K Number
K231318

Validate with FDA (Live)

Date Cleared
2023-07-07

(60 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated to be used for:

  • · Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
  • Strengthening, Toning and Firming of buttocks and thighs.
Device Description

The Electromagnetic stimulator device consists of a host, hand tools and power cord. The host contains a power supply unit, a control unit, and a cooling unit. The control unit includes a control element and a liquid crystal display. The hand tools include electromagnetic induction coils and cooling fans, it is a non-invasive therapeutic device.
The subject device has two models, Tesla Duet and EM contouring. The two models are exactly the same except for the color of the host shell, one is black and other is white.
The device produces electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the Electromagnetic stimulator device helps to strengthen, tone and firm the abdomen, buttocks and thighs. The device two outputs enable simultaneous treatment by two applicators.
The Electromagnetic stimulator device is equipped with a large color touch screen with wide view angle that significantly facilitates the use of the device. The on-screen information guides the user step by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen and buttons on the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

AI/ML Overview

The provided document is a 510(k) summary for a medical device (Electromagnetic Stimulator Device). It details the device's characteristics and its equivalence to a predicate device. However, it explicitly states "No clinical study is included in this submission" under section 8.0 Clinical Test Conclusion and that "Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness." Therefore, the document does not contain the information required to answer your request regarding acceptance criteria and a study proving the device meets those criteria.

The submission focuses primarily on demonstrating substantial equivalence to a predicate device through non-clinical testing and comparison of technological characteristics, rather than reporting on a standalone clinical study with acceptance criteria.

Therefore, I cannot extract the requested information as it is not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 7, 2023

Beijing ADSS Development Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China

Re: K231318

Trade/Device Name: Electromagnetic Stimulator Device (Models: EM Contouring and Tesla Duet) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: May 8, 2023 Received: May 8, 2023

Dear Mr. Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorywebsite the DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

TusharBansal -SDigitally signed byTushar Bansal -SDate: 2023.07.0714:27:15-04'00'
-----------------------------------------------------------------------------------------------------

for Heather Dean, Ph.D. Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231318

Device Name

Electromagnetic Stimulator Device (Model: EM Contouring, Tesla Duet)

Indications for Use (Describe)

The device is indicated to be used for:

  • · Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
    · Strengthening, Toning and Firming of buttocks and thighs.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 submitter's Information

Name:Beijing ADSS Development Co., Ltd.
Applicant Address:Room 609, F6, Building 13, Yard 5 Tianhua Street,Daxing District, Beijing, 102600, P. R. China
Production Address:Fuda Road - Tongsheng Road, Southern Area of IndustrialPark,Gu'an County, 065599 Langfang City, Hebei Province,P.R. China
Tel:86-13051615111
Contact:Song Ying

Designated Submission Correspondent

Contact:Mr. Boyle Wang
Name:Shanghai Truthful Information Technology Co., Ltd.
Address:Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai,200120 China
Tel:+86-21-50313932
Email:Info@truthful.com.cn

Date of Preparation: May.5th,2023

2.0 Device Information

Trade name:Electromagnetic Stimulator Device
Common name:Powered muscle stimulator
Classification name:Stimulator, Muscle, Powered, For Muscle Conditioning
Model(s):EM Contouring, Tesla Duet
Production code:NGX
Regulation number:21CFR 890.5850
Classification:Class II
Panel:Physical Medicine

3.0 Predicate and Reference Devices

Predicate Device:

Manufacturer: BTL Industries, Inc.

{4}------------------------------------------------

Device:BTL 799-2
510(k) number:K180813

Reference Device:

Manufacturer:BTL Industries, Inc.
Device:BTL 799-2L
510(k) number:K190456

4.0 Device Description

The Electromagnetic stimulator device consists of a host, hand tools and power cord. The host contains a power supply unit, a control unit, and a cooling unit. The control unit includes a control element and a liquid crystal display. The hand tools include electromagnetic induction coils and cooling fans, it is a non-invasive therapeutic device.

The subject device has two models, Tesla Duet and EM contouring. The two models are exactly the same except for the color of the host shell, one is black and other is white.

The device produces electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the Electromagnetic stimulator device helps to strengthen, tone and firm the abdomen, buttocks and thighs. The device two outputs enable simultaneous treatment by two applicators.

The Electromagnetic stimulator device is equipped with a large color touch screen with wide view angle that significantly facilitates the use of the device. The on-screen information guides the user step by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen and buttons on the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

5.0 Indication for Use Statement

The device is indicated to be used for:

  • Improvement of abdominal tone, strengthening of the abdominal muscles, . development of firmer abdomen.
  • Strengthening, Toning and Firming of buttocks and thighs.

