K Number
K232454

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-05-03

(263 days)

Product Code
Regulation Number
866.2390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ALB Luz Viral Transport Media is intended for the collection and transport of upper respiratory clinical specimens containing Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), and Rhinovirus from the collection site to the testing laboratory. The Viral Transport Media is a culture-based media that is intended to be used with standard diagnostic/identification techniques that utilize stable recoverable infectious viral particles.

Device Description

The Viral Transport Medium (VTM) is a non-propagating transport device composed of a culturebased media without swabs. The VTM is designed to preserves upper respiratory samples collected from a patient by placing the sample into the polymer tube containing 3 mL of media. The sample and media are then secured with a leak-proof screwcap for transportation.

The VTM maintains cellular integrity and preservation of viruses when properly stored. Prior to use, vials should be stored at 2℃ to 35℃. After specimen collection, the transport tube containing the specimen can be stored for up to 48 hours at either 2-8°C or 20-25°C, for transportation to the laboratory and storage. The medium has been evaluated for storage of the following respiratory viruses, Influenza A, Respiratory Syncytial Virus (RSV), and Rhinovirus, for viral recovery.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the ALB Luz Viral Transport Media (VTM). This document details the device's characteristics, intended use, and performance testing to demonstrate substantial equivalence to a predicate device.

However, the provided text does not contain the kind of information typically associated with acceptance criteria and clinical study results for an AI-powered medical device, such as:

  • Accuracy metrics: Sensitivity, specificity, AUC, F1 score.
  • Study design details: Retrospective/prospective, number of patients/cases in the test set.
  • Expert ground truth: Number of experts, their qualifications, adjudication methods.
  • Comparative effectiveness: MRMC study details, human reader improvement with AI.
  • Standalone algorithm performance.
  • Training set details: Size, ground truth establishment.

Instead, the document focuses on performance testing relevant to a viral transport medium, specifically:

  • Shelf-life studies: Physical stability (appearance, volume) and pH stability over 18 months.
  • Sterilization checks: Absence of microbial growth.
  • Viral recovery studies: Using plaque forming assays to show the device maintains viral viability for specific viruses (Influenza A, RSV, Rhinovirus) at different temperatures and time points (0, 24, 48 hours).

Therefore, I will interpret the request in the context of the provided document, defining "acceptance criteria" and "device performance" based on the characteristics and tests described for the Viral Transport Media, rather than an AI-powered device.


Acceptance Criteria and Study Proving Device Meets Criteria

The ALB Luz Viral Transport Media (VTM) is a Class I, reserved medical device (Product Code JSM) intended for the collection and transport of upper respiratory clinical specimens containing specific viruses. The study presented in the 510(k) summary (K232454) demonstrates the device's ability to maintain the viability of these viruses over time and under specified storage conditions, thereby proving its suitability for its intended use.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this Viral Transport Media are based on its ability to maintain physical and chemical stability, minimize contamination, and crucially, maintain the viability of specific viruses over a defined transport/storage period. The performance is assessed against these criteria.

