K Number
K232597

Validate with FDA (Live)

Date Cleared
2024-01-09

(134 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LYHER® Urine Multi-Drug Test Kit (Cup), LYHER® Urine Multi- Drug Test Kit (Cassette), and LYHER® Urine Multi-Drug Test Kit (Dipcard) are rapid lateral flow immunoassays for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
Amphetamine (AMP)d-Amphetamine1000
Cocaine (COC)Benzoylecgonine300
Marijuana (THC)11-nor-Δ9-THC-9-COOH50
Methamphetamine (MET)d-Methamphetamine1000
Opiates(OPI)Morphine2000
Phencyclidine (PCP)Phencyclidine25

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Device Description

The LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The LYHER® Urine Multi-Drug Test Kit (Cup) device consists of 25 or 40 cup devices and a package insert. The LYHER® Urine Multi-Drug Test Kit (Dipcard) device consists of 10/15/20/25 Dip Card devices, a package insert. The LYHER® Urine Multi-Drug Test Kit (Cassette) device consists of 10/15/20/25 cassette devices, 10/15/20/25 droppers, a package insert.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details from the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the device are implicitly defined by the performance observed in the precision and method comparison studies. The document states that for precision, all test results for samples at and below -25% of the cut-off were negative, and all results for samples at and above +25% of the cut-off were positive. This indicates the device accurately identifies samples substantially below and above the cut-off.

For the method comparison, the device's performance is presented through the concordance with GC/MS results, with discordant results explicitly listed. The overall performance is demonstrated by the number of samples correctly identified as negative or positive across various concentration ranges relative to the cut-off.

Since the document does not explicitly state quantitative acceptance criteria (e.g., minimum sensitivity, specificity, or overall agreement percentage), the performance reported in the tables below represents how the device met the implied expectation of accurate qualitative detection around the cut-off.

Test (Analyte)Cut-off (ng/mL)Acceptance Criteria (Implicit)Reported Device Performance Range (Cassettes, Dipcards, Cups)
Precision Study
AMP, COC, THC, MET, OPI, PCP(Various, see doc)- Samples at/below -25% cut-off: All Negative - Samples at/above +25% cut-off: All Positive- Samples at/below -25% cut-off: All Negative - Samples at/above +25% cut-off: All Positive
Method Comparison (Qualitative Performance around Cut-off)
AMP1000Concordance with GC/MS for negative, positive, and near-cut-off samples.Cassettes: - Negative Urine, <-50% cut-off: 43 Neg, 0 Pos - -50% cut-off ~ cut-off: 12 Neg, 0-2 Pos - cut-off ~ +50% cut-off: 22-23 Pos, 0-1 Neg - >+50% cut-off: 17 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.)
COC300Concordance with GC/MS for negative, positive, and near-cut-off samples.Cassettes: - Negative Urine, <-50% cut-off: 42 Neg, 0 Pos - -50% cut-off ~ cut-off: 13-14 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 37-38 Pos, 0-1 Neg - >+50% cut-off: 5 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.)
MET1000Concordance with GC/MS for negative, positive, and near-cut-off samples.Cassettes: - Negative Urine, <-50% cut-off: 43 Neg, 0 Pos - -50% cut-off ~ cut-off: 10-11 Neg, 1-2 Pos - cut-off ~ +50% cut-off: 23-24 Pos, 0-1 Neg - >+50% cut-off: 16 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.)
OPI2000Concordance with GC/MS for negative, positive, and near-cut-off samples.Cassettes: - Negative Urine, <-50% cut-off: 45 Neg, 0 Pos - -50% cut-off ~ cut-off: 9-10 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 32-33 Pos, 0-1 Neg - >+50% cut-off: 8 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.)
PCP25Concordance with GC/MS for negative, positive, and near-cut-off samples.Cassettes: - Negative Urine, <-50% cut-off: 45 Neg, 2 Pos - -50% cut-off ~ cut-off: 11-12 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 33-34 Pos, 1-2 Neg - >+50% cut-off: 8 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.)
THC50Concordance with GC/MS for negative, positive, and near-cut-off samples.Cassettes: - Negative Urine, <-50% cut-off: 44 Neg, 4 Pos - -50% cut-off ~ cut-off: 16-17 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 23-25 Pos, 1-2 Neg - >+50% cut-off: 15 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.)

