K Number
K231404

Validate with FDA (Live)

Date Cleared
2023-08-04

(81 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button & T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button are to be used for fixation of bone or soft tissue to bone, and are intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. This device is employed in Knee ligament repair and reconstructive procedures, specifically for ACL/PCL repair and reconstruction.

Device Description

The T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button & T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button is a permanently implantable Polyether Ether Ketone (PEEK) cortical tibial suspension fixation device which provides the orthopaedic surgeon a means of accurate fixation in ligament reconstructive surgeries and repair. The fixation device allows for endoscopic or open ligament reconstruction approaches. The Closed button is preloaded with UHMWPE adjustable loop and in Open Button the UHMWPE adjustable loop is non preloaded.

AI/ML Overview

This is a 510(k) premarket notification for a medical device called the T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button and the T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button. This document does not describe a study involving AI or human readers. Instead, it focuses on demonstrating the substantial equivalence of the new device to existing predicate devices through non-clinical performance testing and material comparisons.

Therefore, many of the requested sections related to AI performance, such as sample sizes for test and training sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, and standalone algorithm performance, are not applicable to this type of device submission.

Here's the information that can be extracted from the provided text regarding the device's acceptance criteria and the study proving it meets them:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria / Performance TestReported Device PerformanceComments
Mechanical Performance
Pull-Out TestingThe results were found equipotent (equivalent) to the primary predicate (Arthrex ACL TightRope®) and reference predicate (ANCHORMAN Tibial Ligament Fixation Device), defining the strength of the suture button to hold the loop. The acceptance criteria was met.Similar performance to predicate devices.
Cyclic LoadingThe results were found equipotent (equivalent) to the primary predicate (Arthrex ACL TightRope®) and reference predicate (ANCHORMAN Tibial Ligament Fixation Device). The acceptance criteria was met.Similar performance to predicate devices.
Biocompatibility
Skin Sensitization (Guinea Pig Minimization Test)Non-SensitizerMeets biocompatibility standards.
Intracutaneous Reactivity TestNon-ReactiveMeets biocompatibility standards.
Material Mediated PyrogenicityNon-PyrogenicMeets biocompatibility standards.
Acute Systemic Toxicity TestDid not Induce any Systemic ToxicityMeets biocompatibility standards.
In vitro Cytotoxicity Test: Elution MethodNon-CytotoxicMeets biocompatibility standards.
Bone Implantation Studies (PEEK Rods)Minimal or No ReactionMeets biocompatibility standards.
Bone Implantation Studies (UHMWPE Suture)Minimal or No ReactionMeets biocompatibility standards.
Intramuscular Implant (PEEK)Minimal or No ReactionMeets biocompatibility standards.
Other Safety/Performance Tests
Sterilization ValidationNot explicitly detailed, but stated as "performed to support the demonstration of substantial equivalence." (Sterilization method is ETO, similar to predicate device.)Assumed to meet industry standards.
Shelf-Life5 Years (Similar to predicate device). Not explicitly detailed, but stated as "performed to support the demonstration of substantial equivalence."Assumed to meet industry standards.
Endotoxin MonitoringNot explicitly detailed, but stated as "performed to support the demonstration of substantial equivalence." (Bacterial Endotoxin testing stated for safety data, similar to predicate.)Assumed to meet industry standards.
Pyrogenicity TestingNot explicitly detailed, but stated as "performed to support the demonstration of substantial equivalence." (Material Mediated Pyrogenicity testing stated for safety data.)Assumed to meet industry standards.
Specifications and DimensionsChanges in dimensions (16mm x 7.5mm for subject device button vs. 13mm x 3.3mm x 1.5mm for primary predicate) were observed not to lead to any new risks with performance, and acceptance criteria were met in each performance test. (Loop Length: 90 mm, similar to predicate).Device design differences (larger button) do not negatively impact performance as per testing.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size: Not explicitly stated in the document. The document refers to "bench testing" and "studies" but does not quantify the number of devices tested for mechanical or biocompatibility assessments.
  • Data Provenance: The studies are non-clinical (bench testing, in vitro, and in vivo animal studies for biocompatibility). India (Healthium Medtech Limited is based in Bangalore, India) is the location of the manufacturer and presumably where these tests were conducted or commissioned. The data is prospective, generated specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. This is a physical medical device. "Ground truth" in this context refers to the validated methods and resulting measurements from mechanical and biological tests, not expert interpretation of data. These tests would be conducted by qualified laboratory personnel and validated against international standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. As above, this is not an interpretive diagnostic device. Test results are quantitative measurements against predetermined specifications and international standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a physical implant for orthopedic surgery, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For mechanical performance (Pull-Out, Cyclic Loading): Quantitative measurements of force, displacement, and cycles against predefined specifications and industry standards. The "ground truth" is that the device must withstand certain loads and cycles without failure, demonstrating "equipotent" performance to predicate devices.
  • For biocompatibility: Adherence to international standards (e.g., ISO 10993) for biological evaluation of medical devices, using in vitro and in vivo (animal) models to assess toxicity, sensitization, pyrogenicity, and tissue reaction. The "ground truth" is the observation of "non-sensitizer," "non-reactive," "non-pyrogenic," "non-cytotoxic," and "minimal or no reaction" based on established biological endpoints.

