K Number
K232436

Validate with FDA (Live)

Device Name
Rapid SDH
Manufacturer
Date Cleared
2023-10-25

(72 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Rapid SDH is a radiological computer aided triage and notification software indicated for use in the triage and notification of hemispheric SDH in non-enhanced head images. The device is intended to assist trained radiologists in workflow triage by providing notification of suspected findings of hemispheric Subdural Hemorrhage (SDH) in head CT images. Rapid SDH uses an artificial intelligence algorithm to analyze images and highlight cases with suspected hemispheric SDH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected hemispheric SDH findings include compressed preview images, that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of Rapid SDH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Device Description

Rapid SDH is a radiological computer-assisted triage and notification software device. The Rapid SDH module is a Non-Contrast Computed Tomography (NCCT) processing module which operates within the integrated Rapid Platform to provide triage and notification of suspected hemispheric sub-dural hemorrhage (SDH). The Rapid SDH module is an Al/ML module. The output of the module is a priority notification to clinicians indicating the suspicion of SDH based on positive findings. The Rapid SDH module uses the basic services supplied by the Rapid Platform including DICOM processing, job management, imaging module execution and imaging output including the notification and compressed image.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for iSchemaView, Inc.'s Rapid SDH:

Executive Summary of Device Purpose:
Rapid SDH is a radiological computer-aided triage and notification software that uses an AI algorithm to identify suspected hemispheric Subdural Hemorrhage (SDH) in non-enhanced head CT images. Its primary function is to assist radiologists in workflow triage by providing rapid notifications, not for diagnostic purposes.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Goal)Reported Device Performance (with 95% Confidence Interval)
Primary Endpoint: Exceed 80% performance goal (presumably sensitivity, as it's the most critical for triage of potentially urgent cases)Sensitivity: 0.924 (0.871 - 0.956)
Specificity: 0.987 (0.954 - 0.996)
ROC AUC (using Rapid SDH Volume estimate): 0.995 (0.986, 1.0)
Secondary Endpoint: Median processing time to notify clinician of 45 secondsMedian Processing Time: 45 seconds (min: 33 seconds, max: 107 seconds)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 310 samples (147 positive cases, 163 negative cases).
  • Data Provenance: Retrospective, multinational study. Specific countries are not listed, but various sites are named (e.g., Gradient, Riverside Regional Medical Center, Image Core Lab, Augusta University Medical Center, Ascension, D3, Segmed, Baptist, Hospital de Clinicas de POA, Stanford CA, Ospedale Regionale di Lugano, NYU, Flagler Hospital, MUSC).

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Number of Experts: Three (3)
  • Qualifications of Experts: Neuro-radiologists. No further details on years of experience or other specific qualifications are provided in this document.

4. Adjudication Method for the Test Set

The adjudication method used to establish ground truth is not explicitly stated in the provided document beyond "Truth was established using three (3) expert neuro-radiologists." Common methods like 2+1 or 3+1 (where dissenting opinions require a tie-breaker or consensus review) are not detailed. It implies a consensus approach among the three experts.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to evaluate how human readers improve with AI vs. without AI assistance. The study described focuses on the standalone performance of the AI algorithm.
  • Effect size of human reader improvement: Not applicable, as no MRMC study comparing human readers with and without AI assistance was reported.

6. Standalone Performance Study (Algorithm Only)

  • Was a standalone study done? Yes, the performance data presented is for the standalone (algorithm-only) performance of the Rapid SDH software in identifying SDH in CT scans. The primary endpoints (sensitivity, specificity, AUC) and secondary endpoint (processing time) are all metrics of the algorithm's performance without human intervention in the loop for the performance evaluation itself.

7. Type of Ground Truth Used

  • Type of Ground Truth: Expert consensus. The document states: "Truth was established using three (3) expert neuro-radiologists." This indicates that the ground truth labels for the presence or absence of SDH were determined by the agreement of these medical professionals.

8. Sample Size for the Training Set

  • The document does not specify the sample size used for the training set. It only describes the test set used for performance validation.

