AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn on the hands of healthcare personnel as a barrier for protection against cross-contamination between the healthcare personnel and patient.

These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

This is a disposable natural rubber surgical glove that is tested for use with chemotherapy drugs and fentanyl resistance. The glove is supplied in the following sizes: 5½, 6, 6½, 7, 7½, 8, 8½ and 9.

AI/ML Overview

The provided document describes the acceptance criteria and study results for two types of Natural Rubber Latex Surgical Gloves:

  1. Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green)
  2. Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl

This is a 510(k) Premarket Notification submission, which aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device. The information provided focuses on the performance of the device against established standards (e.g., ASTM standards) rather than a clinical trial with human subjects or an AI-based study. As such, many of the requested points related to AI/MRMC studies, expert ground truth establishment, adjudication methods, and training/test set sample sizes for AI are not applicable to this type of device submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The "Summary of Non-Clinical Testing" (pages 15-16) and "Table 1" (pages 8-14) present the acceptance criteria and the device's performance.

Summary Table of Acceptance Criteria and Reported Device Performance

Characteristic / Test MethodStandard(s)Acceptance CriteriaReported Device Performance
Physical PropertiesASTM D412
Tensile Strength-Before Aging: Min. 24MPa; After Aging: Min. 18MPaBefore Aging: Pass; After Aging: Pass
Ultimate Elongation-Before Aging: Min. 750%; After Aging: Min. 560%Before Aging: Pass; After Aging: Pass
Stress at 500% Elongation-Before Aging: Max. 5.5MPa; After Aging: N/ABefore Aging: Pass; After Aging: N/A
DimensionASTM D3767
LengthEach size satisfies ASTM requirementLength: Each size satisfies ASTM requirementPass (5½: Min. 245mm; 6.0-9.0: Min. 265mm for both devices)
WidthEach size satisfies ASTM requirementWidth: Each size satisfies ASTM requirementPass (e.g., 5½: 70 ± 6 mm; 9.0: 114 ± 6 mm for both devices)
Thickness-Finger: Min. 0.10 mm; Palm: Min. 0.10 mm; Cuff: Min. 0.10 mmPass
WatertightASTM D5151Freedom from holes at AQL 0.65Pass (Meets 21 CFR 800.20 and ASTM D3577 when tested in accordance with ASTM D5151 Inspection Level I, AQL 0.65)
Residual PowderASTM D6124Powder residue limit of ≤2.0 mg per glovePass
Protein ContentASTM D5712≤ 200 µg/dm2Pass
Biocompatibility
Primary Skin IrritationISO 10993-10Not a primary skin irritantNot a primary skin irritation (Passes)
Dermal SensitizationISO 10993-10Not a contact sensitizerNot a contact sensitizer (Passes)
Acute Systemic ToxicityISO 10993-11No biological reactivity shownNo biological reactivity shown (It is concluded that the devices did not induce any acute systemic toxicity)
Material Mediated PyrogenicityISO 10993-11No increase of temperature (Non-pyrogenic)No increase of temperature (It is concluded that the devices are non-pyrogenic)
Cytotoxicity - Direct ContactISO 10993-5Non-cytotoxicNon-cytotoxic for S&C/S&P gloves
Cytotoxicity - MEM ElutionISO 10993-5Non-cytotoxicFor Green gloves: Under the conditions of this study, the test article extract showed potential toxicity to L929 cells. For Strip&Coat/Strip&Pack gloves: Non-cytotoxic for S&C/S&P gloves (Note: This is a discrepancy with the table on page 16, which only shows the Green glove result for MEM Elution. However, the final conclusion states non-cytotoxic for S&C/S&P gloves in the direct contact test, implying potential difference in test methods or interpretation for the two types of gloves).
Bacterial EndotoxinLimulus Amoebocyte Lysate (LAL) test≤20 EU/pair of glovesPass
Chemotherapy Drug PermeationASTM D6978-05Breakthrough Detection Time in Minutes: Generally >240 minutesGreen Glove: Most drugs >240 min. Exceptions: Carmustine (11.9 min), Thiotepa (14.7 min). Strip&Coat/Strip&Pack Glove: Most drugs >240 min. Exceptions: Carmustine (5.7 min), Thiotepa (13.8 min). WARNING labels are provided for Carmustine and Thiotepa.
Fentanyl Citrate PermeationASTM D6978-05Breakthrough Detection Time in Minutes: No breakthrough up to 240 minutesNo breakthrough up to 240 minutes (for both glove types).
Sterility StatusMedical Glove Guidance Manual - LabelingSterileSterile
Sterility MethodISO 11137-1, ISO 11137-2Gamma Radiation (effective)Gamma Radiation (Same as predicate)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the specific number of gloves (sample size) used for each physical, chemical, and biocompatibility test. It refers to compliance with ASTM standards, which typically specify minimum sample sizes for testing (e.g., ASTM D5151 for watertightness requires a specific inspection level and AQL, implying a certain sample size based on lot size). However, the exact sample numbers are not itemized here.
  • Data Provenance: The tests are non-clinical (laboratory tests) performed on the manufactured gloves. The manufacturer is WRP Asia Pacific Sdn. Bhd., located in Malaysia (as per Section 1.0 Submitter). The data is generated from testing of the finished device. It is inherently "prospective" in the sense that the tests were conducted specifically for this submission, validating the device's properties.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not Applicable. This is a medical device submission based on a laboratory evaluation against recognized performance standards (ASTM, ISO, CFR). The "ground truth" for the device's physical and chemical properties is established by these objective, standardized test methods and their defined acceptance criteria, not by human expert interpretation or consensus as would be the case for an AI-enabled diagnostic device.