2. Sample Size Used for the Test Set and Data Provenance

The studies used unaltered clinical samples for the method comparison. The samples were collected and tested as follows for each analyte and device type (Cassette, Dipcard, Cup):

  • AMP: 43 Negative urine samples, 0 samples <-50% cut-off, 12 samples -50% cut-off ~ cut-off, 23 samples cut-off ~ +50% cut-off, 17 samples >+50% cut-off. (Total = 95 samples per operator/device)
  • COC: 42 Negative urine samples, 0 samples <-50% cut-off, 14 samples -50% cut-off ~ cut-off, 38 samples cut-off ~ +50% cut-off, 5 samples >+50% cut-off. (Total = 99 samples per operator/device)
  • MET: 43 Negative urine samples, 0 samples <-50% cut-off, 12 samples -50% cut-off ~ cut-off, 24 samples cut-off ~ +50% cut-off, 16 samples >+50% cut-off. (Total = 95 samples per operator/device)
  • OPI: 45 Negative urine samples, 0 samples <-50% cut-off, 10 samples -50% cut-off ~ cut-off, 33 samples cut-off ~ +50% cut-off, 8 samples >+50% cut-off. (Total = 96 samples per operator/device)
  • PCP: 45 Negative urine samples, 2 samples <-50% cut-off, 12 samples -50% cut-off ~ cut-off, 35 samples cut-off ~ +50% cut-off, 8 samples >+50% cut-off. (Total = 102 samples per operator/device)
  • THC: 44 Negative urine samples, 4 samples <-50% cut-off, 17 samples -50% cut-off ~ cut-off, 25 samples cut-off ~ +50% cut-off, 15 samples >+50% cut-off. (Total = 105 samples per operator/device)

The total number of unique samples is not directly reported, as these numbers represent the distribution of samples across different concentration ranges. Since "unaltered clinical samples" were used, this suggests they were retrospective in nature. The provenance (country of origin) of the data is not explicitly stated, but the applicant company is based in Hangzhou, China, suggesting the studies were likely performed there.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The ground truth for the test set was established using Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS). These are "preferred confirmatory methods" as stated in the Indications for Use. The document does not specify the number of experts or their qualifications for operating these instruments or interpreting their results; it implies the use of standard laboratory practices for these confirmatory methods.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method for the test set results. The comparison is directly between the device's qualitative result (positive/negative) and the quantitative GC/MS or LC/MS result. Discordant results are noted but no further adjudication (e.g., by multiple experts reviewing the same case) is mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study evaluated the standalone performance of the device against a gold standard (GC/MS/LC/MS), not its effectiveness in assisting human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

Yes, the studies conducted were standalone performance evaluations. The device (LYHER® Urine Multi-Drug Test Kit) is a rapid lateral flow immunoassay, which produces a visual result (line present or absent). The "operators" in the study simply interpret this visual result. The results presented are the device's performance as interpreted by an operator, not an AI algorithm. So, while it involves human interpretation of the device's output, it is a standalone performance of the device itself without additional AI assistance beyond the inherent design of the test kit.

7. The Type of Ground Truth Used

The ground truth used was quantitative analytical results from Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS). These are considered highly accurate confirmatory methods for drug detection and concentration.

8. The Sample Size for the Training Set

The document describes "Precision studies" involving spiking drugs into negative samples to achieve concentrations at various percentages around the cut-off. However, this is described as part of the analytical performance characterization and not as a "training set" for an algorithm. This device is a lateral flow immunoassay, not a machine learning or AI-based device, and therefore does not have a "training set" in the conventional sense of an AI model.

9. How the Ground Truth for the Training Set was Established

As this is not an AI/ML device, there isn't a "training set" that requires ground truth establishment. For the precision studies mentioned (which are part of analytical performance validation), the "ground truth" was established by spiking known concentrations of the target drug into negative urine samples. These spiked concentrations were then "confirmed by LC/MS" to ensure accuracy of the prepared samples.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Hangzhou Laihe Biotech Co., Ltd. % Ethan Liu RA specialist Shanghai Thinkwell Consulting Co., Ltd Xinling Rd., 211/6F Shanghai, Shanghai 201100 China

Re: K232597

Trade/Device Name: LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG, DIO, DKZ, LDJ, DJC, LCM Dated: November 27, 2023 Received: December 1, 2023

Dear Ethan Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by Jos Joseph A. A Kotarek -S Date: 2024.01.09 14:04:36 Kotarek -S Joseph Kotarek, PhD Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232597

