K Number
K231310

Validate with FDA (Live)

Date Cleared
2023-10-27

(175 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. The intended wrist circumference is 13.5-21.5 cm.

Device Description

The Electronic Blood Pressure Monitor, is a battery driven automatic non-invasive blood pressure monitor, comprised of the host machine and the wrist cuff. It can measure systolic and diastolic blood pressure and pulse rate of the adult person at wrist via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. The device can store 199 groups of measurement data for two users.

AI/ML Overview

This document describes the regulatory approval of an Electronic Blood Pressure Monitor (Models HWA11, HWA10) by Shenzhen Jumper Medical Equipment Co., Ltd. The approval is based on demonstrating substantial equivalence to a predicate device (Omron Model BP6100, K182127).

The provided text does not contain detailed acceptance criteria or a comprehensive study plan with the specific information requested in the prompt (e.g., specific thresholds for accuracy, sample sizes for test sets, number of experts for ground truth, adjudication methods, MRMC study details, etc.).

However, based on the information provided, here's what can be extracted and inferred regarding acceptance criteria and the supporting study:


Implied Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily implied by conformance to international standards for blood pressure monitors and demonstration of statistical equivalence to a recognized measurement method.

Table of Acceptance Criteria and Reported Device Performance (Inferred from Text):

Acceptance Criterion (Implied)Reported Device Performance (HWA11 & HWA10)
Blood Pressure Accuracy: Conformance to ANSI/AAMI/ISO 81060-2:2013 statistical accuracy requirements for automated measurement type.The clinical investigation "demonstrated that HWA11 performed equivalently to the auscultation method and is in conformance with ANSI/AAMI/ISO 81060-2:2013." Blood Pressure: Within ± 3mmHg (This is likely the manufacturer's specified accuracy, which the clinical study aimed to validate adherence to the standard).
Pulse Rate Accuracy: Conformance to relevant standards.Pulse Rate: Within ± 5 % of reading (This is likely the manufacturer's specified accuracy, which the clinical study aimed to validate adherence to the standard).
Biocompatibility: Meet ISO 10993-1 requirements."Biocompatibility of patient-contacting materials per ISO 10993-1 requirements." (Tested and presumably met).
Electrical Safety, EMC, ESD: Meet relevant standards."Electrical safety, electromagnetic compatibility, and electrostatic discharge testing." (Tested and presumably met).
Software Verification & Validation: Meet V&V requirements."Software verification and validation." (Tested and presumably met).
Performance Verification: Acceptable performance of features."Performance verification testing to confirm acceptable performance of device features and functions" (e.g., measurement range, inflation/deflation methods). The device specifications included: Cuff pressure range 0 to 295mmHg, Pulse Rate: 40 to 199 beats/min.
Cleaning Verification: Retain performance after cleaning."Cleaning verification testing to confirm device retains its performance when cuff is cleaned with household detergents as may be required in home use environment." (Tested and presumably met).
Environmental Performance: Operation/Storage within specified conditionsOperating Conditions: 5 to 40 °C (41 to 104 °F), 15 to 85 %RH (non-condensing), 700 to 1060 hPa. Storage/Transport Conditions: -20 to 55 °C (-4 to 131 °F), 10 to 93 %RH (non-condensing), 700 to 1060 hPa. The text states these comply with IEC 60601-1, IEC 60601-1-11 and IEC 80601-2-30.

Study Information (Extracted and Inferred):

  1. Sample size used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated. The text mentions a "clinical investigation" but does not specify the number of subjects. However, for validation to ANSI/AAMI/ISO 81060-2:2013, a specific number of subjects (usually 85 for the initial phase, and more for specific populations) from a defined demographic (e.g., age and blood pressure ranges) are typically required.
    • Data Provenance: Not explicitly stated, but clinical studies for such devices are typically prospective. The text does not mention the country of origin where the clinical study was conducted.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not explicitly stated.
    • Qualifications of Experts: "Trained medical staff" were used for the auscultation method. ANSI/AAMI/ISO 81060-2:2013 guidelines require specific training and certification for these observers.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • No adjudication method is mentioned. The ground truth was established by "trained medical staff" using the auscultation method, implying a direct comparison rather than a multi-reader consensus process for image interpretation.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was NOT done. This device is an automated blood pressure monitor, not an AI-assisted diagnostic imaging device. The "study" here refers to the clinical validation of its measurement accuracy compared to a standard reference method (auscultation), not a human-in-the-loop diagnostic improvement study.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence, this was a standalone performance study for the device. The study validated the "accuracy of blood pressure measurements by HWA11 based on an oscillometric method as compared to an auscultation method." This is the device's inherent performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Expert Reference Method (Auscultation): The ground truth for blood pressure measurements was established using the "auscultation method using a calibrated sphygmomanometer by trained medical staff." This is considered the gold standard for clinical blood pressure measurement in such validation studies.
  7. The sample size for the training set:

    • This is not applicable for this type of device and study. The device is a non-AI, oscillometric blood pressure monitor. It uses a defined algorithm, not a machine learning model that requires a "training set" in the common AI sense.
  8. How the ground truth for the training set was established:

    • Not applicable (see point 7).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

October 27, 2023

Shenzhen Jumper Medical Equipment Co., Ltd. Danyin Li - Regulatory Affairs Specialist D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen. Guangdong. China 518103

Re: K231310

Trade/Device Name: Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 5, 2023 Received: May 5, 2023

Dear Danyin Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K231310

Device Name Electronic Blood Pressure Monitor

Indications for Use (Describe)

The Electronic Blood Pressure Monitor is intended to measure the systolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. The intended wrist circumference is 13.5-21.5 cm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the text "K231310" twice, once slightly above and to the left of the other. Below this text is the word "JUMPER" in a large, pink, sans-serif font. To the right of "JUMPER" are four Chinese characters, "京柏医疗", in a gray, sans-serif font, which translates to "Jingbai Medical".

Section 5

510(K) Summary

1. Prepared Date: 2023-02-14

2. Submitter Information [807.92(a)(1)]

NameShenzhen Jumper Medical Equipment Co., Ltd.
AddressD Building, No. 71, Xintian Road, Fuyong Street, Baoan,Shenzhen, Guangdong, China, 518103
Tel+86-755-26696279
Fax+86-755-26852025

3. Submission Correspondent [807.92(a)(1)]

Contact personDanyin Li
E-mailep03@jumper-medical.com
Tel+86-755- 26696279

4. Subject Device Information [807.92(a)(2)]

Proprietary NameElectronic Blood Pressure Monitor
Common NameNoninvasive blood pressure measurement system
ModelHWA11, HWA10
ClassificationSystem, Measurement, Blood-Pressure, Non-Invasive
Review PanelCardiovascular
Product CodeDXN
Regulation Number870.1130
Regulatory ClassII

5. Predicate Device [807.92(a)(3)]

K182127, Omron Healthcare, Inc.
Trade/Device NameWrist Blood Pressure Monitor
ModelBP6100
Regulation NameNoninvasive Blood Pressure Measurement System
Regulatory ClassClass II

{4}------------------------------------------------

JUMPER 京相医疗

Product Code DXN

6. Indications for Use [807.92(a)(5)]

The Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. The intended wrist circumference is 13.5-21.5 cm.

7. Device Description [807.92(a)(4)]

The Electronic Blood Pressure Monitor, is a battery driven automatic non-invasive blood pressure monitor, comprised of the host machine and the wrist cuff. It can measure systolic and diastolic blood pressure and pulse rate of the adult person at wrist via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. The device can store 199 groups of measurement data for two users.

8. Summary of technological characteristics of device compared to the predicate devices, see the table

A comparison of key similarities and differences between the proposed devices (Electronic Blood Pressure Monitor, HWA11, HWA10) and the predicate devices (K182127) is provided below:

{5}------------------------------------------------

JUMPER 京柏医疗

|| ' 深圳京柏医疗科技股份有限公司
SHENZHENJUMPERMEDICAL EQUIPMENT CO.,LT

Propose DevicePropose DevicePredicated DeviceComparisonIndications for Use
Product NameElectronic Blood PressureMonitorElectronic Blood PressureMonitorWrist Blood PressureMonitor/The Electronic Blood PressureMonitor is intended tomeasure the systolic anddiastolic blood pressure aswell as the pulse rate of adultperson via non-invasiveoscillometric technique atmedical facilities or at home.The intended wristcircumference is 13.5-21.5 cm.The Electronic Blood PressureMonitor is intended to measurethe systolic and diastolic bloodpressure as well as the pulserate of adult person via non-invasive oscillometric techniqueat medical facilities or at home.The intended wristcircumference is 13.5-21.5 cm.The device is a digitalmonitor intended for use inmeasuring blood pressureand pulse rate in adultpatient population withwrist circumference rangingfrom 5.3 inches to 8.5inches(13.5 cm to 21.5 cm).The device detects theappearance of irregularheartbeats duringmeasurement and gives awarning signal withreadings.DifferentNote1DifferentNote1Same
ManufacturerShenzhen Jumper MedicalEquipment Co., Ltd.Shenzhen Jumper MedicalEquipment Co., Ltd.Omron Healthcare, Inc./Environment ofUseMedical Facilities/Home UseMedical Facilities /Home UseHome UseHome UseSame
AppearanceImage: Electronic Blood Pressure MonitorImage: Electronic Blood Pressure MonitorImage: Wrist Blood Pressure Monitor/PatientPopulationAdultsAdultsAdultsAdultsSame
ModelHWA11HWA10BP6100 (K182127)/Over-The-Counter (OTC)YesYesYesYesSame
Classification21 CFR§870.1130,Noninvasive blood pressuremeasurement system21 CFR§870.1130,Noninvasive blood pressuremeasurement system21 CFR§870.1130,Noninvasive blood pressuremeasurement systemSameMeasurementmethodCuff oscillometric methodCuff oscillometric methodCuff oscillometric methodCuff oscillometric methodSame
Product CodeDXN - Noninvasive bloodpressuremeasurementDXN - Noninvasive bloodpressuremeasurementDXN - Noninvasive bloodpressuremeasurementSameWrist CuffCircumference13.5-21.5 cm.13.5-21.5 cm.13.5-21.5 cm.13.5-21.5 cm.Same

{6}------------------------------------------------

JUMPER 京柏医疗

10 深圳京柏医疗科技股份有限公
SHENZHEN JUMPER MEDICAL EQUIPMENT CO., L

{7}------------------------------------------------

---------

K231310

MeasurementRangeCuff pressure range 0 to295mmHgPulse Rate: 40 to 199beats/min.Cuff pressure range 0 to295mmHgPulse Rate: 40 to 199 beats/min.Cuff pressure range 0 to299 mmHgPulse Rate: 40 to 180beats/min.DifferentNote2
AccuracyBlood Pressure: Within$\pm$ 3mmHgPulse Rate: Within $\pm$ 5 % ofreadingBlood Pressure: Within$\pm$ 3mmHgPulse Rate: Within $\pm$ 5 % ofreadingBlood Pressure: Within$\pm$ 3mmHgPulse Rate: Within $\pm$ 5 % ofreadingSame
PressureSensorPiezo resistance sensorPiezo resistance sensorPiezo resistance sensorSame
InflationMethodAutomatic inflation withelectric pumpAutomatic inflation with electricpumpAutomatic inflation withelectric pumpSame
DeflationMethodAutomatic rapid deflationvalveAutomatic rapid deflation valveAutomatic rapid deflationvalveSame
Display TypeLCD digital displayLCD digital displayLCD digital displaySame
Power SourceTwo "AAA" batteriesTwo "AAA" batteriesTwo "AAA" batteriesSame
OperatingConditions5 to 40 °C (41 to 104 °F)15 to 85 %RH (non-condensing)700 to 1060 hPa5 to 40 °C (41 to 104 °F)15 to 85 %RH (non-condensing)700 to 1060 hPa10 to 40 °C (50 to 104 °F)15 to 90%RH (non-condensing)800 to 1060 hPaDifferentNote3

{8}------------------------------------------------

Storage/Transport ConditionsDimensionsWeight
-20 to 55 °C (-4 to 131 °F)10 to 93 %RH (non-condensing)700 to 1060 hPa61 (W) x 24.1(D) × 87 (H) mmApprox. 111±5g (without batteries)
-20 to 55 °C (-4 to 131 °F)10 to 93 %RH (non-condensing)700 to 1060 hPa61 (W) x 24.1(D) × 87 (H) mmApprox. 111±5g (without batteries)
-20 to 60 °C (-4 to 140 °F)10 to 90 %RH (non-condensing)93 (W) x 20 (D) × 62 (H) mmApprox. 86g (3.0 oz) (not including battery)
DifferentNote3DifferentNote4DifferentNote4

{9}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

The proposed device and the predicate device have similar technological characteristics. Both devices are battery-operated wrist blood pressure monitors intended for home use and employ the oscillometric method for measuring blood pressure and pulse. The devices are intended for the same wrist circumference range of 13.5 cm to 21.5 cm. The devices share the same accuracy range of pressure reading of ±3 mmHg and the same accuracy range for pulse rate reading of ± 5%.