4. Adjudication Method for the Test Set

  • Not Applicable. As the tests are objective, laboratory-based measurements (e.g., tensile strength, breakthrough time, watertightness), there is no need for expert adjudication. The results are quantitative and compared directly to predetermined numerical acceptance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) and the impact of AI assistance on that interpretation. This submission is for a physical barrier device (surgical gloves), not an AI-enabled diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI algorithm.

7. The Type of Ground Truth Used

  • Ground Truth: The ground truth for this device is established by objective, standardized test methods and their defined acceptance criteria. These standards (e.g., ASTM D3577, ASTM D6978-05, ISO 10993 series) specify a methodology for testing and a quantitative or qualitative result that the device must meet (e.g., minimum tensile strength, maximum protein content, no breakthrough of chemicals). The chemical permeation results are based on the ASTM D6978-05 standard for chemotherapy drug permeation.

8. The Sample Size for the Training Set

  • Not Applicable. There is no "training set" as this is not an AI/machine learning device. The glove is a manufactured product whose performance is evaluated against material and safety standards.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. Since there is no training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 28, 2024

WRP Asia Pacific Sdn. Bhd. % Michael Scaglione U.S. Agent WRP USA Inc 3700 Massillon Road, Suite 340 Uniontown, Ohio 44685

Re: K232079

Trade/Device Name: Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green); Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Regulation Number: 21 CFR 878.4460 Regulation Name: Non-powdered surgeon's glove Regulatory Class: Class I. reserved Product Code: KGO, LZC, OPJ, QDO Dated: March 5, 2024 Received: March 26, 2024

Dear Michael Scaglione:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director

{2}------------------------------------------------

DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232079

Device Name

Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy (Green); Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Slove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl

Indications for Use (Describe)

A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn on the hands of healthcare personnel as a barrier for protection against cross-contamination between the healthcare personnel and patient.

These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

NATURAL RUBBER LATEX SURGICAL GLOVE, POWDER FREE, STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS AND FENTANYL (GREEN):

Chemotherapy DrugConcentration Average Breakthrough Detection Time (Minutes)
Cisplatin (1.0 mg/ml)> 240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)> 240
Dacarbazine (10.0 mg/ml)> 240
Doxorubicin Hydrochloride (2.0 mg/ml)> 240
Etoposide (20.0 mg/ml)> 240
Fluorouracil (50.0 mg/ml)> 240
Ifosfamide (50.0 mg/ml)> 240
Methotrexate (25.0 mg/ml)> 240
Mitomycin C (0.5 mg/ml)> 240
Mitoxantrone (2.0 mg/ml)> 240
Paclitaxel (6.0 mg/ml)> 240
Vincristine Sulfate (1.0 mg/ml)> 240
*Thiotepa (10.0 mg/ml)14.7
*Carmustine (BCNU) (3.3 mg/ml)11.9
Opioid DrugFentanyl Citrate 100 mcg/2mL
Breakthrough Detection Time in MinutesNo breakthrough up to 240 minutes

*WARNING: Testing showed an average breakthrough time of 11.9 minutes for Carmusine and an average breakthrough time of 14.7 minutes for Thiotepa. Do not use with Carmustine and Thiotepa.

