K Number
K231448

Validate with FDA (Live)

Date Cleared
2023-07-19

(62 days)

Product Code
Regulation Number
888.3360
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Smith & Nephew TANDEM Unipolar and Bipolar Hip System is intended for partial hip arthroplasty in skeletally mature patients. It is indicated for use in patients not suitable for total hip arthroplasty, with a non-functional femoral head due to femoral neck fracture.

The TANDEM Unipolar and Bipolar head components are intended to articulate against the natural acetabulum, which requires no use of bone cement.

Device Description

The TANDEM Hip System is designed to be used for hemiarthroplasty as implantation of these devices replace diseased, affected, or broken femoral head and/or neck with a prosthetic component, while keeping the natural acetabulum, or hip socket, intact. The TANDEM Hip System includes the TANDEM Unipolar Hip System and TANDEM International Bipolar Hip Systems are used for partial hip replacement procedure, in which the unipolar shell are designed to articulate against the host acetabular articular cartilage and replace the femoral head and/or neck during the implantation procedure. The key difference between a unipolar and a bipolar version of the TANDEM Hip System is number of rotational points that are provided – single (unipolar) or two (bipolar) — within the socket.

AI/ML Overview

This document describes a 510(k) premarket notification for the TANDEM Hip System, which focuses on labeling updates to a previously cleared device. Therefore, the information provided does not include typical performance criteria and studies for a novel device. Instead, the submission argues for substantial equivalence based on the device being functionally and technologically identical to its predicates, with only changes to labeling and indications for use.

Due to the nature of this 510(k) submission (labeling updates only), there are no acceptance criteria and associated performance data from a study designed to prove the device meets these criteria in the traditional sense of a new device clearance. The "performance" being assessed here is the continued safety and effectiveness under the updated labeling, which is justified by the equivalence to existing cleared devices.

Here's an breakdown based on the provided text, addressing your points where applicable:

  1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated as quantifiable performance metrics for a new device. The implicit acceptance criterion is that the device, with its updated labeling, remains substantially equivalent to its cleared predicates in terms of safety and effectiveness.
    • Reported Device Performance: Not applicable in the context of a new performance study. The submission asserts that the devices are "identical in function, design features, materials, sterilization, manufacturing methods and operational principles to what was previously 510(k) cleared." Therefore, their performance is presumed to be consistent with the performance already established for the predicate devices.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • No test set was used for performance testing, as per the submission: "Therefore, since there are no changes to the design features, materials, or manufacturing methods of the TANDEM Hip System devices, no performance testing (bench, animal, clinical) was required."
    • The submission mentions that "post market surveillance data" was used to inform the labeling updates, particularly those related to the European Medical Device Regulation 2017/745. However, specific details about the sample size, provenance, or whether it was retrospective or prospective are not provided.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No new test set required ground truth establishment by external experts for performance evaluation. The justification relies on the existing clearance of the predicate devices.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No test set for performance evaluation was conducted.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted device. No MRMC study was conducted.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm-based device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the purpose of this submission (labeling updates), the "ground truth" for the device's safety and effectiveness relies on the previously established substantial equivalence of the predicate devices to which the subject device is identical, and implicitly, the outcomes data and clinical experience gathered since their original clearance. The submission explicitly states "post market surveillance data" informed the labeling changes, which would involve real-world outcomes.
  8. The sample size for the training set

    • Not applicable. No new training set was required for performance testing as there were no changes to device design, materials, or manufacturing requiring new performance data.
  9. How the ground truth for the training set was established

    • Not applicable, as no new training set was used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 19, 2023

Smith & Nephew, Inc. Nikita Deshpande Sr. Regulatory Affairs Specialist 1450 Brooks Road Memphis, Tennessee 38116

Re: K231448

Trade/Device Name: TANDEM Hip System Regulation Number: 21 CFR 888.3360 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metallic Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: LZY, KWY Dated: May 18, 2023 Received: May 18, 2023

Dear Nikita Deshpande:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Limin Sun-S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of JointArthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Submission Number (if known)

K231448

Device Name

TANDEM Hip System

Indications for Use (Describe)

The Smith & Nephew TANDEM Unipolar and Bipolar Hip System is intended for partial hip arthroplasty in skeletally mature patients. It is indicated for use in patients not suitable for total hip arthroplasty, with a non-functional femoral head due to femoral neck fracture.

The TANDEM Unipolar and Bipolar head components are intended to articulate against the natural acetabulum, which requires no use of bone cement.

Type of Use (Select one or both, as applicable)

< | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K231448 Page 1 of 3
VI-\ C- LIGO 100 OUL0000 0- 11

21 CFR 807.92(a)(1)

Applicant NameSmith & Nephew, Inc.
Applicant Address1450 Brooks Road Memphis TN 38116 United States
Applicant Contact Telephone901-825-8527
Applicant ContactMs. Nikita Deshpande
Applicant Contact Emailnikita.deshpande@smith-nephew.com
21 CFR 807.92(a)(2)
Device Trade NameTANDEM Hip System
Common NameHip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis
Classification NameProsthesis, Hip, Hemi-, Femoral, Metal Ball
Regulation Number888.3360
Product CodeLZY
21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K896580TANDEM UNI-POLAR Hip SystemLZY
K023743TANDEM International Bipolar Hip SystemKWY
K211176Smith & Nephew Hip SystemLPH
21 CFR 807.92(a)(4)
The purpose of this traditional 510(k) is to notify FDA of Smith & Nephew's intent to request clearance for labeling updates to Smith &

Nephew's TANDEM Hip System.

