Powder Free Nitrile Examination Glove, Pink Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Orange Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Black Colored, Non-Sterile, Low Dermatitis
K232461 · Kossan International Sdn Bhd · LZA · Nov 14, 2023 · General Hospital
Device Facts
Record ID
K232461
Device Name
Powder Free Nitrile Examination Glove, Pink Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Orange Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Black Colored, Non-Sterile, Low Dermatitis
Applicant
Kossan International Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Nov 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable nitrile examination gloves; non-sterile; available in pink, orange, blue, and black. Worn by healthcare professionals during patient examinations to provide a barrier against contamination. Tested for permeation resistance against various chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Blue and black variants also feature low dermatitis potential. Gloves act as a physical barrier; intended to protect both patient and examiner. Not for use with Carmustine or Thiotepa due to low permeation breakthrough times.
Clinical Evidence
Bench testing only. Permeation testing performed per ASTM D6978-05 (Reapproved 2019) to determine breakthrough detection times for various chemotherapy drugs, Fentanyl Citrate, Xylazine HCl, and simulated gastric acid.
Technological Characteristics
Material: Nitrile. Form factor: Non-sterile, powder-free examination gloves. Colors: Pink, Orange, Blue, Black. Features: Low dermatitis potential (Blue/Black variants). Testing standards: ASTM D6978-05 (Reapproved 2019) for chemotherapy drug permeation.
Indications for Use
Indicated for use as a disposable patient examination glove worn by medical examiners to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs and Fentanyl Citrate. Warning: Do not use with Carmustine (BCNU) and Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K251141 — Powder Free Nitrile Patient Examination Glove, Blue colored, Non-sterile Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Patient Examination Glove, Black colored, Non-sterile Tested for Use with Chemotherapy Drugs and Fentanyl Citrate · Mercator Medical (Thailand), Ltd. · Nov 17, 2025
K183287 — Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate · Kossan International Sdn Bhd · Apr 5, 2019
K222714 — Nitrile Patient Examination Gloves Blue Tested For Use With Chemotherapy Drugs, Nitrile Patient Examination Gloves Blue Violet Tested For Use With Chemotherapy Drugs, Nitrile Patient Examination Gloves Black Tested For Use With Chemotherapy Drugs · Guangdong Kingfa Sci. & Tech.Co., Ltd. · Oct 7, 2022
K230121 — Nitrile Examination Gloves Powder Free Tested For Use With Chemotherapy Drugs & Fentanyl Citrate (Blue & Black) · Careglove Global Sdn. Bhd. · May 15, 2023
K220609 — Powder-Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl (White, Blue, Black, Orange and Green Color) · Encompass Industries Sdn. Bhd. · May 27, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2023
Kossan International Sdn Bhd Cho Sow Fong Senior Manager Regulatory Affairs Wisma Kossan, Lot 782, Jalan Sungai Putus, Off Batu 3 3/4, Jalan Kapar Klang, Selangor 42100 Malaysia
## Re: K232461
Trade/Device Name: Powder Free Nitrile Examination Glove, Pink Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Orange Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate: Powder Free Nitrile Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate: Powder Free Nitrile Examination Glove, Black Colored, Non-Sterile, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate. Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved Product Code: LZA, LZC, QDO, OPJ Dated: August 11, 2023 Received: August 15, 2023
Dear Cho Sow Fong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
{1}------------------------------------------------
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K232461
Device Name
Powder Free Nitrile Examination Glove, Pink Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
Powder Free Nitrile Examination Glove, Orange Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
Powder Free Nitrile Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis
Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
Powder Free Nitrile Examination Glove, Black Colored, Non-Sterile, Low Dermatitis
Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes | Pink | Orange | Blue | Black |
|-----------------------------------------|------------------------------------------------|------|------------|------|-------|
| Carmustine (BCNU) (3.3 mg/ml) | | 12.4 | 11.7 | 10.1 | 14.4 |
| Cisplatin (1.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Cytarabine(100 mg/ml) | | >240 | Not tested | >240 | >240 |
| Dacarbazine (DTIC) (10.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Etoposide (20.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Fluorouracil (50.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Ifosfamide (50.0 mg/ml) | | >240 | Not Tested | >240 | >240 |
| Methotrexate (25.0 mg/ml) | | >240 | Not Tested | >240 | >240 |
| Mitomycin C (0.5 mg/ml) | | >240 | Not Tested | >240 | >240 |
| Mitoxantrone (2.0 mg/ml) | | >240 | Not Tested | >240 | >240 |
| Paclitaxel (Taxol) (6.0 mg/ml) | | >240 | >240 | >240 | >240 |
| Thiotepa (10.0 mg/ml) | | 43.1 | 43.2 | 30.2 | 29.2 |
| Vincristine Sulfate (1.0 mg/ml) | | >240 | Not Tested | >240 | >240 |
Please note that Carmustine (BCNU) and Thiotepa has low permeation times.
Warning: Do Not Use with Carmustine (BCNU) and Thiotepa
| Opioid Drug and Concentration | Minimum Breakthrough Detection Time in Minutes | | | |
|------------------------------------------|------------------------------------------------|--------|------|-------|
| | Pink | Orange | Blue | Black |
| Fentanyl Citrate Injection (100 mcg/2ml) | >240 | >240 | >240 | >240 |
| Xylazine HCl (100 mg/ml) | Not Tested | >240 | >240 | >240 |
{4}------------------------------------------------
| Simulated Gastric Acid | Minimum Breakthrough Detection Time in Minutes |
|------------------------|------------------------------------------------|
|------------------------|------------------------------------------------|
| | Pink | Orange | Blue | Black |
|------------------------|------|--------|------|-------|
| Simulated Gastric Acid | >240 | >240 | >240 | >240 |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.