NATURAL RUBBER LATEX (STRIP&COAT)/(STRIP&PACK) SURGICAL GLOVE, POWDER FREE, STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS AND FENTANYL:

Chemotherapy DrugAverage Breakthrough Detection Time (Minutes)
Cisplatin (1.0 mg/ml)> 240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)> 240
Dacarbazine (10.0 mg/ml)> 240
Doxorubicin Hydrochloride (2.0 mg/ml)> 240
Etoposide (20.0 mg/ml)> 240

{4}------------------------------------------------

Fluorouracil (50.0 mg/ml)> 240
Ifosfamide (50.0 mg/ml)> 240
Methotrexate (25.0 mg/ml)> 240
Mitomycin C (0.5 mg/ml)> 240
Mitoxantrone (2.0 mg/ml)> 240
Paclitaxel (6.0 mg/ml)> 240
Vincristine Sulfate (1.0 mg/ml)> 240
*Thiotepa (10.0 mg/ml)13.8
*Carmustine (BCNU) (3.3 mg/ml)5.7

Opioid Drug Fentanyl Citrate 100 mcg/2mL Breakthrough Detection Time in Minutes No breakthrough up to 240 minutes

  • WARNING: Testing showed an average breakthrough time of 5.7 minutes for Carmustine and an average breakthrough time of 13.8 minutes for Thiotepa. Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

1.0 Submitter:

Name: Muhammad Ameer Arief bin Mohd Mujab
AddressWRP Asia Pacific Sdn. Bhd.Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru SalakTinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA
Phone No.: +60 3 8706 1486
Fax No.: +60 3 8706 1557
Date of SummaryPrepared: 28th March 2024

2.0 Identification of the Subject Device:

Trade Name:Natural Rubber Latex Surgical Glove, Powder Free,Sterile, Tested for Use with Chemotherapy Drugs andFentanyl (Green); Natural Rubber Latex(Strip&Coat)/(Strip&Pack) Surgical Glove, PowderFree, Sterile, Tested for Use with ChemotherapyDrugs and Fentanyl
Common Name:Surgical Gloves
Classification Name:Non-powdered surgeon's glove
Device Classification:I
Regulation Number:21 CFR 878.4460
Product Code:KGO, LZC, OPJ, QDO

3.0 Predicate Device:

Predicate
ManufacturerWRP Asia Pacific Sdn. Bhd.
Device namePowder-Free Polymer Coated Latex SurgicalGloves, Sterile with Protein Labelling Claim (50Micrograms or Less)
510(k) NumberK020019
Regulatory ClassI
Product CodeKGO

{6}------------------------------------------------

4.0 Description of The Device:

This is a disposable natural rubber surgical glove that is tested for use with chemotherapy drugs and fentanyl resistance. The glove is supplied in the following sizes: 5½, 6, 6½, 7, 7½, 8, 8½ and 9.

5.0 Indication for use:

A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn on the hands of healthcare personnel as a barrier for protection against cross-contamination between the healthcare personnel and patient.

These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Chemotherapy DrugConcentrationMinimumBreakthroughDetection Time(Minutes)
*Carmustine (BCNU)3.3 mg/ml11.9
Cisplatin1.0 mg/ml>240
Cyclophosphamide (Cytoxan)20.0 mg/ml>240
Dacarbazine10.0 mg/ml> 240
Doxorubicin Hydrochloride2.0 mg/ml> 240
Etoposide20.0 mg/ml> 240
Fluorouracil50.0 mg/ml> 240
Ifosfamide50.0 mg/ml> 240
Methotrexate25.0 mg/ml> 240
Mitomycin C0.5 mg/ml> 240
Mitoxantrone2.0 mg/ml> 240
Paclitaxel6.0 mg/ml> 240
*Thiotepa10.0 mg/ml14.7
Vincristine Sulfate1.0 mg/ml> 240

5.1 Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green)

{7}------------------------------------------------

*WARNING: Please note the following drugs have extremely low permeation times: Carmustine (BCNU): 11.9 minutes and Thiotepa: 14.7. Do not use with Carmustine and Thiotepa.

