K Number
K232405

Validate with FDA (Live)

Device Name
ENDOFIX EXO
Manufacturer
Date Cleared
2024-04-25

(259 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ENDOFIX EXO is indicated for use in interventions where a rigid endoscope is indicated for use. The device is intended to hold and position a rigid endoscope / exoscope during general and neurological diagnostic and therapeutic procedures.

Device Description

The ENDOFIX EXO is a manually operated surgical device intended to hold endoscopes during general and neurological diagnostic and therapeutic procedures. It emulates an arm which is capable of guiding and positioning an endoscopic camera with several degrees of movement.

lt ensures stable and sensitive positioning of the endoscope and allows the surgeon to work one-handed and/or two-handed. The system provides surgeons with a stable and rocksteady field of vision, even in extreme endoscopic positions. The wide scope of movement provides surgeons with full visibility of the situs. The ENDOFIX exo can be used with both, an endoscope and exoscope.

An integrated DEACTIVATION PUSHBUTTON permits manual movement of the ENDOFIX EXO at the push of a button. In order to reposition a guided endoscopic camera, the ENDOFIX EXO is unlocked by a sterilizable CONTROL at the push of a button. The ENDOFIX EXO is operated by unlocking all 6 axes and subsequent manual positioning using the CONTROL. As soon as the button on the CONTROL is released, all 6 axes are locked again.

Despite its large scope of movement, the ENDOFIX EXO is light-weight and compact and is fastened directly to the OR table by means of a quick-acting clamp. Thus, repositioning with reference to the patient is not necessary even if the OR table is moved in the meantime.

On the rigid endoscope, there is a camera linked that sends a video to a monitor. Endoscope and camera as well as the monitor are not part of the ENDOFIX EXO.

To allow the surgeon to convert to an open surgery in case of an emergency, the ENDOFIX EXO can be completely mounted or removed by a quick fastener.

During use, the ENDOFIX EXO is covered with a STERILE COVER. The ENDOFIX EXO can be removed from the OR table and stored on the TROLLEY.

AI/ML Overview

The provided text is a 510(k) Summary for the medical device ENDOFIX EXO. It details the device's characteristics, comparison to a predicate device, and the non-clinical tests performed. However, it does not describe any acceptance criteria or studies involving AI / human reader performance.

Here's an analysis based on the provided text for the points that can be addressed:

The document clearly states that the ENDOFIX EXO is a manually operated surgical device. This means the device itself does not involve AI or algorithms for diagnosis or analysis, but rather for physical positioning of an endoscope. Therefore, many of the questions related to AI performance, such as acceptance criteria for algorithm performance, sample size for test/training sets, ground truth establishment for AI, MRMC studies, or standalone algorithm performance, are not applicable to this device as described.

The document discusses non-clinical performance tests to demonstrate the safety and effectiveness of the device compared to its predicate.

Here's a breakdown of the relevant information provided:

1. A table of acceptance criteria and the reported device performance
The document does not provide a formal table of acceptance criteria with reported performance values for specific metrics. Instead, it lists the types of non-clinical tests performed and broadly states that these tests were successful in demonstrating safety and effectiveness.

Non-Clinical Tests Performed:

  • Lifetime test
  • Lifetime test CONTROL
  • Moving after fixation test
  • Usability test
  • Package drop test

Reported Device Performance (Implicit):
The summary states, "The non-clinical tests have shown that the ENDOFIX EXO device is as safe and effective as the predicate device." It also notes specific findings related to safety:

  • Electromagnetic locks were mechanically tested according to IEC 60601-1:2005 + A1:2012, with no risks regarding safety observed.
  • Displacement after fixation of the arm was tested, with no risks regarding safety observed.
  • The device was tested with 5kg load, though its safe working load is specified as 1kg, and "no problems occurred."
  • The absence of a battery in ENDOFIX EXO ensures continuous power operation safety. In case of power failure, the arm continues to hold position, and quick-release fasteners allow safe removal.
  • Electrical safety tested according to IEC 60601-1, posing no danger to the user.
  • Reprocessing and sterilization process validated, confirming no increased risk.
  • Usability test with 15 American physicians did not negatively mention weight, despite the device being heavier than the predicate.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Sample Size: Not specified for most tests (e.g., lifetime test, moving after fixation test, package drop test).
    • For the Usability Test: "15 American physicians."
  • Data Provenance:
    • Country of Origin: The usability test involved "15 American physicians," suggesting US-based data. For other tests, the origin is not explicitly stated but the applicant is based in Germany.
    • Retrospective or Prospective: Not explicitly stated but implied to be prospective testing for device validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This question is not applicable as this relates to AI/diagnostic performance where expert consensus is needed for ground truth. This device is a mechanical endoscope holder.
  • However, for the Usability Test, "15 American physicians" were involved. Their specific qualifications beyond being "physicians" are not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This question is not applicable as there is no diagnostic or interpretative task requiring adjudication. The tests described are mechanical, electrical, and usability in nature.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable as the device is a mechanical endoscope holder and does not involve AI for diagnostic assistance or human-in-the-loop performance improvement.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as there is no algorithm or AI component in this device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • This question is not applicable as the "ground truth" concept as typically applied to diagnostic AI is not relevant here. The "truth" for the mechanical/electrical tests would be adherence to engineering specifications and safety standards (e.g., proper functioning after a lifetime test, no displacement after fixation, meeting electrical safety standards).

8. The sample size for the training set

  • This question is not applicable as there is no AI/algorithm component requiring a training set.

9. How the ground truth for the training set was established

  • This question is not applicable as there is no AI/algorithm component requiring ground truth for a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 25, 2024

AKTORmed GmbH Hanna Kafurke Regulatory Affairs/Quality Management Neugablonzer Strasse 13 Neutraubling, BY 93073 Germany

Re: K232405

Trade/Device Name: Endofix Exo Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCV Dated: August 1, 2023 Received: August 10, 2023

Dear Hanna Kafurke:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.04.25 16:18:53
-04'00' Mark Trumbore Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K232405

Device Name

ENDOFIX EXO (171943)

Indications for Use (Describe)

The ENDOFIX EXO is indicated for use in interventions where a rigid endoscope is indicated for use. The device is intended to hold and position a rigid endoscope / exoscope during general and neurological diagnostic and therapeutic procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" in white, followed by the word "AKTORmed" in gray, and the words "ROBOTIC SURGERY" in a smaller font below. The letters "TM" are in the upper right corner of the word "AKTORmed".

Dokumentennummer / Document Number Q002-244988 Version / Version 00 Ersteller / Created by Erstellt Datum / Created Date 22.04.2024 Dokumentenname / Document name Ablageort / File

Hanna Kafurke 510(k) Summary 1DA9486B9E2CC68

freigegeben / released 22.04.2024

FunktionNameDatumSignaturVermerk
QM / RegulatoryAffairsHanna Kafurke22.04.2024Hanna KafurkeQ102-244968-00erstellt/created
GeschäftsführerManaging DirectorRobert Geiger22.04.202422.04.2024 15:20:27Robert GeigerQ002-244988-0022.04.2024 15:20:36freigegeben/released
Version /versionÄnderung / ChangesName /nameDatum / Date
00Erstellung / createdHK22.04.2024

510(k) Summary

Inhaltsverzeichnis / contents

1.510(k) Summary
2.Contact Details
3.Device Name
4.Legally Marketed Predicate Devices
5.Device Description Summary
6.Intended use/Indications for Use
7.Indications for Use Comparison
8.Technological Comparison
9. Non-Clinical and/or Clinical Tests Summary & Conclusions

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" in white on the left side. To the right of the circle is the word "AKTORmed" in gray, with the words "ROBOTIC SURGERY" in a smaller font below it. The letters "TM" are in superscript to the right of the word "AKTORmed".

1. 510(k) Summary

510(k) Number: K232405

2. Contact Details

21 CFR 807.92(a)(1)

Applicant NameAKTORmed GmbH
Applicant AddressNeugablonzer Strasse 13, Neutraubling, BY, 93073, Germany
Applicant Contact Telephone+49 9401 9320 150
Applicant ContactMrs. Hanna Kafurke
Applicant Contact EmailHanna.kafurke@aktormed.com

3. Device Name

21 CFR 807.92(a)(2)

Device Trade NameENDOFIX EXO (171943)
Common NameEndoscope and accessories
Classification NameEndoscope Holder
Regulation Number876.1500
Product CodeOCV

4. Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #K190576
Predicate Trade Name(Primary Predicate is listedfirst)WalterLorenz Surgical Assist Arm Scope Holder
Product CodeOCV

5. Device Description Summary

21 CFR 807.92(a)(4)

The ENDOFIX EXO is a manually operated surgical device intended to hold endoscopes during general and neurological diagnostic and therapeutic procedures. It emulates an arm which is capable of guiding and positioning an endoscopic camera with several degrees of movement.

lt ensures stable and sensitive positioning of the endoscope and allows the surgeon to work one-handed and/or two-handed. The system provides surgeons with a stable and rocksteady field of vision, even in extreme endoscopic positions. The wide scope of movement provides surgeons with full visibility of the situs. The ENDOFIX exo can be used with both, an endoscope and exoscope.

