K Number
K231719

Validate with FDA (Live)

Date Cleared
2024-03-19

(280 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Medical Surgical Mask is single use, flat-pleated masks that are provided in blue. The Medical Surgical Mask is available in two types, which are Level 2 and Level 3 based on ASTM F2100-19. The inner and outer layers of the mask are made of polypropylene non-woven fabric, the middle layer is made of polypropylene meltblown fabric, and the nose clip is made of polypropylene and iron wire. Users can adjust the nose clip according to the shape of the bridge of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The ear loops are made of spandex. The ear loops are held in place over the users' mouth and nose by two ear loops welded to the mask. This is a single use, disposable device(s), provided non-sterile.

AI/ML Overview

The provided FDA 510(k) summary for the Azur Medical Company Inc. Medical Surgical Mask (K231719) details the acceptance criteria and study results for the device.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The device is a Medical Surgical Mask with two types: Level 2 and Level 3, based on ASTM F2100-19.
The table below combines information from the "VII Comparison of technological characteristics with the predicate device" and "VIII Summary of Non-Clinical Tests" sections. Note that for some performance characteristics, the specific numerical results are given in comparison to the predicate, while for others, only "Pass" is stated against the acceptance criteria.

Test ItemAcceptance Criteria (from ASTM F2100-19, or other standards)Reported Device Performance (K231719)
Performance Testing
Particulate Filtration Efficiency (PFE)$\ge$ 98% (ASTM F2299/F2299M-03)Level 2 and Level 3: average 99.96% (Pass)
Differential Pressure$<$ 6.0 mmH2O/cm² (ASTM F2100: 2019)Level 2 and Level 3: average 2.8 mmH2O/cm² (Pass)
Bacterial Filtration Efficiency (BFE)$\ge$ 98% (ASTM F2101: 2019)Level 2 and Level 3: average 99.9% (Pass)
FlammabilityClass I (16 CFR Part 1610)Class 1 (Pass)
Synthetic Blood Penetration ResistanceLevel 2: Pass at 120mmHg; Level 3: Pass at 160mmHg (ASTM F1862/F1862M-17)Level 2: Pass at 120mmHg; Level 3: Pass at 160mmHg (Pass)
Biocompatibility Testing
In Vitro CytotoxicityNon-cytotoxic (ISO10993-5: 2009)Non-cytotoxic (Pass)
Skin IrritationNon-irritating (ISO10993-23: 2021)Non-irritating (Pass)
Skin SensitizationNon-sensitizing (ISO10993-10: 2021)Non-sensitizing (Pass)
Other Characteristics
ASTM F2100 LevelMeet requirements for Level 2 and Level 3Level 2 and Level 3
Shelf lifeNot explicitly stated as an acceptance criterion in the provided section for novel data, but predicate was "Unknown"2 years (based on accelerated aging, real-time ongoing)

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample sizes used for each specific non-clinical test (e.g., how many masks were tested for PFE, BFE, etc.). It only reports the overall "Test results" as "Pass" or an average value for the proposed device.

The data provenance is not specified in terms of country of origin, but the testing would typically be performed by accredited labs. The studies described are non-clinical (laboratory-based performance testing and biocompatibility testing), not clinical trials, so the terms "retrospective" or "prospective" do not apply in the same way they would for patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable to this document. The "ground truth" for non-clinical testing of a surgical mask is established by validated standardized test methods (e.g., ASTM, ISO standards) and their defined acceptance criteria. It does not involve expert consensus in the way clinical diagnostic studies might.

4. Adjudication Method for the Test Set

This is not applicable to this document. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where expert readers independently review cases and then resolve discrepancies. Non-clinical performance tests follow predefined protocols and acceptance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This is not applicable to this document. This submission is for a physical medical device (surgical masks), not an AI/software as a medical device. Therefore, no MRMC comparative effectiveness study involving human readers or AI assistance was conducted or would be relevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

This is not applicable to this document. This is not an algorithm or AI device.

7. The Type of Ground Truth Used

The ground truth for this device's evaluation is based on standardized test methods and their defined performance criteria, primarily from ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). For example:

  • Particulate Filtration Efficiency: ASTM F2299/F2299M-03
  • Differential Pressure: ASTM F2100: 2019
  • Bacterial Filtration Efficiency: ASTM F2101: 2019
  • Flammability: 16 CFR Part 1610
  • Synthetic Blood Penetration Resistance: ASTM F1862/F1862M-17
  • Biocompatibility: ISO 10993 series (ISO10993-5, ISO10993-23, ISO10993-10)

These standards define the methodology and the quantitative or qualitative criteria for "passing" the test, which serves as the ground truth.