9. How Ground Truth for the Training Set Was Established

  • The document does not detail how the ground truth for the training set was established. It only describes the ground truth establishment for the test set. Given it's an AI/ML module, it's highly likely that a similar expert review process would have been used for training data, but it's not explicitly stated.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

iSchemaView, Inc. Jim Rosa SVP Regulatory and Quality 1120 Washington Ave. Suite 200 Golden, Colorado 80401

October 25, 2023

Re: K232436

Trade/Device Name: Rapid SDH Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: OAS Dated: October 5, 2023 Received: October 5, 2023

Dear Jim Rosa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,
Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232436

Device Name Rapid SDH

Indications for Use (Describe)

Rapid SDH is a radiological computer aided triage and notification software indicated for use in the triage and notification of hemispheric SDH in non-enhanced head images. The device is intended to assist trained radiologists in workflow triage by providing notification of suspected findings of hemispheric Subdural Hemorrhage (SDH) in head CT images. Rapid SDH uses an artificial intelligence algorithm to analyze images and highlight cases with suspected hemispheric SDH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected hemispheric SDH findings include compressed preview images, that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of Rapid SDH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Contraindications/Exclusions/Cautions:

· Rapid SDH is one input to physician diagnosis for patients >21 years old undergoing screening for hemispheric SDH, both acute and chronic.

· Rapid SDH is validated against hemispheric SDH ≥ 1ml.

  • Subdural hygroma, subdural empyema, and subdural effusion mimics were not included in the validation data set.
  • · Excessive patient motion may lead to artifacts that make the scan technically inadequate.

· For use with non-contrast scans. Presence of intravenous contrast may lead to false positive indication of suspected hemispheric SDH.

• Identification of suspected findings is not for diagnostic use beyond notification. Images that are previewed through email and the mobile application are compressed and are for informational purposes only and not intended for diagnostic use beyond notification.

• Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary

iSchemaView, Inc.'s Rapid SDH

This document contains the 510(k) summary for the iSchemaView Rapid SDH. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

Applicant Name and Address:

Name:iSchemaView, Inc.
Address:1120 Washington AveSte. 200Golden, CO 80401
Official Contact:Jim RosaPhone: (303) 704-3374Email: rosa@ischemaview.com

August 11, 2023 Summary Preparation Date:

Device Name and Classification:

Trade Name:Rapid SDH
Common Name:Radiological Computer-Assisted Triage AndNotification Software (CADt)
Classification:II
Product Code:QAS
Regulation No:21 C.F.R. §892.2080
ClassificationPanel:Radiology Devices

Predicate Devices:

The iSchemaView Rapid is claimed to be substantially equivalent to the following legally marketed predicate device: Rapid ICH (K221456).

Device Description:

Rapid SDH is a radiological computer-assisted triage and notification software device. The Rapid SDH module is a Non-Contrast Computed Tomography (NCCT) processing module which operates within the integrated Rapid Platform to provide triage and notification of suspected hemispheric sub-dural hemorrhage (SDH). The Rapid SDH module is an Al/ML module. The output of the module is a priority notification to clinicians indicating the suspicion of SDH based on positive findings. The Rapid SDH module uses the basic services supplied by the Rapid Platform including DICOM processing, job management, imaging module execution and imaging output including the notification and compressed image.

{6}------------------------------------------------

iSchemaView - Traditional 510(k) Rapid SDH 510(k) Summary

Indications for Use:

Rapid SDH is a radiological computer aided triage and notification software indicated for use in the triage and notification of hemispheric SDH in non-enhanced head images. The device is intended to assist trained radiologists in workflow triage by providing notification of suspected findings of hemispheric Subdural Hemorrhage (SDH) in head CT images.

Rapid SDH uses an artificial intelligence algorithm to analyze images and highlight cases with suspected hemispheric SDH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected hemispheric SDH findings. Notifications include compressed preview images, that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of Rapid SDH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Contraindications/Exclusions/Cautions:

  • Rapid SDH is one input to physician diagnosis for patients >21 years old undergoing screening for . hemispheric SDH, both acute and chronic.
  • . Rapid SDH is validated against hemispheric SDH ≥ 1ml.
  • . Subdural hygroma, subdural empyema, and subdural effusion mimics were not included in the validation data set.
  • . Excessive patient motion may lead to artifacts that make the scan technically inadequate.
  • . For use with non-contrast scans. Presence of intravenous contrast may lead to false positive indication of suspected hemispheric SDH.
  • . Identification of suspected findings is not for diagnostic use beyond notification. Images that are previewed through email and the mobile application are compressed and are for informational purposes only and not intended for diagnostic use beyond notification.
  • Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer . and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.