Acceptance Criteria CategorySpecific Acceptance CriterionReported Device Performance and Confirmation
Shelf-life (Product)Physical Stability: No changes in appearance (color, turbidity) or volume over specified shelf life.Confirmed: Three lots tested at 2°C and 35°C showed no changes in color (remained clear with slight precipitation and orange-pink color) and no changes in liquid media volume over 18 months.
pH Stability: pH remains within the acceptable range (7.2 to 7.6) over specified shelf life.Confirmed: Three lots tested at 2°C and 35°C showed pH measurements within the acceptable range of 7.2 to 7.6 over 18 months.
Contamination ControlSterility/Low Contamination: No microbial growth after manufacturing process (filtration and aseptic transfer).Confirmed: Microbial contamination check conducted by incubating tubes for 48 hours at 35°C ± 2°C, followed by transfer to BHI media and incubation for 24 hours at 35°C ± 2°C. "No growth was observed on any of the tested media." (Note: The device is not claimed to be sterile, but this test confirms effective contamination reduction post-filtration processes).
Viral RecoveryMaintain Viral Viability: Average viral recovery demonstrates percent changes within ±90% (i.e., 1 log change) compared to 0-hour recovery for tested viruses after 24 and 48 hours of storage.Confirmed for all tested viruses (Influenza A, RSV, Rhinovirus) at both storage temperatures (2-8°C and 20-25°C):Influenza A:- 2-8°C: Percent changes for 24 hours range from 0% to -9%; for 48 hours, -35% to -38%.- 20-25°C: Percent changes for 24 hours range from -7% to -20%; for 48 hours, -33% to -39%.Respiratory Syncytial Virus (RSV):- 2-8°C: Percent changes for 24 hours range from -11% to 7%; for 48 hours, -38% to -50%.- 20-25°C: Percent changes for 24 hours range from -15% to 0%; for 48 hours, -36% to -43%.Rhinovirus:- 2-8°C: Percent changes for 24 hours range from 0% to 5%; for 48 hours, -33% to -52%.- 20-25°C: Percent changes for 24 hours range from -21% to -18%; for 48 hours, -29% to -46%.All reported percent changes are well within the ±90% acceptance criterion.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): For the viral recovery studies, "Three lots of VTM with newly manufactured, mid-range lot and close to expiry lots were used." For each lot, virus stock was serially diluted, transferred into VTM, and stored at two temperatures (2-8°C and 20-25°C) for three time points (0, 24, 48 hours). For each condition, viral samples were serially diluted and "added to the monolayer in triplicate."
    • This implies a sample size of 3 lots x 2 temperatures x 3 time points x 3 replicates = 54 measurements per virus strain. Since three virus strains were tested, the total number of primary viral recovery measurements would be 162 (excluding the initial virus stock titrations and cell culture maintenance).
  • Data Provenance: The study appears to be an in-vitro laboratory validation study rather than a clinical trial with patient data. The specific origin of the "pooled negative clinical nasal matrix" is not explicitly stated in terms of country, but the sponsor is based in Brazil.
  • Retrospective or Prospective: The study design described (testing newly manufactured, mid-range, and close-to-expiry lots, and specific time points) indicates a prospective laboratory validation study for performance characteristics, particularly viral recovery and shelf-life.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This section is not applicable in the context of this device and study. The "ground truth" for Viral Transport Media performance is established through direct laboratory measurements of viral viability (Plaque-Forming Units/mL) and physical/chemical properties (pH, appearance). No human expert interpretation of images or clinical data for diagnosis is involved.

4. Adjudication Method for the Test Set

This section is not applicable for the reasons stated above. There is no human interpretation or diagnostic decision-making that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. An MRMC study is relevant for AI (or other diagnostic) devices where human readers interpret medical images or data. This is a laboratory performance validation for a transport medium.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No. This section pertains to AI algorithms. The device under review is a physical transport medium, not an algorithm.

7. The Type of Ground Truth Used

The ground truth used for the performance evaluation of the Viral Transport Media is based on direct laboratory measurements of viral viability (Plaque-Forming Units/mL) through plaque assays and physical/chemical measurements (pH, visual inspection for appearance/volume). This is an objective, quantitative, and directly observable ground truth.

8. The Sample Size for the Training Set

This section is not applicable. This is not an AI/machine learning device that requires a "training set." The performance evaluation is based on laboratory testing of the manufactured product.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION".

May 3, 2024

ALB Luz % Graziela Brum Regulatory Affairs Specialist PR Servicos Regulatorios Administrativos Ltda Rua Alice Alem Saadi, 855/2402 Ribeirao Preto, Sao Paulo 14096-570 Brazil

Re: K232454

Trade/Device Name: Viral Transport Media (VTM) Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I, reserved Product Code: JSM Dated: April 4, 2024 Received: April 4, 2024

Dear Graziela Brum:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Please note that if you modify your IVD in the future to exceed any of the limitations to the exemption found in 21 CFR 866.9(c), your device will require a new 510(k) prior to marketing this device in the United States and will not be exempt from the premarket notification requirements so long as it exceeds the limitations to the exemption found in 21 CFR 866.9.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch

{2}------------------------------------------------

Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232454

Device Name Viral Transport Media (VTM)

Indications for Use (Describe)

The ALB Luz Viral Transport Media is intended for the collection and transport of upper respiratory clinical specimens containing Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), and Rhinovirus from the collection site to the testing laboratory. The Viral Transport Media is a culture-based media that is intended to be used with standard diagnostic/identification techniques that utilize stable recoverable infectious viral particles.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

VIRAL TRANSPORT MEDIA

510(k) Summary: K232454

ADMINISTRATIVE INFORMATION

SponsorALB LUZ Ltda.Rua Um, 437Valinhos, São Paulo, Brazil 13273-200Telephone: +55 (19) 3849-7499
Contact Person and PreparerGraziela Brum and Tatiana Jabor BoturaRegulatory Affairs SpecialistPassarini Regulatory Affairse-mail: graziela@passarini.com.brTelephone: +55 (16) 3421 8488
Data PreparedApril, 02,2024

DEVICE NAME AND CLASSIFICATION

Trade NameViral Transport Media - VTM
Common NameSpecimen Collection and Transport System
Regulation Number21 CFR 866.2390 – Transport Culture Medium
Regulatory ClassClass I, Reserved
Product CodeJSM
Classification PanelMI -Microbiology

PREDICATE DEVICE INFORMATION

Predicate Device: K212856 - iClean Viral Transport System (VTM-RT kit)

Indications For Use

The ALB Luz Viral Transport Media is intended for the collection and transport of upper respiratory clinical specimens containing Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), and Rhinovirus from the collection site to the testing laboratory. The Viral Transport Media is a culture-based media that is intended to be used with standard diagnostic/identification techniques that utilize stable recoverable infectious viral particles.

Device Description

The Viral Transport Medium (VTM) is a non-propagating transport device composed of a culturebased media without swabs. The VTM is designed to preserves upper respiratory samples collected from a patient by placing the sample into the polymer tube containing 3 mL of media. The sample and media are then secured with a leak-proof screwcap for transportation.

The VTM maintains cellular integrity and preservation of viruses when properly stored. Prior to

{5}------------------------------------------------

use, vials should be stored at 2℃ to 35℃. After specimen collection, the transport tube containing the specimen can be stored for up to 48 hours at either 2-8°C or 20-25°C, for transportation to the laboratory and storage. The medium has been evaluated for storage of the following respiratory viruses, Influenza A, Respiratory Syncytial Virus (RSV), and Rhinovirus, for viral recovery.

Technological Characteristics

The Viral Transport Media - VTM has a formulation that preserves viral viability. The VTM contains the following ingredients.

Hank's balanced salt solution (HBSS) Amphotericin B Colistin sulfate HEPES buffer Phenol red Bovine serum albumin fraction V (BSA) Dextrose Penicillin G potassium Gelatin powder

Hank's Balanced Salt Solution (HBSS) is further enriched with proteins and sugars for stabilization, and antimicrobial agents to inhibit the overgrowth of bacteria, fungi, and yeasts. A buffer system to maintain neutral pH and a pH indicator. Polymer screw-cap top tubes are filled with 3mL of VTM.

Substantial Equivalence Information

The subject and predicate devices comparison is described in the table below:

