K Number
K232388

Validate with FDA (Live)

Date Cleared
2023-11-21

(104 days)

Product Code
Regulation Number
870.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

High Pressure Tubing is indicated for use in PTCA surgery to connect the angiographic catheter and the angiographic syringe to deliver contrast media.

Device Description

High Pressure Tubing is designed as a connecting line for injection of contrast media during coronary angiographic procedures. The device has a pressure resistance of 1200psi (8.3MPa); It is divided into two series according to the luer connector combination. ETM series consists of tubing and rotating male luer connecters, and the ETF series consists of tubing and rotating male luer connecter; The rotating male luer connector and female luer connector are made of polycarbonate and silica gel; The tubing is made of polyurethane, and the middle of the tube wall also contains nylon braided wire, which does not contact with outside. The High is for single use only, non-pyrogenic, sterilized by EO gas and the sterilization process is validated.

AI/ML Overview

The question asks to describe the acceptance criteria and the study that proves the device meets the acceptance criteria based on the provided text. However, the provided text is a 510(k) Pre-market Notification for a "High Pressure Tubing" device, which describes its comparison to a predicate device and includes information about performance testing, but it does not contain information typically found in a study proving an AI/ML device meets acceptance criteria.

Specifically, the document discusses a medical device that is a physical product (High Pressure Tubing) used in PTCA surgery. It does not mention any AI or machine learning components. Therefore, information regarding "sample sizes used for the test set," "data provenance," "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," "sample size for training set," or "how ground truth for training set was established" is completely absent and not applicable to this document.

The document primarily focuses on establishing substantial equivalence to a predicate device by comparing technological characteristics and presenting results of biocompatibility and bench testing.

Based on the provided text, the following information can be extracted regarding the physical device's performance testing, which is analogous to "acceptance criteria" and "device performance" for a non-AI medical device:

1. Table of Acceptance Criteria and Reported Device Performance (as inferred from the bench testing section):

Acceptance Criteria (Test)Reported Device Performance (Conclusion)
Biocompatibility (ISO 10993-1: 2018)Performed and ensured biocompatibility (Cytotoxicity, Sensitization, Irritation, Pyrogenicity, Acute systemic toxicity, Hemocompatibility)
Visual InspectionPerformed and implied satisfactory
Dimension AccuracyPerformed and implied satisfactory
Particulate ContaminationPerformed and implied satisfactory
LeakagePerformed and implied satisfactory
Tensile StrengthPerformed and implied satisfactory
6% Luer Connector compliancePerformed and implied satisfactory (presumably to ISO 80369-7: 2021)
EO ResidualPerformed and implied satisfactory
ECH ResidualPerformed and implied satisfactory
SterilityPerformed and implied satisfactory
EndotoxinPerformed and implied satisfactory
Pouch VisualPerformed and implied satisfactory
Pouch Seal Peel StrengthPerformed and implied satisfactory
Pouch IntegrityPerformed and implied satisfactory

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Not Applicable. The document describes performance and bench testing of a physical medical device, not a software or AI/ML device. Specific sample sizes for each test are not provided, nor is data provenance in the context of clinical studies (e.g., country of origin, retrospective/prospective). The tests were conducted to assure reliable design and performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. This information is relevant for AI/ML device validation where human experts establish ground truth. For this physical device, the "ground truth" is established by adherence to recognized international standards (e.g., ISO 10993, ISO 8536-4, ISO 80369-7, ISO 80369-20).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. Adjudication methods are used in clinical studies or expert reviews for AI/ML devices to resolve discrepancies in ground truth establishment. This document pertains to physical device testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. MRMC studies are specific to AI/ML devices involving human readers. This document is about a physical high-pressure tubing.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This concept is only relevant for AI/ML algorithms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Not Applicable. For this physical device, "ground truth" is adherence to established engineering specifications and international standards for medical device safety and performance.

8. The sample size for the training set:

  • Not Applicable. This device is a physical product and does not involve AI/ML models that require training data.

9. How the ground truth for the training set was established:

  • Not Applicable. This device is a physical product and does not involve AI/ML models that require training data.

In summary, the provided document is a 510(k) Pre-market Notification for a conventional (non-AI/ML) medical device. Therefore, most of the questions related to AI/ML device validation are not applicable. The "acceptance criteria" and "device performance" for this specific device are demonstrated through adherence to relevant international standards and successful completion of biocompatibility and bench testing as listed above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 28, 2024

Ningbo DIZEGENS Medical Science Co., Ltd. Zhang Wenwei Vice General Manager Floor 3&4, Building A4, No. 777, Binhai 4th Road Qianwan New Area. Ningo, Zhejiang 315336 China

Re: K232388

Trade/Device Name: High Pressure Tubing Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II Product Code: DXT

Dear Zhang Wenwei:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 21, 2023. Specifically, FDA is updating this SE Letter due to a typographical error in the company name listed in the letter as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Shruti Mistry, Assistant Direction Devices at 301-796-6605 or Shruti.Mistry@fda.hhs.gov.

