K Number
K231274

Validate with FDA (Live)

Device Name
Natural Cycles
Date Cleared
2023-08-24

(114 days)

Product Code
Regulation Number
884.5370
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Natural Cycles is a stand-alone software application, intended for women 18 years and older, to monitor their fertility. Natural Cycles can be used for preventing a pregnancy (contraception) or planning a pregnancy (conception).

Device Description

Natural Cycles is an over-the-counter web and mobile-based standalone software application that monitors a woman's menstrual cycle using information entered by the user and informs the user about her past, current, and future fertility status. The following information is used by the Natural Cycles software:

  • daily body temperature measurements
  • information about the user's menstruation cycle (i.e., start date, number of days)
  • optional ovulation or pregnancy test results
    A proprietary algorithm evaluates the data and returns the user's fertility status.
    Natural Cycles is available in three modes: Contraception (NC° Birth Control), Conception (NC° Plan Pregnancy), and Pregnancy (NCº Follow Pregnancy). For NCº Birth Control mode, the device provides predictions of "not fertile," shown as green days, and "use protection," shown as red days, that allow the user to determine the days on which her risk of conception is highest, and then make choices about either abstaining from sex or using a barrier method of contraception to prevent pregnancy.
    In addition to measuring daily basal body temperature with an oral thermometer with two decimal points, the predicate submission cleared the Oura Ring for automatic temperature input to the Natural Cycles algorithm, and the current submission expands to allow the device to utilize automatic temperature inputs from the Apple Watch.
    Natural Cycles can be used by women 18 years and older. Women who have been on hormonal birth control within 60 days prior to using Natural Cycles have a higher risk of becoming pregnant when compared to women who have not been on hormonal birth control within the 12 months prior to using the device may not be right for women who have a medical condition where pregnancy would be associated with a significant risk to the mother or the fetus.
AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary:

This device, Natural Cycles, is a software application for contraception. The 510(k) submission seeks to add the Apple Watch as an accepted source for daily basal body temperature (BBT) input, to be used by the existing Natural Cycles fertility algorithm. The core algorithm remains unchanged.

1. Table of Acceptance Criteria and Reported Device Performance

The core acceptance criterion for this submission revolves around the comparability of the algorithm's performance when using Apple Watch temperature data versus a traditional oral thermometer, specifically in the context of identifying ovulation and providing "green days" (not fertile). The study did not define explicit acceptance criteria in terms of specific sensitivity/specificity thresholds, but rather focused on demonstrating that the Apple Watch input did not negatively impact the safety and effectiveness as compared to the predicate device.

Acceptance Criteria (Implicit)Reported Device Performance
Algorithm's ability to identify ovulation with Apple Watch temperature data is comparable to oral thermometer data."The results of the clinical study demonstrated that when temperature was inputted from either the Apple Watch or the two-decimal place oral thermometer, the Natural Cycles algorithm was able to identify that ovulation had occurred, which was confirmed in the study by comparison to the 153 positive LH test results."
Use of Apple Watch temperature data does not increase the risk of unintended pregnancy (e.g., by incorrectly assigning "green days" in the fertile window)."Compared to the two-decimal place oral thermometer, the Natural Cycles algorithm provided 0.6 fewer green days (not fertile) in the luteal phase of the menstrual cycle when the input temperature was from the Apple Watch without increasing the risk of unintended pregnancy." (Note: A reduction in "green days" in the luteal phase, post-ovulation, would generally be considered a more cautious or neutral outcome in terms of pregnancy risk, as the luteal phase, while generally infertile, should not be incorrectly identified as fertile.)
The change in temperature input source does not raise different questions of safety and effectiveness.The summary concludes: "A comparison of intended use and technological characteristics combined with performance data demonstrates that Natural Cycles is as safe and effective as the predicate device and supports a determination of substantial equivalence." The 0.6 fewer green days with Apple Watch vs. 1.6 additional green days with Oura Ring (predicate) are noted, but explicitly stated: "however, this difference does not impact the safety and effectiveness of Natural Cycles for its intended use." This implicitly means the performance variations were within acceptable safety margins for contraceptive effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • 272 users (de-identified information) met criteria for the main study.
    • A subset of 104 users was assessed for the ovulation Key Performance Indicator (KPI), requiring a positive luteinizing hormone (LH) test.
    • Data were collected over a total of 1918 menstrual cycles.
    • 505 of these cycles were considered complete and met study criteria.
    • 153 of the complete cycles had at least one positive LH test.
  • Data Provenance:
    • Country of Origin: The majority (58%) of the women were located in the European Union (EU) and the United Kingdom (UK), with the remainder (42%) located in the US.
    • Retrospective or Prospective: Prospective. The study involved women actively wearing the Apple Watch nightly, recording temperatures, and providing LH test results over time.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This summary does not specify the number or qualifications of experts used. The ground truth for ovulation was established through Luteinizing Hormone (LH) test results. These are over-the-counter medical tests used by individuals to detect the LH surge that precedes ovulation, and their interpretation is generally self-reported by the user.