Device & PredicateDevice(s):Device: K232454Predicate: K212856
Device Trade NameViral Transport Media - VTMiClean Viral TransportSystem (VTM-RT kit)
Product Code andClassificationJSM, Class IJSM, Class I
General DeviceCharacteristicSimilarities
Intended Use/IndicationsFor UseThe ALB Luz Viral TransportMedia is intended for thecollection and transport of upperrespiratory clinical specimenscontaining Influenza A, InfluenzaB, Respiratory Syncytial Virus(RSV), and Rhinovirus from thecollection site to the testinglaboratory. The Viral TransportMedia is a culture-based mediaiClean Viral TransportSystem (VTM-RT) isintended for the collectionand transport of clinicalspecimens containingrespiratory viruses,Chlamydiae, or Mycoplasmahominis from the collectionsite to the testing laboratory.The collection system is a
Device & PredicateDevice(s):Device: K232454Predicate: K212856
that is intended to be used withstandard diagnostic/identificationtechniques that utilize stablerecoverable infectious viralparticles.culture-based media that isintended to be used withstandard laboratoryexamination, culture or withother assays that utilize stablerecoverable infectious viralparticles or bacteria.
ValidationCulture MediaCulture Media
Tube MaterialPlastic Screw-Cap TubePlastic Screw-Cap Tube
Single Use DeviceYesYes
General DeviceCharacteristicDifferences
Shelf Life18 months12 months
Storage Temperature2 to 35°C20 to 25°C
pH7.4±0.27.4±0.4
Supported StrainsInfluenza A,Influenza B,RespiratorySyncytial Virus (RSV),RhinovirusAdenovirus,Cytomegalovirus,Echovirus Type 30,Herpes SimplexVirus Type 1,Herpes Simplex VirusType 2,Influenza A, Parainfluenza 3,Respiratory Syncytial Virus,Chlamydia pneumoniae,Chlamydia trachomatis,Mycoplasma hominis
List of IngredientsHank's balanced salt solution(HBSS)Amphotericin BColistin sulfateHEPES bufferPhenol redBovine serum albumin fraction V(BSA)DextrosePenicillin G potassiumGelatin powderHank's balanced salt solution(HBSS)Amphotericin BColistinHEPES bufferPhenol redBovine Serum Albumin(BSA)Gentamicin sulfateL-Glutamic acid

{6}------------------------------------------------

{7}------------------------------------------------

Shelf life

The shelf-life for the ALB LUZ Viral Transport Media was established to be 18 months from the date of manufacture when stored in a clean, dry, ventilated environment at 2 - 35°C. The shelf-life study was conducted by using real-time aging performance test for 18 months. In this study, the shelf life was evaluated for physical stability of appearance and volume, and pH stability during the period of storage.

    1. Physical Stability: To evaluate physical stability, any changes in appearance in color, turbidity and volume of the ALB LUZ Viral Transport Media were physically or visually examined at storage time points T = 0, 3, 6, 9, 12, 15, and 18 months. Three lots were tested for physical stability at storage conditions of 2°C and 35°C. At each time point, appearance of the product was inspected visually which appeared clear with slight precipitation and maintained an orange-pink color (i.e., no changes in color) and no changes in liquid media volume. All results support the claim that the ALB LUZ Viral Transport Media is physically or visually stable for 18 months.
    1. pH Stability: To evaluate product stability, the pH of the ALB LUZ Viral Transport Media was tested at storage time points T = 0, 3, 6, 9, 12, 15, and 18 months. Three lots were tested for pH stability at storage condition of 2℃ and 35℃. At each time point, the pH was measured by immersing the pH meter electrode into each ALB LUZ Viral Transport Media by adjusting the temperature to 25°C. For all the tubes tested, the pH measurement was within the acceptable pH range of 7.2 to 7.6. All results support the claim for pH stability of the ALB LUZ Viral Transport Media for 18 months.

The results for physical stability and pH stability collectively support the 18-month stability claim for the ALB LUZ Viral Transport Media.

    1. Sterilization:
      The ALB LUZ Viral Transport Media is not claimed to be sterile nor is it intended to be sterilized by the end user. To reduce the contamination, the transport media is filtered using a 0.22 um sterile filter membrane and then 3 ml aliquot is aseptically transferred into individual sterile conical screw-capped tubes. A microbial contamination check was conducted by incubating the tubes for 48 hours at 35°C ± 2°C followed by transferring 100 µL of transport media into BHI media and incubating for 24 hours at 35°C ± 2°C. No growth was observed on any of the tested media.

Performance testing

Performance of ALB LUZ Viral Transport Media was evaluated by Culture-Based Recovery Studies for viral test strains using plaque forming assay. The viral strains and cell lines used for the plaque forming assay are shown in the Table 1.

