K Number
K232440

Validate with FDA (Live)

Device Name
Adravision Perio
Manufacturer
Date Cleared
2023-12-05

(113 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Adravision Perio is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.

The intended users of Adravision Perio are trained professionals including, but not limited to, dental hygienists.

The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Adra Corporation has not evaluated the performance of the device on primary dentition.

Device Description

Adravision Perio is a Software as a Medical Device (SaMD intended to aid dentists in the measurement of mesial and dista bone levels associated with each tooth on bitewing and periapical dental radiographs of permanent teeth in patients 22 years of age or older.

The device provides assistance to dentists by suggesting the measurement of interproximal bone levels. The device recommendations are for reference only and the user should review associated radiographs, view annotations and modify annotations if needed. The device is not intended as a replacement for a complete clinician's review or their clinical judgment because users can take into account other relevant information from the image, intraoral exam or patient history.

AI/ML Overview

The Adravision Perio device provides semi-automated image processing software to aid dental professionals in measuring mesial and distal bone levels from bitewing and periapical radiographs. This analysis details the acceptance criteria and the study proving the device meets these criteria.

1. Acceptance Criteria and Reported Device Performance

The core performance of the device was assessed using two main metrics related to bone level measurements from radiographs: Object Key Similarity (OKS) and Length Measurement.

OKS Assessment Acceptance Criteria & Performance:

MetricAcceptance Criteria (Not Explicitly Stated, Inferred from Performance)Reported Device Performance (Bitewing)Reported Device Performance (Periapical)
PrecisionHigh (e.g., >80-85%)91.0%84.8%
RecallHigh (e.g., >85-90%)94.0%89.3%

Note: The document doesn't explicitly state quantitative acceptance criteria for Precision and Recall, but the reported performance values indicate that the device met internal thresholds for demonstrating sufficient accuracy in key point detection.

Length Measurement Assessment Acceptance Criteria & Performance:

MetricAcceptance Criteria (Not Explicitly Stated, Inferred from Performance)Reported Device Performance (Bitewing)Reported Device Performance (Periapical)
SensitivityHigh (e.g., >85-90%)90.7 %92.5 %
SpecificityHigh (e.g., >85-90%)94.3 %86.8 %
Mean Absolute ErrorLow (e.g., < 0.6 mm)0.434 mm0.504 mm

Note: Similar to the OKS assessment, specific numeric acceptance criteria are not provided, but the reported values demonstrate strong performance across these metrics, suggesting the device met the manufacturer's internal performance goals for clinical utility.

Overall Tooth Number Classification Accuracy:

MetricAcceptance Criteria (Not Explicitly Stated)Reported Device Performance
Overall Tooth Number Classification Accuracy (OKS Study)High (e.g., >90%)91.8%
Overall Tooth Number Classification Accuracy (Length Assessment)High (e.g., >90%)93.2%

2. Sample Sizes and Data Provenance

  • OKS Assessment Test Set: 340 dental radiographs (bitewing and periapical radiographs).
  • Length Measurement Assessment Test Set: 78 dental radiographs.
  • Data Provenance: The radiographs were "collected retrospectively." No specific country of origin is mentioned, but the intended patient population is stated as "patients living in the United States."

3. Number of Experts and Qualifications for Ground Truth

  • OKS Assessment: "a ground truther" (singular). No specific qualifications are listed for this individual.
  • Length Measurement Assessment: "three (3) ground truthers." No specific qualifications are listed for these individuals.

4. Adjudication Method for the Test Set

  • OKS Assessment: Not explicitly stated, but implies a single ground truther, so no formal adjudication method (e.g., 2+1, 3+1) is indicated.
  • Length Measurement Assessment: The document states "consensus CAL length measurements labeled by three (3) ground truthers," implying an adjudication process to reach a consensus, but the specific method (e.g., simple majority, discussion to resolve discrepancies) is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was mentioned or performed. The studies described are standalone performance evaluations against established ground truths, not studies comparing human reader performance with and without AI assistance.

6. Standalone Performance

Yes, a standalone (algorithm only without human-in-the-loop performance) was performed. The "Bench Testing" section describes evaluations of the Adravision Perio device's performance against ground truth data, focusing on precision, recall, sensitivity, specificity, and mean absolute error, which are metrics typically used for standalone algorithm performance.

7. Type of Ground Truth Used

The ground truth for both OKS and Length Measurement assessments was based on expert labeling/consensus.

