K Number
K232568

Validate with FDA (Live)

Date Cleared
2024-03-21

(210 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and Accessories is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indication: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

Urology: • Ablation ofBenign Prostatic Hyperplasia (Hypertrophy) [BPH] • Laser Resection of the Prostate (LRP) • Laser Enucleation of the Prostate (LEP) ● Laser Ablation of the Prostate (LAP) ● Transurethral Incision of the Prostate (TUIP) ● Condylomas ● Urethral strictures ● Lesions of External Gentalia ● Bladder Neck Incisions (BNI) ● Ablation and Resection of Bladder Tumors, and Ureteral Tumors . Endoscopic Fragmentation of Urethral, Ureteral, Bladder, and Renal Calculi ● Treatment of distal impacted fragments remaining in the ureters following lithotripsy

Lithotripsy and Percutaneous Urinary Indications: ● Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dihydrate stones ● Endoscopic fragmentation of calculi ● Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed

Gastroenterology: Open and endoscopic gastroentery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: • Appendectomy • Angiodysplasia • Polyps • Colorectal Cancer • Biopsy • Telangiectasias • Gall Bladder Calculi • Telangiectasias of the Osler-Weber-Renu Disease • Biliary/Bile duct Calculi • Vascular Malformation ● Ulcers ● Gastric Ulcers ● Esophagitis ● Duodenal Ulcers ● Esophageal Ulcers ● Non Bleeding Ulcers . Varices . Pancreatits . Colitis . Hemorrhoids . Mallory-Weiss Tear . Cholecystectorny . Gastric Erosions . Benign and Malignant Neoplasm

Gynecology: Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.

Device Description

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories are portable thulium fiber lasers intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories generates non-ionizing radiation which is transferred to the relevant anatomical treatment site through surgical fibers that serve as the delivery vehicle from the laser energy generator to the treatment site.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories includes wireless (optional) and wired, three-pedal footswitches for hands-free control. The system can utilize an auxiliary video monitor to display operating parameters.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories front panel includes a surgical fiber connection port and a touchscreen with a user interface. The back panel includes electrical and interface cable ports. The Laser Console may be placed on an optional mobile cart.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories includes:

  • Laser Console ●
  • Footswitches, wired or wireless (optional) ●
  • Surgical Fibers ●
  • . Accessories:
    • Power cord O
    • O Video output cable
    • Video adapter O
    • Safety glasses O
    • Fiber stripper O
    • Fiber cleaver o
    • Blast shield O
    • O Interlock connector
    • Mobile cart (optional) о

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories on emission wavelengths between 1920 and 1960 nm.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories can operate at a maximum power output of 60 watts and a maximum frequency output of 2400 Hz with a maximum pulse energy of 6 Joules.

The IPG Medical Family of Thulium Fiber Lasers must be operated with the IPG Family of Surgical Fibers. The surgical fibers are flexible optical fibers that deliver the laser energy to the target tissue. The IPG Family of Surgical Fibers are offered in a range of core diameters from 150 microns - 940 microns and include single use, reusable and ball tip options. The surgical fibers are the only patient-contacting components of the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories.

AI/ML Overview

The provided text is a 510(k) Summary for the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and Accessories. It focuses on demonstrating substantial equivalence to predicate devices based on non-clinical performance and design verification testing. It does not describe an AI/ML-driven medical device, and therefore the specific questions regarding acceptance criteria, study details, and ground truth for an AI device cannot be answered from the provided text.

Specifically, there is no information about:

  • A table of acceptance criteria and reported device performance for an AI/ML component.
  • Sample sizes for a test set or its provenance for an AI/ML component.
  • Number of experts or their qualifications to establish ground truth for an AI/ML component.
  • Adjudication method for an AI/ML component test set.
  • A multi-reader multi-case (MRMC) comparative effectiveness study or related effect sizes for AI assistance.
  • Standalone (algorithm-only) performance of an AI/ML component.
  • The type of ground truth used for an AI/ML component (expert consensus, pathology, outcomes data, etc.).
  • Sample size for the training set of an AI/ML component.
  • How the ground truth for an AI/ML component's training set was established.