An integrated DEACTIVATION PUSHBUTTON permits manual movement of the ENDOFIX EXO at the push of a button. In order to reposition a guided endoscopic camera, the ENDOFIX EXO is unlocked by a sterilizable CONTROL at the push of a button. The ENDOFIX EXO is operated by unlocking all 6 axes and subsequent manual positioning using the CONTROL. As soon as the button on the CONTROL is released, all 6 axes are locked again.

Despite its large scope of movement, the ENDOFIX EXO is light-weight and compact and is fastened directly to the OR table by means of a quick-acting clamp. Thus, repositioning with reference to the patient is not necessary even if the OR table is moved in the meantime.

On the rigid endoscope, there is a camera linked that sends a video to a monitor. Endoscope and camera as well as the monitor are not part of the ENDOFIX EXO.

To allow the surgeon to convert to an open surgery in case of an emergency, the ENDOFIX EXO can be completely mounted or removed by a quick fastener.

Dokumentennummer / document numberQ002-244988
Version / version00
Datum / date22.04.2024
Seite / page2 / 4

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for AKTORmed Robotic Surgery. The logo consists of a blue circle with the letters "am" in white inside of it. To the right of the circle is the word "AKTORmed" in gray, with the words "ROBOTIC SURGERY" in a smaller font below it.

During use, the ENDOFIX EXO is covered with a STERILE COVER. The ENDOFIX EXO can be removed from the OR table and stored on the TROLLEY.

6. Intended use/Indications for Use

21 CFR 807.92(a)(5)

The ENDOFIX EXO is indicated for use in interventions where a rigid endoscope is indicated for use. The device is intended to hold and position a rigid endoscope / exoscope during general and neurological diagnostic and therapeutic procedures.

7. Indications for Use Comparison

21 CFR 807.92(a)(5)

The indications for use are very similar for both devices have the same intended use, which is the fixation of an endoscope in a position. The ENDOFIX EXO is for use during general and neurological diagnostic and therapeutic procedures. The WalterLorenz Surgical Assist Arm can be used in general and neurological, diagnostic and therapeutic procedures. However, this does not create a new intended use, since the is identical for both devices. For WalterLorenz Surgical Assist Arm Scope Holder the user is a trained medical professional, for ENDOFIX EXO the user is defined as a trained surgeon performing the operation. Both statements mean the same thing, they are just worded differently. The Patient group for the ENDOFIX EXO is all patients, who are suitable within the scope of indications for use. There are no known patient limitations for the WalterLorenz Surgical Assist

Arm Scope Holder. Both devices should only be used in professional healthcare environment.

8. Technological Comparison

21 CFR 807.92(a)(6)

Both products have for the most part the same technological characteristics:

Both devices have a permanently installed unlock button, which releases all locks so that the arm can move freely. The unlock button of the ENDOFIX exo is called DEACTIVATION and is located at the distal end of the arm. In contrast, the button of the WalterLorenz Surgical Assist arm is located on the Control Box with User Interface Keypad. However, both devices have a second control panel with an unlock button that is connected to the base unit of the arm via a cable. The control unit of the WalterLorenz Surgical Assist Arm is a sterile, disposable accessory that is supplied separately and can be attached to the distal end of the arm as a release button. The CONTROL of the ENDOFIX EXO is slid over the magnetic bracket of the endoscope clamp and secured from slipping off by a magnetic attraction. When the status LED next to the DEACTIVATION PUSHBUTTON indicates, the CONTROL is ready to operate. This does not happen until the CONTROL is on the magnetic bracket of the endoscope clamp, because there is a sensor in the CONTROL that detects the magnet and only then activates the CONTROL. This increases safety for the patient, as the locks cannot be released if the CONTROL is not on the endoscope clamp. For both devices the Energy Source is Electricity and both do not emit energy. The control units are in the devices. The ENDOFIX exo as well as the WalterLorenz Surgical Assist Arm can be fixed on U.S. table rails. They have no software and do not include biologics, drugs, coatings and additives. For both devices, the sterilizable accessories can be sterilized by steam sterilization and cleaned by wiping disinfection. Regarding the power supply, both devices have the identical input voltage of 100-240 VAC; 50-60Hz and the operating mode is continuous.