Opioid DrugConcentrationBreakthrough DetectionTime in Minutes
Fentanyl Citrate Injection100mcg/2mLNo breakthrough up to 240minutes

5.2 Natural Rubber Latex Surgical Glove (Strip&Coat)/(Strip&Pack), Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl

Chemotherapy DrugConcentrationMinimumBreakthroughDetection Time(Minutes)
*Carmustine (BCNU)3.3 mg/ml5.7
Cisplatin1.0 mg/ml>240
Cyclophosphamide (Cytoxan)20.0 mg/ml>240
Dacarbazine10.0 mg/ml> 240
Doxorubicin Hydrochloride2.0 mg/ml> 240
Etoposide20.0 mg/ml> 240
Fluorouracil50.0 mg/ml> 240
Ifosfamide50.0 mg/ml> 240
Methotrexate25.0 mg/ml> 240
Mitomycin C0.5 mg/ml> 240
Mitoxantrone2.0 mg/ml> 240
Paclitaxel6.0 mg/ml> 240
*Thiotepa10.0 mg/ml13.8
Vincristine Sulfate1.0 mg/ml> 240

*WARNING: Please note the following drugs have extremely low permeation times: Carmustine (BCNU): 5.7 minutes and Thiotepa: 13.8. Do not use with Carmustine and Thiotepa.

Opiod DrugConcentrationBreakthrough DetectionTime in Minutes
Fentanyl Citrate Injection100mcg/2mLNo breakthrough up to 240minutes