The TANDEM Hip Systems have been initially cleared by FDA and the information related to the clearance of these devices has been provided below:

  1. TANDEM Unipolar Hip System (CoCr Heads and Taper Sleeves) – K896580 (S.E Date – 02/15/1990) 2. TANDEM INTL (International) Bipolar Hip System – K023743 Letter To File (Date – 01/21/2004) Note: The TANDEM International Bipolar Hip System is a modified version of the TANDEM Bipolar System which was cleared for market under 510(k) submission number K023743.

The subject Smith & Nephew TANDEM Hip System devices are identical in function, design features, materials, sterilization, manufacturing methods and operational principles to what was previously 510(k) cleared. These labeling updates do not affect the safety and effectiveness of the subject devices when used as labeled.

The labeling updates include changes made to the (IFU), device labels, and surgical technique. The indications for use and target patient population have been limited for using the subject TANDEM Hip System devices and post market surveillance data. These updates have been made to the European Medical Device Regulation

{4}------------------------------------------------

2017/745 and standardization of verbiage across Smith & Nephew's hip implants.

The TANDEM Hip System is designed to be used for hemiarthroplasty as implantation of these devices replace diseased, affected, or broken femoral head and/or neck with a prosthetic component, while keeping the natural acetabulum, or hip socket, intact. The TANDEM Hip System includes the TANDEM Unipolar Hip System and TANDEM International Bipolar Hip Systems are used for partial hip replacement procedure, in which the unipolar shell are designed to articulate against the host acetabular articular cartilage and replace the femoral head and/or neck during the implantation procedure. The key difference between a unipolar and a bipolar version of the TANDEM Hip System is number of rotational points that are provided – single (unipolar) or two (bipolar) — within the socket.

Materials

The TANDEM Unipolar Hip System comprises of unipolar femoral head and taper sleeve. The TANDEM Unipolar femoral head components are manufactured using Cobalt-Chromium alloy conforming to ASTM F75 and ISO 5832-4. The Taper Sleeve components are manufactured from titanium (Ti644V) alloy conforming to ASTM F1472 and ISO 5832-3. The Taper Sleeve components are also manufactured from annealed ELI (extra low interstitial) grade Titanium alloy (Ti6Al4V) conforming to ASTM F136.

The TANDEM International Bipolar Hip System is manufactured from cobalt chrom cobalt chromium (CoCr) alloy conforming to ASTM F75 and ISO 5832-4, Ultra High Molecular Weight Polyethylene (UHMWPE) conforming to ASTM F648 and ISO 5834-2, titanium (Ti6Al4V) alloy conforming to ASTM F1472 and ISO 5832-3.

Intended Use/Indications for Use

The Smith & Nephew TANDEM Unipolar and Bipolar Hip System is intended for partial hip arthroplasty in skeletally mature patients. It is indicated for use in patients not suitable for total hip arthroplasty, with a non-functional femoral neck fracture.

The TANDEM Unipolar and Bipolar head components are intended to articulate against the natural acetabulum, which requires no use of bone cement.

Indications for Use Comparison

The purpose of this traditional 510(k) is to notify FDA of Smith & Nephew's intent to request clearance for labeling updates to Smith & Nephew's TANDEM Hip System. The labeling updates include changes to the indications for use.

The Smith & Nephew TANDEM Unipolar and International Bipolar Hip Systems are identical in function, materials, sterilization, manufacturing methods and operational principles to the commercially available predicate devices TANDEM Unipolar Hip System (K896580, 02/15/1990) and TANDEM International Bipolar Hip System (K0237343, 01/21/2004).

The labeling updates include changes made to the (IFU), device labels, and surgical technique. The indications for use and target patient population have been limited for using the subject TANDEM Hip System devices and post market surveillance data. These updates have been made to the European Medical Device Regulation 2017/745 and standardization of verbiage across Smith & Nephew's hip implants.

Technological Comparison

The devices that comprise the TANDEM Unipolar and Bipolar Hip System have the same technological characteristics, including device design and material, as the predicate devices previously cleared by the FDA. Therefore, no additional testing was performed for the subject devices to support the proposed labeling updates. There are no new issues related to safety and effectiveness of the subject devices.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The purpose of this traditional 510(k) is to request clearance from FDA for labeling changes to the subject Smith & Nephew TANDEM Hip System implants.

The subject Smith & Nephew TANDEM Hip System devices are identical in function, design features, materials, sterilization, manufacturing methods and operational principles to what was previously 510(k) cleared. These labeling updates do not affect the safety and effectiveness of the subject devices when used as labeled.

Therefore, since there are no changes to the design features, materials, or manufacturing methods of the TANDEM Hip System devices, no performance testing (bench, animal, clinical) was required.

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

{5}------------------------------------------------

Clinical data was not needed to support the safety and effectiveness of the subject devices.

No new or modified hip implant components are being introduced as a result of this filing. The subject devices are substantially equivalent to the previously 510(k) cleared predicate devices.

N/A