K Number
K231303

Validate with FDA (Live)

Date Cleared
2024-01-26

(266 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm.

Device Description

The Arm-type Fully Automatic Digital Blood Pressure Monitor (BPM) series, noninvasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, the electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min.

The pulse rate measurement compares the longest and the shortest time intervals of detected pulse waves to the mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.

Meanwhile, this blood pressure monitor device can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display measurement results from the blood pressure monitor device, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

AI/ML Overview

The provided text describes the 510(k) submission for the "Arm-type Fully Automatic Digital Blood Pressure Monitor, Model: BM 92". The main study used to demonstrate the device meets acceptance criteria is a clinical investigation following the ISO 81060-2:2018+AMD2020 standard, which focuses on the clinical investigation of intermittent automated measurement type non-invasive sphygmomanometers.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by compliance with the ISO 81060-2:2018+AMD2020 standard, which specifies accuracy requirements for blood pressure monitors. The document states:
"The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020."

While specific numerical acceptance criteria (e.g., mean difference and standard deviation between device and reference measurements) are not explicitly stated in the document, the adherence to this ISO standard implies meeting its criteria. The general accuracy specifications are given in the comparison table:

Criteria CategoryAcceptance Criteria (Implied by ISO 81060-2:2018+AMD2020)Reported Device Performance (Summary)
Blood Pressure AccuracyWithin acceptable scope specified in ISO 81060-2:2018+AMD2020."The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020."
Static Pressure Accuracy± 3mmHg± 3mmHg
Pulse Accuracy± 5%± 5%

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: "A total of 90 patients participated in the clinical study (2 were excluded because they were part of other clinical trials)." So, the effective sample size for the test set was 88 patients.
  • Data Provenance: The document does not explicitly state the country of origin. However, the manufacturer is JOYTECH Healthcare Co., Ltd. located in Hangzhou, Zhejiang, China. Given this, it's highly probable the study was conducted in China. The study was a prospective clinical investigation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states: "A manual Mercury Sphygmomanometer was used as a reference device." This implies that the ground truth was established by human readers taking measurements using a mercury sphygmomanometer. However, the number of experts and their qualifications are not specified in the provided text. The ISO 81060-2 standard typically requires trained observers for reference measurements, but the details are omitted here.

4. Adjudication Method for the Test Set

The text mentions: "Same arm sequential method was adopted during the clinical testing. A manual Mercury Sphygmomanometer was used as a reference device." This suggests a comparative method, but it does not describe an adjudication method for discrepancies, which is usually relevant for subjective interpretations (e.g., image reading). For blood pressure measurements, the reference device readings are typically considered the objective ground truth; thus, a formal "adjudication" between multiple observers might not be explicitly detailed if the standard protocol for reference measurement was followed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-assisted image interpretation devices to assess the impact of AI on human reader performance. The device described is an automated blood pressure monitor, not an AI interpretation system. Its function is to directly measure and display blood pressure, not to assist human readers in interpreting complex data.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, in essence, the device's accuracy was tested in a standalone manner. The device "determines" systolic and diastolic pressures using its internal algorithms (oscillometric method). The clinical study directly evaluated the accuracy of these measurements against a reference standard (manual mercury sphygmomanometer). The performance cited ("The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020") refers to the device's direct measurement capability.

7. The Type of Ground Truth Used

The ground truth used was comparative measurements against a manual Mercury Sphygmomanometer. The text states: "A manual Mercury Sphygmomanometer was used as a reference device." This is a standard method for establishing ground truth for automated blood pressure monitors.

8. The Sample Size for the Training Set

The document does not specify the sample size for a training set. This is common for traditional medical devices like blood pressure monitors, where "training" in the machine learning sense isn't explicitly detailed. The device likely relies on established oscillometric principles and algorithms developed over time, not a specific "training set" of patient data in the context of deep learning models.

9. How the Ground Truth for the Training Set Was Established

As no specific "training set" for an AI model is mentioned or implied, the question of how its ground truth was established is not applicable in the context of the provided document. The device utilizes an "oscillometric method" and "software algorithms within the microprocessor," which are based on known physiological principles and traditional signal processing rather than data-driven machine learning training requiring a designated ground-truth dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 26, 2024

Joytech Healthcare Co., Ltd Ren Yunhua General Manager No.365, Wuzhou Road, Yuhang Economic Development Zone Hangzhou, Zhejiang 311100 China

Re: K231303

Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor, Model: BM 92 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 26, 2023 Received: December 26, 2023

Dear Ren Yunhua:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231303

Device Name

Arm-type Fully Automatic Digital Blood Pressure Monitor

Model: BM 92

Indications for Use (Describe)

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm.

Type of Use (Select one or both, as applicable)

□ Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1/1

PSC Publishing Services (301) 443-6740

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for TECH-I, with the letters in blue and red. The first part of the logo is a blue arrow pointing to the right. The letters 'TEC' are in blue, and the letter 'H' is in red. To the right of 'TECH-I' are some Chinese characters in blue.