8. The sample size for the training set:

  • Not Applicable. This is a physical medical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established:

  • Not Applicable. As above, no training set for an AI model. For the device itself, the "ground truth" for its design and manufacturing is established through engineering specifications, material standards, and compliance with quality management systems.

{0}------------------------------------------------

August 4, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Healthium Medtech Limited Pankaj Dawar, PhD Deputy General Manager Regulatory Affair 472-D, 13th Cross, 4th Phase, Peenya Industrial Area Bangalore, Karnataka 560058 India

Re: K231404

Trade/Device Name: T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button, T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 15, 2023 Received: May 15, 2023

Dear Dr. Dawar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jesse Muir -S

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button &

T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button

Indications for Use (Describe)

T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button & T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button are to be used for fixation of bone or soft tissue to bone, and are intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. This device is employed in Knee ligament repair and reconstructive procedures, specifically for ACL/PCL repair and reconstruction.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in bold, black letters, with the "O" in "SIRONIX" being a red circle. Below "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, black font.

K231404

Image /page/3/Picture/3 description: The image contains the word "Healthium" in a bold, blue font. To the right of the word is a logo that consists of a blue circle with a white cross inside. The top right corner of the logo has an orange section.

T-Button® Adjustable Loop UHMWPE Suture PEEK Button

1. 510(k) Summary

Submitter Information: 1.1.

Application Correspondent:PANKAJ DAWARHealthium Medtech Limited472-D, 13th Cross, 4th Phase,Peenya Industrial Area,Bangalore Karnataka 560058, India
Phone:+91-9886529934
E-mail:pankaj.d@healthiummedtech.com
Specification Developer:Healthium Medtech Limited472-D, 13th Cross, 4th Phase,Peenya Industrial Area,Bangalore, Karnataka 560058, India
Phone:+91 - 80 - 41868000
Contact Person:PANKAJ DAWAR
E-mail:pankaj.d@healthiummedtech.com
Date Prepared:01-08-2023
1.2. Device Identification:
Device Trade Name:T-Button® A Adjustable Loop UHMWPE Suture PEEKButton, Close ButtonT-Button® S Adjustable Loop UHMWPE Suture PEEKButton, Open Button
Device Common Name:Non-Absorbable Cortical Fixation
Classification Name:Smooth or threaded metallic bone fixation fastener
Device Class:Class II
Regulation Number:21 CFR 888.3040

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, sans-serif font.

Image /page/4/Picture/2 description: The image shows the Healthium logo. The logo consists of the word "Healthium" in blue, with two overlapping circles to the right. The top circle is orange, and the bottom circle is blue. A white cross is formed where the circles overlap.

1.3. Predicate Devices:

Device NamePredicate510(k) Number
Arthrex ACL TightRope®, PCLTightRope®, and TightRope® IIPrimary PredicateK221128*
ANCHORMAN Tibial LigamentFixation DeviceReference PredicateK170388

Table 1 - List of Predicate Devices

* In K221128, Arthrex ACL TightRope® has been selected as the Primary Predicate.

1.4. Device Description

The T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button & T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button is a permanently implantable Polyether Ether Ketone (PEEK) cortical tibial suspension fixation device which provides the orthopaedic surgeon a means of accurate fixation in ligament reconstructive surgeries and repair. The fixation device allows for endoscopic or open ligament reconstruction approaches.

The Closed button is preloaded with UHMWPE adjustable loop and in Open Button the UHMWPE adjustable loop is non preloaded.

