K Number
K231839

Validate with FDA (Live)

Date Cleared
2023-07-19

(27 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization of the spine as an adjunct to fusions in patients with:

  • · degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
  • · spondylolisthesis,
  • · trauma (i.e. fractures or dislocations),
  • · tumors,
  • · deformity (defined as kyphosis, lordosis, or scoliosis),
  • · pseudoarthrosis.
  • · failed previous fusion,
  • · spinal stenosis.

The AccelFix Lumbar Plate System is indicated for use via the anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spinal stenosis, or a failed previous spine surgery.

Device Description

The CastleLoc-P Anterior Cervical Plate System is composed of plates, screws and lockers which are made from titanium alloy per ASTM F136. These plates attach to the anterior cervical spine with a minimum of four screws per plate. The plates are offered in one-level, two-level, four-level fusion configurations (13~97mm). The plate screws are 3.5mm and 4.0mm diameter head screws. They are self-tapping and self-drilling threaded. This device can be provided both as non-sterile.

The AccelFix Lumbar Plate System's implants are Lumbar Plate System intended for use as an aid in spinal fixation. They are made from titanium alloy per ASTM F136. The AccelFix Lumbar Plate System consists of a variety of shapes and sizes of plates and screws. The plate has been designed to include spikes for added stability and alignment during screw insertion. The plates feature a curvature for anatomic fit. The diameter of screw is available from 5.5 to 6.0 mm and the length from 20 to 55 mm.

AI/ML Overview

This document is a 510(k) Pre-Market Notification from the FDA, which declares substantial equivalence for medical devices to legally marketed predicate devices. It typically does not contain acceptance criteria or study results demonstrating performance in the way a clinical trial or AI device validation would.

Instead, the FDA 510(k) process focuses on demonstrating substantial equivalence to a predicate device. This means showing that the new device is as safe and effective as a legally marketed device that is not subject to pre-market approval (PMA).

Based on the provided text, here's what can be extracted and inferred regarding "acceptance criteria" and "study":

1. A table of acceptance criteria and the reported device performance:

This document does not contain a table of acceptance criteria or reported device performance in the context of diagnostic accuracy or clinical outcomes for the CastleLoc-P Anterior Cervical Plate System or the AccelFix Lumbar Plate System. The submission is a 510(k) for spinal intervertebral body fixation orthoses, which are mechanical implants, not AI/diagnostic devices.

The "performance testing" section, specifically for the CastleLoc-P Anterior Cervical Plate System and AccelFix Lumbar Plate System, states:

"The additional components to be added through this submission do not require additional mechanical testing. None of the additional components is the worst case of the CastleLoc-P Posterior Anterior Cervical Plate System."
"Therefore, we substitute mechanical test data of CastleLoc-P Anterior Cervical Plate System (predicate devices K113509, K143279, K190425) and AccelFix Lumbar Plate System (predicate device K 192481) for additional components with the predicate device data."

This indicates that instead of new performance tests for the new components, the company relied on the existing mechanical test data of the predicate devices. Therefore, the "acceptance criteria" implied are that the new components perform no worse than the predicate devices' mechanical properties. However, these specific mechanical properties (e.g., fatigue strength, torsional rigidity) and their quantitative acceptance limits are not detailed in this document.

2. Sample size used for the test set and the data provenance:

Not applicable. As explained above, no new performance study with a test set of data (e.g., patient cases) was conducted for these devices as part of this 510(k). The evaluation relies on the substantial equivalence to predicate devices, including their prior mechanical testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This is not a study involving ground truth established by experts for a diagnostic or AI device.

4. Adjudication method for the test set:

Not applicable. No such test set or adjudication process is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

Not applicable. This is not a diagnostic device for which an MRMC study would typically be performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a spinal implant, not an algorithm or AI device.

7. The type of ground truth used:

Not applicable. "Ground truth" in the context of diagnostic or AI studies is not relevant here. The evaluation instead relies on conformity to material specifications (ASTM F136) and the mechanical performance previously established for the predicate devices.

