K Number
K232020

Validate with FDA (Live)

Date Cleared
2023-10-16

(101 days)

Product Code
Regulation Number
890.5525
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Iontophoresis Electrodes are intended to be used to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.

Device Description

The proposed device is indicated for the administration of soluble salts or other drugs into the body for medical purposes. The Iontophoresis Electrodes utilize adhesive electrodes have both anode and cathode electrodes that adhere to the skin. The electrodes may either be connected by leads (wires) to a separate controller Forming a current loop. The portion of the active electrode that contacts the skin also has a drug/solution reservoir. End users saturate the reservoir with the ionic solution to be delivered. The reservoir contacts the electrode and the drug/solution is delivered to the patient transdermally.

Five models (Small, Medium, Large, Butterfly and Return Electrode) are included in this series products, and they have the same intended purposes, working theories and specifications.

AI/ML Overview

The provided document describes the acceptance criteria and the study that proves the device meets those criteria for the Iontophoresis Electrodes (K232020).

Here's the breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The document presents a comparison to a predicate device (Iontophoresis Drug Delivery Electrodes, K040495) to demonstrate substantial equivalence, rather than setting explicit numerical acceptance criteria for a new device's performance. The "acceptance criteria" are implied by the predicate device's specifications and the non-clinical test conclusions.

Acceptance Criteria (Implied by Predicate & Standards Compliance)Reported Device Performance (Proposed Device)Remark
Product Code: EGJEGJSame
Regulation Number: 21 CFR 890.552521 CFR 890.5525Same
Indications for Use: Introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.Introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.Same
Patient Population: Patients requiring iontophoresis treatmentPatients requiring iontophoresis treatmentSame
Usage Environment: Hospitals, medical clinicsHospitals, medical clinicsSame
Design (Electrode Size): Comparable to predicate sizes (e.g., Small, Medium, Large, Butterfly, Return Electrode)ION-T01 (69x69 mm), ION-T02 (82x82 mm), ION-T03 (88x88 mm), ION-T04 (84x94 mm), ION-T05 (64x64 mm)Same (Note 1 - differences in design requirements, not affecting safety/performance)
Main Materials: Silver/Silver Chloride (Ag/AgCl), Buffering Agent, PU Conductive carbon filmSilver/Silver Chloride (Ag/AgCl), Buffering Agent, PU Conductive carbon filmSame
Anatomical Sites: For body surface skinFor body surface skinSame
Working mode: Must be used as a complete set, consisting of an active drug delivery electrode and a passive return electrode.Must be used as a complete set, consisting of an active drug delivery electrode and a passive return electrode.Same
Compatibility with other devices: Designed for use with iontophoresis devices only.Designed for use with iontophoresis devices only.Same
Impedance: 2MΩ ~ 10MΩ2MΩ ~ 10MΩSame
System current: 4.0mA4.0mASame
Maximum Dosage: 80 mA-min80 mA-minSame
Fill Volume: Comparable to predicate fill volumes (e.g., 1.5cc, 2.5cc, 4.0cc, 2.0cc for specific models)1.5cc (ION-T01), 2.5cc (ION-T02), 4.0cc (ION-T03), 2.0cc (ION-T04)Same
Biocompatibility (Cytotoxicity): No CytotoxicityNo CytotoxicitySame
Biocompatibility (Skin sensitization): No evidence of sensitizationNo evidence of sensitizationSame
Biocompatibility (Irritation): No evidence of irritationNo evidence of irritationSame
EMC, Electrical Safety (Electrical safety): Comply with IEC 60601-1Comply with IEC 60601-1Same
EMC, Electrical Safety (EMC): Comply with IEC 60601-1-2Comply with IEC 60601-1-2Same

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document explicitly states: "No clinical study is included in this submission."
Therefore, there is no sample size for a clinical test set or information on its provenance. The testing was non-clinical.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical study was conducted that would require expert ground truth. The evaluation focused on non-clinical performance and safety based on recognized standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical study was conducted.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an Iontophoresis Electrode, which is a physical device for drug delivery, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm or software.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no clinical study was conducted. The "ground truth" for the non-clinical tests was established by compliance with international standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10, ISO 10993-23) and comparison to a legally marketed predicate device.

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model.

