K Number
K231780

Validate with FDA (Live)

Device Name
Perifit Care+
Manufacturer
Date Cleared
2023-12-20

(187 days)

Product Code
Regulation Number
884.1425
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Perifit Care+ is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.

Perifit Care+ is indicated for an adult female.

Device Description

The Perifit Care+ device consists of a rigid probe covered in a silicone sheath that is temporarily inserted into the vagina. Sensors located under the sheath measure the strength of the user's pelvic floor muscles. This information is then transmitted wirelessly to a smartphone application in order to provide real-time feedback to the user. It is a single patient, reusable device to be supplied over-the-counter.

AI/ML Overview

The provided document is a 510(k) summary for the Perifit Care+ device. It outlines the device's indications for use, technological characteristics, and comparison to a predicate device (Perifit K221476) to establish substantial equivalence.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" in the format of specific thresholds for performance metrics. Instead, it relies on demonstrating substantial equivalence to a predicate device (Perifit K221476) through similarity in indications for use, technological characteristics, and safety testing.

The "reported device performance" is largely described through the completion of various non-clinical performance tests, indicating the device met the standards for safety and functionality, rather than achieving specific quantitative performance metrics against a defined acceptance criterion.

However, based on the provided text, we can infer some "acceptance criteria" from the testing performed and the comparison to the predicate.

Acceptance Criteria (Inferred from testing/comparison)Reported Device Performance (from Perifit Care+ section)
Material BiocompatibilityTested in accordance with ISO 10993 standards (Cytotoxicity, Sensitization, Vaginal Irritation, Systemic Toxicity) and found to be safe for intended purpose.
Electrical SafetyTested in accordance with IEC 60601-1 and IEC 60601-1-2.
Electromagnetic Compatibility (EMC)Tested in accordance with IEC 60601-1-2 and FDA guidance documents "Electromagnetic Compatibility (EMC) of Medical Devices" and "Radio Frequency Wireless Technology in Medical Devices".
Software Verification & Validation (V&V)Conducted in accordance with FDA guidance (Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304:2006. (Perifit Care+ includes updates to the software of the Perifit App and embedded firmware, implying these updates were subject to V&V).
CybersecurityEvaluated in accordance with FDA guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
Mechanical Integrity/DurabilityVarious mechanical tests conducted, including mechanical drop testing and durability testing. (New tail attribute also tested for bending and resistance).
Sensor FunctionalitySensor behavior testing conducted.
User Comprehension/OperabilityUser testing through a questionnaire supports that users understand the key labeling provisions and how to operate the device.
Functional Equivalence"The Perifit Care+ has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate device." "Performance data demonstrate that the Perifit Care+ is as safe and effective as the Perifit." (This is the overarching "acceptance criterion" for substantial equivalence, implying all relevant functionalities are comparable/equivalent to the predicate).

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes for each non-clinical performance test (e.g., number of units dropped, number of users for the questionnaire). It also does not explicitly state the provenance (country of origin, retrospective/prospective) for these tests, except that they were conducted to a standard (e.g., ISO, IEC). User testing via a questionnaire is mentioned, implying a human subject component, but details are absent.

This type of information is usually detailed in the full 510(k) submission, not typically in the public summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The studies described are primarily non-clinical, involving engineering and materials testing, and software validation, rather than clinical studies requiring expert ground truth for diagnostic accuracy. The "user testing through a questionnaire" is a form of user experience or usability testing, not typically a "ground truth" establishment in a diagnostic context.

4. Adjudication Method for the Test Set

Since there's no mention of a diagnostic accuracy study or establishment of "ground truth" by experts, an adjudication method (like 2+1 or 3+1) is not applicable and not mentioned in this summary.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a MRMC comparative effectiveness study was not done. The Perifit Care+ is a patient-operated perineometer providing biofeedback, not an AI-assisted diagnostic tool that would typically involve human readers interpreting output with or without AI assistance. The effectiveness is primarily linked to the patient's own exercise and biofeedback, not an external reader's interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

The device is a perineometer that provides biofeedback for pelvic floor muscle exercise via a smartphone app. Its "performance" is based on its ability to accurately sense muscle contractions and transmit that data, along with the functionality of the app itself. The software verification and validation (and "sensor behavior testing") would assess its standalone performance in these aspects. There's no "algorithm only" performance reported in the sense of an AI model making a diagnosis or prediction without human input, as this is not the nature of the device.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical tests, the "ground truth" would be established by:

  • Engineering specifications and standards: For mechanical, electrical, and sensor performance (e.g., a drop test either meets or fails a predetermined physical integrity standard).
  • Reference materials/methods: For biocompatibility testing (e.g., cell cultures, animal models for irritation).
  • Software requirements and design specifications: For software V&V (the software performs as designed and meets requirements).
  • User feedback/comprehension: For the user questionnaire (users correctly understand labeling and operation).

There is no mention of ground truth established by expert consensus, pathology, or outcomes data related to diagnostic accuracy, as it's not a diagnostic device.

8. The Sample Size for the Training Set

This information is not provided and is generally not applicable in this context. The device is not an AI/ML model that is "trained" on a dataset in the conventional sense for making diagnostic predictions. Its software includes an application and embedded firmware, which are developed and validated against requirements, not "trained."

9. How the Ground Truth for the Training Set Was Established

As there's no mention of a "training set" in the context of machine learning, this information is not applicable and not provided. Software "ground truth" would be established by its adherence to design specifications and user requirements during the development and validation phases.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 20, 2023

X6 Innovations % Lina Kontos Partner Hogan Lovells US LLP 555 13th Street NW Washington, DC 20004

Re: K231780 Trade/Device Name: Perifit Care+ Regulation Number: 21 CFR§ 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: November 21, 2023 Received: November 21, 2023

Dear Lina Kontos:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen -S

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known)

K231780

Device Name

Perifit Care+

Indications for Use (Describe)

The Perifit Care+ is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.

Perifit Care+ is indicated for an adult female.

Type of Use (Select one or both, as applicable)

□ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

X6 Innovations' Perifit Care+

Submitter

X6 Innovations 128 rue de la Boétie, 75008 Paris, France Phone: +33 6 51 66 55 94 Contact Person: Robin Reynaud Date Prepared: December 19, 2023

Date of Submission: June 16, 2023

Name of Device: Perifit Care+

Common or Usual Name: Perineometer

Classification Name: (21 CFR 884.1425) Perineometer

Regulatory Class: Class II

Product Code: HIR

Predicate Devices:

Perifit (K221476)

Intended Use / Indications for Use

The Perifit Care+ is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.

Perifit Care+ is indicated for an adult female.

Device Description

The Perifit Care+ device consists of a rigid probe covered in a silicone sheath that is temporarily inserted into the vagina. Sensors located under the sheath measure the strength of the user's pelvic floor muscles. This information is then transmitted wirelessly to a smartphone application in order to provide real-time feedback to the user. It is a single patient, reusable device to be supplied over-the-counter.

Comparison of Technological Characteristics

Pelvic floor muscle contraction and instantaneous feedback to the user's smartphone is the technological principle for both the subject and predicate devices. An intravaginal device with embedded force sensors is used to monitor the contraction force of the user's pelvic floor muscles.

{5}------------------------------------------------

This information is transmitted to the patient's smartphone via bluetooth and is displayed on the screen. The subject and predicate devices are based on the following same technological elements:

  • an intravaginal probe with external silicone shell
  • · force sensors and bluetooth transmitter embedded inside the probe
  • a smartphone with a dedicated App

The technological differences between the Perifit Care+ and the predicate device do not raise different questions of safety or effectiveness. A table comparing the key features of the subject and predicate devices is provided below.

X6 Perifit Care+Perifit (K221476)Comments
Indications forUseThe Perifit Care+ is aperineometer designed totreat stress, mild-moderateurge and mixed urinaryincontinence in women, bystrengthening of the pelvicfloor muscles throughexercise. This deviceprovides biofeedback viasmart phone technology.The Perifit is aperineometer designed totreat stress, mild-moderateurge and mixed urinaryincontinence in women, bystrengthening of the pelvicfloor muscles throughexercise. This deviceprovides biofeedback viasmart phone technology.Identical
User PopulationAdult females with urinaryincontinence; availableover-the-counterAdult females with urinaryincontinence; availableover-the-counterIdentical
Technological Characteristics
Mode of UseReusable for single patientReusable for single patientIdentical
Principle ofOperationA probe inserted into thevagina to determine thestrength of the pelvic floormuscles. Probe sendssignals to external deviceto indicate musclecontraction strength toencourage and assist userA probe inserted into thevagina to determine thestrength of the pelvic floormuscles. Probe sendssignals to external deviceto indicate musclecontraction strength toencourage and assist userIdentical

