K Number
K231353

Validate with FDA (Live)

Device Name
AccuCTP Pro
Date Cleared
2023-09-14

(127 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AccuCTP Pro is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices.

AccuCTP Pro provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CTP), which can visualize and analyze dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

Device Description

AccuCTP Pro is an extension to legally cleared device AccuCTP (K220663).

AccuCTP Pro is a standalone software package that provides visualization and study of changes of tissue perfusion in digital images captured by CT (Computed Tomography). The software provides viewing, quantification, analysis and reporting capabilities, and it allows repeated use and continuous processing of data and can be deployed on a supportive customer's PC or a virtual platform that meets the minimum system requirements.

AccuCTP Pro works with the DICOM compliant medical image data. AccuCTP Pro provides tools for performing the following types of analysis:

  • Volumetry of threshold maps
  • Time intensity plots for dynamic time courses
  • Measurement of mismatch between rCBF and Tmax threshold volumes obtained from the same scan.
AI/ML Overview

The provided text describes the AccuCTP Pro device, which is an extension of a previously cleared device, AccuCTP. The submission is a "Special 510(k) Summary," indicating that it relies on substantial equivalence to a predicate device and focuses on minor changes.

Based on the provided information, the document does not contain details about a clinical study with acceptance criteria and a human reader performance study (MRMC) to prove the device meets specific performance metrics. Instead, it states that the algorithm and output settings of the AccuCTP Pro are the same as the predicate device (AccuCTP). Therefore, it relies on the previous predicate device's validation and general software verification and validation tests for the updated features.

Here's an analysis of the provided text with respect to your requested information:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of acceptance criteria for clinical performance and reported device performance metrics in the way one might expect for a new or significantly modified device requiring de novo validation. It primarily focuses on software verification and validation for the changes made.

The closest statement regarding "acceptance criteria" for performance is: "All tests met the pre-defined acceptance criteria and were passed." However, these refer to software verification tests, not clinical performance metrics.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Test Set: No specific sample size for a clinical test set is mentioned for AccuCTP Pro's validation. The document explicitly states: "No additional pre-clinical or clinical data is being provided with this submission." It defers to the predicate device's validation.
  • Data Provenance: Not applicable as no new clinical data is presented.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Not applicable, as no new clinical study data is presented requiring expert ground truth establishment for a test set.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable, as no new clinical study data is presented requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such MRMC study was performed or presented in this submission for AccuCTP Pro. The document focuses on software updates and relies on the predicate device's previous validation.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is implicitly addressed by the statement: "Phantom test and validation study were completed and reviewed as part of the predicate review (K220663), and the results concluded AccuCTP was safe and effective. As all algorithm and output of the subject device AccuCTP Pro is same as the predicate device AccuCTP, it can be concluded that AccuCTP Pro is acceptable for use." This suggests that the predicate device had a standalone evaluation, and since the algorithm hasn't changed, the standalone performance is assumed to be the same.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the predicate device's "phantom test and validation study," the ground truth would likely be established through physical phantom measurements or a gold standard comparison. The document does not specify the type of ground truth for the predicate.

8. The sample size for the training set

Not mentioned. The document describes the device as an "extension" with the "same fundamental scientific technology" and "same operation principle and algorithm embedded in the software" as the predicate. This suggests the training would have occurred for the original AccuCTP, and no new training data is discussed for AccuCTP Pro given that the algorithm is the same.

9. How the ground truth for the training set was established

Not mentioned, for the same reasons as in point 8.


Summary of what the document does provide regarding device acceptance and performance:

The acceptance of AccuCTP Pro is based on its substantial equivalence to the predicate device AccuCTP (K220663). The core argument is:

  • The intended use, intended user, intended patient population, fundamental scientific technology, operation principle, and algorithm of AccuCTP Pro are the same as the predicate device.
  • The changes in AccuCTP Pro are primarily related to data transmission mode, simplified manual operation, and updated software installation environments (e.g., compatibility with Linux and Windows, and virtual platforms like VMware, and basic PACS functions).
  • These minor differences were evaluated through software verification and validation tests (import/export of DICOM, automatic selection/calculation, software management, case management, recalculation/adjustment, operating environment tests).
  • Cybersecurity testing was performed according to FDA guidance.
  • Human factors testing was conducted with fifteen qualified participants to validate usability and the user manual, concluding safe and effective use without residual use-related risks.
  • The manufacturer relies on the original phantom test and validation study conducted for the predicate device (AccuCTP), arguing that since the algorithm and output are identical, the conclusions of safety and effectiveness for the predicate device extend to AccuCTP Pro.