Comparison of Technological Characteristics:

Rapid SDH does not raise new questions of safety or effectiveness compared to the previously cleared predicate, Rapid ICH (K221456). Both devices use machine learning algorithms to determine the presence of intracranial hemorrhage and notify clinicians to the suspicion without removing the case from normal workflow processing. Rapid SDH has a minor difference from Rapid ICH, regarding a singular indication for the hemispheric sub-

{7}------------------------------------------------

510(k) Summary

dural hemorrhage (SDH) subtype vs multiple hemorrhage types within the predicate. Based on the comparison the subject device is substantially equivalent to the predicate device. The features are compared in the following table:

Substantial Equivalence Table
ComparisonFeatureRapid ICH (K221456)Rapid SDH
Indicationsfor UseRapid ICH is a radiologicalcomputer aided triage andnotification software indicated foruse in the analysis of non-enhancedhead CT images. The device isintended to assist hospital networksand trained radiologists inworkflow triage by flagging andcommunication of suspectedpositive findings of pathologies inhead CT images, namelyIntracranial Hemorrhage (ICH).Rapid ICH uses an artificialintelligence algorithm to analyzeimages and highlight cases withdetected ICH on a standalonedesktop application in parallel tothe ongoing standard of care imageinterpretation. The user ispresented with notifications forcases with suspected ICH findings.Notifications include compressedpreview images that are meant forinformational purposes only andnot intended for diagnostic usebeyond notification. The devicedoes not alter the original medicalimage and is not intended to beused as a diagnostic device.The results of Rapid ICH areintended to be used in conjunctionwith other patient information andbased on professional judgment, toassist with triage/prioritization ofmedical images. Notified cliniciansare responsible for viewing fullimages per the standard of care.Rapid SDH is a radiologicalcomputer aided triage andnotification software indicated foruse in the triage and notification ofhemispheric SDH in non-enhancedhead images. The device isintended to assist trainedradiologists in workflow triage byproviding notification of suspectedfindings of hemispheric SubduralHemorrhage (SDH) in head CTimages.Rapid SDH uses an artificialintelligence algorithm to analyzeimages and highlight cases withsuspected hemispheric SDH on aserver or standalone desktopapplication in parallel to theongoing standard of care imageinterpretation. The user ispresented with notifications forcases with suspected hemisphericSDH findings. Notificationsinclude compressed previewimages, that are meant forinformational purposes only andnot intended for diagnostic usebeyond notification. The devicedoes not alter the original medicalimage and is not intended to beused as a diagnostic device.The results of Rapid SDH areintended to be used in conjunctionwith other patient information andbased on professional judgment, toassist with triage/prioritization of
medical images. Notifiedclinicians are responsible forviewing full images per thestandard of care.Note: See limitations in IFUstatement above
Stroke/HeadHemorrhagic Stroke/HeadHemispheric Sub-DuralHemorrhage/Head
Removal ofcases fromworklistqueueNoNo
PrimaryImagingModalitiesNCCTNCCT
TechnicalImplementationML/AI/Neural NetworkML/AI/Neural Network
Segmentationof ROINo, the device does not highlight ordirect a user's attention to a specificlocation in the image file.No, the device does not highlight ordirect a user's attention to a specificlocation in the image file.
PreviewImagesPresentation of a preview of thestudy for initial assessment notmeant for diagnostic purposes.The device operates in parallel withthe standard of care, which remains.Presentation of a preview of thestudy for initial assessment notmeant for diagnostic purposes.The device operates in parallel withthe standard of care, which remains.
PrimaryUser(s)RadiologistRadiologist
Alteration oforiginalimage databaseNoNo
AltersStandard ofCareWorkflowIn parallel toIn parallel to
Notification/PrioritizationYes - PACS, WorkstationYes - PACS, Workstation, email,mobile

{8}------------------------------------------------

510(k) Summary

{9}------------------------------------------------

510(k) Summary

Performance Standards:

Rapid has been developed in conformance with the following standards, as applicable:

ISO 14971:2019Application of Risk Management to Medical Devices
IEC 62304:2015Medical device software – Software lifecycle processes
IEC 62366:2015Application of Usability Engineering to Medical Devices
NEMA PS 3.1 - 3.20Digital Imaging and Communications in Medicine (DICOM)

Rapid has been designed to meet the cybersecurity requirements using design Vulnerability Assessments, SBOM's, and PEN Testing.

Performance Data:

Rapid complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.

iSchemaView conducted a retrospective, blinded, multinational study with Rapid SDH with the primary endpoint to evaluate the software's performance in identifying CT scans containing sub-dural intracranial hemorrhage (SDH). The performance data is derived from 310 samples with 147 positives and 163 negatives. Truth was established using three (3) expert neuro-radiologists.

The primary endpoint of the study was to exceed 80% performance goal. Sensitivity (Se) was measured at Se: 0.924 (95% CI: 0.871 - 0.956) and Sp: 0.987 (95% CI: 95477 - 0.996). The RoC/AUC analysis using Rapid SDH Volume estimate as a predictor of Suspected SDH is AUC: 0.995 (0.986, 1.0):

Image /page/9/Figure/10 description: The image is a plot of sensitivity versus 1-specificity, also known as a receiver operating characteristic (ROC) curve. The x-axis represents 1-specificity or the false positive rate, ranging from 0 to 1.00. The y-axis represents sensitivity or the true positive rate, also ranging from 0 to 1.00. The curve is close to the top left corner, indicating a high level of accuracy.