Nevertheless, there are few differences of technological characteristics:

The first difference is the mechanism of the two devices. While the WalterLorenz Surgical Assist Arm uses electromechanical brakes in the axes, the ENDOFIX EXO uses electromagnetic brakes to lock movement in all axes. However, there are no safety or effectiveness implications or other safety issues arising from this difference, as the electromagnetic locks have been mechanically tested according to IEC 60601-1:2005 + A1:2012. In addition, the displacement after fixation of the arm has also been tested. From both tests, there are no risks regarding the safety of the arm.

Dokumentennummer / document numberQ002-244988
Version / version00
Datum / date22.04.2024
Seite / page3 / 4

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Aktormed logo. The logo consists of a blue circle with the letters "am" in white, followed by the word "AKTORmed" in gray. Below the word "AKTORmed" is the phrase "ROBOTIC SURGERY" in a smaller, gray font. The logo is clean and modern, and the colors are professional and trustworthy.

The ENDOFIX EXO is compatible with endoscopes. For the WalterLorenz Surgical Assist Arm Scope Holder, no exact specification of which scopes are compatible could be found. By specifying exactly which scopes are compatible, the ENDOFIX EXO has at least an equivalent level of safety in terms of compatibility with devices as the WalterLorenz Surgical Assist Arm Scope Holder.

The ENDOFIX EXO's safe working load is specified as 1kg, it was even tested with 5kg and no problems occurred. No defined safe working load was found for the WalterLorenz Surgical Assist Arm.

The WalterLorenz Surgical Assist Arm is battery powered. The ENDOFIX EXO, on the other hand, does not have a battery, which means that there is no risk of the battery failing during an operation and the safety of continuous power operation is ensured. However, this means that the ENDOFIX EXO can no longer be moved in the event of a power failure, but it has been assured that the arm will continue to hold the endoscope in a fixed position. In the event that power cannot be restored by reconnection, both the endoscope itself and the device can be safely and quickly removed from the patient using quick-release fasteners as described in the instructions for use.

The weights of the two devices differ significantly. The ENDOFIX EXO weighs 26.5lbs, the WalterLorenz Surgical Assist Arm without the Scope Holder 8lbs. However, the higher weight does not lead to an increased safety risk, as the ENDOFIX EXO is equipped with a counterbalance to compensate for the endoscope and the camera head. In addition, the ENDOFIX EXO is mounted to the operating table so that no weight has to be carried by one person. If the ENDOFIX EXO has to be transported over a long distance, it can be mounted on the trolley provided for this purpose and then only needs to be pushed. A usability test was also conducted with 15 American physicians, where weight was not mentioned negatively.

The output voltage of the two devices varies slightly. The ENDOFIX EXO has an output voltage of 24V DC and the WalterLorenz Surgical Assist Arm has an output voltage of 18V DC. However, the ENDOFIX EXO has been sufficiently tested according to IEC 60601-1 for electrical safety to state that it poses no danger to the user.

The last difference is that the distal unlock button of the WalterLorenz Surgical Assist Arm is a single-use product and comes sterile packed. The control unit of the ENDOFIX EXO, on the other hand, can be used more often, but must be cleaned and sterilized before the first use. The complete validation of the reprocessing and sterilization process confirms that there is no increased risk.

9. Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

The following tests were performed:

    1. Lifetime test
    1. Lifetime test CONTROL
    1. Moving after fixation test
    1. Usability test
    1. Package drop test

Nonclinical performance tests of the substantial device are unknown. Therefore own nonclinical performance tests were successfully performed to demonstrate the safety and efficiency of the ENDOFIX EXO.

The non-clinical tests have shown that the ENDOFIX EXO device is as safe and effective as the predicate device. For the WalterLorenz Surgical Assist Arm there are unfortunately no which non-clinical performance tests have been performed. However, since the WalterLorenz Surgical Assist Arm is only a Class 1 device, it can be assumed that all non-clinical performance tests for the ENDOFIX EXO are sufficient to ensure at least equivalent safety.

Dokumentennummer / document numberQ002-244988
Version / version00
Datum / date22.04.2024
Seite / page4 / 4

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.