In essence, AccuCTP Pro's "acceptance" is not demonstrated by new clinical performance data but by proving that its software updates do not introduce new questions of safety or effectiveness, due to the unchanged core algorithm and the successful completion of software engineering, cybersecurity, and usability tests.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

ArteryFlow Technology Co., Ltd. % Ashley Fu RA Specialist 459 Qianmo Road, Suite C1-501, Binjiang District Hangzhou, Zhejiang 310051 CHINA

September 14, 2023

Re: K231353

Trade/Device Name: AccuCTP Pro Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: August 31, 2023 Received: August 31, 2023

Dear Ashley Fu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231353

Device Name AccuCTP Pro

Indications for Use (Describe)

AccuCTP Pro is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of brain images are acquired through DICOM compliant imaging devices.

AccuCTP Pro provides both viewing and analysis capabilities for functional and dynamic imaging datasets acqured with CT Perfusion (CTP), which can visualize and analyze dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue blood volume.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Special 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1. Submitter's Information

Submitter: ArteryFlow Technology Co., Ltd.

Address: 459 Qianmo Road, Suite C1-501, Binjiang District, 310051 Hangzhou City, Zhejiang Province, China

Phone Number: +86-571-86772567

Primary correspondent: XIANG Jianping, PhD, General Manager

Email: jianping.xiang@arteryflow.com

Secondary correspondent: Ashley Fu, RA Specialist

Email: fang.fu@arteryflow.com

Date of preparation: August 31, 2023

2. Device Information

Trade/ Device Name: AccuCTP Pro

Common Name: Radiological Image Processing Software

Regulatory Class: Class II

Regulation Name: Medical image management and processing system

Regulation Number: 21 CFR 892.2050

Classification Product Code: LLZ

510(k) number: K231353

3. Predicate Device Information

Manufacturer: ArteryFlow Technology Co., Ltd.

Device Name: AccuCTP

Common Name: Radiological Image Processing Software

Regulatory Class: Class II

Regulation Number: 21 CFR 892.2050

Classification Product Code: LLZ

510(k) number: K220663

{4}------------------------------------------------

4. Device Description

AccuCTP Pro is an extension to legally cleared device AccuCTP (K220663).

AccuCTP Pro is a standalone software package that provides visualization and study of changes of tissue perfusion in digital images captured by CT (Computed Tomography). The software provides viewing, quantification, analysis and reporting capabilities, and it allows repeated use and continuous processing of data and can be deployed on a supportive customer's PC or a virtual platform that meets the minimum system requirements.

AccuCTP Pro works with the DICOM compliant medical image data. AccuCTP Pro provides tools for performing the following types of analysis:

  • Volumetry of threshold maps
  • Time intensity plots for dynamic time courses
  • Measurement of mismatch between rCBF and Tmax threshold volumes obtained from the same scan.

5. Indications for Use

AccuCTP Pro is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices.

AccuCTP Pro provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CTP), which can visualize and analyze dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

6. Technological Characteristic Comparison

AccuCTP Pro is the first update of AccuCTP. Compared to the predicate device AccuCTP (K220663), the intended use of the subject device is equivalent to the intended use of the previously cleared device. The intended user and intended patient population of AccuCTP Pro are not changed. Most important, AccuCTP Pro has the same fundamental scientific technology as the predicate device. The subject and predicate device are based on the same operation principle and algorithm embedded in the software.

Compared to AccuCTP. AccuCTP Pro mainly updated the data transmission mode, simplified manual operation, updated software installation environment (operating systems and virtual platforms).

The technological characteristic comparisons are listed in the following table.

