYes, printable, with analysismetrics for recorded ECG signal

{7}------------------------------------------------

HiCardi+ H100 510(k) Submission

PERFORMANCE DATA

Verification and validation activities established the safety and performance characteristics of the subject device with respect to the predicate. The following performance data were provided in support of the substantial equivalence determination.

Sterilization and Shelf-Life Testing

The HiCardi+ H100 and its standard accessories are not sterilized and do not have a shelf-life.

Biocompatibility Testing

ECG electrode

The biocompatibility assessment for ECG electrode was performed, included in-vitro cytotoxicity, irritation and sensitization according to the ISO 10993-1 standard. Please refer to Section 15. Biocompatibility in this submission for details.

Software Verification and Validation Testing

Software verification and validation activities were performed in accordance with IEC 62304 and FDA's quidance, General Principles of Software Validation, and documentation is provides as recommended by FDA's quidance, Guidance for the Content of Premarket Submissions for Software in Medical Devices and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Please refer to Section 16. Software in this submission for details.

Electrical Safety and Electromagnetic Compatibility (EMC) Testing

The HiCardi+ H100 Wearable ECG ambulatory recorder substantially has been tested in accordance with the IEC 60601-1, IEC 60601-1-11, ISO 60601-2-47 standards for safety and the IEC 60601-1-2 standard for EMC. Please refer to Section 17. Electromagnetic Compatibility and Electrical Safety in this submission for details.

Bench Testing

Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following test report:

  • . General Safety (IEC 60601-1)
  • . Electromagnetic Compatibility (IEC 60601-1-2)
  • . Usability / Human factor (IEC 60601-1-6, IEC 62366-1)
  • . Home Healthcare (IEC 60601-1-11)
  • . Ambulatory ECG (IEC 60601-2-47)
  • . Li-ion Rechargeable Battery (IEC 62133-2)
  • . Protection against water and solid foreign objects (IEC 60529)
  • . Shock & Vibration Test Report (IEC 60068-2-27, 2-64)
  • Temperature and Pressure for Operation and Storage (IEC 60068-2-13. 2-39) .

{8}------------------------------------------------

  • . Packaging Performance (ISTA 3A)
  • Biocompatibility (ISO 10993-1, -5, -10, -23)
  • . Disposable ECG electrodes Performance (ANSI/AAMI EC12)
  • . Software Verification and Validation (IEC 62304)
  • . Validation of heart rate detection and development of arrhythmia detection algorithm (IEC 60601-2-47, EC57)

Compliance to applicable standards includes IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, IEC 60601-1-11, IEC 60601-2-47 (EC57), IEC 62133-2, IEC 60068-2-27, IEC 60068-2-27, IEC 60068-2-64, IEC 60068-2-13, IEC 60068-2-39, ISTA 3A, ISO 10993-1, -5, -10 -23, ANSI/AAMI EC12, IEC 62304 requirements.

The bench test report of this submission showed that substantial equivalence has demonstrated to be suitable for the subject device. Please refer to Section 18. Performance Testing - Bench in this submission for details.

Animal Study

Animal performance testing was not required for this device.

Clinical Study

Clinical data was reviewed to demonstrate the substantial equivalence of the subject device.

CONCLUSIONS

The comparison and performance data presented above support a determination of substantial equivalence of the MEZOO CO., LTD., HiCardi+ H100 to its predicate device, the iRhythm Technologies, Inc., ZIO® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service.

The minor differences between the current subject device (MEZOO CO., LTD., HiCardi+ H100) and the predicate devices (iRhythm Technologies, Inc., ZIO® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service) do not adversely affect the safety or effectiveness.

Therefore, it is concluded that the HiCard+ H100 is substantially equivalent to the identified predicate devices.

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).