In addition, a secondary endpoint was to show median processing time to notify the clinician of 45 seconds with min of 33 seconds and maximum of 107 seconds.

{10}------------------------------------------------

510(k) Summary

Performance across subtype differentiation shows:

GenderMeasureNEstimateLower 95% CIUpper 95% CI
FemaleSensitivity430.9300.8140.976
FemaleSpecificity670.9850.9200.997
MaleSensitivity1030.9320.8660.967
MaleSpecificity840.9880.9360.998

Performance Metrics by Gender

Performance Metrics by Age Groups

Age GroupMeasureNEstimateLower 95% CIUpper 95% CI
Age ≤ 50Sensitivity180.7780.5480.910
Specificity261.0000.8711.000
50 < Age < 70Sensitivity510.9220.8150.969
Specificity551.0000.9351.000
Age ≥ 70Sensitivity800.9500.8780.980
Specificity620.9680.8910.991

Performance by SDH Subtype: Chronic, Acute, Mixed

TypeMeasureNEstimateLower 95% CIUpper 95% CI
ChronicSensitivity470.9150.8010.966
AcuteSensitivity480.8750.7530.941
Mixed†Sensitivity620.9680.8900.991

Performance by Hemorrhage Mix

Hemorrhage SubtypeMeasureNEstimateLower 95% CIUpper 95% CI
Subdural OnlySensitivity1190.9330.8730.966
Subdural + OtherSensitivity380.8950.7590.958
Other, non-subduralSpecificity501.000.9291.00

Performance Metrics by Volume

VolumeMeasureNEstimateLower 95% CIUpper 95% CI
< 10 mlSensitivity290.7240.5430.853
≥ 10.0 mlSensitivity1280.9690.9220.988

{11}------------------------------------------------

510(k) Summary

Slice ThicknessMeasureNEstimateLower 95% CIUpper 95% CI
Slice Thickness ≤ 2.5Sensitivity470.9360.8280.978
Specificity671.0000.9461.000
2.5 < Slice Thickness < 5Sensitivity390.9740.8680.995
Specificity341.0000.8981.000
Slice Thickness = 5Sensitivity670.9850.9200.997
Specificity601.0000.9401.000

Performance Metrics by Slice Thickness

Performance Metrics by Manufacturer

ManufacturerMeasureNEstimateLower 95% CIUpper 95% CI
GESensitivity520.9420.8440.980
GESpecificity351.0000.9011.000
PHILIPSSensitivity190.7890.5670.915
PHILIPSSpecificity411.0000.9141.000
TOSHIBASensitivity470.9150.8010.966
TOSHIBASpecificity351.0000.9011.000
SIEMENSSensitivity390.9740.8680.995
SIEMENSSpecificity420.9520.8420.987

Site based information.

Many of the sites have inclusion of either positive or negative cases (blinded information during validation), for the positive case biased sites Se is provided; for the non-positive case sites Sp is provided in those sites where enough samples are included to provide statistical relevance:

SourceTPFPFNTNTotalMeasureEstimate95%LCI95%UCI
Gradient5904164Se0.9370.8480.975
Riverside Regional MedicalCenter5103743Sp0.9740.8650.995
Image Core Lab8032536Sp1.0000.8671.000
Augusta University MedicalCenter0023133Sp1.0000.8901.000
Ascension3100031Se1.0000.8901.000
D31002728Sp1.0000.8751.000
Segmed2201023Se0.9570.7900.992
Baptist0011314Sp1.0000.7721.000
Hospital de Clinicas de POA4001014Sp1.0000.7221.000
Stanford CA701311Se0.875

{12}------------------------------------------------

510(k) Summary

SourceTPFPFNTNTotalMeasureEstimate95% LCI95% UCI
Ospedale Regionale di Lugano71008Se1.000
NYU00033
Flagler Hospital00011
MUSC10001

The performance validation for achievement of effective triage by the Rapid SDH image analysis algorithm as well as effective notification functionality of the Rapid SDH application, as compared to the standard of care for improved time-to-exam-open of a notified case was met.

Prescriptive Statement:

Caution: Federal law restricts this device to sale by or on the order of a physician.

Safety & Effectiveness:

Rapid SDH has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2012 (risk management) and the software development process conforms to ISO 62304:2015.

Conclusion:

In conclusion, iSchemaView's Rapid SDH is substantially equivalent in technological characteristics, safety, and performance characteristics to the legally marketed predicate device, Rapid SDH (K221456).

N/A