6.0 Comparison to the Predicate Device

Table 6.1: Technological Characteristics between Subject and Predicate Device

{5}------------------------------------------------

ItemSubject DevicePredicate DeviceReference DeviceComparison
ManufacturerBeijing ADSS DevelopmentCo., Ltd.BTL Industries, Inc.BTL Industries, Inc.——
510(k) No.K231318K180813K190456——
Trade nameElectromagnetic StimulatorDeviceBTL 799-2BTL 799-2L——
ModelEM Contouring, Tesla DuetBTL 799-2BTL 799-2L——
Regulationnumber21 CFR 890.585021CFR 890.585021CFR 890.5850Same
Regulation NamePowered muscle stimulatorPowered muscle stimulatorPowered muscle stimulatorSame
Product codeNGXNGXNGXSame
ClassIIIIIISame
Indications foruse/Intended useThe device is indicated tobe used for:· Improvement ofabdominal tone,strengthening of theabdominal muscles,development of firmerabdomen.· Strengthening, Toningand Firming of buttocksBTL 799-2 is indicated to be used for:· Improvement of abdominal tone,strengthening of the abdominalmuscles, development of firmerabdomen.· Strengthening, Toning andFirming of buttocks and thighs.BTL 799-2L is indicated to beused for:· Improvement of abdominaltone, strengthening of theabdominal muscles,development of firmerabdomen.· Strengthening, Toning andFirming of buttocks, thighsand calves.Same
and thighs.Improvement of muscletone and firmness, forstrengthening muscles inarms.
Location for usePrescription UsePrescription UsePrescription UseSame
Primary FunctionMuscle stimulationMuscle stimulationMuscle stimulationSame
Principle of ActionInitiating action potential of nerves results in muscle contractionInitiating action potential of nerves results in muscle contractionInitiating action potential of nerves results in muscle contractionSame
Electrical ProtectionClass II, BFClass II, BFClass II, BFSame
User InterfaceTouch screenTouch screenTouch screenSame
Touch screen size15.6" (39.6 cm)15.6" (39.6 cm)15.6" (39.6 cm)Same
Firmware ControlledYesYesYesSame
Type of EnergyMagnetic fieldMagnetic fieldMagnetic fieldSame
Number of outputs222Same
Number of Magnetic Coils in the Applicator111Same
Number of Magnetic coils111Same
in the
Applicator
ApplicatorDimensions252.3167.396mm(±0.2mm)Not publicly availableNot publicly available
Surface AreaSurface Area of theTreatmentprobe:13471.4mm²Surface Area of theApplicator: 33257.7mm²Not publicly availableNot publicly available
Magnetic FieldIntensity (onthe coilsurface)0.5-1.8T,±20%299-6 applicator: 0.5-1.8 TBTL 299-6 applicator:0.5 - 1.8 T , ±20%BTL 299-7 applicator:0.7 - 2.0 T , ±20%Same
MaximumMagnetic FieldIntensity atApplicatorCenterSurface1.0T±20%BTL 299-6 applicator:1.154 T±20%;BTL 299-6 applicator:1.154 T±20%;BTL 299-7 applicator:1.173 T±20%.The maximum field intensity atApplicatorCenter Surface is 1.0T ascompared to the 1.154T of thepredicate device. However, themagnetic field intensity is within therange of the predicate device.
Induced Currentin the Tissue20.6mA28-30 mANot publicly availableThe induced current on the tissuefor the subject device is less thanthe predicate device and thereforesafer with the same effect.
Type ofOperationContinuousContinuousContinuousSame
Pulse RepetitionRate1 – 150 Hz1 – 150 Hz1 – 150 HzSame
Pulse Duration270 ±20% μs280 ± 20% μsBTL 299-6 applicator:280 ± 20% μs;BTL 299-7 applicator:190 ± 20% μsThe subject device has little shorter pulse width when compared to the predicate device but in the range between the predicate and reference device.
Pulse Amplitude0 – 100%, step 1%0 – 100%Not publicly availableSame
Selection ofparameters(Intensity, Time)YesYesYesSame
Shape ofStimulationPulseSymmetrical Biphasic SineWaveSymmetrical Biphasic SineWaveSymmetrical Biphasic SineWaveSame
Therapy Time1-60minUp to 60 minUp to 60 minSame
ApplicationHands-free, applicator fixed by fixation beltHands-free, applicator fixed by fixation beltHands-free, applicator fixed by fixation beltSame
Energy Source100–240 VAC, 50–60 Hz100 - 240 V AC, 50-60 Hz100 - 240 V AC, 50-60 HzSame
SystemDimensions(W×H×D)525×433×1279mm/640×560×1390mm500×1380×580 mm(20×55×23 in)580×1380×580 mm(23×55×23 in)Different dimensions have no influence on the safety or effectiveness of the device.
AmbientTemperature-20°C~+55°C-10°C to +55°C-10°C to +55°CThis difference has no influence on the safety or effectiveness of the device.
EnvironmentalSpecificationsFor indoor use onlyFor indoor use onlyFor indoor use onlySame
ConformanceStandardIEC 60601-1;IEC 60601-1-2;IEC 60601-1-6;IEC 60601-1-9;IEC 60601-2-10IEC 60601-1;IEC 60601-1-2;IEC 60601-1-6;IEC 60601-2-10IEC 60601-1;IEC 60601-1-2;IEC 60601-1-6;IEC 60601-2-10Same
Patient contactmaterialHand Tool: ABS&PCNot publicly availableNot publicly availableBiocompatibility evaluation for bothsubjectdevicethetheandpredicate device. The differenceaffectdoesnotsafetyandeffectiveness.
BiocompatibilityISO 10993-5ISO 10993-10ISO 10993-5ISO 10993-10ISO 10993-5ISO 10993-10Same

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

7.0 Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the subject devices met all design specifications as were Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

● ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and essential performance

● IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

● IEC60601-2-10:2016, Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators

· IEC 60601-1-6: 2013, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

● IEC 60601-1-9:2007+A1:2013, Medical electrical equipment- Part 1-9: General requirements for basic safety and essential performance Collateral Standard: Requirements for environmentally conscious design

● ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity

● ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

Software Information:

Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"

8.0 Clinical Test Conclusion

No clinical study is included in this submission.

9.0 Conclusion

Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. And based on the performance testing and compliance with acceptable voluntary standards, we believe the subject device is substantially equivalent to its predicate device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).