K Number
K232250

Validate with FDA (Live)

Date Cleared
2024-01-11

(167 days)

Product Code
Regulation Number
880.2750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss.

The Triton AI configuration is intended to be used with surgical sponges, software, hardware and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entent Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use.

The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites.

The Triton QBL (Quantification of Blood Loss) configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items.

Device Description

The SurgiCount+ [SC+] System Software is a multi-functional software application that is intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other surgical substrates. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton Quantitative Blood Loss [QBL]. Additionally, combined workflows (SC+AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. Two of the five software configurations (Triton Al and SC+Al) are Class II functions. The remaining configurations are Class I functions.

The Class II Triton Al software configuration estimates the hemoglobin (Hb) mass contained on used surgical sponges using an AI algorithm that analyzes the image of each sponge. It also calculates an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The Triton Al software configuration's sponge counting functionality has been modified to enhance the product's surgical sponge counting/management functionality, compared to the predicate device. New workflow steps allow users to scan, identify, and count RFID-tagged surgical sponges and other absorbent items, and to locate missing surgical sponges inside the operating room and, noninvasively, in surgical sites.

The SaMD product includes the following nonmedical device and Class I consumable and hardware accessories: a mobile device (Apple iPad Pro), RFID-tagged surgical sponges/absorbent items, an RFID reader, a bluetooth-enabled scale, and a stand (or optional wall mount) that houses the hardware accessories and connects the accessory devices to electrical power.

AI/ML Overview

The provided text describes the Stryker SurgiCount+ System, an image processing device for estimating external blood loss. Here's a breakdown of the acceptance criteria and study details based on the provided FDA 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Hemoglobin (Hb) Algorithm Performance Validation: Limits of agreement between actual hemoglobin mass and sHbL measured by the SurgiCount+ software within an acceptance limit of ±1.99 g Hb.The limits of agreement between the actual hemoglobin mass and sHbL measured by the SurgiCount+ software fell within the acceptance limit of ±1.99 g Hb. (Implicitly, this means it met the criteria).
Sponge Recognition Algorithm (SRA) Performance Validation: Failure rate (sum of false images and failed detection) of less than or equal to 6.5% for a representative 18x18 inch sponge type.The SRA had a failure rate (sum of false images and failed detection) of 0.19%, when used to detect a representative 18x18 inch sponge type. (Well below the acceptance criterion).
Electromagnetic Compatibility (EMC) Testing: Conformance with IEC 60601-1-2 and demonstration of electromagnetic compatibility (EMC) safety and effectiveness in the hospital environment.The SurgiCount+ System conforms with IEC 60601-1-2 and demonstrates electromagnetic compatibility (EMC) safety and effectiveness in the hospital environment.
Wireless Coexistence Testing: Functioned as designed in the presence of Wi-Fi and Bluetooth interferers that were no closer than 30 cm away from the SurgiCount+ System's iPad (Implicitly, no significant interference).The SurgiCount+ System functioned as designed in the presence of Wi-Fi and Bluetooth interferers that were no closer than 30 cm away from the SurgiCount+ System's iPad.
Human Factors Validation: The system is reasonably safe and effective for its intended users and use environment, and all use risks were effectively mitigated. No significant residual or new usability risks were found. (Implicitly, the study findings indicated this).The validation study concluded that the SurgiCount+ System is reasonably safe and effective for its intended users and use environment and that all use risks were effectively mitigated. No significant residual or new usability risks were found.

2. Sample Size Used for the Test Set and Data Provenance

  • Hb Algorithm Performance Validation: The text states, "Testing was conducted to evaluate the accuracy of the Hb algorithm in estimating the Hb mass on surgical sponges, compared to the known Hb mass on each sponge as determined by a reference assay." This implies that a set of surgical sponges was used, each with a "known Hb mass." However, the exact sample size (number of sponges) for this test set is not explicitly stated.
  • Sponge Recognition Algorithm (SRA) Performance Validation: The text states it was "used to detect a representative 18x18 inch sponge type." The specific sample size (number of images or instances) for this test is not explicitly stated.
  • Human Factors Validation: "Fifteen nurses participated in a complete, end-to-end, usability validation."
  • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It describes non-clinical (bench) testing and human factors validation, suggesting controlled environments rather than real-world patient data collection for the primary performance validations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The text does not provide information regarding the number or qualifications of experts used to establish ground truth for the Hb Algorithm or SRA performance validations. The ground truth for the Hb algorithm was established by a "reference assay."
  • For Human Factors Validation, "Fifteen nurses" participated as users, but they were not establishing ground truth, rather their performance and feedback were observed and evaluated.