Sincerely,

Juliane C. Lessard -S

Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 21, 2023

Ningbo DIZENGENS Medical Science Co., Ltd. Zhang Wenwei Vice General Manager Floor 3&4, Building A4, No. 777, Binhai 4th Road Qianwan New Area. Ningo, Zhejiang 315336 China

Re: K232388

Trade/Device Name: High Pressure Tubing Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II Product Code: DXT Dated: October 27, 2023 Received: October 27, 2023

Dear Zhang Wenwei:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{2}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Juliane C. Lessard -S

Juliane C. Lessard Director, Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices

{3}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K232388

Device Name High Pressure Tubing

Indications for Use (Describe)

High Pressure Tubing is indicated for use in PTCA surgery to connect the angiographic catheter and the angiographic svringe to deliver contrast media.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary

1. Submitter/510(k) Holder

Submission: Traditional 510(k) Premarket Notification Submitter: Ningbo DIZEGENS Medical Science Co., Ltd. Address: Floor 3&4, Building A4, No. 77, Binhai 4th Road, Qianwan New Area, Ningbo, Zhejiang Province, P.R. China. 315336 Contact Person: Zhang Wenwei, Vice General Manager Telephone: +86-574-82359758 Telefax: +86-574-82359059 Email: edwardzhang@dizegens.com Date prepared: Oct 27, 2023

2. Subject Device

Device Trade Name: High Pressure Tubing Device Common Name: High Pressure Tubing Regulatory Name: Angiographic injection and syringe Regulation Number: 21 CFR 870.1650 Product Code: DXT Regulatory Class: II Review Panel: Cardiovascular 510(k) Number: K232388

3. Predicate Device

The High Pressure Tubing is substantially equivalent to the following legally marketed predicate device.

510(k) NumberTrade NameManufacturer
K170014High Pressure TubingShanghai Kindly Medical Instruments Co., Ltd.

4. Device Description

High Pressure Tubing is designed as a connecting line for injection of contrast media during coronary angiographic procedures. The device has a pressure resistance of 1200psi (8.3MPa); It is divided into two series according to the luer connector combination. ETM series consists of tubing and rotating male luer connecters, and the ETF series consists of tubing and rotating male luer connecter; The rotating male luer connector and female luer connector are made of polycarbonate and silica gel; The tubing is made of polyurethane, and the middle of the tube wall also contains nylon braided wire, which does not contact with outside. The High is for single use only, non-pyrogenic, sterilized by EO gas and the sterilization process is validated. The High Pressure Tubing has 12 models and the differences for each model are listed in the following table 1.

{6}------------------------------------------------

Table 1 Models of High Pressure Tubing
No.ModelLength(cm)Luer ConnectorCombinationPressureResistance(psi)
1ETF-02525Rotating Male LuerConnector + Female LuerConnector1200
2ETF-05050
3ETFSeriesETF-07575
4ETF-100100
5ETF-125125
6ETF-150150
7ETM-02525Rotating Male LuerConnector +Rotating MaleLuer Connector
8ETM-05050
9ETMSeriesETM-07575
10ETM-100100
11ETM-125125
12ETM-150150

Table 1 Models of High Pressure Tubing

5. Indications for Use

High Pressure Tubing is indicated for use in PTCA surgery to connect the angiographic catheter and the angiographic syringe to deliver contrast media.

6. Comparison of Technological Characteristics with the Predicate Device

The subject device, High Pressure Tubing, and the predicate device, High Pressure Tubing of Shanghai Kindly medical instruments Co., Ltd. have different intended use, indications for use, principle of operation but the differences do not raise additional concerns of safety and effectiveness. The subject device and predicate device have identical characteristics including components, technology and design, pressure rating, conical fitting, flexible, biocompatibility, sterilization, sterile, single use.

Difference in technology characteristics between the subject device, High Pressure Tubing, and predicate device, High Pressure Tubing, are provided in Table 2.

