(58 days)
The NorthStar OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, cervical spine (C1-C7), and upper thoracic spine (T1-T3):
- · Traumatic spinal fractures and/or traumatic dislocations,
- · Instability or deformity,
- · Failed previous fusion (e.g., pseudoarthrosis),
- · Tumors involving the cervical/thoracic spine,
- · Degenerative disease. including intractable radiculopathy and neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and
- · Degenerative disease of the facets with instability.
The NorthStar OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The NorthStar OCT System can also be linked to other FDA-cleared SeaSpine/Orthofix screw systems with the use of transitional rods and/or transitional rod connectors.
The NorthStar OCT System is a posterior occipital cervical thoracic (OCT) system that consists of implants and the associated instruments used to build constructs within the body to act as a temporary or permanent posterior fixation system to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusion to occur. The system includes a variety of non-sterile, single-use implants manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136), commercially pure titanium (Grade 2 per ASTM F67), and/or cobalt chrome alloy (Co-28Cr-6Mo per ASTM F1537) and is comprised of polyaxial screws, rods, hooks, connectors, occipital plates, occipital screws, and set screws that can be rigidly locked together in a variety of configurations.
The NorthStar OCT System includes the associated instruments that are designed to facilitate the placement, adjustment, final locking, and removal, if necessary, of the system implants. The implants and instruments are provided in trays and caddies for storage, protection, and organization prior to and during the steam sterilization process.
The provided text describes the NorthStar OCT System, a posterior occipital cervical thoracic (OCT) system, and its premarket notification (510(k)) to the FDA. However, the document does not contain any information about acceptance criteria or a study proving the device meets these criteria in the context of an AI/ML-driven device.
The document discusses:
- The device's intended use and indications for use.
- Its components, materials, and overall design.
- Non-clinical mechanical testing (dynamic compression bending, static compression bending, static torsion, and dynamic torsion testing per ASTM F2706) to demonstrate substantial equivalence to predicate devices.
- A explicit statement that clinical testing was not applicable and the determination of substantial equivalence is not based on an assessment of clinical performance data.
Therefore, I cannot fulfill your request to describe acceptance criteria and a study proving the device meets them, as the provided text pertains to a traditional orthopedic implant (hardware) that relies on mechanical testing for 510(k) clearance, not an AI/ML diagnostic or assistive device that would typically involve a test set, ground truth experts, or MRMC studies.
If you have a document describing an AI/ML device, please provide that, and I would be happy to analyze it according to your requested criteria.
{0}------------------------------------------------
SeaSpine Orthopedics Corporation Mr. Jesse Albright Manager, Regulatory Affairs 5770 Armada Dr. Carlsbad, California 92008
Re: K231654
Trade/Device Name: NorthStar OCT System Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG, KWP Dated: June 5, 2023 Received: June 6, 2023
Dear Mr. Jesse Albright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
August 3, 2023
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2023.08.03 for
17:04:39 -04'00'
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
Device Name
NorthStar OCT System
Indications for Use (Describe)
The NorthStar OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, cervical spine (C1-C7), and upper thoracic spine (T1-T3):
- · Traumatic spinal fractures and/or traumatic dislocations,
- · Instability or deformity,
- · Failed previous fusion (e.g., pseudoarthrosis),
- · Tumors involving the cervical/thoracic spine,
- · Degenerative disease. including intractable radiculopathy and neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and
- · Degenerative disease of the facets with instability.
The NorthStar OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The NorthStar OCT System can also be linked to other FDA-cleared SeaSpine/Orthofix screw systems with the use of transitional rods and/or transitional rod connectors.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Contact Details
| Applicant Name: | SeaSpine Orthopedics Corporation | ||
|---|---|---|---|
| Address:Phone number:Fax number: | 5770 Armada Drive, Carlsbad, CA 92008(760) 216-5176(760) 683-6874 | ||
| Contact Person: | Jesse Albright, Manager, Regulatory Affairs | ||
| Date Prepared: | June 5, 2023 | ||
| Device Name | |||
| Trade Name: | NorthStar OCT System | ||
| Common Name: | Posterior Cervical Screw SystemSpinal Interlaminal Fixation System | ||
| Classification Name: | Posterior Cervical Screw System (21 CFR 888.3075)Appliance, Fixation, Spinal Interlaminal (21 CFR 888.3050) | ||
| Product Code(s): | NKG, KWP |
Legally Marketed Predicate Devices
2
| 510(k) Number | Product Code(s) | Trade Name(s) | Manufacturer |
|---|---|---|---|
| Primary Predicate Device | |||
| K193615 | NKG; KWP | NorthStar OCT System | SeaSpineOrthopedicsCorporation |
| Additional Predicate Device(s) | |||
| K080526 | KWP | Sierra OCT System | SeaSpineOrthopedicsCorporation |
Device Description
Device Class:
The NorthStar OCT System is a posterior occipital cervical thoracic (OCT) system that consists of implants and the associated instruments used to build constructs within the body to act as a temporary or permanent posterior fixation system to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusion to occur. The system
{4}------------------------------------------------
includes a variety of non-sterile, single-use implants manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136), commercially pure titanium (Grade 2 per ASTM F67), and/or cobalt chrome alloy (Co-28Cr-6Mo per ASTM F1537) and is comprised of polyaxial screws, rods, hooks, connectors, occipital plates, occipital screws, and set screws that can be rigidly locked together in a variety of configurations.
The NorthStar OCT System includes the associated instruments that are designed to facilitate the placement, adjustment, final locking, and removal, if necessary, of the system implants. The implants and instruments are provided in trays and caddies for storage, protection, and organization prior to and during the steam sterilization process.
Intended Use/Indications for Use
The NorthStar OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, cervical spine (C1-C7) and upper thoracic spine (T1-T3):
- Traumatic spinal fractures and/or traumatic dislocations,
- Instability or deformity,
- . Failed previous fusion (e.g., pseudoarthrosis),
- Tumors involving the cervical/thoracic spine,
- . Degenerative disease, including intractable radiculopathy and/or myelopathy and neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and
- . Degenerative disease of the facets with instability.
The NorthStar OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The NorthStar OCT System can also be linked to other FDA-cleared SeaSpine/Orthofix screw systems with the use of transitional rods and/or transitional rod connectors.
Summary of Technological Characteristics
The NorthStar OCT System is identical or similar to the cited predicate systems in regard to intended use/indications for use, device description, technological characteristics (e.g., operating principle, design, components, materials, manufacturing, labeling, sterility, etc.), and non-clinical performance (i.e., mechanical testing).
{5}------------------------------------------------
Non-Clinical Testing
The NorthStar OCT System demonstrated substantially equivalent mechanical performance to the predicate systems through dynamic compression bending, static compression bending, static torsion, and dynamic torsion testing per ASTM F2706.
Clinical Testing
Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data.
Conclusions
The submitted data demonstrate that the NorthStar OCT System is substantially equivalent to the cited legally marketed predicates.
N/A