1.5. Intended Use & Indications for Use

The T-Button® A Adjustable Loop UHMWPE Suture PEEK Button, Close Button & T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button are to be used for fixation of bone to bone or soft tissue to bone, and are intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. This device is employed in Knee ligament repair and reconstructive procedures, specifically for ACL/PCL repair and reconstruction.

1.6. Comparison of Technological Characteristics

The fundamental technology characteristics such as Design and Materials of construction are compared between the T-Button® A Adjustable Loop UHMWPE Suture PEEK Button. Close Button & T-Button® S Adjustable Loop UHMWPE Suture PEEK Button, Open Button with the primary predicate device Arthrex ACL Tightrope.

The Reference Predicate device (K170388) "ANCHORMAN Tibial Ligament Fixation Device" has the same material of construction as that of the subject devices i.e. PEEK

Table 2 summarizes the comparison of technological characteristics between the subject and predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Sironix. The word "SIRONIX" is written in a sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller font.

T-Button® Adjustable Loop UHMWPE Suture PEEK Button

Image /page/5/Picture/3 description: The image contains the logo for Healthium. The logo consists of the word "Healthium" in a blue sans-serif font. To the right of the word is a graphic consisting of two overlapping circles, one orange and one blue. The blue circle is positioned slightly below and to the left of the orange circle.

Table 2 – Substantial Equivalence Table

S. NoParametersArthrex ACL TightRope®(K221128)(Primary Predicate)T-Button® A Adjustable Loop UHMWPESuture PEEK Button, Close Button&T-Button® S Adjustable Loop UHMWPESuture PEEK Button, Open Button(Subject Device)Comments
1.ManufacturerArthrex Inc.Healthium Medtech Limited-
2.Product CodeMAI (Classification Product Code)MBI (Subsequent Product Code)MBIAs the subject device isnondegradable so onlyMBI code is applicable.
3.Regulation Number21 CFR 888.303021 CFR 888.304021 CFR 888.3040Similar to primarypredicate device accordingto MBI code.
4.ClassificationClass IIClass IISimilar to predicate device
5.IntendedUse/Indications forUseThe Arthrex ACL TightRope ®,device is intended to be used forfixation of bone to bone or softtissue to bone, and are intended asfixation posts, a distribution bridge,or for distributing suture tensionover areas of ligament or tendonrepair. Specifically, Arthrex will beoffering these devices for ACL/PCLrepair andThe Adjustable Loop UHMWPE Suture PEEKButton (Close Button & Open Button) are to beused for fixation of bone to bone or soft tissue tobone, and are intended as fixation posts, adistribution bridge, or for distributing suturetension over areas of ligament or tendon repair.This device is employed in Knee ligament repairand reconstructive procedures, specifically forACL/PCL repair and reconstruction.Similar to primarypredicate device excludingthe pediatric patientindications
Healthium Medtech Limited
Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, Indiawww.healthiummedtech.com CIN: U03311KA1992PLC013831Page 3 of 10

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" represented by a red circle. Below the company name, the words "Arthroscopy Solutions" are written in a smaller, lighter font.

T-Button® Adjustable Loop UHMWPE Suture PEEK Button

S. NoParametersArthrex ACL TightRope®(K221128)(Primary Predicate)T-Button® A Adjustable Loop UHMWPESuture PEEK Button, Close Button&T-Button® S Adjustable Loop UHMWPESuture PEEK Button, Open Button(Subject Device)Comments
reconstruction for the adult andpediatric patient population.
6.Button MaterialTitaniumPEEKSE Analysis 1
7.Loop MaterialUHMWPEClosed Button Loop: UHMWPEOpen Button Loop: Non-absorbable, UHMWPE(UHMWPE 100.0%) +Suture Button: (Non-absorbable, UHMWPE(UHMWPE ≥ 88.0%) +Additive: Loctite 4014 (<12.0% in the Suture Button))SE Analysis 2

He

Healthium Medtech Limited
Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, Indiawww.healthiummedtech.com CIN: U03311KA1992PLC013831Page 4 of 10

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Sironix. The word "SIRONIX" is written in a sans-serif font, with the "O" replaced by a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller font.

T-Button® Adjustable Loop UHMWPE Suture PEEK Button

Image /page/7/Picture/3 description: The image contains the logo for Healthium. The logo consists of the word "Healthium" in a dark blue, sans-serif font. To the right of the word is a graphic consisting of two overlapping circles, one orange and one blue, with a white plus sign where they intersect.