Device Name

L YHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard)

Indications for Use (Describe)

The LYHER® Urine Multi-Drug Test Kit (Cup), LYHER® Urine Multi- Drug Test Kit (Cassette), and LYHER® Urine Multi-Drug Test Kit (Dipcard) are rapid lateral flow immunoassays for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
Amphetamine (AMP)d-Amphetamine1000
Cocaine (COC)Benzoylecgonine300
Marijuana (THC)11-nor-Δ9-THC-9-COOH50
Methamphetamine (MET)d-Methamphetamine1000
Opiates(OPI)Morphine2000
Phencyclidine (PCP)Phencyclidine25

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image displays the logo for "LYHER" in green with the "Y" in yellow. A registered trademark symbol is next to the "R". Below the logo, the text "Hangzhou Laihe Biotech Co., Ltd." is written in black.

510(k) Summary 510(k) Number: K232597

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

5.1 Submitter

Submitted by:Hangzhou Laihe Biotech Co., Ltd.
Address: Room 401-406, F1-3, Building 1, No.425Miaohouwang Road, Xixing Street, Binjiang District,Hangzhou, 310051 Zhejiang P.R. China
Tel.: +86 571 8765 3090
Fax: +86 571 8665 8000
www.lyher.com
Contact Person:Ethan Liu
Shanghai Thinkwell Consulting Co., Ltd
Address:Xinling Rd., 211/6F Shanghai, 201100,P.R. China
Phone:0086-15216699240
Email:xtdeepwater@126.com
Date Prepared:Dec 29, 2023

5.2 Device

LYHER® Urine Multi-Drug Test Kit(Cup) LYHER® Urine Multi-Drug Test Kit(Cassette) LYHER® Urine Multi-Drug Test Kit(Dipcard)

Classification:

TestProduct CodeCFR #Panel
Opiates(OPI)DJG862.3650/Opiate test system.Toxicology
Cocaine (COC)DIO862.3250/Cocaine andcocaine metabolite testsystem.Toxicology
Amphetamine (AMP)DKZ862.3100/Amphetamine TestSystemToxicology
Marijuana (THC)LDJ862.3870/Cannabinoid testsystem.Toxicology
Methamphetamine(MET)DJC862.3610/Methamphetaminetest system.Toxicology
Phencyclidine (PCP)LCMUnclassifiedToxicology

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for LYHER, a company named Hangzhou Laihe Biotech Co., Ltd. The LYHER logo is in green and yellow. The company name is in black text and is located below the logo.

5.3 Predicate Device:

ATTEST Drug Screen Cup, ATTEST Drug Screen Dip Card, K182123

5.4 Device Description

The LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The LYHER® Urine Multi-Drug Test Kit (Cup) device consists of 25 or 40 cup devices and a package insert. The LYHER® Urine Multi-Drug Test Kit (Dipcard) device consists of 10/15/20/25 Dip Card devices, a package insert. The LYHER® Urine Multi-Drug Test Kit (Cassette) device consists of 10/15/20/25 cassette devices, 10/15/20/25 droppers, a package insert.

5.5 Indication for Use:

The LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are rapid lateral flow immunoassays for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
Amphetamine (AMP)d-Amphetamine1000
Cocaine (COC)Benzoylecgonine300
Marijuana (THC)11-nor-\u03949-THC-9-COOH50
Methamphetamine (MET)d-Methamphetamine1000
Opiates(OPI)Morphine2000
Phencyclidine (PCP)Phencyclidine25

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive positive result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/Mass Spectrometry (LC/MS) and their tandem massspectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

5.8 Substantial Equivalence

A summary comparison of features of the LYHER® Urine Multi-Drug Test Kit(Cup/Cassette/Dipcard) and the predicate devices is provided in following tables.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "LYHER" in green and yellow letters. The "L", "H", "E", and "R" are green, while the "Y" is yellow. There is a registered trademark symbol next to the "R".

Section 5 510(k) Summary

Hangzhou Laihe Biotech Co., Ltd.

ItemProposed DevicePredicate Device-K182123
Indicationsfor UseFor the qualitative determination of drugs ofabuse and/or their metabolites in human urine.Same
MethodologyLateral flow immunochromatographic assaybased on competitive bindingSame
Type of TestQualitativeSame
SpecimenTypeHuman UrineSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for LYHER. The "LY" is in yellow, and the "HER" is in green. There is a registered trademark symbol next to the "R".