9. How the ground truth for the training set was established

Not applicable. This is a physical medical device, not a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 16, 2023

TOP-RANK Health Care Co., Ltd. % Heather Wang Consultant Microkn Medical Technology Service (Shanghai) Co., Ltd. Room 901, No. 889, Pinglu Road, Jing'an District, Shanghai (Shanghai Jing'an HUAFA Center) Shanghai, Shanghai 200435 China

Re: K232020

Trade/Device Name: Iontophoresis Electrodes Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis Device Regulatory Class: Class II Product Code: EGJ Dated: August 18, 2023 Received: August 18, 2023

Dear Heather Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices

{2}------------------------------------------------

OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232020

Device Name Iontophoresis Electrodes

Indications for Use (Describe)

Iontophoresis Electrodes are intended to be used to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1. Contact information

1.1. Applicant

Applicant Name: Top-Rank Health Care Co., Ltd.

Address: Mcshan Village, Dongguan Street, Shangyu District, Shaoxing City,

Zhejiang Province, P.R. China.

Contact Person: Yao fei jie

Title: Product Manager

Tel: +86-575-82912999

Email: yaofeijie@etop-rank.cn

1.2. Consultant

Company: Microkn Medical Technology Service (Shanghai) Co., Ltd.

Address: Room 901, No. 889, Pinglu Road, Jing'an District, Shanghai

(Shanghai Jing'an HUAFA Center)

Contact Person: Heather. Wang

Telephone: +86 13166194697

Email: heather.wang@microkn.com

2. Device information

{5}------------------------------------------------

  • Common Name: Iontophoresis Electrodes ●
  • Model(s): ION-T01, ION-T02, ION-T03, ION-T04, ION-T05 ●
  • Classification: II o
  • Product Code: EGJ 0
  • Regulation Number: 21CFR 890.5525 o

3. Indications for Use

Iontophoresis Electrodes are intended to be used to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.

4. Legally Marketed Predicate Device

Product name: Iontophoresis Drug Delivery Electrodes

510(k) Number: K040495

Product Code: EGJ

Manufacture: NAIMCO, INC.

5. Description of the device

The proposed device is indicated for the administration of soluble salts or other drugs into the body for medical purposes. The Iontophoresis Electrodes utilize adhesive electrodes have both anode and cathode electrodes that adhere to the skin. The electrodes may either be connected by leads (wires) to a separate controller Forming a current loop. The portion of the active electrode that contacts the skin also has a drug/solution reservoir. End users saturate the

{6}------------------------------------------------

reservoir with the ionic solution to be delivered. The reservoir contacts the electrode and the drug/solution is delivered to the patient transdermally.

Five models (Small, Medium, Large, Butterfly and Return Electrode) are included in this series products, and they have the same intended purposes, working theories and specifications.

{7}------------------------------------------------

The main components of proposed device shown in Table 1.

Component nameDescriptionApplied to
Conductive button(Upper)It serves as a connecting component to the output terminal ofthe controller, completing mechanical and electricalconnection functionsALLModels
Conductive button(Lower)Conductive button (Lower) is used in conjunction with theupper part and mechanically riveted to achieve a firmconnection between materialsALLModels
POLY film layerIt is a thin film made of polyethylene material that is used tosupport the surface of the product and achieve electricalinsulation functions with other internal componentsALLModels
Conductive carbonfilm layerIt is a conductive film made of PU material and carbonmaterial, which realizes the transmission and uniformdistribution of current by connecting with metal buckle andcovering gel layer in a large areaALLModels
Self-adhesive GELlayerIt is used to adhere return electrode to the skin surface, and itsconductive properties can form a conductive circuit betweenthe controller and the body surface skinALLModels
GEL ProtectivelayerIt is used to store return electrode and has isolation andprotection effects on Self-adhesive GELALLModels
Foma layerIt is used to adhere active drug delivery electrode to the skinsurface,at the same time, it ensures a firm fit and prevents theoutflow of drug/solutionALLModels
AgAgCl film layerAgAgCl is mainly used to alleviate the electrolytic reaction ofwater, thereby preventing pH related corrosive skin reactionsALLModels
drug/solutionreservoir layerIt is a medical porous material with good water absorptionand is used for temporary storage of drug/solutionALLModels
Release paper layerUsed for long-term storage of active drug delivery electrodesto ensure stable electrode performanceALLModels

Table 1 Main Components of Proposed Device

{8}------------------------------------------------

6. Non-Clinical Test conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1 Medical electrical device Part1: General requirements for o basic safety and essential performance
  • IEC 60601-1-2 Medical electrical equipment- Part 1-2: General ● requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests
  • ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests ● for In Vitro cytotoxicity
  • ISO 10993-10 Biological evaluation of medical devices Part 10: ● Tests for skin sensitization
  • ISO 10993-23 Biological evaluation of medical devices Part 23: ● Tests for irritation

7. Clinical Test Conclusion

No clinical study is included in this submission.