DescriptionHigh Pressure Tubing(Predicate Device)High Pressure Tubing(Subject Device)Comparison topredicate device
510(k) NumberK170014K232388N/A
RegulationNumber21 CFR 870.165021 CFR 870.1650Same
Regulatory NameAngiographic injection and syringeAngiographic injection and syringeSame
Product CodeDXTDXTSame
Regulatory ClassIIIISame
Intended UseThe High Pressure Tubing is indicated foruse as a connecting line for injection ofcontrast media or saline during coronaryangiographic procedures.High Pressure Tubing is indicated for usein PTCA surgery to connect theangiographic catheter and theangiographic syringe to deliver contrastmedia.Different1
Indications for UseThe High Pressure Tubing is indicated foruse as a connecting line for injection ofcontrast media or saline during coronaryangiographic proceduresHigh Pressure Tubing is indicated for usein PTCA surgery to connect theangiographic catheter and theangiographic syringe to deliver contrastmedia.Different1
Principle ofOperationManual connect the male luer or rotatingmale luer of high pressure with othercatheter device.1. Check whether the package iscomplete; Do not use if the unitpackage has been damaged;2. Open the package and check theproduct., do not use if the product isdamaged or incomplete;3. Connect the luer connector with thecorresponding device;4. After the infusion of liquid, checkwhether there is leakage at theconnecting position; If there isleakage, it is necessary to checkwhether the connection of the deviceis loose and reconnect it firmly, orreplace it with a new product.Different2
Technology andDesignDesigned to multiple length and connectortypes. PU tubing with polyamides liner for1200psiDesigned to multiple length andconnector types. PU tubing with nylonbraided wire for 1200psi.Different3
ComponentsTubing, female/male luer, rotating adapterTubing, Female Luer Connector, RotatingMale Luer ConnectorSame
Size30cm,60cm,90cm,120cm,150cm,25cm,50cm,75cm,100cm,125cm,150cm,Different4
Pressure Rating1200psi1200psiSame
Conical FittingComply with ISO 80369-7-2021Comply with ISO 80369-7: 2021Same
FlexibleYesYesSame
BiocompatibilityISO 10993 seriesISO 10993 seriesSame
SterilizationEO sterilizationEO sterilizationSame
SterileYesYesSame
Single UseYesYesSame

Table 2 Technological Characteristics Comparison

T

{7}------------------------------------------------

Note': The subject device is intended to deliver contrast media in PTCA surgery same as predicate device, but not deliver saline which have significant decrease in risk compared to predicate device. This difference does not raise new or different questions of safety and effectiveness.

Note2: In Principle of Operation, the step 3 of subject device is identical with the operation of predicate device. In addition, more check steps have introduced to the operations of subject device, including step 1, 2 and 4, which

{8}------------------------------------------------

could decrease the risks in the operation. Therefore, this difference does not raise new or different questions of safety and effectiveness.

Noted: In Technology and Design, PU tubing with Nylon braided wire is the identical material with PU tubing with Polyamides liner that it just is the different description. This difference does not raise new or different questions of safety and effectiveness.

Note4: The sizes of predicate device are 30 to 150cm with 30cm intervals, and the sizes of subject device are 25 to 150cm with 25cm intervals. The minimum length of 25cm is suitable for potential clinical demand, and less length is meaning lower risk. The two devices have same longest configuration, 150cm, which is the most challenge size for characteristics. Therefore, this difference does not raise new or different questions of safety and effectiveness.

Conclusion: The differences between the subject and the predicate device do not raise additional questions of safety and effectiveness.

7. Performance Data

The subject device, High Pressure Tubing, was subjected to the following applicable testing to assure reliable design and performance under the specified testing parameters:

Biocompatibility Testing:

Per ISO 10993-1: 2018 and FDA guidance, the following tests were performed to ensure the biocompatibility of the subject device.

  • . Cytotoxicity
  • Sensitization
  • . Irritation or intracutaneous reactivity
  • Material mediated pyrogenicity ●
  • Acute systemic toxicity ●
  • . Hemocompatibility

Bench Testing:

Per ISO 8536-4: 2019, ISO 80369-7: 2021, ISO 80369-20: 2015, etc., the following tests were performed to ensure performance/functionality of the subject device.

  • . Visual
  • . Dimension
  • Particulate Contamination
  • Leakage ●
  • Tensile Strength ●
  • . 6% Luer Connector
  • . EO Residual
  • ECH Residual
  • Sterility
  • Endotoxin
  • . Pouch Visual
  • Pouch Seal Peel Strength .

{9}------------------------------------------------

. Pouch Integrity

Clinical Tests:

Clinical tests were not required to demonstrate performance of High Pressure Tubing. Product functionality has been adequately assessed by non-clinical tests.

Animal Tests:

Animal tests were not required to demonstrate performance of High Pressure Tubing. Product functionality has been adequately assessed by non-animal tests.

8. Conclusions

Based on the information presented in this 510(k) premarket notification, the subject device, High Pressure Tubing is considered substantially equivalent to the predicate device, High Pressure Tubing of Shanghai Kindly Medical Instruments Co., Ltd.

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.