K Number
K232589

Validate with FDA (Live)

Device Name
Horos MD™
Manufacturer
Date Cleared
2024-01-03

(131 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Horos MD™ is a software device intended for viewing images acquired from computed tomography (CT), computed radiography (CR), magnetic resonance (MR), ultrasound (US) and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be stored, communicated, processed, and displayed within the system. It is intended for use as a diagnostic and review tool by trained healthcare professionals.

This device is not to be used for mammography.

It is the User's responsibility to operate the device in accordance with the software and hardware requirements listed in the instructions for use, in particular ensuring that monitor (display) quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

Device Description

The Horos MD™ is an interactive image display and navigation software device for diagnosis of medical images. It provides both 2D and 3D image visualization tools for CT and MRI scans from different makes and models of image acquisition hardware. It does not produce any original medical images and does not contain controls for the direct operation of a diagnostic imaging system. Horos MD™ conforms to the DICOM standard to allow the sharing of medical images with other digital imaging systems such as PACS (Picture Archiving and Communication System).

The Horos MD™ software device runs on the macOS X platform, taking advantage of its optimized 3D graphic capabilities, which are provided by the METAL framework developed and maintained by Apple Inc.

The user interface of the software follows Apple's Human Interface Guidelines (HIG) in order to create a user interface that is intuitive and easy to use for users who are familiar with other Apple products. Typical users of this device are radiologists and clinicians who are familiar with 2D scan images.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

Function/MetricAcceptance Criteria (Not explicitly stated, but implied by reported performance)Reported Device Performance
Distance (1-10mm)High accuracy (implied > 99%)99.68%
Distance (>10mm)High accuracy (implied > 99%)99.82%
Area (1-10mm)High accuracy (implied > 99%)99.19%
Area (>10mm)High accuracy (implied > 99%)99.66%
Angle (1-190mm)100% accuracy100%
Point (1-190mm)100% accuracy100%
Software FunctionalityOperates according to intended use and cybersecurity requirements.Verified and validated

Note: The acceptance criteria for accuracy are not explicitly defined as thresholds (e.g., "must be greater than 99%"). Instead, the performance data is presented, implying that these values are acceptable and demonstrate sufficient accuracy for the intended use. The overall acceptance criterion is "substantial equivalence" to the predicate device, demonstrated through non-clinical bench testing.

Study Details

The provided text describes a non-clinical bench performance test to demonstrate the accuracy of measurement functions and software verification and validation to ensure overall functionality and cybersecurity.

2. Sample size used for the test set and the data provenance:

  • Test set sample size: For the measurement accuracy testing, 75 Digital Reference Objects were used.
  • Data provenance: Not explicitly stated, but "Digital Reference Objects" implies synthetically generated data with known values, not clinical patient data from a specific country. This is a common approach for bench testing. The study is retrospective in the sense that these objects were created for the purpose of testing the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of experts: Not applicable. The ground truth for the measurement accuracy tests was established using known values of the Digital Reference Objects, not expert interpretation.
  • Qualifications of experts: Not applicable.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Adjudication method: Not applicable. The ground truth for the measurement accuracy tests was deterministic (known values of Digital Reference Objects). For software verification and validation, the method would typically involve testing against predefined requirements, not expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC study: No. The provided text explicitly states, "No clinical testing was required to demonstrate safety or effectiveness for the subject device as the device’s non-clinical (bench) testing was sufficient to support the intended use of the device." This device is a medical image management and processing system, not an AI-powered diagnostic aid that enhances human reader performance.
  • Effect size: Not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Standalone study: Yes, in a way. The measurement accuracy tests were performed on the algorithm's output against known values. The software verification and validation also represent standalone testing of the software's functionality and security. The device itself is "software device intended for viewing, storing, communicating, processing, and displaying medical images," which operates in a 'standalone' computational manner, though it is a tool for human healthcare professionals.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Type of ground truth: For the measurement accuracy tests, the ground truth was based on known values derived from the design of the Digital Reference Objects. For software verification, the ground truth is adherence to predefined software requirements and specifications.