{8}------------------------------------------------

Summary of the Technological Characteristics of the Device: See table 6.0

Table 1

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATECURRENT
510(k) NumberManufacturer(s)-K020019WRP AsiaPacific Sdn.Bhd.K232079WRP AsiaPacific Sdn.Bhd.Same
MaterialASTM D3577Natural RubberLatexNatural RubberLatexSame
Color-Natural WhiteGreen andNaturalDifferent
Texture-MicroroughenedsurfaceMicroroughenedsurfaceSame
Physical PropertiesASTM D3577MeetsMeetsSame
Before AgingTensile Strength:Ultimate Elongation:Stress at 500%Elongation:Min. 24MPaMin. 750%Max. 5.5MPa
After AgingTensile Strength:Ultimate Elongation:Stress at 500%Elongation:MeetsMin. 18MPaMin. 560%N/AMeetsSame
Thickness:- Finger- Palm- CuffASTM D3577MeetsMin. 0.10 mmMin. 0.10 mmMin. 0.10 mmMeetsMin. 0.10 mmMin. 0.10 mmMin. 0.10 mmSame
Dimension, LengthASTM D3577Meets5½: Min.245mm6.0: Min. 265mm6½: Min. 265mm7.0: Min. 265mm7½: Min. 265mm8.0: Min. 265mm8½: Min. 265mm9.0: Min. 265mmMeets5½: Min.245mm6.0: Min. 265mm6½: Min. 265mm7.0: Min. 265mm7½: Min. 265mm8.0: Min. 265mm8½: Min. 265mm9.0: Min. 265mmSame
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATECURRENT
510(k) NumberDimension, WidthASTM D3577K020019Meets5½: 70 ± 6 mm6.0: 76 ± 6 mm6½: 83 ± 6 mm7.0: 89 ± 6 mm7½: 95 ± 6 mm8.0: 102 ± 6 mm8½: 108 ± 6 mm9.0: 114 ± 6 mmK232079Meets5½: 70 ± 6 mm6.0: 76 ± 6 mm6½: 83 ± 6 mm7.0: 89 ± 6 mm7½: 95 ± 6 mm8.0: 102 ± 6 mm8½: 108 ± 6 mm9.0: 114 ± 6 mmSame
Powder FreeASTM D6124Meets requirements of ≤2.0 mg/glove for Powder-Free designation per ASTM D3577Meets requirements of ≤2.0 mg/glove for Powder-Free designation per ASTM D3577Same
BiocompatibilityPrimary Skin Irritation - ISO 10993-10Passes (Not a primary skin irritant)Passes (Not a primary skin irritant)Same
Dermal Sensitization- ISO 10993-10Passes (Not a contact sensitizer)Passes (Not a contact sensitizer)Same
Acute Systemic Toxicity, ISO 10993-11N/AIt is concluded that the devices did not induce any acute systemic toxicityDifferent
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATECURRENT
510(k) Number-K020019K232079-
MaterialMediatedPyrogenicity,ISO 10993-11N/AIt is concludedthat the devicesare non-pyrogenicDifferent
Cytotoxicity-Direct Contact,ISO 10993-5N/ANon-cytotoxic forS&C/S&P glovesDifferent
Cytotoxicity-MEM Elution,ISO 10993-5N/AUnder theconditions ofthis study, thetest articleextract showedpotentialtoxicity to L929cells for GreenglovesDifferent
Watertight (1000ml)ASTM D5151Meets 21 CFR800.20 andASTM D3577when tested inaccordancewith ASTMD5151InspectionLevel I, AQL0.65Meets 21 CFR800.20 andASTM D3577when tested inaccordance withASTM D5151Inspection LevelI, AQL 0.65Same
Intended use /Indications for Use-A powder-freesterilesurgeon's gloveis a disposabledevice made ofnatural rubberintended to beworn on thehands ofhealthcarepersonnel as abarrier forprotectionagainst cross-contaminationbetween thehealthcarepersonnel andpatient.A powder-freesterilesurgeon's gloveis a disposabledevice made ofnatural rubberintended to beworn on thehands ofhealthcarepersonnel as abarrier forprotectionagainst cross-contaminationbetween thehealthcarepersonnel andpatient.These gloveswere tested forDifferent
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATECURRENT
510(k) Number-K020019K232079-
ChemotherapyDrugs andFentanyl Citrateas per ASTMD6978StandardPractice forAssessment ofMedical Glovesto PermeationbyChemotherapyDrugs.
SizeMedical GloveGuidanceManual -Labeling5½6.06½7.07½8.08½9.05½6.06½7.07½8.08½9.0Same
Single UseMedical GloveGuidanceManual -LabelingSingle useSingle useSame
Sterility StatusMedical GloveGuidanceManual -LabelingSterileSterileSame
Sterility MethodISO 11137-1ISO 11137-2GammaRadiationGammaRadiationSame
Chemotherapy DrugPermeation TestASTM D6978-05
1. Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with ChemotherapyDrugs and Fentanyl (Green)
* Carmustine (BCNU)(3.3 mg/ml)N/A11.9Different
Cisplatin (1.0 mg/ml)N/A>240Different
Cyclophosphamide(Cytoxan) (20.0 mg/ml)N/A>240Different
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISON ANALYSIS
PREDICATECURRENT
510(k) Number-K020019K232079-
Dacarbazine (10.0 mg/ml)N/A> 240Different
DoxorubicinHydrochloride (2.0 mg/ml)N/A> 240Different
Etoposide (20.0 mg/ml)N/A> 240Different
Fluorouracil (50.0 mg/ml)N/A> 240Different
Ifosfamide (50.0 mg/ml)N/A> 240Different
Methotrexate (25.0 mg/ml)N/A> 240Different
Mitomycin C (0.5 mg/ml)N/A> 240Different
Mitoxantrone (2.0 mg/ml)N/A> 240Different
Paclitaxel (6.0 mg/ml)N/A> 240Different
* ThioTepa (10.0 mg/ml)N/A14.7Different
Vincristine Sulfate (1.0 mg/ml)N/A> 240Different
Warning StatementN/A*WARNING:Please note thefollowing drugshave extremelylow permeationtimes:Carmustine(BCNU): 11.9minutes andThiotepa: 14.7minutes. Do notuse withCarmustine andThiotepa.Different
Fentanyl Citrate (100 mcg/2mL)ASTM D6978-05N/A> 240Different
2. Natural Rubber Latex Surgical Glove (Strip&Coat)/(Strip&Pack), Powder Free, Sterile, Testedfor Use with Chemotherapy Drugs and Fentanyl
* Carmustine (BCNU)(3.3 mg/ml)N/A5.7Different
Cisplatin (1.0 mg/ml)N/A>240Different
CharacteristicsPREDICATECURRENTComparison
510(k) NumberK020019K232079
Cyclophosphamide(Cytoxan) (20.0mg/ml)N/A>240Different
Dacarbazine (10.0mg/ml)N/A> 240Different
DoxorubicinHydrochloride (2.0mg/ml)N/A> 240Different
Etoposide (20.0mg/ml)N/A> 240Different
Fluorouracil (50.0mg/ml)N/A> 240Different
Ifosfamide (50.0mg/ml)N/A> 240Different
Methotrexate (25.0mg/ml)N/A> 240Different
Mitomycin C (0.5mg/ml)N/A> 240Different
Mitoxantrone (2.0mg/ml)N/A> 240Different
Paclitaxel (6.0 mg/ml)N/A> 240Different
* ThioTepa (10.0mg/ml)N/A13.8Different
Vincristine Sulfate (1.0mg/ml)N/A> 240Different
Warning StatementN/A*WARNING:Please note thefollowing drugshave extremelylow permeationtimes:Carmustine(BCNU): 5.7minutes andThiotepa: 13.8minutes. Do notuse withCarmustine andThiotepa.Different
Fentanyl Citrate (100mcg/2mL)ASTM D6978-05N/A> 240Different