S. NoParametersArthrex ACL TightRope®(K221128)(Primary Predicate)T-Button® A Adjustable Loop UHMWPESuture PEEK Button, Close Button&T-Button® S Adjustable Loop UHMWPESuture PEEK Button, Open Button(Subject Device)Comments
8.Design FeaturesArthrex ACL TightRope®:Image: Arthrex ACL TightRopeT-Button®A - Closed button preloaded withUHMWPE adjustable loopImage: Closed ButtonImage: Closed button preloaded with UHMWPE adjustable loopSE Analysis 3
Healthium Medtech LimitedPage 5 of 10
Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, Indiawww.healthiummedtech.com CIN: U03311KA1992PLC013831

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Sironix. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" being a red circle. Below the company name, the words "Arthroscopy Solutions" are written in a smaller, sans-serif font.

T-Button® Adjustable Loop UHMWPE Suture PEEK Button

S. NoParametersArthrex ACL TightRope®(K221128)(Primary Predicate)T-Button® A Adjustable Loop UHMWPESuture PEEK Button, Close Button&T-Button® S Adjustable Loop UHMWPESuture PEEK Button, Open Button(Subject Device)Comments
T-Button® S - Open button non preloaded withUHMWPE adjustable loopImage: Open ButtonImage: Suture Button and UHMWPE Adjustable Loop

He

Healthium Medtech Limited
Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, IndiaPage 6 of 10
www.healthiummedtech.com CIN : U03311KA1992PLC013831

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" represented by a red circle. Below the company name, the words "Arthroscopy Solutions" are written in a smaller, lighter font.

T-Button® Adjustable Loop UHMWPE Suture PEEK Button

S. NoParametersArthrex ACL TightRope®(K221128)(Primary Predicate)T-Button® A Adjustable Loop UHMWPESuture PEEK Button, Close Button&T-Button® S Adjustable Loop UHMWPESuture PEEK Button, Open Button(Subject Device)Comments
9.Specifications andDimensions• ACL TightRope® RT, double-loaded passing suturesButton - 13mm x 3.3mm x1.5mm(Length x width x Height)Loop Length: 90 mm• T-button A - Closed button:Button Dimension: 16mm x 7.5mm (Diameterx Height)Loop Length: 90 mm• T-button S - Open button:Button Dimension: 16mm x 7.5mm(Diameter x Height)Loop Length: 90 mmSE Analysis 4
10.Single Use/ReuseSingle useSingle useSimilar as predicate device
11.Shelf Life5 Years5 YearsSimilar as predicate device
12.SterilizationETOETOSimilar as predicate device
13.Performance Data• Ultimate load testing• Cyclic displacement• Pull-Out Testing• Cyclic LoadingSimilar as predicate device
14Safety Data• Bacterial Endotoxin• Skin Sensitization,• Intracutaneous reactivity• Material Mediated Pyrogenicity• Acute Systemic Toxicity• In vitro cytotoxicity• Bone Implantation studies• Intramuscular Implantation Study• Bacterial Endotoxin-

Healthium Medtech Limited Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India Page 7 of 10 www.healthiummedtech.com | CIN : U03311KA1992PLC013831

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Sironix. The word "SIRONIX" is written in a sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, sans-serif font.

Image /page/10/Picture/2 description: The image shows the logo for Healthium. The logo consists of the word "Healthium" in a blue, sans-serif font. To the right of the word is a graphic of two overlapping circles, one orange and one blue. The circles are divided into quadrants by a white cross.

Substantial Analysis

SE Analysis 1 Button Material

The primary predicate Arthrex ACL TightRope® button is made of Titanium whereas the reference predicate ANCHORMAN Tibial Ligament Fixation Device material of construction is PEEK. The subject device material of construct for button is PEEK (Polyether Ether Ketone) which is equivalent to the reference predicate.

The reference predicate has already proven safety and efficacy of PEEK material in reference to the titanium based button in their K170388 submission.

PEEK is chemically inert and insoluble, has a modulus of elasticity closer to human cortical bone. Hence, PEEK represents a stable and biocompatible material that may address the issues that are present with titanium screws, such as graft damage because of material hardness and interference with imaging. PEEK has poor osseointegration. PEEK has excellent bio-compatibility and biological stability, and is a very favorable implant material.

Also, the bio-compatibility testing of the subject devices has been performed to substantiate the biocompatible nature of the subject device. The summary of the biocompatibility testing performed has been shown in the below table.