The document describes performance testing for a laser system and its accessories, not an AI/ML algorithm. The listed tests include:

  • Electrical Safety and EMC (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1)
  • Biocompatibility (e.g., ISO 10993-1, -5, -10, -11, -12)
  • Packaging and Sterile Barrier (e.g., ISO 11607-1, -2, ISO 14644-1, -2, ISO 14698-1, -2, ISO 15223-1, F1886/F1886M-16, F1980-16, F88/F88M-15, F2096-11, ASTM D4169-16, ASTM D4332-14, ISTA 3A, 3B)
  • Sterilization (Surgical Fibers) (e.g., ISO 11135, ISO 10993-7, ISO 11138-1, ISO 11737-1, -2, ISO 17665)
  • Human Factors (e.g., IEC 62366-1, IEC 60601-1-6)
  • Risk Management (e.g., ISO 14971, ISO 22442-1)
  • Software (e.g., IEC 62304)
  • Measuring Systems (e.g., BS EN ISO 10012:2003)
  • Design Verification Testing for the laser system and surgical fibers
  • Shelf-Life testing for the Surgical Fibers
  • Side-by-Side Comparison Testing (bench and animal tissue testing and analysis)

The conclusion states that "All testing demonstrated that the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories met its design requirements." and "Comparative testing (bench and animal tissue testing and analysis) between the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories and the primary predicate and additional predicate devices showed equivalent device performance and tissue response." This indicates that the acceptance criteria were met by passing these various non-clinical engineering and performance tests, rather than through a clinical study or AI/ML performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 13, 2024

IPG Medical Corporation Alexander Wybornov Vice Predisent 225 Cedar Hill Street Marlborough, Massachusetts 01752

Re: K232568

Trade/Device Name: IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 20, 2024 Received: February 20, 2024

Dear Samuel Wade:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232568

Device Name

IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and Accessories

Indications for Use (Describe)

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and Accessories is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indication: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

Urology: • Ablation ofBenign Prostatic Hyperplasia (Hypertrophy) [BPH] • Laser Resection of the Prostate (LRP) • Laser Enucleation of the Prostate (LEP) ● Laser Ablation of the Prostate (LAP) ● Transurethral Incision of the Prostate (TUIP) ● Condylomas ● Urethral strictures ● Lesions of External Gentalia ● Bladder Neck Incisions (BNI) ● Ablation and Resection of Bladder Tumors, and Ureteral Tumors . Endoscopic Fragmentation of Urethral, Ureteral, Bladder, and Renal Calculi ● Treatment of distal impacted fragments remaining in the ureters following lithotripsy

Lithotripsy and Percutaneous Urinary Indications: ● Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dihydrate stones ● Endoscopic fragmentation of calculi ● Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed

Gastroenterology: Open and endoscopic gastroentery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: • Appendectomy • Angiodysplasia • Polyps • Colorectal Cancer • Biopsy • Telangiectasias • Gall Bladder Calculi • Telangiectasias of the Osler-Weber-Renu Disease • Biliary/Bile duct Calculi • Vascular Malformation ● Ulcers ● Gastric Ulcers ● Esophagitis ● Duodenal Ulcers ● Esophageal Ulcers ● Non Bleeding Ulcers . Varices . Pancreatits . Colitis . Hemorrhoids . Mallory-Weiss Tear . Cholecystectorny . Gastric Erosions . Benign and Malignant Neoplasm

Gynecology: Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Manufacturer Identification:

IPG Medical Corporation 225 Cedar Hill Street Marlborough, MA 01752 Phone: 508-506-2800 Establishment Registration #3020221089 Contact Person: Alexander Vybornov (508-506-2718)

Date Prepared:

August 18, 2023

Device Identification:

