K Number
K231149

Validate with FDA (Live)

Device Name
qXR-CTR
Date Cleared
2023-09-22

(154 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans.

qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm.

The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities.

The system is suitable for adults > 22 years of age.

The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.

Device Description

The qXR-CTR is a non-invasive software medical device designed to be installed on the computer with specific system requirements. It is a radiological computer-assisted software system that automatically analyzes DICOM chest X-ray images in PA view and outputs the cardiac diameter, thoracic diameter, and CTR through an artificial intelligence algorithm. The structured report includes a preview of the compressed chest X-ray image with the automatically derived CTR result and annotation line, indicating the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity.

AI/ML Overview

Acceptance Criteria and Study Details for qXR-CTR

1. Table of Acceptance Criteria and Reported Device Performance

MeasureAcceptance Criteria (Predicate)Reported Device Performance (qXR-CTR)
Cardiac Diameter RMSE8.81 mm7.55 (6.95, 8.34) mm
Thoracic Diameter RMSE14.4 mm5.43 (4.95, 6.11) mm
Cardiac Diameter MAENot Reported5.66 (5.0) mm
Thoracic Diameter MAENot Reported4.04 (3.63) mm
CTR0.03 (0.02, 0.03)0.02 (0.02)

2. Sample Size and Data Provenance for Test Set

  • Sample Size: 435 scans.
  • Data Provenance: Retrospectively collected chest X-rays from various parts of the US.

3. Number and Qualifications of Experts for Ground Truth (Test Set)

  • Number of Experts: 3
  • Qualifications: ABR (American Board of Radiology) thoracic radiologists with a minimum of 10 years of experience.

4. Adjudication Method for Test Set

The adjudication method is not explicitly stated, but with 3 experts establishing ground truth, a consensus method (e.g., majority vote or discussion to agreement) is implied. It's not explicitly stated as 2+1, 3+1, or none.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. The document describes a standalone performance evaluation of the qXR-CTR against ground truth established by experts.

6. Standalone Performance

Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The reported performance metrics (RMSE, MAE) are intrinsic to the device's measurements compared to expert-established ground truth.

7. Type of Ground Truth Used

Expert consensus. The ground truth was established by three ABR thoracic radiologists.

8. Sample Size for Training Set

The sample size for the training set is not provided in the given text.

9. How Ground Truth for Training Set was Established

The method for establishing ground truth for the training set is not described in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

Qure.ai Technologies % Srinidhi Ragunathan Regulatory Clinical Affairs Manager Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon(E) Mumbai, Maharashtra 400063 INDIA

September 22, 2023

Re: K231149

Trade/Device Name: qXR-CTR Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: August 24, 2023 Received: August 24, 2023

Dear Srinidhi Ragunathan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231149

Device Name qXR-CTR

Indications for Use (Describe)

qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans.

qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm.

The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities.

The system is suitable for adults > 22 years of age.

The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Qure.ai's qXR-CTR

1 SUBMITTER

Qure.ai Technologies Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon (E), Mumbai 400 063 Phone: +91-9768123013 Primary Contact Person: Srinidhi Ragunathan Secondary contact person: Bunty Kundnani

Date Prepared: August 23, 2023

2 DEVICE

Name of Device:qXR-CTR
Common or Usual Name:Automated radiological image processing software
Classification Name:Medical image management and processing system
Regulatory Class:Class II
Regulation Number:21 CFR 892.2050
Product Code:QIH

3 PREDICATE DEVICE

Name of Device:EFAI Intelligent Cardiothoracic Ratio (iCTR) Assessment System
Manufacturer:Ever Fortune.Al Co., Ltd.
510(k) Number:K212624

{4}------------------------------------------------

4 INTENDED USE / INDICATIONS FOR USE:

qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans.

qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm.

The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities.

The system is suitable for adults ≥ 22 years of age.

The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.