Table 1: Substantial Equivalence Table

{6}------------------------------------------------

X6 Perifit Care+Perifit (K221476)Comments
with voluntary kegel exercises.with voluntary kegel exercises.
Sensing methodOutput from force sensing resistors (wireless).Output from force sensing resistors (wireless).Identical
Sensor's placementInside the rigid plastic enclosureInside the rigid plastic enclosureIdentical
MaterialsRigid plastic (PC/ABS) structure enclosed within a medical grade silicone outer layerRigid plastic (PC/ABS) structure enclosed within a medical grade silicone outer layerSimilar, new materials supplier
Parameter monitoredAnalogue to digital output of uncalibrated force exerted against external walls of device by pubococcygeus and puborectalis muscles.Analogue to digital output of uncalibrated force exerted against external walls of device by pubococcygeus and puborectalis muscles.Identical
User InterfaceSmartphone GUISmartphone GUIIdentical
Anatomical SitesFemale Pubococcygeus muscle areaFemale Pubococcygeus muscle areaIdentical
External shapeTwo egg shaped sensing areas with a tailTwo egg shaped sensing areasSimilar
Shaft length87.5 mm90 mmSimilar
WeightProbe weight: 40gProbe weight: 54gSimilar
Power SourceNon-rechargeable CR2032 Panasonic batteries, Voltage 3.0VDCNon-rechargeable CR2032 Panasonic batteries, Voltage 3.0VDCIdentical
Safety FeaturesElectronics and internal parts sealed in a medical grade silicone shellElectronics and internal parts sealed in a medical grade silicone shellIdentical
X6 Perifit Care+Perifit (K221476)Comments
Electrical SafetyTested in accordance withIEC 60601-1-2 and IEC60601-1Tested in accordance withIEC 60601-1-2 and IEC60601-1Perifit Care+includes updates tointernal electronicsto accommodatethe modified probeshape and anaddedaccelerometer.
BiocompatibilityBiocompatible - tested inaccordance with ISO10993standardsBiocompatible - tested inaccordance with ISO10993standardsIdentical
SoftwareSmartphone appcompatible for iOS andAndroidSmartphone appcompatible for iOS andAndroidPerifit Care+includes updatesto the software ofthe Perifit App andembeddedfirmware
SterilizationNon-sterile deviceNon-sterile deviceIdentical

{7}------------------------------------------------

Summary of Non-Clinical Performance Testing

The patient contacting materials in the Perifit Care+ have been tested in accordance with ISO 10993 standards and was found to be safe for the intended purpose. Biocompatibility testing included Cytotoxicity (ISO 10993-5, 2009), Sensitization (ISO 10993-10, 2010), Vaginal Irritation (ISO 10993-10, 2010), and Systemic Toxicity (ISO 10993-11, 2006). Electrical safety and electromagnetic compatibility testing have been conducted in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11 to establish the safety of the device and FDA guidance documents "Electromagnetic Compatibility (EMC) of Medical Devices" and "Radio Frequency Wireless Technology in Medical Devices" . Software verification and validation testing has also been conducted in accordance with FDA guidance (Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304:2006. Cybersecurity has been evaluated in accordance with FDA guidance document Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". In addition, various mechanical tests have been conducted which support substantial equivalence including mechanical drop testing, durability testing, and sensor behavior testing. The new device attribute, i.e., the tail for use in removing the device was tested tail bending and resistance testing. User testing through a questionnaire supports that users understand the key labeling provisions and how to operate the device.

Conclusions

The Perifit Care+ is as safe and effective as the Perifit Care+ has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate device. In addition, the technological differences between the Perifit Care+ and the predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Perifit Care+ is as safe and effective as the Perifit. Thus, the Perifit Care+ is substantially equivalent.

§ 884.1425 Perineometer.

(a)
Identification. A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.(b)
Classification. Class II (performance standards).