4. Adjudication Method for the Test Set

No explicit adjudication method is mentioned. The ground truth appears to be based on user-reported LH test results, which are biochemical markers rather than subjective expert interpretations requiring adjudication. The comparison was statistical, between the algorithm's output using Apple Watch data versus oral thermometer data, benchmarked against LH test results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC study was not performed. This device is a standalone software application, and the study focused on the performance of the algorithm with different temperature inputs, not on human reader performance or improvement with AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, this was effectively a standalone algorithm performance evaluation. The study compared how the Natural Cycles algorithm performed in identifying ovulation and predicting "green days" when fed with temperatures from the Apple Watch versus an oral thermometer. The algorithm's output was then confirmed against LH test results. Humans were involved in providing the data (wearing the watch, taking oral temperature, performing LH tests) but not in interpreting images or making diagnoses that the AI would then assist with.

7. The Type of Ground Truth Used

The primary ground truth used for validating the ovulation detection performance was Luteinizing Hormone (LH) test results. This is a form of biochemical marker or outcomes-related data (in the sense that an LH surge is a biological event preceding ovulation).

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for the training set. It refers to the Natural Cycles algorithm and states "There have been no changes to how the Natural Cycles algorithm determines the daily fertility status." This implies the algorithm was already trained and validated as part of prior 510(k) submissions (e.g., K202897, the predicate device). The current study is an evaluation of new input data rather than a re-training effort.

9. How the Ground Truth for the Training Set Was Established

Since the training set details are not provided in this specific document (as the algorithm itself wasn't re-trained for this submission), this information cannot be definitively extracted from the provided text. However, given that Natural Cycles has been cleared for contraception previously, its algorithm would have been extensively validated with large datasets of menstrual cycle data, BBT readings, and confirmed ovulation events (likely through similar biochemical markers like LH, or possibly other clinical methods).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 24, 2023

Natural Cycles Nordic AB % Shelia Hemeon-Heyer, MS, JD, FRAPS President Heyer Regulatory Solutions, LLC 125 Cherry Lane Amherst, MA 01002

Re: K231274

Trade/Device Name: Natural Cycles Regulation Number: 21 CFR§ 884.5370 Regulation Name: Software Application for Contraception Regulatory Class: II Product Code: PYT Dated: July 27, 2023 Received: July 27, 2023

Dear Shelia Hemeon-Heyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael T. Bailey -S

For Monica D. Garcia, Ph. D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231274

Device Name Natural Cycles

Indications for Use (Describe)

Natural Cycles is a stand-alone software application, intended for women 18 years and older, to monitor their fertility. Natural Cycles can be used for preventing a pregnancy (contraception) or planning a pregnancy (conception).

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K231274 - Natural Cycles

  • A. Submitter: Natural Cycles Nordic AB St Eriksqatan 63B 112 34 Stockholm, Sweden Contact: Raoul Scherwitzl, PhD, CEO and Co-Founder
    Phone: +46 707174866 E-mail: Raoul.scherwitzl@naturalcycles.com

  • B. Correspondent: Hever Requlatory Solutions LLC 125 Cherry Lane Amherst, MA 01002
    Contact: Sheila Hemeon-Heyer E-mail: Sheila@heyer-requlatory.com

  • C. Date Prepared: August 23, 2023

Device Name and Classification Information: D.

Trade Name:Natural Cycles
Common Name:Software application for contraception
Regulation Name:Software application for contraception
Regulation Number:21 CFR 884.5370
Regulatory Class:II
Product Code:PYT (Device, fertility diagnostic, contraceptive, software application)

E. Predicate Device(s): K202897 Natural Cycles

The predicate device has not been subject to a design-related recall.