  • For OKS, it was "key points labeled by a ground truther."
  • For Length Measurement, it was "consensus CAL length measurements labeled by three (3) ground truthers."

8. Sample Size for the Training Set

The document does not provide information on the sample size used for the training set. It only describes the test sets used for performance evaluation.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth was established for the training set. The descriptions are limited to the establishment of ground truth for the test sets.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 5, 2023

Adra Corporation % Moeini Golnaz Regulatory Consultant GM Consulting 1425 Golden Meadow Square SAN JOSE, CA 95117

Re: K232440

Trade/Device Name: Adravision Perio Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: November 9, 2023 Received: November 9, 2023

Dear Moeini Golnaz:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232440

Device Name Adravision Perio

Indications for Use (Describe)

Adravision Perio is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The intended users of Adravision Perio are trained professionals including, but not limited to, dental hygienists.

The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Adra Corporation has not evaluated the performance of the device on primary dentition.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "adravision". The logo consists of a stylized, abstract symbol to the left of the word "adravision". The symbol is a teal-colored, curved shape that resembles a stylized letter "C" or a crescent moon. The word "adravision" is written in a sans-serif font, with all letters in lowercase.

510(k) Summary (K232440)

This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

Date:Dec 1, 2023
Applicant:MedEdge ConsultingGolnaz Moeinigolnaz.moeini@gmail.com408-504-3187
Trade Name:Adravision Perio
Common Name:Dental Image Processing Software
Regulation Name:Medical Image Management and Processing System
Regulation Number:21 CFR 892.2050
Product Code: DeviceQIH
Classification: DeviceClass II
Description:Adravision Perio is a Software as a Medical Device (SaMD intended to aid dentists in the measurement of mesial and dista bone levels associated with each tooth on bitewing and periapical dental radiographs of permanent teeth in patients 22 years of age or older.The device provides assistance to dentists by suggesting the measurement of interproximal bone levels. The device recommendations are for reference only and the user should

device should review associated radiographs, view annotations and modify annotations if needed. The device is not intended as a replacement for a complete clinician's review or their clinical judgment because users can take into account other relevant information from the image, intraoral exam or patient history.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Adravision. The logo consists of a stylized, abstract symbol to the left of the word "adravision". The symbol is a teal-colored, curved shape that resembles a stylized letter "C" or a crescent moon. The word "adravision" is written in a simple, sans-serif font and is colored black.

Indications for Use:Adravision Perio is a radiological semi-automated imageprocessing software device intended to aid dental professionalsin the measurements of mesial and distal bone levels associatedwith each tooth from bitewing and periapical radiographs. Itshould not be used in-lieu of full patient evaluation or solelyrelied upon to make or confirm a diagnosis.The intended users of Adravision Perio are trained professionalsincluding, but not limited to, dentists and dental hygienists.The intended patient population of the device is patients livingin the United States, who are 22 years old or older, and that donot have any remaining primary teeth. Adra Corporation has notevaluated the performance of the device on primary dentition.
Predicate Device:Overjet Inc., Overjet Dental Assist, 510(k): K210187Trade Name: Overjet Dental AssisstClassification Name: Medical Image Management and Processing SystemRegulation number: 21 CFR 892.2050, Product Code: LLZ, Class II

Substantial Equivalence:

Feature/FunctionProposed DeviceAdravision Perio DevicePrimary PredicateOverjet Dental AssistK210187Identical?
Data sourceBitewing and periapicalradiographsBitewing and periapicalradiographsYes
Image inputsources1Images imported from theradiographic device, orfrom the practicemanagement system, ormanually uploaded into thesystemImages imported from theradiographic device, or fromthe practice managementsystemNo
Imageformat2DICOM, JPEG, JPG, PNG,TIFF, TIF, JFIF or BMPfilejpg, png, jfif, eop, etp, jifNo
Platform3Web- Edge, Chrome,Firefox, Brave, SafariWeb - Edge, Chrome, FirefoxNo
End userDentist, dental hygienistDentist, dental hygienistYes
PatientpopulationThe intended patientpopulation of the device isThe intended patientpopulation of the device isYes

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Adravision. The logo consists of a stylized, abstract shape in a light blue color, resembling a crescent or a curved ribbon. To the right of the shape is the word "adravision" in a sans-serif font, with the letters in black.