ഗ DEVICE DESCRIPTION

The qXR-CTR is a non-invasive software medical device designed to be installed on the computer with specific system requirements. It is a radiological computer-assisted software system that automatically analyzes DICOM chest X-ray images in PA view and outputs the cardiac diameter, thoracic diameter, and CTR through an artificial intelligence algorithm. The structured report includes a preview of the compressed chest X-ray image with the automatically derived CTR result and annotation line, indicating the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity.

ട COMPARISON OF THE PREDICATE DEVICE

qXR-CTR is technologically similar to the predicate device, EFIA iCTR Assessment in regard to intended use and technological characteristics. Both are medical image management and processing system intended to read chest X-rays to measure target structures (cardiac and thoracic diameter). The algorithms function similarly and there are no notable technological differences between the subject and predicate devices.

In terms of establishing substantial equivalence, the subject and predicate device have the same intended use, as an image processing tool that measures anatomical structures and produces caselevel output. The indications for use proposed for the subject device are similar those of the predicate device.

Predicate DeviceSubject Device
Device NameEFAI Intelligent CardiothoracicRatio (iCTR) AssessmentSystemqXR-CTR
510(k) NumberK212624NA
Regulation21 CFR 892.205021 CFR 892.2050
Regulation Descriptionradiological computer-assisted softwareradiological computer-assistedsoftware
Table 1 Comparison between gXR-CTR and the Predicate Device
Predicate DeviceSubject Device
Product CodeQIHQIH
Device typeQuantifying softwareQuantifying software
ManufacturerEver Fortune.Al Co., Ltd.Qure.ai Technologies
Intended use /Indications for UseEFAI Intelligent CardiothoracicRatio Assessment System (oriCTR) is a software for use byhospitals and clinics toautomatically assess thecardiothoracic ratio (CTR) of achest X-ray image from the X-ray imager subject. The iCTR isdesigned to measure themaximal transverse diameterof heart and maximal innertransverse diameter ofthoracic cavity and calculatethe CTR of a chest X-ray imagein posterior-anterior (PA)chest view using an artificialintelligence algorithm.Intended users of thesoftware are aimed to thephysicians or other licensedpractitioners in the healthcareinstitutions, such as clinics,hospitals, healthcare facilities,residential care facilities andlong-term care services. Thesystem is suitable for adultsbetween 20 - 80 years of age.Its results are not intended tobe used on a stand-alonebasis for clinical decisionmaking or otherwise precludeclinical assessment ofcardiothoracic ratio (CTR)cases.qXR-CTR is a deep-learning basedsoftware for use by hospitals andclinics for automated assessment ofthe CTR on chest X-ray (CXRs) scans.qXR-CTR is designed to measure theratio of the maximal transversediameter of the heart (CD) and themaximal inner transverse diameter(TD) of the thoracic cavity andcalculate the CTR value on posterior-anterior view chest view using anartificial intelligence algorithm.The intended users of this device arephysicians or licensed practitioners inhealthcare institutions, such as clinics,hospitals, residential care facilities,long-term care services, andhealthcare facilities.The system is suitable for adults ≥ 22years of age.The device is used to aid the intendedusers and results are not intended tobe used on a stand-alone basis forclinical decision making or otherwisepreclude clinical assessment of CTRcases.
Algorithm formeasurement of targetstructuresAI AlgorithmAI Algorithm
ModalityChest X-ray in DigitalRadiography (DR)Chest X-ray in Digital Radiography (DR)
InputPost-anterior (PA) view chestX-ray imagePost-anterior (PA) view chest X-rayimage
Input formatDICOMDICOM
Output FormatReports, DICOM SecondaryCapture seriesPdf report is generated, DICOMSecondary capture is returned to thePACS
Report StructureReport will be output in theDICOM and JSON file formatReport will be output in the DICOMand JSON file format which isstructured with following information
Predicate DeviceSubject Device
following information andfunction tool: 1) CTR 2)adjustable annotation line(maximal inner borderdiameter of thoracic cavityand maximal diameter ofheart) 3) the trajectory of CTR(including chest X-ray)and function tool: 1) CTR 2) annotationline (maximal inner border diameter ofthoracic cavity and maximal diameterof heart)
Intended UserPhysicians or other licensedpractitioners in the healthcareinstitutionsPhysicians or other licensedpractitioners in the healthcareinstitutions
Target Population20-80 years≥ 22 years
ModalityChest X-ray in DigitalRadiographyChest X-ray in Digital Radiography
Intended UseEnvironmentHealthcare institutions(Clinics, hospitals, healthcarefacilities, residential carefacilities and long-term careservices)Healthcare institutions (Clinics,hospitals, healthcare facilities,residential care facilities and long-termcare services)
Software device thatYesYes
operates on off the shelf
hardware
Software devices usesYesYes
software algorithms forimage
Diameter MeasurementYesYes
StorageSaved in JSON and DICOM fileformat text for DICOM andDICOM file formatThe text report is generated with the CXRscan on a Structured Report (SR) format forPACS-based mode. For the Web-basedmode, the user can download a PDFversion of the report with the sameinformation
Software RequirementUbuntu 18.04 (Web browser:chrome 88.0.4324.182 orabove)Ubuntu 20.04+
Performance metrics of the predicate device and qXR-CTR
Performance levelRMSE:RMSE:
Cardiac Diameter = 8.81 mmThoracic Diameter = 14.4 mmCardiac Diameter = 7.55 mmThoracic Diameter = 5.43 mm
Mean Absolute Error:NRMean Absolute Error:5.66 (5.0) mm
MeasurementRoot Mean Squared Error (95%CI) in mmMean Absolute Error (SD) in mm
Cardiac Diameter7.55 (6.96, 8.38)5.66 (5.0)
Thoracic Diameter5.43 (4.94, 6.09)4.04 (3.63)
CTR0.03 (0.02, 0.03)0.02 (0.02)