F. Device Description:

Natural Cycles is an over-the-counter web and mobile-based standalone software application that monitors a woman's menstrual cycle using information entered by the user and informs the user about her past, current, and future fertility status. The following information is used by the Natural Cycles software:

  • daily body temperature measurements
  • information about the user's menstruation cycle (i.e., start date, number of days)
  • optional ovulation or pregnancy test results ●

A proprietary algorithm evaluates the data and returns the user's fertility status.

Natural Cycles is available in three modes: Contraception (NC° Birth Control), Conception (NC° Plan Pregnancy), and Pregnancy (NCº Follow Pregnancy). For NCº Birth Control mode, the

{4}------------------------------------------------

device provides predictions of "not fertile," shown as green days, and "use protection," shown as red days, that allow the user to determine the days on which her risk of conception is highest, and then make choices about either abstaining from sex or using a barrier method of contraception to prevent pregnancy.

In addition to measuring daily basal body temperature with an oral thermometer with two decimal points, the predicate submission cleared the Oura Ring for automatic temperature input to the Natural Cycles algorithm, and the current submission expands to allow the device to utilize automatic temperature inputs from the Apple Watch.

Natural Cycles can be used by women 18 years and older. Women who have been on hormonal birth control within 60 days prior to using Natural Cycles have a higher risk of becoming pregnant when compared to women who have not been on hormonal birth control within the 12 months prior to using the device may not be right for women who have a medical condition where pregnancy would be associated with a significant risk to the mother or the fetus.

G. Indications for Use Statement

Natural Cycles is a stand-alone software application, intended for women 18 years and older, to monitor their fertility. Natural Cycles can be used for preventing a pregnancy (contraception) or planning a pregnancy (conception).

Comparison with Predicate Device H.

A detailed comparison of the intended use and technological features of the subject and predicate device are described in the table below:

ParameterSubject DeviceNatural CyclesK231274Predicate DeviceNatural CyclesK202897Comparison
Indications forUse StatementNatural Cycles is a stand-alone software application,intended for women 18 yearsand older, to monitor theirfertility.Natural Cycles can be usedfor preventing a pregnancy(contraception) or planning apregnancy (conception).Natural Cycles is a stand-alone software application,intended for women 18 yearsand older, to monitor theirfertility.Natural Cycles can be usedfor preventing a pregnancy(contraception) or planninga pregnancy (conception).Identical
UseEnvironmentApp is downloaded touser's smartphone andused in the homeenvironmentApp is downloaded touser's smartphone andused in the homeenvironmentIdentical
ParameterSubject DeviceNatural CyclesK231274Predicate DeviceNatural CyclesK202897Comparison
InputInformation• Manual input of two-decimal daily basal bodytemperature (BBT)measurements, orautomatic input from theOura ring or Apple Watch.• Manual input ofinformation about theuser's menstruationcycle, i.e., start date,number of days.• Optional manual input ofovulation or pregnancytest results.• Manual input of two-decimal daily basalbody temperature(BBT) measurementsor automatic input fromthe Oura ring.• Manual input ofinformation about theuser's menstruationcycle, i.e., start date,number of days.• Optional manual input ofovulation or pregnancytest results.Expandedoption for anew source ofdaily basalbodytemperatureinformation.
OutputInformation• NC° Birth Control: Foreach day, whether thewoman is fertile (red) ornon- fertile (green), withdescriptive texts.• NC° Plan Pregnancy:Fertility status results aredisplayed as a scale forfertile days, and green fornon-fertile days, togetherwith description texts.• NC° FollowPregnancy: Provideseducationalinformation about theprogress of thepregnancy.• All uses in Preventmode or Plan modereceive the ovulationdate for the month anddaily statement offertility status. Historicdata is available for all• Prevent mode: For eachday, whether the womanis fertile (red) or non-fertile (green), withdescriptive texts.• Plan mode: Fertilitystatus results aredisplayed as a scale forfertile days, and greenfor non-fertile days,together withdescription texts.• Pregnancy mode:Provides educationalinformation about theprogress of thepregnancy.• All uses in Preventmode or Plan modereceive the ovulationdate for the month anddaily statement offertility status. Historicdata is available for allIdentical; titlesfor the modeshave beenchanged

{5}------------------------------------------------

The subject and predicate device have identical indications for use statements and have the same intended use - to predict fertile and non-fertile days for use in providing patient-specific recommendations related to contraception. The subject and predicate device have different technological characteristics (i.e., temperature data input sources). This difference does not raise different questions of safety and effectiveness as compared to the predicate device.