Device Trade Name:IPG Medical Family of Thulium Fiber Lasers, SurgicalFibers, and Accessories
Device Common Name:Laser Instrument for Use in General Surgery
Classification Name:Laser Instrument, Surgical, Powered
Classification Number:21 CFR 878.4810
Product Code:GEX
Device Class:Class II
Review Panel:Gastroenterology / Urology

Identification of Predicate Devices:

Trade NameManufacturer510(k)
SOLTIVE Laser System (primary predicate)Gyrus, ACMI, Inc.K211401

{5}------------------------------------------------

Device Description:

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories are portable thulium fiber lasers intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories generates non-ionizing radiation which is transferred to the relevant anatomical treatment site through surgical fibers that serve as the delivery vehicle from the laser energy generator to the treatment site.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories includes wireless (optional) and wired, three-pedal footswitches for hands-free control. The system can utilize an auxiliary video monitor to display operating parameters.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories front panel includes a surgical fiber connection port and a touchscreen with a user interface. The back panel includes electrical and interface cable ports. The Laser Console may be placed on an optional mobile cart.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories includes:

  • Laser Console ●
  • Footswitches, wired or wireless (optional) ●
  • Surgical Fibers ●
  • . Accessories:
    • Power cord O
    • O Video output cable
    • Video adapter O
    • Safety glasses O
    • Fiber stripper O
    • Fiber cleaver o
    • Blast shield O
    • O Interlock connector
    • Mobile cart (optional) о

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories on emission wavelengths between 1920 and 1960 nm.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories can operate at a maximum power output of 60 watts and a maximum frequency output of 2400 Hz with a maximum pulse energy of 6 Joules.

The IPG Medical Family of Thulium Fiber Lasers must be operated with the IPG Family of Surgical Fibers. The surgical fibers are flexible optical fibers that deliver the laser energy to the target tissue. The IPG Family of Surgical Fibers are offered in a range of core diameters from 150 microns - 940 microns and include single use,

IPG Medical Family of Thulium Fiber Lasers 510(k) Summary Confidential

{6}------------------------------------------------

reusable and ball tip options. The surgical fibers are the only patient-contacting components of the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories.

Indications for Use:

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

Urology:

  • Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
  • Laser Resection of the Prostrate (LRP) ●
  • Laser Enucleation of the Prostate (LEP) ●
  • Laser Ablation of the Prostate (LAP) ●
  • Transurethral Incision of the Prostate (TUIP) ●
  • Condylomas
  • . Urethral strictures
  • . Lesions of external genitalia
  • Bladder neck incisions (BNI) ●
  • Ablation and resection of bladder tumors, urethral tumors, and ureteraltumors ●
  • Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi ●
  • Treatment of distal impacted fragments remaining in the ureters following lithotripsy ●

Lithotripsy and Percutaneous Urinary Lithotripsy Indications:

  • Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including . cystine, calcium oxalate, monohydrate and calcium oxalate dihydrate stones
  • Endoscopic fragmentation of calculi ●
  • . Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.

Gastroenterology:

Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

  • Appendectomy ●
  • Angiodysplasia ●
  • Polyps ●
  • Colorectal cancer ●
  • Biopsy ●
  • Telangiectasias
  • Gall Bladder calculi
  • Telangiectasias of the Osler-Weber-Renu disease ●
  • . Biliary/Bile duct calculi
  • Vascular Malformation ●

{7}------------------------------------------------

  • Ulcers ●
  • Gastritis ●
  • Gastric ulcers .
  • . Esophagitis
  • Duodenal ulcers ●
  • . Esophageal ulcers
  • Non Bleeding Ulcers ●
  • . Varices
  • Pancreatitis
  • Colitis ●
  • Hemorrhoids
  • Mallory-Weiss tear .
  • Cholecystectomy
  • Gastric Erosions ●
  • Benign and Malignant Neoplasm .

Gynecology: Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.

