K Number
K231176

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-05-25

(30 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stingray LP Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the coronary vasculature.

When used as part of the system consisting of the CrossBoss™ Catheter, and Stingray Guidewire, the Stingray LP Catheter is indicated for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions [CTOs]) prior to PTCA or stent intervention.

Device Description

The Stingray LP Catheter facilitates the placement, support and steering of a guidewire into discrete regions of the coronary vasculature through the central guidewire lumen or through one of two side-ports. These side-ports are on opposite sides of the balloon and are identified by radiopaque markers. The side-ports communicate with the central guidewire lumen and facilitate guidewire steering (at an angle to the central lumen) by allowing the quidewire to exit the Stingray LP Catheter. The Stingray LP Catheter contains a small balloon used for fluoroscopic orientation on the distal tip of a flexible shaft. The distal end of the catheter is hydrophilic coated. The Stingray LP Catheter is compatible with 6F guide catheters with minimum inner diameter of 0.070 in (1.7 mm), and may be used with guidewires ≤0.014 in (0.36 mm).

AI/ML Overview

This document describes a 510(k) premarket notification for the Stingray LP Catheter. It is a submission for a medical device and, as such, does not contain information related to AI/ML or software performance evaluation studies. Therefore, I cannot provide information on acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, adjudication methods, or MRMC studies, as the provided text does not contain any of these details in the context of an AI/ML device study.

The document focuses on explaining that the device is substantially equivalent to a previously cleared predicate device based on its indications for use, technological characteristics, and non-clinical performance data (bench and biocompatibility testing).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

May 25, 2023

Boston Scientific Gabrielle Reynolds Regulatory Affairs Specialist II Two Scimed Place Maple Grove, Minnesota 55311

Re: K231176

Trade/Device Name: Stingray LP Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: April 25, 2023 Received: April 25, 2023

Dear Gabrielle Reynolds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Lydia S Lydia S. Glaw -S. Glaw -S. Glaw -S
Date: 2023.05.25 15:21:58

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231176

Device Name Stingray LP Catheter

Indications for Use (Describe)

The Stingray LP Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the coronary vasculature.

When used as part of the system consisting of the CrossBoss™ Catheter, and Stingray Guidewire, the Stingray LP Catheter is indicated for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions [CTOs]) prior to PTCA or stent intervention.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

per 21 CFR §807.92

SponsorBoston Scientific Corporation300 Boston Scientific WayMarlborough, Massachusetts 01752USA
Contact Name andInformationGabrielle ReynoldsTwo Scimed PlaceMaple Grove, MinnesotaEmail: gabrielle.reynolds@bsci.com
Date PreparedApril 25 , 2023
Proprietary NameStingray™ LP Catheter
Common NamePercutaneous Catheter
Product CodeDQY
ClassificationClass II, 21 CFR Part 870.1250
Predicate DeviceStingray™ LP Catheter, K152401, cleared December 4, 2015

Device Description

The Stingray LP Catheter facilitates the placement, support and steering of a guidewire into discrete regions of the coronary vasculature through the central guidewire lumen or through one of two side-ports. These side-ports are on opposite sides of the balloon and are identified by radiopaque markers. The side-ports communicate with the central guidewire lumen and facilitate guidewire steering (at an angle to the central lumen) by allowing the quidewire to exit the Stingray LP Catheter. The Stingray LP Catheter contains a small balloon used for fluoroscopic orientation on the distal tip of a flexible shaft. The distal end of the catheter is hydrophilic coated. The Stingray LP Catheter is compatible with 6F guide catheters with minimum inner diameter of 0.070 in (1.7 mm), and may be used with guidewires ≤0.014 in (0.36 mm).

Indications for Use / Intended Use

The Stingray LP Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the coronary vasculature.

When used as part of the system consisting of the CrossBoss™ Catheter, Stingray LP Catheter, and Stingray Guidewire, the Stingray LP Catheter is indicated for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions [CTOs]) prior to PTCA or stent intervention.

Comparison of Technological Characteristics

The proposed Stingray LP Catheter incorporates substantially equivalent materials, packaging, operating principles, fundamental scientific technology, manufacturing processes, sterilization process, and intended use as the predicate Stingray LP Catheter (K152401).

{4}------------------------------------------------

Non-clinical Performance Data

Bench and Biocompatibility testing were performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed Stingray LP Catheter has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the device testing.

The following bench tests were completed in support of the proposed Stingray LP Catheter:

  • Balloon Preparation, Deployment and Retraction .
  • . Balloon Rated Burst Pressure (RBP)
  • Catheter Bond Strength ●
  • Flexibility and Kink Test ●
  • Torque Strength Test .

The following biocompatibility tests were completed in support of the proposed Stingray LP Catheter:

  • . Cytotoxicity
  • . Sensitization
  • Intracutaneous Reactivity ●
  • Acute Systemic Toxicity ●
  • Materials Mediated Pyrogenicity ●
  • Hemolysis (Direct Contact)
  • Hemolysis (Extract Method) ●
  • Partial Thromboplastin Time ●
  • Platelet and Leukocyte Count
  • Complement Activation ●
  • USP <661> Physiochemical ●
  • . Genotoxicity

Conclusion

Based on the indications for use, technological characteristics, and safety and performance testing, the proposed Stingray™ LP Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate Stingray™ LP Catheter (K152401).

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).