{6}------------------------------------------------

l. Summary of Non-Clinical Performance Testing

  • . Software documentation provided in accordance with the 2005 FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices to support device software with a major level of concern.
  • . Cybersecurity information provided in accordance with the 2014 FDA guidance document Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
  • . Human factors testing in DEN170052 was used to support that device users could safely use the device.

Summary of Clinical Performance Testing J.

The purpose of this 510(k) is to add the option for the user to obtain the daily basal body temperature information used by the Natural Cycles fertility algorithm from the Apple Watch. There have been no changes to how the Natural Cycles algorithm determines the daily fertility status.

A clinical study was conducted to compare the daily temperatures generated by a wearable temperature monitor, the Apple Watch, to a two decimal place oral thermometer for use by the Natural Cycles algorithm. The study included data of de-identified information from 272 users after all applied criteria were met. A subset of the data, 104 users, was assessed for the ovulation key performance indicator (KPI), which required a positive result from a luteinizing hormone (LH) test. The mean age of the study participants was 30.5 ± 5.1 years, and the mean menstrual cycle length for complete cycles during the study was 28.8 ± 5.5 days. The majority (58%) of the women were located in the European Union and the United Kingdom, with the remainder located in the US (42%).

The women wore the Apple Watch nightly, with the user's temperature information being transferred to the subject device each morning and recorded their temperature each morning as taken using a two-decimal place oral thermometer. Data were collected over a total of 1918 menstrual cycles, 505 of which were considered complete cycles and met the applied criteria, per the study definition. The participants were also asked to record the results of luteinizing hormone (LH) tests each cycle. At least one positive LH test was submitted in 153 of the complete study cycles.

The results of the clinical study demonstrated that when temperature was inputted from either the Apple Watch or the two-decimal place oral thermometer, the Natural Cycles algorithm was able to identify that ovulation had occurred, which was confirmed in the study by comparison to the 153 positive LH test results. Compared to the two-decimal place oral thermometer, the Natural Cycles algorithm provided 0.6 fewer green days (not fertile) in the luteal phase of the menstrual cycle when the input temperature was from the Apple Watch without increasing the risk of unintended pregnancy. This is in contrast with the predicate device, which provided an additional 1.6 green days when the input temperature was from the Oura Ring; however, this difference does not impact the safety and effectiveness of Natural Cycles for its intended use.

{7}------------------------------------------------

K. Conclusion

A comparison of intended use and technological characteristics combined with performance data demonstrates that Natural Cycles is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5370 Software application for contraception.

(a)
Identification. A software application for contraception is a device that provides user-specific fertility information for preventing a pregnancy. This device includes an algorithm that performs analysis of patient-specific data (e.g., temperature, menstrual cycle dates) to distinguish between fertile and non-fertile days, then provides patient-specific recommendations related to contraception.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate the contraceptive effectiveness of the software in the intended use population.
(2) Human factors performance evaluation must be provided to demonstrate that the intended users can self-identify that they are in the intended use population and can correctly use the application, based solely on reading the directions for use for contraception.
(3) Software verification, validation, and hazard analysis must be performed. Documentation must include the following:
(i) A cybersecurity vulnerability and management process to assure software functionality; and
(ii) A description of the technical parameters of the software, including the algorithm used to determine fertility status and alerts for user inputs outside of expected ranges.
(4) Labeling must include:
(i) The following warnings and precautions:
(A) A statement that no contraceptive method is 100% effective.
(B) A statement that another form of contraception (or abstinence) must be used on days specified by the application.
(C) Statements of any factors that may affect the accuracy of the contraceptive information.
(D) A warning that the application cannot protect against sexually transmitted infections.
(ii) Hardware platform and operating system requirements.
(iii) Instructions identifying and explaining how to use the software application, including required user inputs and how to interpret the application outputs.
(iv) A summary of the clinical validation study and results, including effectiveness of the application as a stand-alone contraceptive and how this effectiveness compares to other forms of legally marketed contraceptives.