K Number
K232159

Validate with FDA (Live)

Device Name
QEVO System
Date Cleared
2023-09-21

(63 days)

Product Code
Regulation Number
882.1480
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QEVO System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior spinal procedures, such as nucleotomy, discectory, and foraminotomy.

Device Description

The QEVO System comprises of the QEVO ECU (Endoscope Control Unit) and QEVO endoscope. The system is intended for viewing internal surgical sites and for use in visualization during general and certain neurosurgical and spinal procedures. The QEVO System has to be installed and integrated with a host display device (surgical microscope, a monitor, etc). Requirements for physical integration, connectivity, power supply, display resolution, and software integration are established and tested.

AI/ML Overview

The provided document is a 510(k) premarket notification summary for the QEVO System, declaring its substantial equivalence to a predicate device (QEVO System with KINEVO 900). This type of submission focuses on demonstrating that a new device is as safe and effective as a legally marketed device, primarily by showing similar technological characteristics and intended use.

Crucially, this document does NOT contain information about acceptance criteria or a study proving the device meets those criteria in the context of AI/ML performance testing. The "Summary of Studies" section only mentions:

  • Sterilization and Shelf Life: The device is reusable and the reprocessing instructions are identical to the predicate device.
  • Biocompatibility: Testing was done in accordance with ISO 10993 for the patient-contacting component (insertion tube).
  • Performance Testing - Bench: Optical safety was assessed according to IEC 62471:2006. It explicitly states: "The determination of substantial equivalence was not based on an assessment of performance data." This indicates that no clinical performance study (like an MRMC study or standalone algorithm performance) was submitted or required for this 510(k) clearance, as the device is an endoscopic visualization system, not an AI/ML diagnostic tool.

Therefore, I cannot extract the information required to answer your prompt because the provided text pertains to a traditional medical device (an endoscope system) clearance, not an AI/ML-driven device that would involve the rigorous testing methodologies you've asked about (e.g., ground truth, reader studies, test set sizes, etc.).

To summarize why I cannot provide the requested information based on the given text:

  • No AI/ML Component: The QEVO System is described as a visualization system (endoscope) that displays images. There's no mention of an embedded AI/ML algorithm for image analysis, diagnosis, or decision support.
  • No Performance Data for Clinical Effectiveness: The submission explicitly states that the substantial equivalence determination was not based on performance data. This implies a reliance on technological similarity to the predicate and standard bench testing for safety (electrical, optical) and functionality (image resolution, field of view, etc.).
  • Focus on Substantial Equivalence: The entire document is about demonstrating that the new QEVO System is "substantially equivalent" to an existing predicate device, primarily by comparing their specifications and intended use, rather than proving a new diagnostic capability through clinical performance studies.

If you have a document related to an AI/ML medical device, please provide that, and I would be able to address your specific questions about acceptance criteria, study design, and ground truth establishment.

{0}------------------------------------------------

September 21, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Carl Zeiss Meditec, Inc. % Chaitali Gawde Senior Regulatory Affairs Specialist 5300 Central Parkway Dublin, California 94568

Re: K232159

Trade/Device Name: QEVO System Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: August 23, 2023 Received: August 24, 2023

Dear Chaitali Gawde:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Adam D. Pierce - S" in a large font on the left side of the image. On the right side of the image, the text "Digitally signed by Adam D. Pierce - S" is shown. Below this text, the date "2023.09.21" and time "08:32:09 -04'00'" are shown.

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232159

Device Name QEVO System

Indications for Use (Describe)

The QEVO System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological surgical procedures as well as for viewing internal surgical sites during anterior spinal procedures, such as nucleotomy, discectory, and foraminotomy.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 807 Subpart C)

| > Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.92 the 510(k) Summary for the QEVO System is provided below.

1. SUBMITTER

Applicant:Carl Zeiss Meditec AGGoeschwizer Strasse 51-52D-07745 JenaGermany
Primary CorrespondentChaitali GawdeSenior Regulatory Affairs SpecialistCarl Zeiss Meditec, Inc.5300 Central Parkway Dublin, CA 94568(224) 300-3992 PhoneE-mail: chaitali.gawde@zeiss.com (preferred)
Secondary CorrespondentMaria GolovinaHead of Regulatory AffairsCarl Zeiss Meditec, Inc.5300 Central Parkway Dublin, CA 94568(925) 216-1078 PhoneE-mail: maria.golovina@zeiss.com(preferred)
Date Prepared:

September 21, 2023

{4}------------------------------------------------

DEVICE 2.

Device Trade Name:QEVO System
Common Name:Neurological endoscope
Classification:21 CFR 882.1480 Neurological endoscope
Regulatory Class:II
Product Code:GWG

3. PREDICATE DEVICE (K170667)

Predicate Device:QEVO System with KINEVO 900
Manufacturer:Carl Zeiss Meditec AG
Classification:21 CFR 882.1480 Neurological endoscope
Regulatory Class:II
Product Code:GWG

4. DEVICE DESCRIPTION

The QEVO System comprises of the QEVO ECU (Endoscope Control Unit) and QEVO endoscope. The system is intended for viewing internal surgical sites and for use in visualization during general and certain neurosurgical and spinal procedures.