8. The sample size for the training set:

Not applicable. There is no "training set" as this is not an AI/machine learning device.

9. How the ground truth for the training set was established:

Not applicable. There is no "training set" or corresponding ground truth establishment process described.

In summary:

This 510(k) submission for the CastleLoc-P Anterior Cervical Plate System and AccelFix Lumbar Plate System focuses on demonstrating substantial equivalence to existing predicate devices. The "study" mentioned for performance is the reliance on prior mechanical test data of the predicate devices. The document does not provide details on specific acceptance criteria and detailed performance data from new studies in the manner typically expected for AI/diagnostic software.

{0}------------------------------------------------

July 19, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

L & K Biomed Co., Ltd. Katherine Kim RA #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil. Giheung-gu Yongin-si, Gyeonggi-do 17015 South Korea

Re: K231839

Trade/Device Name: CastleLoc-P Anterior Cervical Plate System, AccelFix Lumbar Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 20, 2023 Received: June 22, 2023

Dear Katherine Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'neill -S

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K231839

Device Name

CastleLoc-P Anterior Cervical Plate System

Indications for Use (Describe)

The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization of the spine as an adjunct to fusions in patients with:

  • · degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
  • · spondylolisthesis,
  • · trauma (i.e. fractures or dislocations),
  • · tumors,
  • · deformity (defined as kyphosis, lordosis, or scoliosis),
  • · pseudoarthrosis.
  • · failed previous fusion,
  • · spinal stenosis.
Type of Use (Select one or both, as applicable)
---------------------------------------------------

|X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K231839

Device Name

AccelFix Lumbar Plate System

Indications for Use (Describe)

The AccelFix Lumbar Plate System is indicated for use via the anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spinal stenosis, or a failed previous spine surgery.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "K231839" and "Page 1 of 3" in the upper right corner. In the lower left corner is the logo for L&K BIOMED. The logo includes the text "LEARNING & KNOWLEDGE BIOMED" above the letters "L&K BIOMED" in a larger font. To the right of the text is a circular graphic made of blue dots.

510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

1. MANUFACTURER

Submitter's Name:L&K Biomed Co., Ltd.
Submitter's Address:#101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gilGiheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea.
Submitter's Telephone:82-2-6717-1983
Contact Person:Katherine Kim/82-2-6717-1983khkim@lnkbiomed.com
Prepared Date:June 20, 2023

2. DEVICE IDENTIFICATION

Device Trade NameCastleLoc-P Anterior Cervical Plate System
Common/Usual NameAppliance, Fixation, Spinal Intervertebral Body
Regulation Class /NumberClass II / 21 CFR 888.3060
Regulation NameSpinal Intervertebral Body Fixation Orthosis
Product CodeKWQ
Classification PanelSpinal Devices (DHT6B)
Device Trade NameAccelFix Lumbar Plate System
Common/Usual NameLumbar Spinal Plate
Regulation Class /NumberClass II/ 21 CFR 888.3060
Regulation NameSpinal Intervertebral Body Fixation Orthosis
Product CodeKWQ
Classification PanelSpinal Devices (DHT6B)

3. PREDICATE OR LEGALLY MARKETED DEVICES WHICH ARE SUBSTANTIALLY EQUIVALENT.

The subject devices are identical to the predicate devices in all characteristics.

Subject Device Name510K NO.Trade or Proprietary or Model NamePredicate Type
CASTLELOC-PAnterior CervicalPlate SystemK 143271CastleLoc-P Anterior Cervical Plate SystemPrimary
K113509K143279LnK Anterior Cervical Plate SystemAdditional
K 210035K 223719Paeon Anterior Plate System
AccelFix LumbarPlate SystemK 192481AccelFix Lumbar Plate SystemPrimary
K 210035K 223719Elatus Lumbar Plate SystemAdditional

The design feature, indications for use, material and manufacturing process for the subject devices are substantially equivalent to the predicate devices (CastleLoc-P Anterior Cervical Plate System K113509, K143279, K143271, K190425/ AccelFix Lumbar Plate System K192481/200794).