8. The sample size for the training set:

  • Training set sample size: Not applicable. This device is described as a "Medical Image Management and Processing System" that provides tools for image visualization and measurement. It is not an AI/ML model that would typically require a training set in the context of learning from data to perform a diagnostic task. The software functionality is based on deterministic algorithms for image manipulation and measurement.

9. How the ground truth for the training set was established:

  • How ground truth was established: Not applicable, as there is no training set mentioned for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

iCat Solutions Ltd % Bhoomika Joyappa Senior Regulatory Consultant Medical Device Academy, Inc. 345 Lincoln Hill Rd Shrewsbury, Vermont 05738

January 3, 2024

Re: K232589

Trade/Device Name: Horos MD™M Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: December 13, 2023 Received: December 13, 2023

Dear Bhoomika Joyappa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

Device Name

Horos MD TM

Indications for Use (Describe)

Horos MD™ is a software device intended for viewing images acquired from computed tomography (CT), computed radiography (CR), magnetic resonance (MR), ultrasound (US) and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be stored, communicated, processed, and displayed within the system. It is intended for use as a diagnostic and review tool by trained healthcare professionals.

This device is not to be used for mammography.

It is the User's responsibility to operate the device in accordance with the software and hardware requirements listed in the instructions for use, in particular ensuring that monitor (display) quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46

1. DESCRIPTION

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

2. SUBMITTER

Company NameiCat Solutions Ltd
Address51-59 Rose Lane, NR1 1BY, Norwich, United Kingdom
tel+44(0)7517456306
Contact PersonGeorgios Michalopoulos

3. DATE PREPARED

2023/07/21

4. DEVICE

Device Trade NameHoros MD™
Classification NameMedical Image Management and Processing System
Regulation21 CFR 892.2050
Regulatory ClassClass II
Device PanelRadiology
Product Classification CodeLLZ

{4}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46

5. PREDICATE DEVICE

Predicate ManufacturerPixmeo, Sarl
Predicate Trade NameOsirix MD™
Predicate 510(k)K101342

No reference devices were used in this submission.

6. DEVICE DESCRIPTION

The Horos MD™ is an interactive image display and navigation software device for diagnosis of medical images. It provides both 2D and 3D image visualization tools for CT and MRI scans from different makes and models of image acquisition hardware. It does not produce any original medical images and does not contain controls for the direct operation of a diagnostic imaging system. Horos MD™ conforms to the DICOM standard to allow the sharing of medical images with other digital imaging systems such as PACS (Picture Archiving and Communication System).

The Horos MD™ software device runs on the macOS X platform, taking advantage of its optimized 3D graphic capabilities, which are provided by the METAL framework developed and maintained by Apple Inc.

The user interface of the software follows Apple's Human Interface Guidelines (HIG) in order to create a user interface that is intuitive and easy to use for users who are familiar with other Apple products. Typical users of this device are radiologists and clinicians who are familiar with 2D scan images.

7. INDICATIONS FOR USE

Horos MD™ is a software device intended for viewing images acquired tomography (CT), computed radiography (CR), magnetic resonance (MR), ultrasound (US) and other DICOM compliant medical imaging systems when institude commercial standard hardware. Images and data can be stored, communicated, processed, and displayed within the system. It is intended for use as a diagnostic and review tool by trained healthcare professionals. This device is not to be used for mammography.

{5}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46

It is the User's responsibility to operate the device in accordance with the software requirements listed in the instructions for use, in particular ensuring that monitor (display) quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

8. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:

CharacteristicsSubject deviceHoros MD™Primary PredicateOsirix MD™Justification fordifferences
SubmitteriCat Solutions LtdPixmeo, Sarl-
510(k) NumberK232589K101342-
Product classificationClass IIClass II
Regulation Number21 CFR 892.205021 CFR 892.2050Same
Product codeLLZLLZSame
Type of usePrescription (RX Only)Prescription (RX Only)Same
ClassIIII
Regulations number21 CFR 892.205021 CFR 892.2050
Regulation Name^Medical Image Management andProcessing SystemPicture archiving andcommunications system21 CFR892.2050 name hasbeen amended inApril 2021
Classification NameSystem, image processing, radiologicalSystem, image processing,radiologicalSame
Product codeLLZLLZSame
Indications for UseHoros MD™ is a software deviceintended for viewing images acquiredOsiriX MD"M isa software deviceintended for viewing of imagesSame.