The QEVO System has to be installed and integrated with a host display device (surgical microscope, a monitor, etc). Requirements for physical integration, connectivity, power supply, display resolution, and software integration are established and tested.

INTENDED USE/INDICATIONS FOR USE న్.

The QEVO System is intended for viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological procedures as well as for viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy.

{5}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE 6.

Table 1. Subject to Predicate Device Comparison Table
---------------------------------------------------------
Device ComparisonTable AttributeQEVOSystem(K232159)Subject DeviceQEVO System with KINEVO 900(K170667)Predicate DeviceEquivalency Analysis
ManufacturerCarl Zeiss Meditec AGCarl Zeiss Meditec AGIdentical
510(k) NumberK232159K170667N/A
Device NameQEVO SystemQEVO System with KINEVO 900N/A
Intended UseThe QEVO System is intended forviewing internal surgical sites duringgeneral surgical procedures and for usein visualization of ventricles andstructures within the brain duringneurological surgical procedures as wellas for viewing internal surgical sitesduring anterior and posterior spinalprocedures, such as nucleotomy,discectomy, and foraminotomy.The KINEVO 900 with QEVO Systemis intended for viewing internal surgicalsites during general surgical proceduresand for use in visualization ofventricles and structures within thebrain during neurological surgicalprocedures as well as for viewinginternal surgical sites during anteriorand posterior spinal procedures, such asnucleotomy, discectomy, andforaminotomy.EquivalentSince, system isdisconnected from aspecific model ofsurgical microscopewhile the QEVO Systemportion remainsidentical.
Classification RegulationProduct Code882.1480, Class IINeurological endoscopeGWG882.1480, Class IINeurological endoscopeGWGIdentical
System ComponentsRigid endoscope, ECURigid endoscope, ECUIdentical
Compatible Host DisplayDeviceAny Zeiss device that is validated to meetthe QEVO requirements for physicalintegration, connectivity, power supply,display resolution, and softwareintegrationKINEVO 900EquivalentConnectivity parametersidentical to those on theKINEVO 900 areavailable on other Zeissdevices.
Light TransmissionLight source in endoscope main body,light transmission through insertion tubevia fiber opticsLight source in endoscope main body,light transmission through insertiontube via fiber opticsIdentical
Device ComparisonTable AttributeQEVOSystem(K232159)Subject DeviceQEVO System with KINEVO 900(K170667)Predicate DeviceEquivalency Analysis
Light SourceIntegrated LED (intensity adjustable)Integrated LED (intensity adjustable)Identical
Image TransmissionRigid rod lenses + CMOS imaging sensorin endoscope main bodyRigid rod lenses + CMOS imaging sensorin endoscope main bodyIdentical
Direction of View45°45°Identical
Field of View100°100°Identical
Depth of Field5-30mm5-30mmIdentical
Image Resolution2 Mega Pixel (Full HD imager)642 TV lines (optical resolution at 15% MTF)2 Mega Pixel (Full HD imager)642 TV lines (optical resolution at 15% MTF)Identical
Image DisplayExternal monitorExternal monitorIdentical
2D / 3D Imaging2D Only2D OnlyIdentical
RecordingVia USB-portVia USB-portIdentical
Insertion Tube WorkingLength120mm120mmIdentical
Insertion Tube OuterDiameter3.6mm3.6mmIdentical
Single Use/ ReusableReusableReusableIdentical
ReprocessingManual and automated cleaning,sterilizationManual and automated cleaning,sterilizationIdentical
Electrical SafetyIEC 60601-1, IEC 60601-1-2, IEC60601-2-18 compliantIEC 60601-1, IEC 60601-1-2, IEC60601-2-18 compliantIdentical

{6}------------------------------------------------

{7}------------------------------------------------

7. SUMMARY OF STUDIES

Sterilization and Shelf Life

The QEVO endoscope is used sterile but is not provided sterilized before first use. It must also be end user cleaned and sterilized between uses. The reprocessing instructions have not been modified and are identical to the predicate device.

Biocompatibility

The only patient contacting component of the subject device is the Insertion tube. The contact category for this component is Tissue/Bone/Dentin Communicating, < 24 hours. The materials were testing in accordance with ISO 10993.

Performance Testing - Bench

There have been no changes in the System Requirements and Verification since the predicate. The following performance testing was provided, to support the substantial equivalence of the subject device:

  • . Optical Safety - The QEVO Endoscope and ECU were assessed for conformity with the relevant requirements of IEC 62471:2006: Photobiological safety of lamps and lamp systems and were found to comply.
    The determination of substantial equivalence was not based on an assessment of performance data.

8. CONCLUSION

The OEVO System and the predicate device are both intended to be used for viewing internal surgical sites during surgical procedures. The indications for use are identical to those of the predicate device with an exception, that the subject device isn't intended to be used with a specific model of surgical microscope.

The technological characteristics and risk profile of the subject device is identical to the predicate device; and therefore, are identical in their relationship to safety and effectiveness.

Testing methods are identical to those of the predicate device; are identical in their relationship to safety and effectiveness.

Therefore, the subject device meets the requirements for substantial equivalence to the predicate device.

§ 882.1480 Neurological endoscope.

(a)
Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.(b)
Classification. Class II (performance standards).