{5}------------------------------------------------

  • MATERIALS- And the additional components material is the same material used in the predicate devices. 4.
CastleLoc-P Anterior Cervical Plate SystemTi-6Al-4V ELI titanium alloy (ASTM F136)
AccelFix Lumbar Plate SystemTi-6Al-4V ELI titanium alloy (ASTM F136)

ર. DESCRIPTION OF THE DEVICE

The CastleLoc-P Anterior Cervical Plate System is composed of plates, screws and lockers which are made from titanium alloy per ASTM F136. These plates attach to the anterior cervical spine with a minimum of four screws per plate. The plates are offered in one-level, two-level, four-level fusion configurations (13~97mm). The plate screws are 3.5mm and 4.0mm diameter head screws. They are self-tapping and self-drilling threaded. This device can be provided both as non-sterile.

The AccelFix Lumbar Plate System's implants are Lumbar Plate System intended for use as an aid in spinal fixation. They are made from titanium alloy per ASTM F136. The AccelFix Lumbar Plate System consists of a variety of shapes and sizes of plates and screws. The plate has been designed to include spikes for added stability and alignment during screw insertion. The plates feature a curvature for anatomic fit. The diameter of screw is available from 5.5 to 6.0 mm and the length from 20 to 55 mm.

6. INDICATION FOR USE

The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization and stabilization of the spine as an adjunct to fusions in patients with:

  • degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
  • . spondylolisthesis,
  • trauma (i.e., fractures or dislocations), "
  • . tumors.
  • . deformity (defined as kyphosis, lordosis, or scoliosis),
  • . pseudoarthrosis,
  • . failed previous fusion,
  • spinal stenosis.

The AccelFix Lumbar Plate System is indicated for use via the anterior, lateral, or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spine, spinal stenosis, or a failed previous spine surgery.

7. PERFORMANCE TESTING

CastleLoc-P Anterior Cervical Plate System

The additional components to be added through this submission do not require additional mechanical testing. None of the additional components is the worst case of the CastleLoc-P Posterior Anterior Cervical Plate System.

AccelFix Lumbar Plate System

The additional components to be added through this submission do not require additional mechanical testing. None of the additional components is the worst case of the AccelFix Lumbar Plate System.

Therefore, we substitute mechanical test data of CastleLoc-P Anterior Cervical Plate System (predicate devices K113509, K143279, K190425) and AccelFix Lumbar Plate System (predicate device K 192481) for additional components with the predicate device data.

{6}------------------------------------------------

SUMMARY OF TECHNOLOGY CHARACTERISTICS 8.

Subject devices are identical to the predicate devices in all (design feature, indications for use, surgical approach and manufacturing process) characteristics.

510K no.Indicationfor useDesignSurgicalapproachMaterial
Subject Device-SameSimilarSameTi-6Al-4V ELI(ASTM F136)
CastleLoc-P Anterior CervicalPlate SystemPrimaryK143271SameSimilarSameSame
LnK Anterior Cervical PlateSystemCastleLoc-P Anterior CervicalPlate SystemK113509K143279K190425SameSimilarSameSame
Paeon Anterior Plate SystemK210035K223719SameSimilarSameSame

CastleLoc-P CastleLoc-P Anterior Cervical Plate System

AccelFix Lumbar Plate System

510K no.Indicationfor useDesignSurgicalapproachMaterial
Subject Device-SameSimilarSameTi-6Al-4V ELI(ASTM F136)
AccelFix Lumbar Plate SystemPrimaryK192481SameSimilarSameSame
Elatus Lumbar Plate SystemK210035K223719SameSimilarSameSame

9. SUBSTANTIAL EQUIVALENCE

Subject devices are shown to be substantially equivalent to the predicate devices in indications for use, design, function, and materials used.

10. CONCLUSION

The overall technology characteristics lead to the conclusion that the CastleLoc-P Anterior Cervical Plate System, and AccelFix Lumbar Plate System is substantially equivalent to the predicate devices (CastleLoc-P Anterior Cervical Plate System K113509, K143271, K190425/ AccelFix Lumbar Plate System K192481/200794).

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.