Table 1: Comparison of Horos MD™ with Osirix MD™ (K101342).

{6}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46
from computed tomography (CT),computed radiography (CR), magneticresonance (MR), ultrasound (US) andother DICOM compliant medical imagingsystems when installed on suitablecommercial standard hardware. Imagesand data can be stored, communicated,processed, and displayed within thesystem.It is intended for use as a diagnostic andreview tool by trained healthcareprofessionals.This device is not to be used formammography.It is the User's responsibility to operatethe device in accordance with thesoftware and hardware requirementslisted in the instructions for use, inparticular ensuring that monitor (display)quality, ambient light conditions, andimage compression ratios are consistentwith the clinical application.acquired from CT, MR, CR, DR, USand other DICOM compliant medicalimaging systems when installed onsuitable commercial standardhardware.Images and data can be captured,stored, communicated, processed,and displayed within the system andor across computer networks atdistributed locations.Lossy compressed mammographicimages and digitized film screenimages must not be reviewed forprimary diagnosis or imageinterpretation. For primary diagnosis,post processDICOM "for presentation" imagesmust be used. Mammographicimages should only be viewed with amonitor approved by FDA for viewingmammographic images.It is the User's responsibility toensure monitor quality, ambient lightconditions, and image compressionratios are consistent with the clinicalapplication.The subject deviceis contraindicatedfor Mammography.Therefore, thesubject deviceeliminates the risksassociated with theanalysis ofmammographicimages.
UsersTrained healthcare professionals such asradiologists, clinicians and other qualifiedphysiciansTrained healthcare professionalsSame
Mammography diagnostic useNoYes (Mammographic images shouldonly be viewed with a monitorapproved by FDA for viewingmammographic images)The subject deviceis contraindicatedfor Mammography.Therefore, the

{7}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46
subject device
subject deviceeliminates the risksassociated with thediagnosis andanalysis ofmammographicimages.
Imaging modalities• CT• MR• CR• DR• US• CT• MR• CR• DR• USSame
CommunicationsDICOMDICOMSame
InstallationLocal computersLocal computersSame
Operating system fordiagnostic viewingMacOS (Mac)MacOS (Mac)Same
User authentication (uniquelogin)YesYesSame
Access to DICOM studies• Connect to existing onsite orremote PACS• Local computer/server• USB drives / CD / DVD• Connect to existing onsite orremote PACS• Local computer/server• USB drives / CD / DVDSame
Basic visualization tools• Grayscale display• Rotate• Pan• Zoom• Magnification• Fit to screen.• Scroll• Flip horizontal / vertical• Axial / coronal / sagittal views• Cross-hair• Grayscale display• Rotate• Pan• Zoom• Magnification• Fit to screen• Scroll• Flip horizontal / vertical• Axial / coronal / sagittal views• Cross-hairSame

{8}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46
Adjustable windowing (WW/WL) HU (Hounsfield Unit) Slice indicator ResetAdjustable windowing (WW/WL) HU (Hounsfield Unit) Slice indicator ResetAdjustable windowing (WW/WL) HU (Hounsfield Unit) Slice indicator Reset
Advanced toolsVolume Rendering (3D) MPRVolume Rendering (3D) MPRVolume Rendering (3D) MPR SSDThe subject devicedoes not includeSSD. This limitationdoes not alter thedevice's intendeduse or indicationsfor use, nor does itintroduce any newsafety risks.Therefore, there isno risk to thedevice's safety orperformance.
MeasurementsLine Angle Area (circular, irregular shape, polygonal) Point Cobb Angle CTRLine Angle Area (circular, irregular shape, polygonal) PointLine Angle Area (circular, irregular shape, polygonal) PointThe subject devicehas additionalmeasuringfunctions. Subjectdevice features andfunctions aredescribed in theuser's manual andtested with intendedusers for clinicaluse. Additionalmeasurementfunctions do notchange the device'sintended use or