Section 5 510(k) Summary

Hangzhou Laihe Biotech Co., Ltd.

Cut-off Values
TestCalibratorCut-off ng/mL
Opiates(OPI)Morphine2000
Cocaine (COC)Benzoylecgonine300
Amphetamine (AMP)d-Amphetamine1000
Marijuana (THC)11-nor-Δ9-THC-9-COOH50
Methamphetamine (MET)d-Methamphetamine1000
Phencyclidine (PCP)Phencyclidine25
TestCalibratorCutoff ng/ml
6-Acetylmorphine6-monoacetylmorphine10
Amphetamine 500d-amphetamine500
Amphetamine 1000d-amphetamine1000
BarbituratesSecobarbital300
BenzodiazepineOxazepam300
BuprenorphineBuprenorphine10
Cocaine 150Benzoylecgonine150
Cocaine 300Benzoylecgonine300
EDDP2-ethylidene-1,5-dimethyl-3-3-diphenyl-pyrrolidine300
Ecstasyd/l-methylenedioxy-methamphetamine500
Methamphetamine 500d-methamphetamine500
Methamphetamine 1000d-methamphetamine1000
Methadoned/l-methadone300
MorphineMorphine300
OpiatesMorphine2,000
OxycodoneOxycodone100
PhencyclidinePhencyclidine25
Propoxyphened-propoxyphene300
TricyclicsNortriptyline1,000
MarijuanaTHC-COOH5
Intended UseFor prescription uses.Same
ConfigurationsCup,Dip Card, CassetteCup, Dip Card

5.9 Test Principle

Each device employs lateral flow immunochromatographic technology and is based on the principle of competitive binding. The product contains membrane strips coated with drug-protein conjugates on the T zone, goat anti-rabbit polycolonal antibody at the C zone, and a conjugate pad which 5-4

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" is in yellow, while the "HER" is in green. There is a registered trademark symbol next to the "R".

contains colloidal gold particles coated with mouse monoclonal antibodies specific against to the corresponding drugs and colloidal gold particles coated with rabbit IgG.

If the level of drug in the urine specimen is below the cutoff concentration, the T line appears as a visible burgundy line. If the level of drug in the urine specimen is above the cutoff, no T line develops.The control line (C line) serves as an internal quality control line should always appear as a burgundy-colored band regardless of the drug, if enough sample volume has been added to the test and if the sample has correctly migrated up the test strip. Testing is based on the principle of a competitive immunochemical reaction between a chemically labeled drug (drug-protein conjugate) and the drug or drug metabolites which may be present in the urine sample competing for the limited antibody binding sites.

5.10 Performance Characteristics

5.10.1 Analytical Performance

  • a. Precision
    Precision studies were carried out for samples with concentrations of -100% cut off, -75%cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off, These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. Each concentration has three lots, and 6 operators have joined the performance. Each specimen was analyzed in 50 replicates by each of the testing personnel at each Lab site by using three lots of the product separately.

The test results of the specimens at the concentrations at and below -25% of the cut off obtained by all the three personnel were all negative while the test results of the specimens at the concentration at and above +25% of the cut off value were all positive.

  • b. Linearity
    Not applicable

Stability ﻥ

The devices are stable at 2-30 °C for 24 months based on the accelerated stability study.

d. Cut-off

Dilute the solution by negative urine to the concentration of +50% cut off, +25%cut off, -25%cut off, -50%cut off respectively. To test all these above specimens by three different batches of the candidate product and each concentration will be tested for 25 replicates.

The following cut-off values for the candidate devices have been verified.

TestCalibratorCut-off(ng/mL)
Opiates(OPI)Morphine2000
Cocaine (COC)Benzoylecgonine300
Amphetamine(AMP)d-Amphetamine1000

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "Y" is orange, while the rest of the letters are green. There is a small registered trademark symbol to the right of the "R".