8. The Sample Size for the Training Set

This is not applicable to this document. "Training set" refers to data used to train machine learning models. This submission is for a physical medical device, not an AI/ML device, so there is no training set in this context.

9. How the Ground Truth for the Training Set Was Established

This is not applicable to this document, as there is no training set for an AI/ML model for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 19, 2024

Azur Medical Company Inc. % Esther Zhang Official Correspondent Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, Shanghai 201102 China

Re: K231719

Trade/Device Name: Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 8, 2024 Received: March 8, 2024

Dear Esther Zhang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231719

Device Name Medical Surgical Mask

Indications for Use (Describe)

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) summary K231719

l Submitter

Device submitter: Azur Medical Company Inc. 6710 Everglades Dr., Richmond, Virginia, VA 23225, USA

Contract manufacturer: Wepon Medical Technology CO., LTD. Floor 4, Building A, No. 58, Jinhu Road, Chengdong Street, Wenling Zhejiang, CN 317500

Contact person: Di Zhao General Manager Phone: 928-5922380 Email: dzhao@azur-ppe.com Date: March 14, 2024

II Device

Trade Name of Device: Medical Surqical Mask Common name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation name: Mask, Surgical Regulatory Class: II Product code: FXX Review Panel: General Hospital

III Correspondent

Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District, Shanghai, P.R.China Contact: Esther ZHANG Email: Esther.zhang@llins-tech.com

IV Predicate Device

510(k) Number:K210030
Trade/Device Name:Medical Surgical Mask
Regulation Number:21 CFR 878.4040
Regulation Name:Surgical Apparel
Classification:Class II
Product Code:FXX

{5}------------------------------------------------

V Device description

The Medical Surgical Mask is single use, flat-pleated masks that are provided in blue. The Medical Surgical Mask is available in two types, which are Level 2 and Level 3 based on ASTM F2100-19.

The inner and outer layers of the mask are made of polypropylene non-woven fabric, the middle layer is made of polypropylene meltblown fabric, and the nose clip is made of polypropylene and iron wire. Users can adjust the nose clip according to the shape of the bridge of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The ear loops are made of spandex. The ear loops are held in place over the users' mouth and nose by two ear loops welded to the mask.

This is a single use, disposable device(s), provided non-sterile.

VI Indications for use

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

VII Comparison of technological characteristics with the predicate device

The Medical Surgical Mask has similar intended use, technology, design and performance specifications to the proposed legally marketed predicate device.

ItemSubject device(K231719)Predicate device(K210030)Discussion
Product nameMedical Surgical MaskDisposable Surgical FaceMaskNA
Product CodeFXXFXXidentical
Intended useThe Medical Surgical Maskis intended to be worn toprotect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial. These surgicalmasks are intended for useThe Medical Surgical Masksare intended to be worn toprotect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial.These surgicalmasks are intended for usein infection control practicesidentical
in infection control practicesto reduce the potentialexposure to blood and bodyfluids. This is a single use,disposable device(s),provided non-sterile.to reduce the potentialexposure to blood and bodyfluids. This is a single use,disposable device(s),provided non-sterile.identical
Mask styleFlat-pleatedFlat-pleatedidentical
ASTM F2100 LevelLevel 2 and Level 3Level 2 and Level 3identical
Design featureLevel 2: Ear loopLevel 3: Ear loopLevel 2: Ear loopLevel 3: Ear loop and Tie-onDifferent, seeexplanationbelow
ColorBlueBlueidentical
Dimension17.5cmx9.5cm17.5cmx9.5cmidentical
SterilityNon-SterileNon-Sterileidentical
Shelf life2 yearsUnknownDifferent, seeexplanationbelow
UseSingle Use, DisposableSingle Use, Disposableidentical
ParticulatefiltrationefficiencyLevel 2 and Level 3:average 99.96%Level 2 mask: average99.71%Level 3 mask: average99.93%Different, seeexplanationbelow
Bacterial filtrationefficiencyLevel 2 and Level 3:average 99.9%Level 2 mask: average99.7%Level 3 mask: average99.9%Different, seeexplanationbelow
Differential pressureLevel 2 and Level 3:average 2.8 mmH2O/cm 2Level 2 mask: average2.8mmH2O/cm 2Level 3 mask: average 4.0mmH2O/cm 2EN 14683Different, seeexplanationbelow
FlammabilityClass 1Class 1identical
Fluid resistanceLevel 2: Pass at 120mmHgLevel 3: Pass at 160mmHgLevel 2: Pass at 120mmHgLevel 3: Pass at 160mmHgidentical
Label/LabelingComplied with 21 CFR part801Complied with 21 CFR part801identical
PatientContactingOuterfacinglayerpolypropylene non-wovenfabricSpunbond PolypropyleneDifferent, seeexplanationbelow
Middle layerpolypropylene meltblown fabricMeltblown Polypropylene Filter
Inner facing layerpolypropylene non-woven fabricSpunbond Polypropylene
Nose clippolypropylene and iron wirePE and Iron
Ear loopsspandexSpandex
BiocompatibilityISO 10993-5 and ISO 10993-10;Under the conditions of thestudy, the proposed deviceextract was determined tobe non-cytotoxic, non-sensitizing, and non-irritating.ISO 10993-5 and ISO 10993-10;Under the conditions of thestudy, the proposed deviceextract was determined tobe non-cytotoxic, non-sensitizing, and non-irritating.
identical