ItemProposed devicePredicate device
AccuCTP ProAccuCTP
Product CodeLLZLLZ
Regulation No.892.2050
Indications for useAccuCTP Pro is an image processingsoftware package to be used bytrained professionals, including butnot limited to physicians and medicaltechnicians. The software runs on astandard off-the-shelf computer or avirtual platform, such as VMware, andcan be used to perform imageviewing, processing and analysis ofbrain images. Data and images areacquired through DICOM compliantimaging devices.AccuCTP Pro provides both viewingand analysis capabilities for functionaland dynamic imaging datasetsacquired with CT Perfusion (CTP),which can visualize and analyzedynamic imaging data, showingproperties of changes in contrast overtime. This functionality includescalculation of parameters related totissue flow (perfusion) and tissueblood volume.AccuCTP is an image processingsoftware package to be used by trainedprofessionals, including but not limitedto physicians and medical technicians.The software runs on a standardoff-the-shelf computer, and can beused to perform image viewing,processing and analysis of brainimages. Data and images are acquiredthrough DICOM compliant imagingdevices.AccuCTP provides both viewing andanalysis capabilities for functional anddynamic imaging datasets acquiredwith CT Perfusion (CTP), which canvisualize and analyze dynamic imagingdata, showing properties of changes incontrast over time. This functionalityincludes calculation of parametersrelated to tissue flow (perfusion) andtissue blood volume.
Operating SystemCompatibilityLinux operating system and Windowsoperating systemLinux operating system
Basic PACSFunctionView, process and analyze medicalimages. Perform standards PACSfunctions with respects to queryingand listing.Not supported
ComputerPlatform●Standard off-the-shelf PCworkstation●Virtual platform, such asVMware●Standard off-the-shelf PCworkstation
FunctionalOverviewAccuCTP Pro is a software packagethat provides for the visualization andstudy of changes of tissue perfusionin digital images captured by CT.AccuCTP Pro provides viewing andquantification.AccuCTP is a software package thatprovides for the visualization and studyof changes of tissue perfusion in digitalimages captured by CT. AccuCTPprovides viewing and quantification.
Data AcquisitionAcquires medical image data fromDICOM compliant imaging devicesand modalitiesAcquires medical image data fromDICOM compliant imaging devices andmodalities
Data/ImageTypesDICOM FormatDICOM Format
Acquisition andModalitiesFeaturesCT (CT perfusion)CT (CT Perfusion)
ComputedParameter Mapsof Perfusion CTCerebral blood flow (CBF) Cerebral blood volume (CBV) Mean transit time (MTT) Tissue residue function time to peak (Tmax)Cerebral blood flow (CBF) Cerebral blood volume (CBV) Mean transit time (MTT) Tissue residue function time to peak (Tmax)
CT ToolsArterial input function (AIF)/Venousoutput function (VOF)Arterial input function (AIF)/Venousoutput function (VOF)
Time-courseTime-course
Brain maskBrain mask
Region of interest (ROI) andVolumetryRegion of interest (ROI) and Volumetry
Motion correctionMotion correction
Export perfusion files to local addressExport perfusion files to local address
Acquire, transmit, process, and storemedical imagesAcquire, transmit, process, and storemedical images

{5}------------------------------------------------

{6}------------------------------------------------

The outcome of this technological characteristics comparison and risk assessment demonstrates that the minor differences in the technological characteristics do not affect the intended use and the safety or effectiveness of the subject device, AccuCTP Pro, when compared to the legally marketed predicate device AccuCTP (K220663), thus demonstrating the substantial equivalence of the subject device with the predicate device.

7. Performance Data to Support Substantial Equivalence

AccuCTP Pro complies with DICOM (Digital Imaging and Communications in Medicine) -Developed by the American college of Radiology and the National Electrical Manufactures Association. NEMA PS 3.1-3.20 (2022d)

The following performance tests were performed on AccuCTP Pro in support of the substantial equivalence determination.

The software modifications to the AccuCTP Pro were implemented under the design controls that are compliant with 21 CFR 820.30. A risk analysis was conducted in accordance with "ISO 14971:2019 Medical devices – Application of risk management to medical devices" to assess the risks and risk mitigations for the device modifications. Based on this risk assessment, the following verification and validation tests were identified and conducted. All tests met the pre-defined acceptance criteria and were passed.

{7}------------------------------------------------

Software Verification Testing: Software requirements derived from the intended use as well as risk control measures are verified by system testing. Requirement coverage matrix of AccuCTP Pro provides traceability between software requirements, software design and the software tests. All requirements are tested, and all results of the tests performed are recorded. The main software verification tests conducted are as follows:

  • Tests of importation and exportation of DICOM data
  • Tests of automatic selection and calculation of medical image data
  • Software management tests
  • Case management tests
  • Tests of image recalculation and manual adjustment
  • Tests of software operating environment

The test results demonstrate that the software is designed to meet the software requirements, and they also demonstrate the performance and safety of the subject device is acceptable.

Cybersecurity Testing: Cybersecurity controls are implemented in accordance with the security control categories and recommendations in guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Device". Cybersecurity testing including requirement verification testing, static code analysis, and malformed input testing, have been carried out to demonstrate the cybersecurity of our software.

Human Factors Testing: Usability testing of the modified AccuCTP Pro and its operator manual was conducted in accordance with IEC 62366-1:2015, "Application of usability engineering to medical devices". Fifteen qualified participants performed all critical tasks necessary to validate the usability and user manual without any use errors. The conclusion of the testing was that the AccuCTP Pro can be used safely and effectively by the intended user population. No residual use-related risks were identified.

Labeling Inspection: Labeling inspection was conducted to in accordance with documentation of the company's quality management system. All inspections were passed.

Phantom test and validation study were completed and reviewed as part of the predicate review (K220663), and the results concluded AccuCTP was safe and effective. As all algorithm and output of the subject device AccuCTP Pro is same as the predicate device AccuCTP, it can be concluded that AccuCTP Pro is acceptable for use. The phantom test and validation study conclusions are not affected by the changes proposed under this 510(k). No additional pre-clinical or clinical data is being provided with this submission.

8. Conclusion

AccuCTP Pro has same intended use and algorithm as the predicate device, and it is as safe and effective as AccuCTP (K220663). The verification and validation testing presented demonstrates that the differences between the subject and predicate device do not raise new questions of safety and effectiveness. Thus, the AccuCTP Pro is substantially equivalent.

N/A