Marijuana (THC)11-nor-Δ9-THC-9-COOH50
Methamphetamine(MET)d-Methamphetamine1000
Phencyclidine(PCP)Phencyclidine25

e. Interference

Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and target drugs urine with concentrations at 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different devices

Amphetamine:

SubstanceSubstanceSubstance
AcetaminophenErythromycinPenicillin-G
AcetoneEthanolPheniramine
Acetylsalicylic acidFurosemidePhenothiazine
AlbuminGabapentinPregablin
AmpicillinGlucoseProcaine
Ascorbic acidGuaiacol glyceryl etherPropoxyphene
AspartameHemoglobinQuinidine
AtropineIbuprofenRanitidine
BenzocaineIsoproterenolRiboflavin
BilirubinIsoproterenolSertraline
CaffeineKetamineSodium chloride
ChloroquineLidocaineSulindac
Chlorpheniraminel-PhenylephrineTheophylline
CreatineMethadoneTyramine
DexbrompheniramineMethylephedrineβ-Phenylethylamine
DextromethorphanNaproxenPhencyclidine(PCP)
DimenhydrinateNiacinamideCocaine
DimethylaminoantipyrineNicotineMorphine
DiphenhydramineNorephedrineΔ9-tetrahydrocannabinol
DopamineOxalic acid
EDDPOxycodone

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for LYHER. The letters are in a sans-serif font, with the "LY" in yellow and the "HER" in green. There is a registered trademark symbol next to the "R".

EphedrinePantoprazole
-------------------------

Cocaine(COC)

SubstanceSubstanceSubstance
AcetaminophenErythromycinPenicillin-G
AcetoneEthanolPheniramine
Acetylsalicylic acidFurosemidePhenothiazine
AlbuminGabapentinPregablin
AmpicillinGlucoseProcaine
Ascorbic acidGuaiacol glyceryl etherPropoxyphene
AspartameHemoglobinQuinidine
AtropineIbuprofenRanitidine
BenzocaineIsoproterenolRiboflavin
BilirubinIsoproterenolSertraline
CaffeineKetamineSodium chloride
ChloroquineLidocaineSulindac
Chlorpheniraminel-PhenylephrineTheophylline
CreatineMethadoneTyramine
DexbrompheniramineMethylephedrineβ-Phenylethylamine
DextromethorphanNaproxenPhencyclidine(PCP)
DimenhydrinateNiacinamideMorphine
DimethylaminoantipyrineNicotineMethylamphetamine
DiphenhydramineNorephedrineMDMA
DopamineOxalic acidΔ9-tetrahydrocannabinol
EDDPOxycodoneAmphetamine
EphedrinePantoprazole

MET

SubstanceSubstanceSubstance
AcetaminophenErythromycinPenicillin-G
AcetoneEthanolPheniramine
Acetylsalicylic acidFurosemidePhenothiazine
AlbuminGabapentinPregablin
AmpicillinGlucoseProcaine
Ascorbic acidGuaiacol glyceryl etherPropoxyphene
AspartameHemoglobinQuinidine
AtropineIbuprofenRanitidine
BenzocaineIsoproterenolRiboflavin
BilirubinIsoproterenolSertraline
CaffeineKetamineSodium chloride
ChloroquineLidocaineSulindac
Chlorpheniraminel-PhenylephrineTheophylline

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "L" and "HER" are in green, while the "Y" is in orange. There is a registered trademark symbol next to the "R".

CreatineMethadoneTyramine
DexbrompheniramineMethylephedrineβ-Phenylethylamine
DextromethorphanNaproxenPhencyclidine(PCP)
DimenhydrinateNiacinamideCocaine
DimethylaminoantipyrineNicotineMorphine
DiphenhydramineNorephedrineΔ9-tetrahydrocannabinol
DopamineOxalic acid
EDDPOxycodone
EphedrinePantoprazole

OPI

OPT
SubstanceSubstanceSubstance
AcetaminophenErythromycinPenicillin-G
AcetoneEthanolPheniramine
Acetylsalicylic acidFurosemidePhenothiazine
AlbuminGabapentinPregablin
AmpicillinGlucoseProcaine
Ascorbic acidGuaiacol glyceryl etherPropoxyphene
AspartameHemoglobinQuinidine
AtropineIbuprofenRanitidine
BenzocaineIsoproterenolRiboflavin
BilirubinIsoproterenolSertraline
CaffeineKetamineSodium chloride
ChloroquineLidocaineSulindac
ChlorpheniramineI-PhenylephrineTheophylline
CreatineMethadoneTyramine
DexbrompheniramineMethylephedrineβ-Phenylethylamine
DextromethorphanNaproxenPhencyclidine(PCP)
DimenhydrinateNiacinamideCocaine
DimethylaminoantipyrineNicotineMethylamphetamine
DiphenhydramineNorephedrineMDMA
DopamineOxalic acidAmphetamine
EDDPOxycodoneΔ9-tetrahydrocannabinol
EphedrinePantoprazole

РСР

SubstanceSubstanceSubstance
AcetaminophenErythromycinPenicillin-G
AcetoneEthanolPheniramine
Acetylsalicylic acidFurosemidePhenothiazine
AlbuminGabapentinPregablin
AmpicillinGlucoseProcaine

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in orange, while the "HER" portion is in green. The letters are bold and have a rounded appearance. There is a registered trademark symbol next to the "R".