Biocompatibility TestResult
Skin Sensitization (Guinea PigMaximization Test)Non-Sensitizer
Intracutaneous Reactivity TestNon-Reactive
Material MediatedPyrogenicityNon-Pyrogenic
Acute Systemic Toxicity TestDid not Induce any SystemicToxicity
In vitro Cytotoxicity Test:Elution MethodNon-Cytotoxic
Bone Implantation Studies(PEEK Rods)Minimal or No Reaction
Bone Implantation Studies(UHMWPE Suture)Minimal or No Reaction
Intra muscular Implant(PEEK)Minimal or No Reaction

In reference to the above mentioned evidences the difference in the material of construction between the primary predicate and the subject device do not raise any new questions of safety and effectiveness.

Healthium Medtech Limited
Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India
www.healthiummedtech.com CIN : U03311KA1992PLC013831
Page 8 of 10

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Sironix. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the text "ArthroscopySolutions" in a smaller, sans-serif font.

Image /page/11/Picture/3 description: The image shows the logo for Healthium. The word "Healthium" is written in a blue sans-serif font on the left side of the image. To the right of the word, there is a graphic consisting of two overlapping circles, one blue and one orange, with a white plus sign in the intersection.

SE Analysis 2 Loop Material

The adjustable loop material construct of the primary predicate device (Arthrex ACL TightRope®) and the adjustable loop of the T-button®A - Closed button (subject device) is UHMWPE; which are similar.

However, the adjustable loop of the T-button® S - Open button (Subject Device) has Loop and Suture Button. The material of construct of loop is UHMWPE (UHMWPE 100.0%) and Suture Button has a material of construct of Non-absorbable, UHMWPE (UHMWPE ≥ 88.0%) + Additive: Loctite 4014 (≤12.0% in the Suture Button)).

The suture button in the adjustable loop tends to hold the loop, the pull out strength of the suture button in the bench testing has been performed and evaluated, the results are perceptible to define the strength of suture button to hold the loop.

The above mentioned addition of Suture Button in the adjustable loop of the T-button® S - Open button (Subject Device) has no impact on the safety and efficacy of the product.

SE Analysis 3 Design Feature

The Arthrex ACL TightRope® and the ANCHORMAN Tibial Ligament Fixation Device offers adjustable cortical fixation for cruciate ligament reconstruction. The ACL TightRope® RT comes with the regular button (13mm (L) x 3.3mm (W) x1.5mm (H)) preloaded with adjustable loop and double loaded passing sutures whereas the subject device comes with Open and Closed Button having dimensions (16mm(dia)x 7.5mm (H)). The Closed button is preloaded with UHMWPE adjustable loop and in Open Button the adjustable loop is non preloaded.

The Pull Out Test and the Cyclical Load tests were performed for the primary predicate, reference predicate and the subject device and the results were compared and found equipotent in terms of safety and efficacy irrespective to the material of construction and design.

SE Analysis 4 Specifications and Dimensions

During the bench testing of subject device, it was observed that the changes in dimensions are not leading to any kind of new risks with the performance of subject device, and the acceptance criteria was met in each of the performance tests. By considering the performance data of the subject device, it is evident that the changes in dimensions are not causing and or does not affect the safety and efficacy of the product when comparing it with Predicate device.

1.7. Summary of Performance Data

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, sans-serif font.

Image /page/12/Picture/2 description: The image contains the word "Healthium" in a bold, blue font. To the right of the word is a logo consisting of two overlapping circles. The top circle is orange, and the bottom circle is blue. Each circle has a white plus sign in the center.

The T-Button® Adjustable Loop UHMWPE Suture PEEK Button, Primary Predicate device and Reference Predicate underwent tests to determine their pull-out strength, cyclical loading. Results demonstrated that the T-Button® Adjustable Loop UHMWPE Suture PEEK Button performs statistically equivalent to the predicate device and in addition to mechanical performance testing, biocompatibility testing, sterilization validation, shelf-life, pyrogenicity testing and endotoxin monitoring were also performed to support the demonstration of substantial equivalence.

1.8. Clinical Testing

Not Applicable

1.9. Conclusion

There are no significant differences between the subject device and the predicate devices that would adversely affect the use of the product. The subject device is substantially equivalent to the predicate devices Arthrex ACL TightRope® in Intended use, Indications for use, function, Sterilization method, Shelf Life, and operational principles. From the data available we can justify that the T-Button®Adjustable Loop UHMWPE Suture PEEK Button is as safe, and as effective and performs the same indications for use as that of already marketed predicate device identified.

N/A