{5}------------------------------------------------

{6}------------------------------------------------

SOFTWARE TESTING 7

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device has a Moderate level of concern.

{7}------------------------------------------------

8 PERFORMANCE TESTING - CLINICAL

The performance of qXR-CTR was validated by clinical tests. All the safety parameters of the device were verified in accordance with the software specifications and applicable performance standards and met the acceptance criteria(passed), demonstrating that the software fulfills all its requirement specifications.

Clinical studies were conducted on retrospectively collected chest X-rays to evaluate the performance of qXR-CTR for measuring the cardiac diameter and thoracic diameter.

The study included 435 scans with PA view from various parts of the US. The ground truth was established by 3 ABR thoracic radiologists with a minimum of 10 years of experience. The RMSE of the device in measuring the target structures exceeded the success criteria with RMSE for cardiac diameter being 7.55 (6.95, 8.34) mm and thoracic diameter being 5.43 (4.95, 6.11) mm. The predicate EFAI iCTR Assessment's performance was cardiac diameter RMSE: 8.81 mm and thoracic diameter: 14.4 mm.

Further, the mean absolute error along with standard deviation for both the cardiac and thoracic diameter was measured at 5.66 (5.0) mm and 4.04 (3.63) mm, respectively.

Table 2 Overall Results of Accuracy Testing of qXR-CTR

ರಿ CONCLUSION

The comparison in Table 1 and the software and performance testing presented above demonstrate that the qXR-CTR device is substantially equivalent to the predicate device. The qXR-CTR is a software only device, similar to the predicate (EFAI iCTR Assessment) and has no new risks related to safety and effectiveness. The qXR-CTR has the same intended users and similar indications, technological characteristics, and principles of operation as the predicate devices operate in parallel to the standard of care workflow. The performance testing demonstrates that the qXR-CTR performs as intended and is therefore substantially equivalent to the predicate. Software and Clinical testing support that the device performs in accordance with the device requirements.

N/A