K Number
K231493

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-08-11

(80 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NITINEX Memory Compression Staple is intended for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

Device Description

The NITINEX Memory Compression Staple is a single-use bone fixation appliance intended to be permanently implanted. The implantation of the NITINEX Memory Compression Staple facilitates hand and foot bone fragment osteotomy fixation and joint arthrodesis. NITINEX Staples are compression staples made of shape memory nickel titanium alloy, nitinol. Vilex LLC will offer NITINEX Staples ranging in width from 8-mm to 30-mm with leg lengths ranging from 8-mm to 30-mm.

AI/ML Overview

The provided text is an FDA 510(k) summary for a medical device called the "NITINEX Memory Compression Staple." It focuses on demonstrating substantial equivalence to a predicate device, as required for 510(k) clearance.

Therefore, the document does not describe acceptance criteria or a study that proves the device meets specific performance criteria in terms of clinical accuracy or diagnostic performance, such as those typically found for AI/ML-based medical devices. The assessments performed are primarily nonclinical (mechanical and corrosion testing) to confirm the device's physical properties are equivalent to the predicate.

Here's why the requested information cannot be extracted from this document:

  • This is not an AI/ML device: The NITINEX Memory Compression Staple is a physical implant (a bone fixation device). The acceptance criteria and performance studies described in your prompt (e.g., diagnostic accuracy, human reader improvement with AI, ground truth establishment) are relevant for software as a medical device (SaMD) or AI-powered diagnostic tools, not for physical implants.
  • 510(k) for physical devices focuses on substantial equivalence: For devices like the NITINEX staple, the 510(k) pathway primarily requires demonstrating that the new device is as safe and effective as a legally marketed predicate device. This is typically done through comparisons of indications for use, technological characteristics, and non-clinical performance data (like mechanical strength and corrosion resistance). Clinical studies for physical implants are often more involved and specific to the device's function (e.g., healing rates, complication rates), but they are not detailed in a 510(k) summary in the way an AI performance study would be.
  • The "performance data" mentioned is nonclinical: The document explicitly states: "The following nonclinical tests were performed to demonstrate the substantial equivalence of the subject device to the predicate device: Static Pullout Testing per ASTM F564-17, Bend Testing per ASTM F564-17, Cyclic Polarization Corrosion Testing per ASTM F2129-19a." These are mechanical and material property tests, not clinical performance studies involving human subjects or expert readers.

Therefore, I cannot fill out the requested table or answer the questions related to AI/ML device performance. The information provided in the document simply does not contain details about:

  • Acceptance criteria for diagnostic accuracy (sensitivity, specificity, AUC)
  • Sample size for test sets in an AI/ML context
  • Data provenance for clinical images or patient data
  • Number/qualifications of experts for ground truth establishment
  • Adjudication methods for ground truth
  • MRMC studies or effect sizes of human reader improvement with AI assistance
  • Standalone algorithm performance
  • Type of ground truth (pathology, outcomes data, expert consensus for diagnosis)
  • Training set sample size or how training set ground truth was established

The document confirms the device passed the specified nonclinical tests and was found to be substantially equivalent to the predicate device in terms of mechanical strength, performance, and corrosion resistance.

If you intended to ask about the acceptance criteria and study for an AI/ML-based medical device, this document is not the correct source of information.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 11, 2023

Vilex, LLC Brock Johnson President 111 Moffitt Street McMinnville, Tennessee 37110

Re: K231493

Trade/Device Name: NITINEX Memory Compression Staple Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDR Dated: August 9, 2023 Received: August 10, 2023

Dear Brock Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K231493

Device Name

NITINEX Memory Compression Staple

Indications for Use (Describe)

The NITINEX Memory Compression Staple is intended for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k)

510(k) SUMMARY

Device Trade NameNITINEX Memory Compression Staple
DateAugust 9, 2023
SponsorVilex, LLC111 Moffitt StreetMcMinnville, TN 37110
Contact PersonBrock JohnsonPresident(801)916-4157brock.johnson@vilex.com
Device Common NameNITINEX Memory Compression Staple
Regulatory ClassClass II
Product CodeJDR
Classification NameStaple, Fixation, Bone
Regulation21 CFR 888.3030
Regulation NameSingle/multiple component metallic bone fixationappliances and accessories
Device Regulation PanelOrthopedic
Predicate DeviceK112837, Vilex eZ-Staple Superelastic Bone FixationStaple

Device Description:

The NITINEX Memory Compression Staple is a single-use bone fixation appliance intended to be permanently implanted. The implantation of the NITINEX Memory Compression Staple facilitates hand and foot bone fragment osteotomy fixation and joint arthrodesis. NITINEX Staples are compression staples made of shape memory nickel titanium alloy, nitinol. Vilex LLC will offer NITINEX Staples ranging in width from 8-mm to 30-mm with leg lengths ranging from 8-mm to 30-mm.

Indications for Use:

The NITINEX Memory Compression Staple is intended for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

The indications for use of the NITINEX Memory Compression Staple are identical to those of the predicate device (K112837).

Technological Characteristics:

The NITINEX Memory Compression Staple is substantially equivalent to the predicate devices cleared under K112837. The NITINEX Memory Compression Staple has identical indications for use, basic designs, fundamental scientific principles of operation, and implant materials as the predicate device.

{4}------------------------------------------------

NITINEX Memory Compression Staple

Traditional 510(k)

The NITINEX Memory Compression Staple has a thicker staple bridge and different teeth size and spacing than the predicate device.

The NITINEX Memory Compression Staple System contains a sizer instrument to aid the surgeon in determining the appropriate size implant to use on the patient. The NITINEX Memory Compression Staple System also contains long drills and short drills to form snowmanshaped holes to aid in insertion. The predicate devices do not contain a sizer or long drills.

The NITINEX Memory Compression Staple System uses instruments that are composed of a nylon material and stainless steel. The predicate devices use instruments composed of stainless steel.

The NITINEX Memory Compression Staple is provided sterile to the end user and is intended for single use only. The predicate devices require end user sterilization prior to use and are intended for single use only.

The differences between the NITINEX Memory Compression Staple and the predicate devices are considered minor and raise no questions of safety or effectiveness. Therefore, the NITINEX Memory Compression Staple is substantially equivalent to the predicate device.

Assessment of performance data:

The following nonclinical tests were performed to demonstrate the substantial equivalence of the subject device to the predicate device:

  • Static Pullout Testing per ASTM F564-17
  • Bend Testing per ASTM F564-17
  • Cyclic Polarization Corrosion Testing per ASTM F2129-19a

The mechanical tests evaluated samples from the NITINEX Memory Compression Staple against the predicate device. All tests demonstrated that the subject device performs equivalently to the predicate device. Therefore, the NITINEX Memory Compression Staple is substantially equivalent to the predicate device with respect mechanical strength and performance.

The corrosion testing demonstrated the corrosion resistance of the subject device. The NITINEX Memory Compression Staple was shown to meet equivalent corrosion resistance criteria as the predicate device.

Conclusion:

Based upon the similarities between the subject and the predicate devices, the NITINEX Memory Compression Staple is substantially equivalent to the predicate device. The technological differences between the subject and predicate devices are considered minor. The similarities in technological characteristics and the performance data demonstrate substantial equivalence.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.