4. Adjudication Method for the Test Set

  • The document does not specify any adjudication methods (e.g., 2+1, 3+1) for the performance validation test sets. The ground truths were established via objective measures (reference assay for Hb, assumed prior knowledge for SRA, and structured observation for Human Factors).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • No, an MRMC comparative effectiveness study was not described. The performance data provided are for the standalone device capabilities (Hb algorithm accuracy, SRA failure rate) and human factors validation (usability and safety), not a comparative analysis of human readers with vs. without AI assistance. The device is intended as an adjunct to blood loss estimation and sponge management, implying it supports human users, but a formal MRMC study to quantify human improvement with AI assistance is not mentioned.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, standalone performance was evaluated for the core AI components:
    • Hemoglobin (Hb) Algorithm Performance Validation: This tested the algorithm's accuracy in estimating Hb mass compared to a known reference, which is a standalone evaluation of the algorithm's output.
    • Sponge Recognition Algorithm (SRA) Performance Validation: This assessed the algorithm's ability to detect sponges and its failure rate, which is also a standalone evaluation.

7. The Type of Ground Truth Used

  • Hemoglobin (Hb) Algorithm: The ground truth was established by a "reference assay" for the "known Hb mass on each sponge." This implies a laboratory measurement or biochemical analysis.
  • Sponge Recognition Algorithm (SRA): The ground truth implicitly was the actual presence/absence and type of sponge, likely determined by controlled experimental setup where the correct sponge type was presented.
  • Human Factors Validation: The "ground truth" here was the observational data of user performance collected through "clinically realistic scenarios" and user input, evaluated against predetermined safety and usability criteria.

8. The Sample Size for the Training Set

  • The document states, "The algorithms generate the same output for a given input (are fixed) and have been trained using machine learning techniques to recognize the sponges and to estimate Hb mass and blood loss volume on the imaged substrates."
  • However, the sample size for the training set is not provided in the given text.

9. How the Ground Truth for the Training Set Was Established

  • The document states the AI algorithms were "trained using machine learning techniques to recognize the sponges and to estimate Hb mass and blood loss volume on the imaged substrates."
  • While it mentions "known Hb mass" for the a test set, the method for establishing ground truth for the training set is not explicitly described. Given the nature of Hb mass estimation, it is highly probable that the training data also included images of sponges with independently measured "known Hb mass" established through venipuncture and lab analysis, similar to the method used for the testing data. For sponge recognition, it would likely involve labeled images indicating the presence and type of sponges.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 11, 2024

Stryker Instruments Ms. Patti Arndt Principal Specialist. Regulatory Affairs 1941 Stryker Way Portage, Michigan 49002

Re: K232250

Trade/Device Name: SurgiCount+ System Regulation Number: 21 CFR 880.2750 Regulation Name: Image Processing Device For Estimation Of External Blood Loss Regulatory Class: Class II Product Code: PBZ Dated: July 28, 2023 Received: November 13, 2023

Dear Ms. Arndt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.01.11

11 -05:52:45 -05:52:45 -05:52:45 -05:00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232250

Device Name SurgiCount+ System

Indications for Use (Describe)

The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss.

The Triton AI configuration is intended to be used with surgical sponges, software, hardware and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entent Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use.

The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites.

The Triton QBL (Quantification of Blood Loss) configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "stryker" in bold, black letters. The font is sans-serif and the letters are closely spaced together. A small registered trademark symbol is located to the right of the letter "r".

510(k) Summary

-Submitter

Stryker Instruments 1941 Stryker Way Portage, MI 49002 Phone: 269-323-7700, 1-800-253-3210 Email: Kim.Wiersema@stryker.com

II. Contact Person

Ms. Patti L. Arndt Principal Specialist, Regulatory Affairs 1941 Stryker Way Portage, MI 49002 Phone: 269-888-0233 Email: Patti.Arndt@stryker.com

Date Prepared: January 9, 2024

lll. Device

Name of Device:SurgiCount+ System
Common or Usual Name:External Blood Loss Estimation System
Classification Name:Image Processing Device for Estimation of ExternalBlood Loss (21 CFR 880.2750)
Regulatory Class:II
Product Code:PBZ
IV. Predicate DevicePixel 3 Sponge SystemK163507

The predicate device has not been subject to a design related recall. No reference devices were used in this submission.

V. Device Description

The SurgiCount+ [SC+] System Software is a multi-functional software application that is intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other surgical substrates. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton Quantitative Blood Loss [QBL]. Additionally, combined workflows (SC+AI and SC+QBL) are provided for

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and the letters are closely spaced together. There is a registered trademark symbol to the right of the letter "r". The word is black and the background is white.

use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. Two of the five software configurations (Triton Al and SC+Al) are Class II functions. The remaining configurations are Class I functions.

