K Number
K231230

Validate with FDA (Live)

Date Cleared
2023-12-05

(221 days)

Product Code
Regulation Number
886.1120
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.

Device Description

Optina-4C™ (model: MHRC-C1N) is a mydriatic fundus camera (also called retinal camera) that presents eye care practitioners (optometrists and ophthalmologists) with a series of the retina obtained sequentially at specific wavelengths (colors) in the spectral range 905 nm to 450 nm in steps of 5 nm.

Pictures of the retina are obtained on a field-of-view of 31.5° without contact with the eye. The patient is positioned in front of the device with the chin on the chinrest and forehead on the forehead rest and a positioning system is used to align the camera relative to the patient's eye. An external fixation target is available to guide the patient's eye. A focus wheel allows for the accommodation for eye refractive error in the range of -15 to +15 diopters. The images are displayed on a monitor and can be saved on the computer for future consultation.

The illumination light of Optina-4C™ is provided by a Tunable Laser Source (TLS). The TLS selects a narrow band of light from a broadband white illumination source. Only a single monochromatic band can output the TLS at a time. The alignment of the retinal camera relative to the patient's eye phase is performed with the illumination light set at a wavelength of 700 nm. The image acquisition phase consists of the sequential acquisition of a series of 92 monochromatic images at wavelengths from 905 nm in steps of 5 nm. Each frame is captured with an exposure time of 10 ms, resulting in a total acquisition time of 920 ms.

The 92 images can be visualized one by one in Optina-4C™ acquisition software. The retinal images captured by Optina-4C™ are monochromatic images having a spectral bandwidth of "10 nm centered on the wavelength indicated in the upper left corner of the visualization pane, with a spectral accuracy of 7.5 nm.

AI/ML Overview

The provided text is a 510(k) summary for the Optina-4C (MHRC-C1N) device, which is a mydriatic fundus camera. The document primarily focuses on demonstrating substantial equivalence to a predicate device (MHRC-C1) and outlines various performance tests conducted to support this claim.

However, the document does not contain specific acceptance criteria for the device's performance regarding its intended use of capturing retinal images, nor does it present a detailed study that proves the device meets such criteria in terms of clinical accuracy or utility.

Instead, the document focuses on:

  • Technological equivalence: Stating that the main elements of the hardware, illumination, and monochromatic images are the same as the predicate.
  • Verification of safety and regulatory compliance: Covering software evaluation, electrical safety, electromagnetic compatibility, biocompatibility, and compliance with recognized consensus standards for ophthalmic cameras (ISO 15004-1:2020, ISO 10940:2009, ANSI Z80.36:2021).
  • Spectral accuracy and reliability of retinal images: This was evaluated in an "eye model using a reference material with tabulated spectral bands," but no specific acceptance criteria or performance metrics (e.g., sensitivity, specificity, accuracy for a particular clinical task) are provided from this evaluation.

Therefore, I cannot populate the requested table and information as the necessary details regarding acceptance criteria, reported device performance related to a specific clinical task, test set characteristics, expert involvement, adjudication, or a multi-reader multi-case study are not present in the provided text. The document describes tests to ensure the device functions safely and in a manner similar to its predicate, but not how well it performs a particular diagnostic task against established clinical benchmarks or ground truth.

Missing Information:

  • No specific acceptance criteria for diagnostic performance (e.g., image quality metrics, sensitivity/specificity for detecting a condition).
  • No reported clinical performance metrics directly related to the device's ability to 'capture images' that translates to a clinical outcome.
  • No details on a clinical study where the device's images were evaluated for a specific purpose against a ground truth.

The document seems to describe the device as an image capture device, and the "performance testing" focuses on its technical specifications and safety rather than its diagnostic accuracy for a particular condition.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Optina Diagnostics Inc. Sarah Lemaire QA/RA Manager 8200 Decarie Boulevard Montreal, QC H4P 2P5 Canada

Re: K231230

Trade/Device Name: Optina-4C (MHRC-C1N) Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: November 6, 2023 Received: November 7, 2023

Dear Sarah Lemaire:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Elvin Y. Ng-S

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231230

Device Name Optina-4C (MHRC-C1N)

Indications for Use (Describe)

Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K231230

Image /page/3/Picture/1 description: The image contains the logo for Optina Diagnostics. The logo consists of a blue circular graphic on the left and the company name on the right. The word "Optina" is in gray, and the word "DIAGNOSTICS" is in a smaller, teal font below it.

Special 510(k): Device Modification

Optina-4C™ Model: MHRC-C1N

510(k) Summary

APPLICANT:Optina Diagnostics, Inc
8200 Decarie Blvd, Suite 220
Montreal (QC)
H4P 2P5
Phone: 514-394-0797
ESTABLISHMENTREGISTRATIONNUMBERRegistration Number: 3014606890
Owner Operator Number: 10081796
DATE PREPARED:August 5, 2022
CONTACT PERSON:Sarah Lemaire
QA/RA Manager
Email: slemaire@optinadx.com
Phone: 579-421-0126
TRADE NAME:Optina-4C™
MODEL:MHRC-C1N
COMMON NAME:Fundus camera
CLASSIFICATIONNAME:Camera, Ophthalmic, Ac-Powered (21 CFR
886.1120, Product Code HKI)
DEVICECLASSIFICATION:Class II
PREDICATE DEVICESMHRC-C1 (K200254)

1. Description of the Modified Device

Optina-4C™ (model: MHRC-C1N) is a mydriatic fundus camera (also called retinal camera) that presents eye care practitioners (optometrists and ophthalmologists) with a series of the retina obtained sequentially at specific wavelengths (colors) in the spectral range 905 nm to 450 nm in steps of 5 nm.