Comparison to Predicate Devices:

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories is compared to the primary predicate, SOLTIVE Laser System, and the additional predicate, the vela XL Laser System, in the table below.

FeatureIPG Medical Family of ThuliumFiber Lasers, Surgical Fibers andAccessoriesSOLTIVE™ Laser System (#K211401)Primary Predicate
Intended UseThe IPG Medical Family ofThulium Fiber Lasers, SurgicalFibers and Accessories isintended for incision, excision,resection, ablation, coagulation,hemostasis, and vaporization ofsoft tissue, with or without anendoscope, in the followingindications: urology, lithotripsy,gastroenterological surgery andgynecological surgery.Urology:• Ablation of BenignProstaticThe SOLTIVE™ Laser System (SOLTIVE™Pro SuperPulsed Laser, SOLTIVE™Premium SuperPulsed Laser, SOLTIVE™Laser Fibers, and Accessories) is intended forincision, excision, resection, ablation,coagulation, hemostasis, and vaporization ofsoft tissue, with or without an endoscope,in the following indications: urology,lithotripsy, gastroenterological surgery andgynecological surgery.Urology:• Ablation of Benign ProstaticHyperplasia (Hypertrophy) [BPH]• Laser Resection of the
Hyperplasia (Hypertrophy) [BPH]
Laser Resection of the Prostate (LRP) Laser Enucleation of the Prostate (LEP) Laser Ablation of the Prostate (LAP) Transurethral Incision of the Prostate (TUIP) Condylomas Urethral Strictures Lesions of External Genitalia Bladder Neck Incisions (BNI) Ablation and Resection of Bladder Tumors, Urethral Tumors, and Ureteral Tumors Endoscopic Fragmentation of Urethral, Ureteral, Bladder, and Renal Calculi Treatment of Distal Impacted Fragments remaining in the Ureters following Lithotripsy Lithotripsy and Percutaneous Urinary Lithotripsy Indications: Endoscopic Fragmentation of Urethral, Ureteral, Bladder, and Renal Calculi including Cystine, Calcium Oxalate, Monohydrate and Calcium Oxalate Dihydrate Stones Endoscopic Fragmentation of CalculiProstrate (LRP) Laser Enucleation of the Prostate (LEP) Laser Ablation of the Prostate (LAP) Transurethral Incision of the Prostate (TUIP) Condylomas Urethral strictures Lesions of external genitalia Bladder neck incisions (BNI) Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi Treatment of distal impacted fragments remaining in the ureters following lithotripsy. Lithotripsy and Percutaneous Urinary Lithotripsy Indications: Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones Endoscopic fragmentation of calculi Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed Gastroenterology: Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: Appendectomy Angiodysplasia Polyps Colorectal cancer Biopsy