The Class II Triton Al software configuration estimates the hemoglobin (Hb) mass contained on used surgical sponges using an AI algorithm that analyzes the image of each sponge. It also calculates an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The Triton Al software configuration's sponge counting functionality has been modified to enhance the product's surgical sponge counting/management functionality, compared to the predicate device. New workflow steps allow users to scan, identify, and count RFIDtagged surgical sponges and other absorbent items, and to locate missing surgical sponges inside the operating room and, noninvasively, in surgical sites.

The SaMD product includes the following nonmedical device and Class I consumable and hardware accessories: a mobile device (Apple iPad Pro), RFID-tagged surgical sponges/absorbent items, an RFID reader, a bluetooth-enabled scale, and a stand (or optional wall mount) that houses the hardware accessories and connects the accessory devices to electrical power.

VI. Indications for Use

The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponqes and other absorbent items. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ Al and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss.

The Triton Al configuration is intended to be used with surgical sponges, software, hardware, and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use.

The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying, and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are closely spaced together, and the word is in black. There is a registered trademark symbol to the upper right of the word. The background is white.

The Triton QBL (Quantification of Blood Loss) configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items.

VII. Comparison of Technology Characteristics with the Predicate Device

The Indications for Use Statement has been updated but is unchanged with respect to its representation of the product's intended use. New indications for use (i.e., associated medical conditions/diseases/patient populations) were not proposed. The modified product remains an adjunctive technology that is intended to be used to estimate external patient blood loss and in the management of surgical sponges. Kev changes include a product name change and an updated description of the modified SaMD product, neither of which impact the safety and effectiveness of the device.

Essential device functions for the SaMD product include the processing of images for the estimation of external patient blood loss and the management of surgical sponges. The SurgiCount+ System uses the same technology as the predicate device to estimate blood loss; its sponge counting functionality, however, is enhanced.

Image Processing for the Estimation of External Blood Loss: Using a qualified mobile device camera, the SaMD product instructs the user to scan/image blood-containing surgical sponges and other absorbent items used during a medical or surgical procedure. The Application then processes the images using proprietary Al Algorithms. The algorithms generate the same output for a given input (are fixed) and have been trained using machine learning techniques to recognize the sponges and to estimate Hb mass and blood loss volume on the imaged substrates.

Sponge Counting and Management: The SurgiCount+ System's sponge counting function uses ultra-high frequency (UHF) RFID technology to scan, count, and locate RFID-tagged surgical sponges. The handheld, reusable, and wireless RFID reader emits an electromagnetic field that activates items labeled with passive RFID tags within its range. The reader's firmware uses a defined communication protocol, and its antenna transmits RF energy between 902-928 MHz. The Application collects and manages the data received from the reader. By comparing the RFID-tagged absorbent articles "counted in" to a surqical procedure to the articles "counted out," the Application is able to display the number and identity of any missing absorbent articles. The predicate software system "counts" sponges by displaying the number of sponges imaged by the device for the primary purpose of blood loss estimation.

The predicate and subject devices are identical with respect to their intended use and essential performance. As shown in the following table, the key difference between the two devices is the augmentation of the predicate system's sponge counting and management functionality.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the first letter, which is uppercase. A registered trademark symbol is located to the right of the word.

Key Feature/CharacteristicPredicateDevicePixel 3SpongeSystem(K163507)SubjectDeviceSurgiCount+System(K232250)Comparison
Product Type:SaMDProductSaMDProductSame
Intended Use:Image-basedestimation ofexternalpatient bloodloss andsurgicalspongemanagementlmage-basedestimation ofexternalpatient bloodloss andsurgicalspongemanagementSame
Clinical WorkflowsSingleWorkflowMultifunctionalSoftwareDevice withFiveWorkflowsSamePredicate devicefunctionality isreplicated oraugmented in themodified device(Triton Al andSC+AlWorkflows).
Essential FunctionEstimation ofHb/BloodVolume onImagedSurgicalSpongesEstimation ofHb/BloodVolume onImagedSurgicalSpongesSameThe incorporatedAl Algorithm isidentical acrossthe two systems.The subjectdevice meets thesame accuracyspecification asdeveloped andrequired for thepredicate system.
Key Feature/CharacteristicPredicateDevicePixel 3SpongeSystem(K163507)SubjectDeviceSurgiCount+System(K232250)Comparison
Implementation ofSpongeCounting/ManagementFunctionalityDevicecounts anddisplays thenumber ofimagedspongesSystem usesRFID-taggedsurgicalsubstratesand an RFIDbar codereader tocount/manageand locatesurgicalspongesSimilarModificationsaugment/enhancethe basic spongecountingfunctionalityprovided by thepredicate system

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the "S", which is uppercase. A registered trademark symbol is located to the right of the "r" in "stryker". The text is black and the background is white.