Table 5-1 Substantial equivalence discussion - Medical Surgical Mask

{6}------------------------------------------------

{7}------------------------------------------------

Different- Design feature

The proposed masks are ear-loop masks. The predicate Level 3 masks are available in two types, ear-loop and Tie-on. The design features of the two levels of masks for the proposed device can be covered by the design features of the predicate device. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

Different-Shelf life

The shelf life of the predicate device is unknown and the proposed mask is 2 years. Shelf life testing has been conducted for the proposed device based on accelerated aging. A real-time aging study is underway. Therefore, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

Different - Particulate filtration efficiency

The test result for particulate filtration efficiency for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask and Level 3 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

Different - Bacterial filtration efficiency

The test result for bacterial filtration efficiency for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask and Level 3 mask based on ASTM F2100-19. Thus, this

{8}------------------------------------------------

difference will not affect the safety and effectiveness between the proposed device and the predicate device.

Different - Differential pressure

The test result for differential pressure for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask and Level 3 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

Different - Patient Contacting Material

The patient contacting material for the proposed device is different from the predicate device. However, biocompatibility test has been performed on the proposed device and the results does not show any adverse effect. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

VIII Summary of Non-Clinical Tests

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate device.

No.Test ItemReference StandardAcceptance criteriaTest results
1.ParticulateFiltrationEfficiency(PFE)ASTM F2299/F2299M-03 (2017)Standard Test Method forDetermining the Initial Efficiency ofMaterial Used in medical FaceMasks to Penetration byParticulates using Latex Spheres.$\ge$ 98%Pass
2.DifferentialPressureASTM F2100: 2019 StandardSpecification for Performance ofMaterials Used in Medical FaceMasks.$<$ 6.0mmH2O/cm²Pass
3.BacterialFiltrationEfficiency(BFE)ASTM F2101: 2019 Standard TestMethod for Evaluating theBacterial Filtration Efficiency(BFE) of Medical Face MaskMaterials, Using a BiologicalAerosol of Staphylococcusaureus.$\ge$ 98%Pass

Performance testing

{9}------------------------------------------------

4.Flammability of Clothing Textiles16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;Class IPass
5.Synthetic Blood Penetration ResistanceASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity);Level 2: Pass at 120mmHgLevel 3: Pass at 160mmHgPass

Biocompatibility testing

Biocompatibility of the Medical Surgical Mask was evaluated in accordance with ISO 10993-1:2018.

No.Test ItemReference StandardAcceptance criteriaTestresults
1.In Vitro CytotoxicityISO10993-5: 2009Under the conditions of thestudy, the proposed deviceextract is determined to benon-cytotoxicPass
Biological evaluation
of medical devices -
Part 5: Tests for in
vitro cytotoxicity
2.Skin IrritationISO10993-23: 2021Under the conditions ofthe study, the proposeddevice extract isdetermined to be non-irritatingPass
Biological evaluation
of medical devices -
Part 23: Tests for
irritation
3.Skin SensitizationISO10993-10: 2021Under the conditionsof the study, theproposed deviceextract is determinedto be non-sensitizingPass
Biological evaluation
of medical devices -
Part 10: Tests for
skin sensitization

IX Summary of Clinical Tests

N/A

X Conclusion

The conclusions drawn from the non-clinical tests demonstrates that the subject device, Medical Surgical Mask, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K210030.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.