510(k) Summary

K231303

The assigned 510(k) number is:

2.1 Subjectter's Identification:

Name: JOYTECH Healthcare Co., Ltd.

Add.: No. 365. Wuzhou Road. Yuhang Economic Development Zone. Hangzhou City.

311100 Zhejiang, China.

Contact Person: Yunhua Ren

Phone: +86-571-81957767

Fax: +86-571-81957750

Email: renyh@sejoy.com

2.2 Name of the Device:

Trade Name: Arm-type Fully Automatic Digital Blood Pressure Monitor

Model: BM 92

Common Name: Blood Pressure Monitor

Classification name: Non-invasive blood pressure measurement System

21 CFR 870-1130, Class II, 74-DXN.

2.3 Classification Information:

Regulation Number: 870.1130

Product Code: DXN

Device Class: II

Panel: 74 Cardiovascular

2.4 Predicate Device Information:

The Arm-Type Fully Automatic Digital Blood Pressure Monitor is substantially equivalent to the following predicate device:

510(k)numberPredicatedevice modelProductcodeManufacturer
----------------------------------------------------------------------------

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of the word "JOYTECH" are three Chinese characters in blue.

K192609GUS622DXNGlobalcare Medical Technology Co., Ltd

2.5 Device Description:

The Arm-type Fully Automatic Digital Blood Pressure Monitor (BPM) series, noninvasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, the electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats/min.

The pulse rate measurement compares the longest and the shortest time intervals of detected pulse waves to the mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.

Meanwhile, this blood pressure monitor device can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display measurement results from the blood pressure monitor device, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for JOYTECH. The logo is in blue and red. The text "JOYTECH" is in blue, except for the "T", which is in red. To the right of "JOYTECH" is some text in another language, also in blue.

The detail functions for BM 92 are listed in table below:

ModelsFeaturesABCDEFGHIJKLMN
BM 92YN2*120memoriesYYYY125954422-42cm71*76YYYY

A = Powered by alkaline battery

B= Powered by AC adaptor

C = Memory Size

D = Time & Date

E = WHO (World Health Organization) Classification Indicator

F = Average of all memories, AM/PM Average for the last 7 days

G = Irreqular Heartbeat Detection

H = Outside Demission (L x W x H in unit mm)

I = Cuff Size

  • J = LCD Size (Viewing Area in unit mm)
    K = LCD Backlight

L = Voice

  • M= Bluetooth Function
    N= Cuff Loose Detection

Note:

Y= Yes N = No

The device is designed and manufactured according to IEC80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

2.6 Intended Use:

The Arm-type Fully Automatic Digital Blood Pressure Monitor is intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm.

2.7 Comparison of Technological Characteristics with predicate device:

The arm-type blood pressure monitor BM 92 manufactured by JOYTECH have the same arm cuff type, features and specifications with the Globalcare blood pressure monitor GUS622 which 510k number is K192609, therefore we choose the device act as

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for JOYTECH-I. The logo is in blue and red. The text "JOYTECH-I" is in English, and there is also Chinese text to the right of the English text.

the predicate device. The detail comparison of technical characteristic as below:

ComparisonitemSubject devices for BM 92Predicate device GUS622(K192609)ComparisonresultExplanation
The trade nameArm-type Fully AutomaticDigital Blood PressureMonitorGlobalcare GUS622 BloodPressure Monitor/
ManufacturerJOYTECH Healthcare Co.,Ltd.Globalcare Medical TechnologyCo., Ltd/
Recommendedclassificationregulation21CFR 870.1130,Noninvasive BloodPressure MeasurementSystem21CFR 870.1130,Noninvasive BloodPressure MeasurementSystemIdentical
RegulatoryclassIIIIIdentical
Panel74 Cardiovascular74 CardiovascularIdentical
Product codeDXNDXNIdentical
forIndicationuseThe Arm-type FullyAutomatic Blood PressureMonitors are intended tomeasure blood pressure(systolic and diastolic) andpulse rate of adults andadolescents over 12 yearsof age with circumferenceranging from 22cm to 42cm.The Globalcare GUS622automatic Blood PressureMonitor is indicated for home usefor the non-invasivemeasurement of diastolic andsystolic blood pressures andpulse rate of adults by means ofan inflatable cuff which iswrapped around the upper arm.The cuff circumference is limitedto 22 to 42 cm.SimilarNote 1
MeasuringprincipleOscillometric methodOscillometric methodIdentical
MeasurementtypeDetermined during inflationDetermined during inflationIdentical
Cuff locationUpper armUpper armIdentical
Specification
MeasuringrangeSystolic Pressure:60mmHg ~ 260 mmHg;Diastolic Pressure:40mmHg ~ 200 mmHg;Pulse:30~180 Beats/MinuteSystolic Pressure:50mmHg ~ 280 mmHg;Diastolic Pressure:30mmHg ~ 220 mmHg;Pulse:40~199 Beats/MinuteSimilar.Note 2
Maxcuffpressure299 mmHg300mmHgSimilar
AccuracyStatic Pressure:± 3mmHgPulse: ± 5%Pressure:±0.4kPa ( 3mmHg)Pulse value: ± 5%Identical
InflationBy air pumpBy air pumpIdentical
PressurereleaseBy solenoid valveBy solenoid valveIdentical
OperatingTemp. &Temp.: 10°C40°CHumidity: 1593%RHTemp.: 10°C~40°CHumidity: < 90% relativeSimilarNote 3
humidityAtmospheric:800hPa~1060hPahumidity (non-condensing)Pressure:800hPa~1050hPa
Storage Temp.& humidityTemp.: -25°C~55°CHumidity:<93% RHTemp.: -20°C55°CHumidity:< 90% relativehumidity (non-condensing)Pressure:800hPa1050hPaSimilarNote 4
Cuffcircumference22-42 cm22-42 cmIdentical
Supply powersource4x1.5V AAA battery4x1.5V AAA battery or ACadaptorSimilarNote 5
BluetoothtransmissionfunctionYesUnclearDifferentNote 6
SterilizationNot applicableNot applicableIdentical
IrregularheartbeatIrregular heartbeatIrregular heartbeatIdentical
Memory2*120 Memories2 x 60 memory spacesSimilar.Note 7
CufflooseindicatorYesYesIdentical