8. Predicate Devices

Product name: Iontophoresis Drug Delivery Electrodes

510(k) Number: K040495

Product Code: EGJ

Manufacture: NAIMCO, INC.

{9}------------------------------------------------

9. Substantially Equivalent (SE) Comparison

The Iontophoresis Electrodes has been carefully compared to legally marketed devices with respect to intended use, configuration, principle of operation (Table 2), performance specifications (Table 3).

Proposed DevicePredicate Device
ItemIontophoresisElectrodes(K232020)Iontophoresis DrugDelivery Electrodes(K040495)Remark
Product CodeEGJEGJSame
RegulationNumber21CFR 890.552521 CFR 890.5525Same
Indications forUseIontophoresisElectrodes are intendedto be used to introducesoluble salts and otherdrugs into the body asan alternative tohypodermic injection.Iontophoresis DrugDelivery Electrodes areintended to be used tointroduce soluble salts andother drugs into the bodyas an alternative tohypodermic injection.Same
PatientPopulationWith patients requiringiontophoresis treatmentWith patients requiringiontophoresis treatmentSame
UsageenvironmentHospitals, medicalclinicsHospitals, medical clinicsSame

Table 2 General Comparison

{10}------------------------------------------------

Proposed DevicePredicate Device
ItemIontophoresisIontophoresis DrugRemark
ElectrodesDelivery Electrodes
Design(ElectrodeSize)ION-T01( $69×69$ mm),ION-T02( $82×82$ mm),ION-T03( $88×88$ mm),ION-T04( $84×94$ mm),ION-T05( $64×64$ mm)Small( $2.75×2.75$ inch),Medium( $3.25×3.25$ inch),Large ( $3.5×3.5$ inch),Butterfly( $3.35×3.75$ inch),Return ElectrodeSame(Note1)
( $2.5×2.5$ inch)
Main materialsSilver/Silver Chloride(Ag/AgCI)Buffering Agent;PU Conductive carbon filmSilver/Silver Chloride(Ag/AgCI)Buffering Agent;PU Conductive carbon filmSame
AnatomicalsitesFor body surface skinFor body surface skinSame
Working modeMust be used as a completesetConsists of an active drugdelivery electrodeandapassive return electrode.Must be used as a completesetConsists of an active drugdelivery electrode and apassive return electrode.Same
Compatibilitywith otherDesigned for use withiontophoresis devices only.Designed for use withiontophoresis devices only.Same

Table 3 Performance Comparison

{11}------------------------------------------------

Proposed DevicePredicate Device
ItemIontophoresisElectrodesIontophoresis DrugDelivery ElectrodesRemark
devices
Impedance$2MΩ~10MΩ$$2MΩ~10MΩ$Same
System current4.0mA4.0mASame
MaximumDosage80 mA-min80 mA-minSame
Fill Volume1.5cc (ION-T01),2.5cc (ION-T02),4.0cc (ION-T03),2.0cc (ION-T04)1.5cc (201-400),2.5cc (201-401),4.0cc (201-402),2.0cc (201-403)Same

Note 1

The new device Iontophoresis Electrodes is substantial equivalence as the predicate device K040495. The differences in Dimensions, model are the differences in design requirements between different devices, which not affect the safety performance of the device, and the new device passed Safety testing, so these differences will not cause any safety and effectiveness issues. Considering the same intended use, Working mode and Product performance ect, they are substantial equivalence.

Safety comparison has been done to validate the EMC, biocompatibility

{12}------------------------------------------------

specification and safety of the device (Table 4).

ItemProposed DevicePredicate DeviceRemark
Biocompatibility
CytotoxicityNo CytotoxicityNo CytotoxicitySame
Skin sensitizationNo evidence of sensitizationNo evidence of sensitizationSame
IrritationNo evidence of irritationNo evidence of irritationSame
EMC, Electrical Safety
Electrical safetyComply with IEC 60601-1Comply with IEC 60601-1Same
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Same

Table 4 Safety Comparison

10.Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

N/A