Ascorbic acidGuaiacol glyceryl etherPropoxyphene
AspartameHemoglobinQuinidine
AtropineIbuprofenRanitidine
BenzocaineIsoproterenolRiboflavin
BilirubinIsoproterenolSertraline
CaffeineKetamineSodium chloride
ChloroquineLidocaineSulindac
ChlorpheniramineI-PhenylephrineTheophylline
CreatineMethadoneTyramine
DexbrompheniramineMethylephedrineβ-Phenylethylamine
DextromethorphanNaproxenCocaine
DimenhydrinateNiacinamideMorphine
DimethylaminoantipyrineNicotineMethylamphetamine
DiphenhydramineNorephedrineMDMA
DopamineOxalic acidAmphetamine
EDDPOxycodoneΔ9-tetrahydrocannabinol
EphedrinePantoprazole

THC

SubstanceSubstanceSubstance
AcetaminophenErythromycinPenicillin-G
AcetoneEthanolPheniramine
Acetylsalicylic acidFurosemidePhenothiazine
AlbuminGabapentinPregablin
AmpicillinGlucoseProcaine
Ascorbic acidGuaiacol glyceryl etherPropoxyphene
AspartameHemoglobinQuinidine
AtropineIbuprofenRanitidine
BenzocaineIsoproterenolRiboflavin
BilirubinIsoproterenolSertraline
CaffeineKetamineSodium chloride
ChloroquineLidocaineSulindac
Chlorpheniraminel-PhenylephrineTheophylline
CreatineMethadoneTyramine
DexbrompheniramineMethylephedrineβ-Phenylethylamine
DextromethorphanNaproxenPhencyclidine(PCP)
DimenhydrinateNiacinamideCocaine
DimethylaminoantipyrineNicotineMorphine
DiphenhydramineNorephedrineMethylamphetamine
DopamineOxalic acidMDMA
EDDPOxycodoneAmphetamine
EphedrinePantoprazole

{12}------------------------------------------------

f. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below. There were no differences observed for different devices.

CompoundConcentration(ng/mL)%Cross Reactivity
Amphetamine (AMP)
D-Amphetamine1000100%
D,L-Amphetamine sulfate250040%
β-Phenylethylamine600001.67%
(±)3,4-Methylenedioxyamphetamine (MDA)150066.7%
Phentermine250040%
Tryptamine600001.67%
Tyramine500020%
L-amphetamine1000001%
COCAINE (COC)ng/ml
Benzoylecgonine300100%
Cocaine75040%
Cocaethylene100003%
Ecgonine200001.5%
Ecgonine methylester500000.6%
MARIJUANA (THC)
Cannabinol200000.25%
11-nor-Δ8-THC-9 COOH30167%
11-nor-Δ9 -THC-9 COOH50100%
Δ8 -THC150000.33%
Δ9 -THC150000.33%
(±)-11-Hydroxy-Δ9-THC50001%
Cannabidiol200000.25%
METHAMPHETAMINE (MET)
p-Hydroxymethamphetamine300003.33%
D-Methamphetamine1000100%
L-Methamphetamine800012.5%
(±)-3,4-Methylenedioxymethamphetamine200050%
Mephentermine500002%
D-Amphetamine1000001%
CompoundConcentration(ng/mL)%Cross Reactivity
L-Amphetamine1000001%
(1R, 2S)-(-)-Ephedrine1000001%
(+/-) 3,4-Methylenedioxy-nethylamphetamine (MDEA)300003.33%
OPIATES (OPI)
Morphine2000100%
Codeine1500133%
Ethylmorphine300006.67%
Hydrocodone800002.5%
Hydromorphone1000020%
Levophanol500040%
6-Monoacetylmorphine2000100%
Morphine 3-β-D-glucuronide500040%
Norcodeine300006.67%
Normorphone1000002%
Oxycodone1000002%
Oxymorphone1000002%
Procaine600003.33%
Thebaine300006.67%
6-Acetylmorphine (6-AM)2000100%
Diacetylmorphine (heroin)500040%
Phencyclidine (PCP)
Phencyclidine25100%
4-hydroxy-PCP125000.2%

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" is in orange, while the "HER" is in green. There is a registered trademark symbol to the right of the "R".