VIII. Performance Data

Extensive software verification and validation activities were completed for the SurqiCount+ System. All test cases passed and fully traced to the product's Risk Analysis and Software Design Specification. Unit, regression, and system-level testing also was conducted and confirmed that the software functioned as designed. Cybersecurity design planning and testing ensured that the software was architected and designed to minimize cybersecurity threats.

Non-clinical (bench) testing was conducted to validate the performance of the SurqiCount+ System and demonstrated compliance with the product's Special Controls.

(1) Hemoglobin (Hb) Algorithm Performance Validation

Testing was conducted to evaluate the accuracy of the Hb algorithm in estimating the Hb mass on surgical sponges, compared to the known Hb mass on each sponge as determined by a reference assay. It was shown that the limits of agreement between the actual hemoglobin mass and sHbL measured by the SurgiCount+ software fell within the acceptance limit of ±1.99 q Hb.

(2) Sponge Recognition Algorithm (SRA) Performance Validation

Test results demonstrated that the SRA had a failure rate (sum of false images and failed detection) of 0.19%, when used to detect a representative 18x18 inch sponge type. This was well below the acceptance criterion of less than or equal to 6.5%.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the "S", which is uppercase. A small registered trademark symbol is located to the right of the "r" in "stryker". The word is in black and the background is white.

(3) Electromagnetic Compatibility (EMC)Testing

The SurgiCount+ System conforms with IEC 60601-1-2 and demonstrates electromagnetic compatibility (EMC) safety and effectiveness in the hospital environment.

  • (4) Wireless Coexistence Testing
    Wireless coexistence testing was conducted on the system in accordance with ANSI C63.27 and AAMI TIR 69. The SurgiCount+ System functioned as designed in the presence of Wi-Fi and Bluetooth interferers that were no closer than 30 cm away from the SurgiCount+ System's iPad.

  • (5) Human Factors Validation
    Fifteen nurses participated in a complete, end-to-end, usability validation of the SurqiCount+ System. The validation study included the SurgiCount+ System's critical tasks and use scenarios, the facilitators' evaluation of the users' performance working through clinically realistic scenarios, users' input on usability and safety of the system, and residual use-error risk analysis. The validation study concluded that the SurgiCount+ System is reasonably safe and effective for its intended users and use environment and that all use risks were effectively mitigated. No significant residual or new usability risks were found.

IX. Conclusions

The SurgiCount+ System has the same intended use as the predicate Triton Sponge System. Both the subject and predicate devices are used to estimate blood loss on used surgical sponges and to count/manage surgical substrates. The SurgiCount+ System shares the same technological characteristics as the predicate device and includes new UFH RFID technology that enhances the product's ability to identify, count, and locate surgical sponges. Software and non-clinical performance testing raised no new questions of safety or effectiveness and support the substantial equivalence of the modified device to its legally marketed predicate.

§ 880.2750 Image processing device for estimation of external blood loss.

(a)
Identification. An image processing device for estimation of external blood loss is a device to be used as an aid in estimation of patient external blood loss. The device may include software and/or hardware that is used to process images capturing externally lost blood to estimate the hemoglobin mass and/or the blood volume present in the images.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. Demonstration of the performance characteristics must include a comparison to a scientifically valid alternative method for measuring deposited hemoglobin mass. The following use conditions must be tested:
(i) Lighting conditions;
(ii) Range of expected hemoglobin concentrations;
(iii) Range of expected blood volume absorption; and
(iv) Presence of other non-sanguineous fluids (
e.g., saline irrigation fluid).(2) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by intended users of the device.
(3) Appropriate analysis and non-clinical testing must validate the electromagnetic compatibility (EMC) and wireless performance of the device.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Software display must include an estimate of the cumulative error associated with estimated blood loss values.
(6) Labeling must include:
(i) Warnings, cautions, and limitations needed for safe use of the device;
(ii) A detailed summary of the performance testing pertinent to use of the device, including a description of the bias and variance the device exhibited during testing;
(iii) The validated surgical materials, range of hemoglobin mass, software, hardware, and accessories that the device is intended to be used with; and
(iv) EMC and wireless technology instructions and information.