{9}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46
AnnotationsyesyesSame
Anonymization functionyesyesSame
Studies comparisonyesyesSame
Share functionyesyesSame
Export functionsYes (JPEG, .dcm, .zip)Yes (JPEG, PNG,.dcm, .zip)The subject device exports files in JPEG, .dcm, and .zip formats but not in PNG format. This limitation does not alter the device's intended use or indications for use, nor does it introduce any new safety risks. Therefore, there is no risk to the device's safety or performance.
Report generationNoNoSame
Device type/ formSoftware device that operates on off-the-shelf hardwareSoftware device that operates on off-the-shelf hardwareSame
510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46
DICOM standard for dataexchangeNEMA PS3.1 2023b (2023) DigitalImaging and Communications InMedicine (Dicom) SetNEMA PS 3.1 - 3.20 (2016) DigitalImaging and Communications InMedicine (Dicom) SetSame
JPEG standard for imagecompressionIEC/ ISO 10918-1, Edition 1 (1994-02-15), Information Technology – DigitalCompression and Coding of Continuous-Tone Still Images: Requirements AndGuidelines [Including: TechnicalCorrigendum 1 (2005)IEC/ ISO 10918-1, Edition 1 (1994-02-15), Information Technology -Digital Compression and Coding ofContinuous-Tone Still Images:Requirements And Guidelines[Including: Technical Corrigendum 1(2005)Same

{10}------------------------------------------------

9. EVALUATION OF SIMILARITIES AND DIFFERENCES

The main differences between the subject device and the primary predicate is in the system installation via the Apple AppStore versus installation from a .dmg binary file for the predicate device) and the support of mammography images (Horos MD™ must not be used for mammography diagnostic use)

Horos MD™ and the primary predicate device have the same technical characteristics, general function, intended use, and does not raise any new potential safety risks and is equivalent in performance. In terms of use and functions, both systems access, upload, and display DICOM images and metadata, and resources to trained healthcare professionals for study review and analysis.

10. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Non-Clinical Data

Software Verification and Validation Testing •

All of the different components of Horos MD™ software have been tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use. Software verification testing were conducted, and documentation was provided as detailed in FDA's Guidance for Industry and FDA Staff: "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices."

{11}------------------------------------------------

510(k) Summary
Horos MD (HOROSMD)Document: DOC-21Date: 2023/08/25 14:27:46

iCat Solutions Ltd has implemented security features for the device and data protection. Cybersecurity requirements, risk analysis, and mitigation was addressed in accordance Content of Premarket Submission for Management of Cybersecurity in Medical Devices (Oct 2014).

Bench Performance Testing .

No clinical testing was required to demonstrate safety or effectiveness for the subject device as the device's non-clinical (bench) testing was sufficient to support the intended use of the device.

The measurement accuracy for the distance (length), and area features was validated using Digital Reference Objects compared against the known values. The Digital Reference Objects (n=75 test cases) created were representative of the clinical range typically encountered in radiology practice (1-190 mm). Inter-operator error (reproducibility), and intra-operator error (repeatability) were assessed.

FunctionAccuracy
Distance (1-10mm)99.68%
Distance (>10mm)99.82%
Area (1-10mm)99.19%
Area (>10mm)99.66%
Angle (1-190mm)100%
Point (1-190mm)100%

11. CONCLUSIONS

There is no difference between the subject device (Horos MD™ ) and the predicate device in terms of its intentions for use and fundamental scientific technology. Neithers in usage, safety, or effectiveness, thus no new questions regarding safety or effectiveness are raised. The non-clinical performance test data and validation demonstrate that the subject device performs comparably to and is as safe and effective as the predicate with the 21 CFR Part 807 and based on the information provided in this premarket notification, Horos MD™ is substantially equivalent to the predicate device.

N/A