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

510(k) SUMMARY

7.0 Summary of Non-Clinical Testing

TestMethodStandardPurpose ofTestingAcceptance CriteriaResults
PhysicalPropertiesASTM D412(Standard TestMethod forVulcanizedRubber andThermoplasticElastomers -Tension)To evaluate thetensile (tension)properties ofgloveTensilestrengthBeforeAgingMin.24MPaAfterAgingMin.18MPaBeforeAgingPassAfterAgingPass
UltimateelongationMin.750%Min.560%PassPass
Stress at500%ElongationMax.5.5MPaN/APassN/A
DimensionASTM D3767(StandardPractice forRubber -Measurementof Dimensions)To measure thelength, width andthickness ofgloveLength: Each size satisfies ASTM requirementWidth: Each size satisfies ASTM requirementPass
DimensionASTM D3767(StandardPractice forRubber -Measurementof Dimensions)To measure thelength, width andthickness ofgloveThicknessFinger - Min. 0.10 mmPalm - Min. 0.10 mmCuff - Min. 0.10 mmPass
WatertightASTM D5151(Standard TestMethod forDetection ofHoles inMedicalGloves)To measureglove integrityFreedom from holes at AQL 0.65Pass
ResidualPowderASTM D6124(Standard TestMethod forResidualPowder onMedicalGloves)To determine theamount ofresidual powderand non-powdersolids found onglovesHave a powder residue limit of2.0 mg per glovePass
ProteinContentASTM D5712(Test Method forAnalysis ofAqueousExtractableProtein in Latex,Natural Rubber,and ElastomericProductsUsing theModified LowryMethod)To determine theamount ofextractableprotein from theglove≤ 200 µg/dm2Pass

{15}------------------------------------------------

510(k) SUMMARY

Test MethodStandardPurpose of TestingResult Summary
BiocompatibilityPrimary Skin Irritation -ISO 10993-10To assess thepotential of glove toproduce dermalirritationNot a primary skin irritation
BiocompatibilityDermal Sensitization-ISO 10993-10To assess thepotential of glove tocause a delayedhypersensitivity(Type IV)immunologicalresponseNot a contact sensitizer
BiocompatibilityAcute Systemic Toxicity,ISO 10993-11To determine theacute systemictoxicity potential ofgloveNo biological reactivity shown
BiocompatibilityMaterial MediatedPyrogenicity, ISO10993-11To determine anypyrogenic responseinduced by the gloveNo increase of temperature
BiocompatibilityCytotoxicity- DirectContact, ISO 10993-5To determine thecytotoxicity potentialof gloveNon-cytotoxic for S&C/S&P gloves
BiocompatibilityCytotoxicity- MEMElution, ISO 10993-5To determine thecytotoxicity potentialof gloveUnder the conditions of thisstudy, the test article extractshowed potential toxicity to L929cells for Natural Rubber LatexGreen
BacterialendotoxinLimulus AmoebocyteLysate (LAL) testTo determine theamount of endotoxinon the glove≤20 EU/pair of gloves

8.0 Summary of Clinical Testing:

No clinical study is included in this submission.

9.0 Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the subject devices Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green) and Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl are as safe, effective, and perform as well as or better than the legally marketed predicate device K020019.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).