Pictures of the retina are obtained on a field-of-view of 31.5° without contact with the eye. The patient is positioned in front of the device with the chin on the chinrest and forehead on the forehead rest and a

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Optina Diagnostics. On the left side of the logo is a blue circular design that is made up of several overlapping circles. To the right of the circular design is the company name, "Optina", in gray text, with the word "DIAGNOSTICS" in smaller, teal text underneath.

positioning system is used to align the camera relative to the patient's eye. An external fixation target is available to guide the patient's eye. A focus wheel allows for the accommodation for eye refractive error in the range of -15 to +15 diopters. The images are displayed on a monitor and can be saved on the computer for future consultation.

The illumination light of Optina-4C™ is provided by a Tunable Laser Source (TLS). The TLS selects a narrow band of light from a broadband white illumination source. Only a single monochromatic band can output the TLS at a time. The alignment of the retinal camera relative to the patient's eye phase is performed with the illumination light set at a wavelength of 700 nm. The image acquisition phase consists of the sequential acquisition of a series of 92 monochromatic images at wavelengths from 905 nm in steps of 5 nm. Each frame is captured with an exposure time of 10 ms, resulting in a total acquisition time of 920 ms.

The 92 images can be visualized one by one in Optina-4C™ acquisition software. The retinal images captured by Optina-4C™ are monochromatic images having a spectral bandwidth of "10 nm centered on the wavelength indicated in the upper left corner of the visualization pane, with a spectral accuracy of 7.5 nm.

2. Indications for Use

Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.

3. Technological Characteristics

Optina-4C™ is similar in fit, form, and function to the predicate mydriatic fundus cameras. The main elements of the hardware including contact areas, the illumination of the retina, and the monochromatic images obtained with Optina-4C™ are the same as for the predicate device.

In comparison with the predicate device, Optina-4C™ has updated software and labeling and features network connectivity. Performance testing was performed to verify that this technological difference raised no new concerns regarding the safety and efficacy of the device.

4. Performance Testing

The following performance data were conducted to support the substantial equivalence determination.

● Software evaluation

Optina-4C™ uses embedded Off-The-Shelf software and Software-of-Unknown-Provenance. This software was evaluated for use in accordance with the FDA's Guidance for Industry "Off-The-Shelf Software Use in Medical Devices" (September 2019) based on a "Moderate Level of Concern" and

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Optina Diagnostics. On the left side of the logo is a blue circular graphic that is made up of many overlapping circles. To the right of the graphic is the company name "Optina" in gray, with the word "DIAGNOSTICS" in a smaller font size and blue color underneath.

IEC 62304:2006 + A1:2015 based on a "Safety class B".

● Electrical safety and electromagnetic compatibility

lt was verified that the Optina-4C™ complies with the standard ANSI AAMI ES60601-1:2005/(R)2012+A1:2012 for electrical safety and with the standard IEC 60601-1-2:2014 for electromagnetic compatibility.

● Evaluation of recognized consensus standards for ophthalmic cameras

Optina-4C™ was found to comply with the recognized consensus standard ISO 15004-1:2020 specifying general requirements for ophthalmic instruments.

The following performance data were conducted on the marketed device (MHRC-C1). No changes that can affect the evaluation or results of these recognized consensus standards were implemented on the Optina-4C™.

● Spectral accuracy and reliability of the retinal images

The spectral accuracy of the illumination light of the MHRC-C1 was verified using a spectrometer. The spectral accuracy and reliability of the retinal images were evaluated in an eye model using a reference material with tabulated spectral bands.

● Biocompatibility

The intact skin from the chin and forehead of the patient is intended to contact the MHRC-C1 for a short period of time. The biocompatibility of the MHRC-C1 was assessed in accordance with ISO 10993-1:2018.

● Evaluation of recognized consensus standards for ophthalmic cameras

The MHRC-C1 was found to comply with the recognized consensus standard ISO 10940:2009 specifying product requirements for fundus camera.

The MHRC-C1 was found to comply with the recognized consensus standard ANSI Z80.36: 2021 specifying fundamental requirements for optical radiation safety for ophthalmic instruments.

5. Substantial Equivalence

In conclusion, the performance testing support that the Optina-4C™ meets the recognized consensus standards for a fundus camera and that retinal imaging at multiple wavelengths (colors) may be accurately and reliably achieved. The proposed device did not raise new concerns regarding its safety and efficacy for its intended use and was therefore deemed substantially equivalent to the predicate device.

3

§ 886.1120 Ophthalmic camera.

(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.