Intended Use

The Arthrex 2.4 mm Knotless Hip SutureTak® Suture Anchor is intended to be used for suture (soft tissue) fixation to bone in the hip. Specifically, acetabular labral repair.

Device Description

The Arthrex Knotless Hip SutureTak® Suture Anchor is a non-absorbable, “tap-in" suture anchor with a barbed profile and a proximally placed external suture eyelet. The anchors are preloaded with Arthrex Suture and are offered preassembled on an inserter. The device is single-use.

AI/ML Overview

This document describes the 510(k) premarket notification for the "Arthrex 2.4mm Knotless Hip SutureTak® Suture Anchor". It is not an AI/ML medical device, but rather a surgical implant.

Therefore, the requested information (acceptance criteria for an AI/ML device, details of an AI/ML study, sample sizes, expert qualifications, etc.) is not applicable to this document.

The "Performance Data" section in the document refers to mechanical strength testing and bacterial endotoxin testing for a physical medical device, not the performance of an AI/ML algorithm.

{0}------------------------------------------------

September 1, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Arthrex Inc. Alex Underberg Regulatory Affairs Specialist, Senior 1370 Creekside Boulevard Naples, Florida 34108

Re: K232340

Trade/Device Name: Arthrex 2.4mm Knotless Hip SutureTak® Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI, HWC, GAT Dated: August 2, 2023 Received: August 4, 2023

Dear Alex Underberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sara S. Thompson -S

For

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232340

Device Name

Arthrex 2.4 mm Knotless Hip SutureTak® Suture Anchor

Indications for Use (Describe)

The Arthrex 2.4 mm Knotless Hip Suture Anchor is intended to be used for suture (soft tissue) fixation to bone in the hip. Specifically, acetabular labral repair.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date PreparedSeptember 1st, 2023
SubmitterArthrex Inc.1370 Creekside BoulevardNaples, FL 34108-1945
Contact PersonName: Alex UnderbergTitle: Regulatory Affairs Specialist, SeniorEmail: Alex.Underberg@Arthrex.com
Trade NameArthrex 2.4 mm Knotless Hip SutureTak® Suture Anchor
Common NameFastener, Fixation, Nondegradable, Soft Tissue
FDA Product CodeMBI (Primary)HWCGAT
Classification Name21 CFR 888.3040: Fastener, Fixation, Nondegradable,Soft Tissue (Primary)21 CFR 888.3040: Smooth or Threaded Metallic BoneFixation Fastener21 CFR 878.5000: Nonabsorbable Poly(Ethyleneterephthalate) Surgical Suture
Regulatory ClassClass II
Primary Predicate DeviceKnotless SutureTak® Suture Anchor (K120155, S.E.02/17/2012)
Reference DevicesArthrex Short Suture Anchor (K151092, S.E.02/23/2016)
Purpose of SubmissionThis Special 510(k) premarket notification is beingsubmitted to obtain clearance for new Arthrex KnotlessHip SutureTak® Suture Anchors as a line extension topreviously cleared Arthrex Knotless SutureTak® SutureAnchors cleared within K120155, S.E. 02/17/2012.
Device DescriptionThe Arthrex Knotless Hip SutureTak® Suture Anchor is anon-absorbable, “tap-in" suture anchor with a barbedprofile and a proximally placed external suture eyelet.The anchors are preloaded with Arthrex Suture and areoffered preassembled on an inserter. The device issingle-use.
Indications for UseThe Arthrex 2.4 mm Knotless Hip SutureTak® SutureAnchor is intended to be used for suture (soft tissue)fixation to bone in the hip. Specifically, acetabular labralrepair.
Performance DataUltimate load testing was performed on the subject device to demonstrate that the minor modifications to the predicate devices does not negatively impact mechanical strength. Bacterial endotoxin adoption, which met the requirements of <20 EU/device of ST72 Bacterial Endotoxin Tests.
Technological ComparisonThe subject, predicate and reference devices are all intended to be used for soft tissue fixation to bone in the hip, including acetabular labral repair, are all manufactured from identical materials, are all anchors and suture preloaded on an inserter, and are all sterile, single-use devices. The subject device and predicate device both utilize knotless fixation and utilize a similar internal mechanism to provide strength to the knotless fixation. As described within the device description section of this submission, the reference device is a similar device; however, it does not utilize knotless fixation whereas both the subject and predicate utilize knotless fixation.
ConclusionThe Arthrex 2.4 mm Knotless Hip SutureTak® Suture Anchor is substantially equivalent to the predicate devices in which the design features, intended use, and surgical technique are the same. Any differences between the subject and the predicate devices do not raise different questions concerning safety and effectiveness. Based on the indications for use, technological characteristics, and the summary of data submitted, Arthrex Inc. has determined that the proposed device is substantially equivalent to the currently marketed predicate devices.

{4}------------------------------------------------

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.