FeatureIPG Medical Family of ThuliumFiber Lasers,Surgical Fibers andAccessoriesSOLTIVETM Laser System (#K211401)Primary Predicate
Treatment of DistalImpacted Fragments ofSteinstrasse when GuideWire cannot be Passed Gastroenterology: Open andendoscopic gastroenterologysurgery (incision, excision,resection, ablation, vaporization,coagulation and hemostasis)including: Appendectomy Angiodysplasia Polyps Colorectal cancer Biopsy Telangiectasias Gall Bladder Calculi Telangiectasias of the Osler-Weber- Renu Disease Biliary/Bile Duct Calculi Vascular Malformation Ulcers Gastritis Gastric Ulcers Esophagitis Duodenal Ulcers Esophageal Ulcers Non Bleeding Ulcers Varices Pancreatitis Colitis Hemorrhoids Mallory-Weiss Tear Cholecystectomy Gastric Erosions Benign and MalignantNeoplasm Gynecology:Telangiectasias Gall Bladder calculi Telangiectasias of the Osler-Weber-Renu disease Biliary/Bile duct calculi Vascular Malformation Ulcers Gastritis Gastric ulcers Esophagitis Duodenal ulcers Esophageal ulcers Non Bleeding Ulcers Varices Pancreatitis Colitis Haemorrhoids Mallory-Weiss tear Cholecystectomy Gastric Erosions Benign and Malignant Neoplasm Gynecology: Open and laparoscopicgynecological surgery (incision, excision,resection, ablation, vaporization, coagulationand hemostasis) of soft tissue.
FeatureIPG Medical Family of ThuliumFiber Lasers,Surgical Fibers and AccessoriesSOLTIVETM Laser System (#K211401)Primary Predicate
Open and laparoscopicgynecological surgery (incision,excision, resection, ablation,vaporization, coagulation andhemostasis) of soft tissue.
FeatureIPG Medical Family of ThuliumFiber Lasers,Surgical Fibers and AccessoriesSOLTIVETM Laser System (#K211401)Primary Predicate
Laser TypeThuliumThulium
Laser ClassClass 4Class 4
LaserEmission ModePulsed and CWPulsed
Average Power2 - 60 W adjustable2 - 60 W adjustable
Peak Optical Power125 W to 500 W125 W to 500 W
Repetition RateUp to 2400 HzUp to 2400 Hz
Operating TempRange15 – 30 deg C15 – 30 deg C
CoolingAir cooledAir cooled
Beam DeliveryFlexible Optical FiberFlexible Optical Fiber
Control Panel10" LCD ColorTouchscreen7" and 12.1" LCD Color TouchscreenDisplays
Electrical100-240V ≈ 1200VA50/60 Hz100-240V ≈ 1200VA50/60 Hz
Weight<40 kg<40 kg
Dimensions43 x 29 x 62 cm30 x 37 x 56 cm
FeatureIPG Medical Family of ThuliumFiber Lasers,Surgical Fibers and AccessoriesSOLTIVETM Laser System (#K211401)Primary Predicate
(HxWxD)
Fiber Sizes150, 200, 365, 550, 940 μm150, 200, 365, 550, 940 μm
FiberSterilizationEtOEtO