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.003 to 1.032 specific gravity or urine samples with pH 3 to 9 were spiked with target

drugs at 25% below and 25% above Cut-off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different devices.

5.10.2 Method comparison

Method comparison studies for the LYHER® Urine Multi-Drug Test were performed in-house with three laboratory assistants for each device. Operators ran samples unaltered clinical samples. The samples were blind labeled and compared to GC/MS results are presented in the tables below:

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for LYHER. The logo is green, except for the Y, which is orange. There is a registered trademark symbol to the right of the word.

AMPCassettesTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4301210
Positive0002217
Operator 2Negative4301100
Positive0012317
Operator 3Negative4301200
Positive0002317

Discordant Results of AMP Cassettes

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 259202300501994Positive
Operator 1752023005171008Negative
AMPDipcardsTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4301210
Positive0002217
Operator 2Negative4301200
Positive0002317
Operator 3Negative4301100
Positive0012317

Discordant Results of AMP Dipcards

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 359202300501994Positive
Operator 1752023005171008Negative
AMP CupsTest resultNegative urine<-50% cut off-50% cut off~cut offCut off~+50%cut off>+50%cut off
Operator 1Negative4301200
Positive0002317
Operator 2Negative4301200
Positive0002317
Operator 3Negative4301100
Positive0012317

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in orange, while the "HER" portion is in green. There is a registered trademark symbol to the right of the word.

Discordant Results of AMP Cups

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 359202300501994Positive

coc

COCCassettesTestresultNegativeurine<-50% cutoff-50% cutoff~cut offCut off~+50%cutoff>+50%cutoff
Operator 1Negative4201310
Positive001375
Operator 2Negative4201400
Positive000385
Operator 3Negative4201300
Positive001385

Discordant Results of COC Cassettes

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 117202300680298Positive
Operator 317202300680298Positive
Operator 121202300684303Negative
COCDipcardsTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4201410
Positive000375
Operator 2Negative4201300
Positive001385
Operator 3Negative4201300
Positive001385

Discordant Results of COC Dipcards

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 217202300680298Positive
Operator 317202300680298Positive
Operator 121202300684303Negative
COC CupsTest resultNegative urine<-50% cut off-50% cut off~cut offCut off~+50%cut off>+50%cut off
Operator 1Negative4201400
Operator 1Positive000385

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "L" and "Y" are in yellow, while the "HER" is in green. There is a small registered trademark symbol to the right of the "R".

Operator 2Negative4201310
Positive001375
Operator 3Negative4201300
Positive001385

Discordant Results of COC Cups

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 217202300680298Positive
Operator 317202300680298Positive
Operator 221202300684303Negative

MET

METCassettesTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4301000
Positive0022416
Operator 2Negative4301100
Positive0012416
Operator 3Negative4301110
Positive0012316

Discordant Results of MET Cassettes

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 115202300331932Positive
Operator 158202300374561Positive
Operator 258202300374561Positive
Operator 358202300374561Positive
Operator 3772023003931007Negative
METDipcardsTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4301100
Positive0012416
Operator 2Negative4301100
Positive0012416
Operator 3Negative4301110
Positive0012316

Discordant Results of MET Dipcards

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the word "LYHER" in green and yellow. The "LY" is green, but the middle of the "Y" is yellow. The rest of the word is green. There is a registered trademark symbol in the upper right corner of the word.

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 158202300374561Positive
Operator 258202300374561Positive
Operator 358202300374561Positive
Operator 3772023003931007Negative
MET CupsTest resultNegative<-50% cut-50%cutCut>+50%cut
urineOffoff~cut offoff~+50%cutેન્દ્રિક રીકેની વિદ્વારા નિર્માન નામાં આવેલું એક ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત
ેન્દ્રિક રીકેની વિદ્વારા નિર્માન નામાં આવેલું એક ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત
Operator 1Negative43O1100
Positive0012416
Operator 2Negative4301100
Positive0012416
Operator 3Negative43O1110
PositiveO012316

Discordant Results of MET Cups

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 158202300374561Positive
Operator 258202300374561Positive
Operator 358202300374561Positive
Operator 3772023003931007Negative

OPI

OPICassettesTest resultNegativeurine<-50% cut off-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4501000
Positive000338
Operator 2Negative450910
Positive001328
Operator 3Negative4501000
Positive000338

Discordant Results of OPI Cassettes

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 2592023001802006Negative
Operator 2802023002011978Positive

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for LYHER. The logo is green, except for the "Y" which is yellow. There is a small registered trademark symbol to the right of the "R" in LYHER.

OPIDipcardsTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4501000
Positive000338
Operator 2Negative450900
Positive001338
Operator 3Negative4501010
Positive000328

Discordant Results of OPI Dipcards

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 3592023001802006Negative
Operator 2802023002011978Positive
OPI CupsTest resultNegative urine<-50% cut off-50% cut off~cut offCut off~+50%cut off>+50%cut off
Operator 1Negative4501010
Operator 1Positive000328
Operator 2Negative450900
Operator 2Positive001338
Operator 3Negative450900
Operator 3Positive001338

Discordant Results of OPI Cups

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 1592023001802006Negative
Operator 2802023002011978Positive
Operator 3802023002011978Positive

РСР

PCPCassettesTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4521110
Positive001348
Operator 2Negative4521210
Positive000348
Operator 3Negative4521220
Positive000338

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in orange, while the "HER" portion is in green. There is a registered trademark symbol next to the "R".

Discordant Results of PCP Cassettes

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 12520230081023.0Positive
Operator 23720230082225.2Negative
Operator 33720230082225.2Negative
Operator 110020230175025.2Negative
Operator 310020230175025.2Negative
PCPDipcardsTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4521200
Positive000358
Operator 2Negative4521120
Positive001338
Operator 3Negative4521200
Positive000358

Discordant Results of PCP Dipcards

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 22520230081023.0Positive
Operator 23720230082225.2Negative
Operator 210020230175025.2Negative
PCP CupsTest resultNegative urine<-50% cut off-50% cut off~cut offCut off~+50%cut off>+50%cut off
Operator 1Negative4521210
Positive000348
Operator 2Negative4521200
Positive000358
Operator 3Negative4521200
Positive000358

Discordant Results of PCP Cups

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 110020230175025.2Negative

THC

THCCassettesTest resultNegativeurine<-50% cutoff-50% cutoff~cut offCut off~+50%cutoff>+50%cutoff
--------------------------------------------------------------------------------------------------------------------------------------

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for LYHER. The "LY" is in orange, and the "HER" is in green. There is a registered trademark symbol to the right of the "R".

Operator 1Negative4441710
Positive0002415
Operator 2Negative4441620
Positive0012315
Operator 3Negative4441700
Positive0002515

Discordant Results of THC Cassettes

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 11220230103650.5Negative
Operator 23620230106052.3Negative
Operator 24520230106949.6Positive
Operator 29220230165151.1Negative
THCDipcardsTest resultNegativeurine<-50% cut off-50% cutoff~cut offCutoff~+50%cutoff>+50%cutoff
Operator 1Negative4441600
Positive0012515
Operator 2Negative4441730
Positive0002215
Operator 3Negative4441710
Positive0002415

Discordant Results of THC Dipcards

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 21220230103650.5Negative
Operator 23620230106052.3Negative
Operator 14520230106949.6Positive
Operator 37520230109952.3Negative
Operator 29220230165151.1Negative
THC CupsTest resultNegative urine<-50% cut off-50% cut off~cut offCut off~+50%cut off>+50%cut off
Operator 1Negative4441600
Positive0012515
Operator 2Negative4441710
Positive0002415

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in yellow, while the "HER" portion is in green. There is a registered trademark symbol to the right of the word.

Operator 3 Negativeﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ1 -
PositiveCJ1 Pﻟ

Discordant Results of THC Cups

OperatorSN.Specimen No.LC/MS resultsResults of Lyher kit
Operator 21220230103650.5Negative
Operator 33620230106052.3Negative
Operator 14520230106949.6Positive
Operator 39220230165151.1Negative

5.11 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou Laihe Biotech Co., Ltd has demonstrated that proposed device LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are substantially equivalent to ATTEST Drug Screen Cup, ATTEST Drug Screen Dip Card, K182123.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).