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for JOYTECH. The logo is blue, except for a red plus sign. To the right of the plus sign are three Chinese characters.

Note 1: The intended use of subject device and predicate device are similar. The clinical accuracy of subject device applied to adolescents over 12 years of age population have been verified and validated, which comply with the requirements of ISO 81060-2. And the clinical safety also have been validated as for without any AE or side-effect in clinical investigation.

Note 2: The measuring range of the subject device have been verified by IEC 60601-1 and 80601-2-30 standard test.

Note 3: The operation environment of the subject device have been verified by IEC 60601-1 and 80601-2-30 standard test.

Note 4: The storage environment of the subject device have been verified by IEC 60601-1 and 80601-2-30 standard test.

Note 5: The different power supply of the subject device have been verified by IEC 60601-1 and IEC 60601-1-2 standard test.

Note 6: The intended use of Bluetooth functionality is to transmit the measurement result from point A to point B only, not intended for active patient monitoring. It not only has been verified and validated through Software verified and validation report, but also passed the IEC 60601-1, IEC60601-1-2 standard test.

{9}------------------------------------------------

Note 7: The quantity of storage readings is different between the subject device and predicate device. The change in the specification is documented, and the change does not affect the intended use or the fundamental scientific technology.

2.8. Performance Data:

Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted.

The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.

The following National and International Standards were utilized for testing the subject device.

Electrical Safety and performance requirements:

  • IEC 60601-1:2005+AMD1:2012+AMD2:2020. Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.
  • AAMI ES60601-1:2005+AMD1:2012+AMD2:2021, Medical Electrical Equipment.
  • IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

Home-used medical equipment requirements and environmental test:

  • IEC 60601-1-11:2015+AMD1:2020 General requirements for basic safety and ● essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
    Electromagnetic compatibility requirements:

  • IEC 60601-1-2:2014+AMD1:2020 , Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    Bio-compatibility Evaluation for patient contacting components:

  • ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red, with the word "JOYTECH" in blue and red letters. To the right of the word "JOYTECH" are three Chinese characters in blue. The logo is simple and modern.

  • . ISO10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

FCC Test

  • FCC Part15 Subpart C
  • RF Exposure Evaluation

Guidance Document:

The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"

The test results all meet or exceed the requirement of these standards.

2.9. Discussion of Clinical Tests Performed:

Clinical Validation:

  • ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers — Part 2: Clinical investigation of intermittent automated measurement type.

The key factors that affect clinical trials, such as the sensors, and the core algorithms are the same between BM92 and DBP-6279B. Therefore, the clinical investigation report for the model for DBP-6279B is applicable for BM92.

A total of 90 patients participated in the clinical study (2 were excluded because they were part of other clinical trials). Same arm sequential method was adopted during the clinical testing. A manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteers to take part in the clinical study. All the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

See Attachment 2-1: ISO 81060-2 Clinical Trial Report

7

{11}------------------------------------------------

2.10. Conclusions:

Based on the information provided in this submission, the subject devices and predicate devices have very similar intended uses and fundamental technological characteristics. It has been tested, documented, and demonstrated that the differences between them do not raise any questions on safety or effectiveness.

This submitted arm-type for BM 92 manufactured by JOYTECH Healthcare Co., Ltd. Has been found to be substantially equivalent to the predicate device (Globalcare GUS622 Blood Pressure Monitor) manufactured by Globalcare Medical Technology Co., Ltd. (K192609).

--- End of this section---

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).