Feature/FunctionProposed DeviceAdravision Perio DevicePrimary PredicateOverjet Dental AssistK210187Identical?
patients living in the UnitedStates, who are 22 years oldor older, that do not haveany remaining primaryteeth.patients living in the UnitedStates, who are 22 years oldor older, and that do not haveany remaining primary teeth.
OperatingsystemAnyAnyYes
UserInterfaceMouse, keyboard, trackpadMouse, keyboard, trackpadYes
PatientDatabaseCompatibilitySQLSQLYes
ImageViewingFull, thumbnailFull, thumbnailYes
FeaturesdetectedBone levelBone levelYes
Model outputLine segmentsLine segmentsYes

Discussion of Differences

Both the proposed and the predicate device are designed to be similar in intended use, indications for use, intended users, and intended use environment. Both devices utilize machine learning technologies to measure interproximal bone levels in dental radiographs and allow users to visualize the radiographs with annotations, add their own annotations and use the information as part of their diagnostic decision.

Other similarities include both software have no direct contact with the patient and both systems utilize standard image types.

A few technological differences between the proposed and the predicate device are discussed below. These differences are deemed to not raise questions of safety and efficacy are detailed below.

Image input sources- - Both the proposed device and predicate support image input sources from the radiographic device and the practice management system. The proposed device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Adravision. The logo consists of a stylized, teal-colored graphic to the left of the word "adravision" in a sans-serif font. The graphic resembles a crescent shape formed by two curved lines.

additionally accepts manual upload of radiographs into the software which the predicate device does not support. The V&V testing for the device demonstrates the device's capability in supporting the analysis of all image input sources. This minor difference does not raise new questions of safety and efficacy compared to the predicate device.

Image formats2 - Both the proposed device and predicate support image formats of JPG, PNG, and JFIF. The proposed device additionally accepts DICOM. JPEG, TIFF, TIF and BMP image formats that the predicate device does not support. The V&V testing for the device demonstrates the device's capability in supporting the analysis of all the designated image formats and demonstrates that this minor difference does not raise new questions of safety and efficacy compared to the predicate device.

Platform2 - Both the proposed device and predicate device support common browsers of Edge, Chrome and Firefox. The proposed device additionally works on Safari and Brave. The V&V testing for the device demonstrates the device's capability to function on designated browsers and demonstrates that this minor difference does not raise new questions of safety and efficacy compared to the predicate device.

Performance Testing:

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a Basic Documentation Level, since a failure or latent flaw could indirectly result in minor injury to the patient or user through incorrect or delayed information through the action of a care provider.

Bench Testing

Adra completed standalone studies according to the predefined protocols to demonstrate the safety and effectiveness of the Adravision Perio device for its indications for use.

Bench testing included evaluation of performance based upon a ground truth data set utilizing (i) Object Key Similarity (OKS) Assessment, and

(ii) Length Measurement Assessment.

In the OKS Assessment. Adravision Perio was evaluated with a total dataset of 340 dental radiographs (bitewing and periapical radiographs) and evaluated the precision and recall against the key points labeled by a ground truther within those radiographs.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for "adravision". The logo consists of a stylized, abstract symbol to the left of the word "adravision". The symbol is a teal-colored, curved shape that resembles a stylized eye or a crescent moon. The word "adravision" is written in a simple, sans-serif font in black.

MetricBitewingPeriapical
Precision91.0%84.8%
Recall94.0%89.3%

In the Length Measurement Assessment, Adravision Perio was evaluated with a dataset of 78 dental radiographs and evaluated the sensitivity and specificity against the consensus CAL length measurements labeled by three (3) ground truthers within those radiographs.

MetricBitewingPeriapical
Sensitivity90.7 %92.5 %
Specificity94.3 %86.8 %
Mean Absolute Error0.434 mm0.504 mm

Analysis of study results for patient age, gender, clinic site, region of the mouth and sensor type were performed as subgroup analysis.

Adravision Perio also obtained an overall tooth number classification accuracy of 91.8% in the OKS study and 93.2% in the Length Assessment demonstrating that it can well locate the CAL measurements on the relevant tooth number and tooth surface.

Our bench testing showed that Adravision Perio software met the prespecified performance requirements when evaluated against ground truth labeling and demonstrated robustness and reliability in drawing CAL line segments in clinically relevant locations with accurate length on the relevant tooth.

No adverse effects or complications were recorded associated with this bench study since all radiographs were collected retrospectively and no treatment decisions were made.

Conclusion:

Adravision Perio is substantially equivalent to the predicate device, Overjet Dental Assist. Differences do not raise any concerns about the safety or efficacy of the device.

N/A