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

The intended use and indications for use are identical between the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories and the primary predicate, the SOLTIVE Laser System. The technological characteristics of the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories and the SOLTIVE Laser System are either identical or very similar; with one key difference.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories is different from the primary predicate, the SOLTIVE Laser System, in that it can be used in continuous wave (CW) mode whereas the SOLTIVE Laser System cannot. Therefore, an additional reference predicate, the vela XL Laser System, was identified. The vela XL Laser System is a thulium fiber laser that is used in CW mode. While the vela XL Laser System has a broader set of indications for use, all the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories indications for use are included within the vela XL Laser System indications for use.

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories is considered substantially equivalent to the primary predicate SOLTIVE Laser System in Pulsed mode and is considered substantially equivalent to the additional reference predicate, the vela XL Laser System, for use in CW mode.

Non-Clinical and Performance Testing

The IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories was tested according to the following standards:

Non-Clinical andPerformanceTestingStandards Used
Electrical Safetyand EMCANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 andA2:2010/(R)2012, Medical electrical equipment - Part 1: General requirements forbasic safety and essential performance
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirementsfor basic safety and essential performance - Collateral Standard: Electromagneticdisturbances - Requirements and tests
IEC 60601-2-22:2012, Medical electrical equipment - Part 2-22: Particularrequirements for basic safety and essential performance of surgical, cosmetic,therapeutic and diagnostic laser equipment
IEC 60825-1:2014 (in accordance with Laser Notice 56), Safety of laser products -Part 1: Equipment classification and requirements
BiocompatibilityISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation andtesting within a risk management process
ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for invitro cytotoxicity
ISO 10993-10:2010, Biological evaluation of medical devices – Part 10: Tests forirritation and skin sensitization
ISO 10993-11:2017, Biological evaluation of medical devices – Part 11: Test forsystemic toxicity
ISO 10993-12:2012, Biological evaluation of medical devices - Part 12: Samplepreparation and reference materials
ISO 10993-12:2021, Biological evaluation of medical devices – Part 12: Samplepreparation and reference materials
Packaging andSterile BarrierISO 11607-1:2019, Packaging for terminally sterilized medical devices – Part 1:Requirements for materials, sterile barrier systems, and packaging systems
ISO 11607-2:2019, Packaging for terminally sterilized medical devices – Part 2:Validation requirements for forming, sealing and assembly processes
ISO 14644-1:2015. Cleanrooms and associated controlled environments - Part 1:Classification of air cleanliness by particle concentration.
ISO 14644-2:2015. Cleanrooms and associated controlled environments.
ISO 14698-1:2003. Cleanrooms and associated controlled environments - Bio-contamination control -- Part 1: General principles and methods.
ISO 14698-2:2003. Cleanrooms and associated controlled environments - Bio-contamination control - Part 2: Evaluation and interpretation of bio-contaminationdata.
ISO 15223-1:2016. Medical devices - Symbols to be used with medical device labels,labelling and information to be supplied - Part 1: General requirements.
F1886/F1886M-16, Standard Test Method for Determining Integrity of Seals forFlexible Packaging by Visual Inspection
F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems forMedical Devices
F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials
F2096-11 (2019), Standard Test Method for Detecting Gross Leaks in Packaging byInternal Pressurization (Bubble Test)
ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containersand Systems
ASTM D4332-14, Standard Practice for Conditioning Containers, Packages, orPackaging Components for Testing
ISTA 3A. Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb.) orLess.
ISTA 3B. Packaged-Products for Less-Than-Truckload (LTL) Shipment.
Sterilization(Surgical Fibers)ISO 11135:2014, Medical devices – Validation and routine control of ethylene oxidesterilization
ISO 10993-7:2008/AMD 1:2019, Biological evaluation of medical devices – Part 7:Ethylene oxide sterilization residuals
ISO 11138-1:2017. Sterilization of health care products - Biological indicators - Part1: General Requirements.
ISO 11737-1:2018. Sterilization of health care products - Microbiological methods -Part 1: Determination of a population of microorganisms on products.
ISO 11737-2:2019. Sterilization of health care products - Microbiological methods -Part 2: Tests of sterility performed in the definition, validation and maintenance of asterilization process.
ISO 17665:2006 Sterilization of health care products -- Moist heat -- Part 1:
Requirements for the development, validation and routine control of a sterilizationprocess for medical devices
Human FactorsIEC 62366-1:2015, Medical devices - Part 1: Application of usability engineering tomedical devices
IEC 60601-1-6:2013, Medical electrical equipment – Part 1-6: General requirementsfor basic safety and essential performance – Collateral standard: Usability
Risk ManagementISO 14971: 2019. Medical devices - Application of risk management to medicaldevices.
ISO 22442-1:2020 Medical devices utilizing animal tissues and their derivatives – Part1: Application of risk management
SoftwareIEC 62304:2006/A1:2016, Medical device software – Software Life Cycle Processes
MeasuringSystemsBS EN ISO 10012:2003. Measurement management systems. Requirements formeasurement processes and measuring equipment.

{13}------------------------------------------------

{14}------------------------------------------------

Performance Testing included:

  • Design Verification Testing: Laser system

  • Design Verification Testing: Surgical fibers

  • A Shelf-Life testing for the Surgical Fibers
  • Side-by-Side Comparison Testing

All testing demonstrated that the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories met its design requirements.

Comparative testing (bench and animal tissue testing and analysis) between the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories and the primary predicate and additional predicate devices showed equivalent device performance and tissue response.

Conclusions

Testing supports a substantial equivalence determination of the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories in comparison to the primary predicate device, the SOLTIVE Laser System, and the additional predicate device, vela XL Laser System. Technological differences between the IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers and Accessories and the predicate devices do not raise any different or new types of safety or effectiveness questions.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.