Virus StrainsCell Lines
Influenza A virus H1N1 (ATCC VR-1520)MDCK (ATCC CCL-34)
Human respiratory syncytial virus (ATCC VR-1540)MRC-5 (Instituto Adolfo Lutz CCIAL 023)
Human rhinovirus (ATCC VR-1178)MRC-5 (Instituto Adolfo Lutz CCIAL 023)

Table 1. Viral strains and Cell Lines used for the plaque forming assay.

{8}------------------------------------------------

To perform the recovery assay, virus stock was 10-fold serially diluted into pooled negative clinical nasal matrix. Each sample dilution was transferred into ALB LUZ Viral Transport Media by using Rayon Tip Swab and stored at 2-8℃ and 20-25℃ for 0, 24 and 48 hours respectively. Three lots of VTM with newly manufactured, mid-range lot and close to expiry lots were used to evaluate viral recovery. For tissue culture, MDCK (ATCC CCL-34) and MRC-5 (Instituto Adolfo Lutz CCIAL 023) were seeded in 24-well plates and grown to a confluent monolayer. To perform the plaque forming assay, the viral samples from ALB LUZ Viral Transport Media were serially diluted 10-fold and added to the monolayer in triplicate and allowed for virus adsorption by incubating the plates for 1 hour at 37°C. Semisolid medium containing DMEM, 1% carboxymethyl cellulose, and 1% penicillin/streptomycin antibiotics were added to each well and incubated at 37°C for 5-7 days. For counting the plaques developed, the cell fixation was performed with 10% paraformaldehyde and the staining of cell monolayer was performed with 1% crystal violet. Plaque-forming units were counted visually. The viral titer of each sample was recorded and calculated in plaque-forming units per mL (PFU/mL) and the results from the triplicates were provided as an average PFU/mL for 0, 24 and 48 hours. Results were considered acceptable if the average viral recovery demonstrates any percent changes within ±90% (i.e.,1 log change).

The results are presented in the Table 2 and Table 3 below. Any reduction in the viral count in the timepoints was shown as percent changes and -ve indicates reduction.

Viral StrainTestedLotsAverage recovery in PFU/mLPercentPercent
0 hr.24 hrs.48 hrs.changes(0 to 24 hrs.)*changes(0 to 48 hrs.) *
Influenza A1133645.56E+045.04E+043.60E+04-9%-35%
1117775.76E+045.40E+043.67E+04-6%-36%
1113074.97E+044.97E+043.07E+040%-38%
RespiratorySyncytial Virus1133642.00E+061.78E+061.11E+06-11%-45%
1117771.78E+061.55E+061.11E+06-13%-38%
1113071.56E+061.67E+067.78E+057%-50%
Rhinovirus1133642.33E+062.44E+061.11E+065%-52%
1117771.89E+061.89E+061.22E+060%-35%
1113072.67E+062.45E+061.78E+06-8%-33%

Table 2. Viral Recovery for specimen storage at 2-8°C.

  • -ve indicates reduction

Table 3. Viral Recovery for specimen storage at 20-25°C.

Viral StrainTestedLotsAverage recovery in PFU/mLPercentPercent
0 hr.24 hrs.48 hrs.changes(0 to 24 hrs.)*changes(0 to 48 hrs.) *
Influenza A1133645.08E+044.70E+043.39E+04-7%-33%
1117774.87E+044.04E+042.98E+04-17%-39%
1113074.73E+043.81E+042.97E+04-20%-37%
RespiratorySyncytial Virus1133642.22E+062.11E+061.33E+06-5%-40%
1117771.56E+061.33E+068.89E+05-15%-43%
1113071.55E+061.56E+069.89E+050%-36%
Rhinovirus1133641.89E+061.55E+061.33E+06-18%-29%

{9}------------------------------------------------

1117772.67E+062.11E+061.44E+06-21%-46%
1113072.11E+062.11E+061.22E+060%-42%
  • -ve indicates reduction

The ALB LUZ Viral Transport Media demonstrated the recovery of tested viruses (Influenza A, RSV and RV), in all replicates at tested incubation times and storage conditions.

Conclusion:

The documentation submitted in this premarket notification demonstrates that the subject device K232454 has comparable features and performance and is substantially equivalent to the predicate device.

§ 866.2390 Transport culture medium.

(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).