There are minor differences in technical specifications and features between the HWA11, HWA10 proposed device and BP6100 predicate device would be discussed below:

SUBSTANTIAL EQUIVALENCE

The proposed Indications for Use for the Electronic Blood Pressure Monitor is like the Indications for Use for the predicate device.

They both intended for use in homecare for blood pressure and pulse rate measurement of adults. And different use environment has been tested by IEC 60601-1 & ISO 80601-2-30 and showed positive result. While the proposed device is designed for household and medical center use and the predicate device is for home use, the difference will not influence the safety and effectiveness of the devices because household environment is more complex. The proposed device can also detect the appearance of irregular heartbeats during measurement and gives a warning signal with readings, which is mentioned in "Features" and stated as "Indication for irregular heartbeats". So, both devices are identical for this. (Note1)

The pressure range is narrower than the predicate device while the pulse rate of the proposed device are larger than the predicate device. Both contain the blood pressure and pulse range of most people, and the measurement range of proposed device is fully verified according to IEC 80601-2-30. Therefore, the differences do not bring additional risks. (Note 2)

Although the operating atmospheric pressure of the proposed device is slightly broader than the predicate device, while the storage temperature and humidity of the proposed device are little narrower, they all comply with IEC 60601-1, IEC 60601-1-11 and IEC 80601-2-30, so this difference will not raise any safety or effectiveness issue. (Note 3)

Minor differences in the dimensions and weight do not impact the safety or performance of blood pressure or pulse rate measurements (Note4). Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the proposed device is substantially equivalent to the predicate device.

PERFORMANCE DATA [807.92(b)]

All necessary bench and clinical testing was conducted on the Electronic Blood Pressure Monitor

{10}------------------------------------------------

Model HWA11, HWA10 to support a determination of substantial equivalence to the predicate device.

[807.92(b)(1)] Nonclinical Testing Summary:

The nonclinical, bench testing included:

  • Comparative blood pressure and pulse rate testing to the predicate device
  • Performance verification testing to confirm acceptable performance of device features and functions
  • Cleaning verification testing to confirm device retains its performance when cuff is cleaned with household detergents as may be required in home use environment
  • Other nonclinical safety testing included:
    • Biocompatibility of patient-contacting materials per ISO 10993-1 requirements
    • Electrical safety, electromagnetic compatibility, and electrostatic discharge testing
    • Software verification and validation

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the HWA11 & HWA10 meet the established specifications necessary for consistent performance during its Intended Use. In addition, the collective bench testing results demonstrate that the HWA10 does not raise different questions of safety or effectiveness for blood pressure and pulse rate measurement in a home use environment when compared to the predicate device.

Clinical Testing Summary [807.92(b)(2)]

A clinical investigation was conducted with the objective of validating the accuracy of blood pressure measurements by HWA11 based on an oscillometric method as compared to an auscultation method using a calibrated sphygmomanometer by trained medical staff. This study was conducted in accordance with guideline per ANSI/AAMI/ISO 81060-2:2013 Noninvasive sphygmomanometers — Part 2: Clinical investigation of automated measurement type. The results demonstrated that BP6100 performed equivalently to the auscultation method and is in conformance with ANSI/AAMI/ISO 81060-2:2013.

CONCLUSIONS [807.92(b)(3)]

The conclusions drawn from the nonclinical and clinical tests performed in support of HWA11, HWA10 demonstrate that the device is safe and effective, and performs at least as safely and effectively as the legally marketed K182127 predicate device.

SUMMARY

The Electronic Blood Pressure Monitor (Model HWA11, HWA10) is substantially equivalent to the Omron Model BP6100 predicate device(K182127).

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).